- Library Home /
- Search Collections /
- Open Collections /
- Browse Collections /
- UBC Faculty Research and Publications /
- Calibration, pressure and temperature sensitivity of...
Open Collections
UBC Faculty Research and Publications
Calibration, pressure and temperature sensitivity of Vaisala GMM 220A CO2 sensors Christen, Andreas; Nesic, Zoran
Abstract
Objective: Calibrate and test performance of four Vaisala CARBOCAP® Carbon Dioxide Module GMM 220A sensors that measure CO2 mixing ratios. The following characteristics were tested: sensor individual zero offset and span (Section 2), temperature sensitivity (Section 3), pressure sensitivity (Section 4), sensor noise (Section 5) and sensor drift over 24 hours (Section 6). The sensors tested were GMM 220 #59, #60, #66, and #67. All sensors are used in the soil respiration set-up of APBI 244 / GEOB 204 at UBC.
Item Metadata
Title |
Calibration, pressure and temperature sensitivity of Vaisala GMM 220A CO2 sensors
|
Creator | |
Date Issued |
2015-10-21
|
Description |
Objective: Calibrate and test performance of four Vaisala CARBOCAP® Carbon Dioxide Module GMM 220A sensors that measure CO2 mixing ratios. The following characteristics were tested: sensor individual zero offset and span (Section 2), temperature sensitivity (Section 3), pressure
sensitivity (Section 4), sensor noise (Section 5) and sensor drift over 24 hours (Section 6). The sensors tested were GMM 220 #59, #60, #66, and #67. All sensors are used in the soil respiration set-up of APBI 244 / GEOB 204 at UBC.
|
Subject | |
Genre | |
Type | |
Language |
eng
|
Date Available |
2017-01-31
|
Provider |
Vancouver : University of British Columbia Library
|
Rights |
Attribution-NonCommercial-NoDerivs 2.5 Canada
|
DOI |
10.14288/1.0221513
|
URI | |
Affiliation | |
Peer Review Status |
Unreviewed
|
Scholarly Level |
Faculty
|
Rights URI | |
Aggregated Source Repository |
DSpace
|
Item Media
Item Citations and Data
Rights
Attribution-NonCommercial-NoDerivs 2.5 Canada