Osteoporotic spine fixation : a biomechanical investigation Tan, Juay Seng 2006

Your browser doesn't seem to have a PDF viewer, please download the PDF to view this item.

Item Metadata

Download

Media
831-ubc_2006-201060.pdf [ 29.55MB ]
Metadata
JSON: 831-1.0080752.json
JSON-LD: 831-1.0080752-ld.json
RDF/XML (Pretty): 831-1.0080752-rdf.xml
RDF/JSON: 831-1.0080752-rdf.json
Turtle: 831-1.0080752-turtle.txt
N-Triples: 831-1.0080752-rdf-ntriples.txt
Original Record: 831-1.0080752-source.json
Full Text
831-1.0080752-fulltext.txt
Citation
831-1.0080752.ris

Full Text

OSTEOPOROTIC SPINE FIXATION: A BIOMECHANICAL INVESTIGATION  by JUAY SENG TAN  A Thesis Submitted in Partial Fulfilment of the Requirements for the Degree of DOCTOR OF PHILOSOPHY in THE FACULTY OF GRADUATE STUDIES (Mechanical Engineering)  THE UNIVERSITY OF BRITISH COLUMBIA Feb 2006 © Juay Seng Tan, 2006  Abstract  Intervertebral fusion of the aging spine is a common surgical procedure for a wide range of clinical problems including trauma, deformity and degeneration. Spinal instrumentation is often used to stabilize the spinal column and help facilitate the fusion. The main problems of using spinal instrumentation in elderly patients (many with osteoporosis) are loss of fixation due to loosening and adjacent segment effects.  The objectives of this thesis were to compare  existing techniques of osteoporotic spine fixation using in vitro  biomechanical models and to  investigate new surgical strategies, so as to provide information to the spine surgeon to decrease the incidence of implant loosening and reduce degenerative changes at the adjacent levels in elderly patients. Pedicle screw loosening is common in patients with poor bone quality. Augmentation of pedicle screws with laminar hooks, sublaminar wires and/or cement are techniques used clinically to minimize pedicle screw loosening. A novel testing configuration was developed to apply physiologic load in relevant magnitudes and directions to study pedicle screw motion in cadaveric vertebrae. Twenty-four lumbar vertebrae (L3-L5) were divided into three groups and instrumented bilaterally with pedicle screws.  A model of loosened screws was created by  overdrilling the pedicle screw trajectory. Two techniques of screw augmentation, using laminar hooks, sublaminar wires and/or calcium phosphate cement, were carried out within each group of specimens. Paired comparisons of screw motion magnitudes at the screw head and screw tip, and overall motion patterns were carried out. Screws augmented with cement exhibited primarily rotational motion patterns with minimal translational motion, and most closely resembled the motion pattern of screws in high density bone.  All three augmentation  techniques resulted in a 50% mean reduction in screw motion. No significant differences in motion magnitudes were found between the different augmentation techniques. The results suggested that augmentation of pedicle screws with calcium phosphate cement might provide enhanced fixation over laminar hook and sublaminar wire augmentation. Anterior interbody device subsidence is another clinical problem that leads to implant loosening.  Ninety-six thoracolumbar cadaveric vertebrae were compressed to failure using  various shaped indentors. The focus was primarily on the superior endplate.  In forty-eight  specimens, the purpose was to determine the effect of cage shape and cage size on cagevertebra interface properties.  In the remaining forty-eight specimens, the effects of pedicle  screw insertion and cement augmentation of screws on cage-vertebra interface properties were determined for various shapes and sizes of devices. Failure load, failure strength and stiffness ii  were compared.  In t h e f i r s t p a r t , l a r g e r s i z e d d e v i c e s ( w i t h 4 0 % e n d p l a t e c o v e r a g e v e r s u s w i t h  2 0 % ) r e s u l t e d in 7 5 % h i g h e r f a i l u r e l o a d .  C l o v e r - l e a f s h a p e d d e v i c e s a l s o r e s u l t e d in a t l e a s t  4 5 % higher failure load a n d failure s t r e n g t h o v e r kidney or elliptical s h a p e d d e v i c e s . f a i l u r e w a s f o u n d t o o c c u r in a s e m i - e l l i p t i c a l z o n e b e n e a t h t h e i n t e r b o d y d e v i c e . part,  pedicle  screw  insertion  vertebral interface strength.  disrupted  the  underlying  trabecular  bone  and  Trabecular  In t h e s e c o n d reduced  cage-  C e m e n t a u g m e n t a t i o n of pedicle s c r e w s structurally r e i n f o r c e d t h e  u n d e r l y i n g t r a b e c u l a e , a n d r e s u l t e d in a n i m p r o v e m e n t in c a g e - v e r t e b r a i n t e r f a c e s t r e n g t h .  No  differences w e r e found between cement a u g m e n t e d anterior vertebral body screws and pedicle screws.  T h e r e w e r e a l s o n o d i f f e r e n c e s in i n t e r f a c e s t r e n g t h b e t w e e n i n d e n t o r s h a p e s f o l l o w i n g  s c r e w insertion, with and without c e m e n t augmentation. cement  augmentation  of vertebral  screws  might  Larger sized interbody devices  reduce  the  incidence  of  interbody  and  device  s u b s i d e n c e in t h e o s t e o p o r o t i c s p i n e . C e m e n t a u g m e n t a t i o n of pedicle s c r e w s a n d extension of posterior instrumentation are t w o t e c h n i q u e s t o i m p r o v e t h e s t a b i l i z a t i o n o f s p i n a l f i x a t i o n in t h e p r e s e n c e o f o s t e o p o r o s i s . U s i n g t h e traditional flexibility protocol a n d a n e w hybrid test t e c h n i q u e , i n t e r s e g m e n t a l r a n g e of motion within the fused  segment  and  at a d j a c e n t  levels following  cement  augmentation  of  p e d i c l e s c r e w s a n d e x t e n s i o n o f p o s t e r i o r r o d s w e r e c o m p a r e d u s i n g a n in vitro b i o m e c h a n i c a l thoracolumbar model.  T w e l v e T 9 - L 3 s e g m e n t s w e r e t e s t e d in a r e p e a t e d m e a s u r e s f a s h i o n t o  d e t e r m i n e the effects of posterior rod extension a n d c e m e n t a u g m e n t a t i o n .  Intact flexibility t e s t s  u n d e r 5 N m p u r e m o m e n t s w e r e f i r s t c a r r i e d o u t in a x i a l r o t a t i o n , l a t e r a l b e n d i n g , a n d f l e x i o n extension.  T h e initial t e s t c o n f i g u r a t i o n i n c l u d e d a T 1 1 c o r p e c t o m y a n d r e c o n s t r u c t i o n w i t h a n  extendable  cage,  and  pedicle  screws  instrumentation  from  T10-T12.  Flexibility tests  were  c a r r i e d o u t r a n d o m l y o n 1) t h e i n i t i a l t e s t c o n f i g u r a t i o n , 2 ) w i t h a d d i t i o n a l p o s t e r i o r r o d e x t e n s i o n to L1 u s i n g s t a n d a r d rigid rods a n d 3) e x t e n s i o n u s i n g flexible a c e t a l rods.  The T12 and  p e d i c l e s c r e w s t r a j e c t o r y w e r e o v e r d r i l l e d a s b e f o r e t o s i m u l a t e l o o s e n i n g in t h e spine.  osteoporotic  F o l l o w i n g flexibility tests u n d e r t h e s e t h r e e fixation m e t h o d s , t h e T 1 2 a n d L1  were augmented with cement. randomized order again.  L1  screws  T h e flexibility tests w e r e r e p e a t e d for t h e t h r e e c o n d i t i o n s  in  T h r e e - d i m e n s i o n a l motion of e a c h vertebra w a s m e a s u r e d using an  optoelectronic camera system.  I n t e r s e g m e n t a l R O M at t h e f u s i o n level w a s n o r m a l i z e d  c o m p a r e d u n d e r t h e s a m e a p p l i e d m o m e n t in t h e f l e x i b i l i t y p r o t o c o l .  and  Vertebral body strains and  i n t e r s e g m e n t a l R O M at t h e a d j a c e n t s e g m e n t s w e r e c o m p a r e d u n d e r t h e s a m e overall T 9 - L 3 ROM  in t h e h y b r i d p r o t o c o l .  U s i n g t h e flexibility protocol, c e m e n t a u g m e n t a t i o n  better fixation at the f u s i o n level t h a n w i t h posterior rod e x t e n s i o n .  resulted  in  Using the hybrid protocol,  p o s t e r i o r r o d e x t e n s i o n r e d u c e d R O M a t t h e a d j a c e n t l e v e l b u t r e s u l t e d in i n c r e a s e d R O M  and  iii  strain  at t h e  increase Flexible  remaining non-instrumented  in R O M a n d s t r a i n a t t h e s e rod extension  extension  might  levels.  T h e use  of flexible  remaining non-instrumented  provide a  findings  better alternative t h a n with  in t h i s t h e s i s s u p p o r t t h e  use  of c e m e n t  rigid  in t h e  i m p r o v e f i x a t i o n s t r e n g t h a t t h e p e d i c l e s c r e w - v e r t e b r a i n t e r f a c e , 2) interbody device-vertebra  interface  and  3)  increase  resulted  in  l e v e l s t h a n with rigid rod extension,  r o d s a r e n e c e s s a r y t o p r e v e n t a d j a c e n t l e v e l e f f e c t at t h e e x t e n s i o n  The  the  rods  lesser rods. when  level.  osteoporotic  to  1)  i m p r o v e fixation strength  at  overall structural stiffness  spine  (decrease  m o t i o n ) at t h e f u s i o n l e v e l .  L a r g e r i n t e r b o d y d e v i c e s s h o u l d b e u s e d in t h e o s t e o p o r o t i c s p i n e t o  further e n h a n c e  stiffness.  construct  E x t e n s i o n of posterior instrumentation with flexible  m i g h t p r e v e n t a d j a c e n t l e v e l d i s e a s e at t h e e x t e n s i o n it m i g h t  aggravate  d i s e a s e s to t h e s e  the  remaining  non-instrumented  rods  l e v e l , b u t it s h o u l d b e u s e d w i t h c a u t i o n levels  and  postpone  the  adjacent  as  level  levels.  iv  Table of Contents Abstract  ii  Table of Contents  v  List of Tables  ix  List of Figures  xii  Acknowledgements  xvii  Co-Authorship Statement  xviii  Chapter 1  Introduction  1  1.1  Overview  2  1.2  The Normal Spine  3  1.2.1  Basic Anatomy and Physiology  3  1.2.2  Basic Biomechanics  7  1.3  1.4  The Aging Spine  12  1.3.1  Epidemiology  12  1.3.2  Age Related Degenerative Pathology  13  1.3.3  Osteoporosis in the Aging Spine  16  1.3.4  Spinal Degeneration and Osteoporosis  21  1.3.5  Biomechanics of the Aged Spine  23  1.3.6  Clinical Problems of the Aging Spine  24  Surgical Management of the Aging Spine  26  1.4.1  Conditions Requiring Spinal Stabilization  26  1.4.2  General Surgical Approach  27  1.4.3  Clinical Problems with Surgical Management  28  1.4.4  Current Surgical Strategies  31  1.4.5  New Surgical Strategies  32  1.4.6  Problem with Device Evaluation  32  1.5  Motivation  33  1.6  Objectives  34  1.7  Scope  34  1.8  References  35  Chapter 2  Importance of Vertebral Bone Density in Pedicle Screw Testing  47  2.1  Introduction  48  2.2  Materials and Methods  49  2.2.1  Specimen Preparation  49  2.2.2  Pedicle Screw Loading  49  v  2.3  2.2.3  Kinematics  51  2.2.4  Statistical A n a l y s i s  53 55  Results 2.3.1  M o t i o n o f Pedicle S c r e w  55  2.3.2  M o t i o n o f Pedicle S c r e w in B o n e ; Effects o f B M D  56  2.3.3  C e n t e r o f Rotation  57  2.4  Discussion  58  2.5  Conclusion  60  2.6  References  60  Chapter 3  A u g m e n t a t i o n o f Pedicle Screws in O s t e o p o r o t i c Spine  62  3T  Introduction  63"  3.2  Methods  64  3.2.1  Specimens  64  3.2.2  Augmentation Methods  64  3.2.3  Mechanical Testing  66  3.2.4  Data Analyses  68  3.2.5  Statistical A n a l y s i s  68  3.3  70  Results 3.3.1  Paired C o m p a r i s o n b e t w e e n A u g m e n t a t i o n M e t h o d s  72  3.3.2  Correlation with B M D  72  3.3.3  Motion Patterns  73  3.4  Discussion  75  3.5  Conclusion  78  3.6  References  78  Chapter 4  Interbody Device Shape and Size o n Bone-Implant Interface S t r e n g t h  81  4l  Introduction  82  4.2  Methods  83  4.2.1  Specimens  83  4.2.2  Test Groups  83  4.2.3  T e s t Protocol  85  4.2.4  Peripheral Quantitative Computer T o m o g r a p h y  87  4.2.5  Statistical A n a l y s i s  88  4.3  Results  88  4.4  Discussion  93  4.5  Conclusion  95  4.6  References  Chapter 5  Cement A u g m e n t a t i o n of Vertebral Screws Enhances the Resistance t o Subsidence of Interbody Devices  95 97  5.1  Introduction  98  5.2  Methods  99  5.2.1  Experimental Protocol  99  5.2.2  Specimens  99  5.2.3  Test Groups  100  5.2.4  Compression Failure Tests  100  5.2.5  Test Procedure  101  5.2.6  Statistical Analysis  104  5.3  Results  105  5.4  Discussion  107  5.5  Conclusion  112  5.6  References  112  Chapter 6  Enhancing Fixation of the Osteoporotic Spine in a C o r p e c t o m y Model: The Effect of Cement A u g m e n t a t i o n and Posterior Instrumentation Extension  116  6l  Introduction  117"  6.2  Methods  119  6.2.1  Specimens  119  6.2.2  Anterior and Posterior Instrumentation Configurations  121  6.2.3  Test Protocol  124  6.2.4  Statistical Analysis  126  6.3  6.4  Results  127  6.3.1  Flexibility across Destabilized Level  127  6.3.2  Flexibility across Rod Extension Level  131  6.3.3  Flexibility across Non-Instrumented Levels  135  Discussion  142  6.4.1  Limitations  142  6.4.2  Destabilized Level  143  6.4.3  Rod Extension Level  143  6.4.4  Other Levels  146  6.4.5  Spinal Fusion and Adjacent Level Effect  146  6.4.6  Testing Methodology  147  6.4.7  Cement Augmentation versus Posterior Rod Extension  147  6.4.8  Advantages and Disadvantages  150  6.5  Conclusion  151  6.6  References  151  Chapter 7  Adjacent Level Effects: Consequences of Extension of Posterior Instrumentation and Cement A u g m e n t a t i o n of Pedicle Screws  156  7.1  Introduction  157  7.2  Methods  164  7.2.1  Hybrid Flexibility-Stiffness Analyses  164  7.2.2  Vertebral Body Strains  166  7.2.3  Statistical Analysis  167  7.3  7.4  Results  168  7.3.1  Overall Flexibility  168  7.3.2  Baseline Range of Motion and Minimum Moment  168  7.3.3  Intersegmental Range of Motion  173  7.3.4  Vertebral Body Strains  182  Discussion  187  7.4.1  Limitations  190  7.4.2  Comparison with Existing Results  191  7.4.3  The Use of Cement Augmentation and Posterior Rod Extension  192  7.4.4  Vertebra Strains  196  7.5  Conclusion  197  7.6  References  197  Chapter 8  General D i s c u s s i o n , Conclusions and Recommendations  201  8.1  General Discussion  202  8.2  Conclusions  206  8.3  Contributions  207  8.4  Recommendations  208  8.5  References  209  Appendix A  2-way ANOVA c o m p a r i s o n on intersegmental ROM at each level,  210  between cement and rod extension using the hybrid test protocol Appendix B  2-way ANOVA c o m p a r i s o n o n vertebra strain at L1 and L2, in FE and LB, between cement and rod extension using the hybrid test protocol  221  List of Tables  Table 2.1  Range and offset of translation and rotation in the sagittal plane.  55  Table 2.2  Correlation between range and offset of translation and rotation against BMD.  56  Table 3.1  P-values analysed using the non-parametric Wilcoxon signed-rank tests between two augmentation methods within the same specimen.  72  Table 3.2  Pearson correlation coefficient, r, of translation offset and range against BMD.  72  Table 3.3  Frequency and percentage of the number of screws that underwent pure rotation, translation, rotation-translation and double rotation.  74  Table 4.1  Average cross-sectional areas (mm ) of each solid indentor.  84  Table 4.2  The 48 vertebral bodies distributed into 6 groups.  84  Table 4.3  Distribution of indentor size used in each test group.  85  Table 4.4  Permanent deformation after 20% applied strain, determined by comparing the vertebral height before and after indentation tests at the mid sagittal plane from pre and post test pQCT scan data.  92  Table 5.1  Distribution of spinal levels and mean bone mineral density (BMD) in each group.  100  Table 5.2  Distribution of indentor size used in each test group percentage area of endplate coverage achieved  102  Table 5.3  Effects of pedicle screws insertion on failure load and failure strength.  109  Table 6.1  Range of motion and neutral zone across destabilized levels for different surgical conditions in the respective test directions.  127  Table 6.2  2-way ANOVA results on normalized ROM and NZ between cement (factor 1) and rod extension (factor 2) at destabilized levels.  128  Table 6.3  Range of motion and neutral zone across extension level for different surgical conditions in the respective test directions.  132  Table 6.4  2-way ANOVA results on normalized ROM and NZ between cement (factor 1) and rod extension (factor 2) at the rod extension level (below destabilized level).  132  Table 6.5  Range of motion and neutral zone across level above destabilized level (T9-T10) for different surgical conditions in the respective test directions.  136  Table 6.6  Range of motion and neutral zone at 2 levels below destabilized level ( L 1 L2) for different surgical conditions in the respective test directions.  136  2  and  mean  IX  Table 6.7  Range of motion and neutral zone at 3 levels below destabilized level (L2L3) for different surgical conditions in the respective test directions.  137  Table 7.1  Mean overall range of motion (across T9-L3 or T8-L2) for different surgical conditions in the respective test directions at 5 Nm applied moment.  169  Table 7.2  Baseline R O M + a n d ROM-for the 12 specimens  169  Table 7.3  Applied moments that created the baseline ROMs in specimen 1  170  Table 7.4  2-way ANOVA results on normalized moment between cement (factor 1) and posterior extension (factor 2) under the same baseline ROM.  170  Table 7.5  Intersegmental ROM+, intersegmental ROM- and overall intersegmental ROM in axial rotation for specimen 1.  173  Table 7.6  Mean ROM across rod extension level for different surgical conditions using the hybrid protocol  177  Table 7.7  Percentage difference in ROM from intact and statistical significance, for rod extension level.  177  Table 7.8  2-way ANOVA results on normalized ROM between cement (factor 1) and posterior rod extension (factor 2) using the hybrid protocol for rod extension level.  177  Table 7.9  Mean ROM at 2 levels below destabilized level (L1-L2) for different surgical conditions using the hybrid protocol.  178  Table 7.10  Percentage difference in ROM from intact and statistical significance, for 2 levels below destabilized level.  178  Table 7.11  2-way ANOVA results on normalized ROM between cement (factor 1) and posterior rod extension (factor 2) using the hybrid protocol, for 2 levels below destabilized level.  178  Table 7.12  Mean ROM at 3 levels below destabilized level (L2-L3) for different surgical conditions using the hybrid protocol.  179  Table 7.13  Percentage difference in ROM from intact and statistical significance, for 3 levels below destabilized level.  179  Table 7.14  2-way ANOVA results on normalized ROM between cement (factor 1) and posterior rod extension (factor 2) using the hybrid protocol, for 3 levels below destabilized level.  179  Table 7.15  Mean ROM across level above destabilized level (T9-T10) for different surgical conditions using the hybrid protocol.  180  Table 7.16  Percentage difference in ROM from intact and statistical significance, for level above destabilized level.  180  Table 7.17  2-way ANOVA results on normalized ROM between cement (factor 1) and posterior rod extension (factor 2) using the hybrid protocol, for level above destabilized level.  180  Table 7.18  Mean ROM across destabilized levels for different surgical conditions using the hybrid protocol.  181  Table 7.19  Percentage difference in ROM from intact and statistical significance, for destabilized level.  181  Table 7.20  2-way ANOVA results on normalized ROM between cement (factor 1) and posterior rod extension (factor 2) using the hybrid protocol, for destabilized level.  181  Table 7.21  Mean (standard deviation) strain gauge values (in microstrain) at baseline ROM in flexion and in extension.  182  Table 7.22  Percentage difference in strain from intact during flexion-extension and statistical significance.  183  Table 7.23  2-way ANOVA results on normalized vertebral strain between cement (factor 1) and posterior rod extension (factor 2) using the hybrid protocol.  184  Table 7.24  Mean (standard deviation) strain gauge values (in microstrain) at baseline ROM in right lateral bending and in left lateral bending.  185  Table 7.25  Percentage difference in strain from intact during lateral bending and statistical significance.  185  Table 7.26  2-way ANOVA results on normalized vertebral strain between cement (factor 1) and posterior rod extension (factor 2) using the hybrid protocol.  186  Table 7.27  Score for different surgical configuration calculated based on comparison to corresponding intact values.  195  xi  List of Figures  Figure 1.1  (Top) Schematic diagram of a vertebra viewed from above, showing the vertebral body and the neural arch.  3  Figure 1.2  Schematic diagram of a vertebral body with an intervertebral disc.  4  Figure 1.3  Schematic diagram of a functional spinal unit (FSU) under compressive load and during bending.  5  Figure 1.4  Schematic diagram showing the ligaments, muscles, spinal canal and intervertebral foramen in the lumbar spine.  7  Figure 1.5  Engineering definition of stability as illustrated by a cone on a flat surface and by a ball on curved or flat surfaces.  8  Figure 1.6  Analogy of the stabilization of the spinal column to an inverted pendulum.  10  Figure 1.7  Free body diagrams of a functional spinal unit (FSU) during extension and during flexion.  11  Figure 1.8  Population distribution of Canada in 1971 and 1996, from Statistics Canada.  12  Figure 1.9  Degenerative changes in the aging spine could result in a smaller intervertebral foramen space, loss of disc height, loss of vertebral height, and a kyphotic posture.  13  Figure 1.10  Typical variation of bone mineral density (BMD) with age in a population (in bold, mean BMD).  16  Figure 1.11  Vertebral compression fracture commonly present with a) concaved endplates, b) decreased vertebral height, c) wedge shaped vertebral body, or in combination (d-g).  18  Figure 1.12  Progression of spinal degeneration in the elderly spine from weak bone or from degenerated disc.  22  Figure 1.13  Radiographs of the spine following corpectomy and insertion of a titanium vertebral body replacement device and pedicle screw and rigid rod fixation.  28  Figure 2.1  (a) Lateral and (b) frontal view of the testing jig.  50  Figure 2.2  A two-dimensional finite element analysis determined the variation in ground reaction moments as a factor of elastic modulus of the lower block.  51  Figure 2.3  Two marker carriers, each with four infrared LEDs, were mounted onto the two rigid bodies, (a) pedicle screw and vertebral specimen, and (b) pedicle screw and UHMWPE block.  52  Figure 2.4  The (a) relative translations and (b) rotations between pedicle screw and bone in all six degrees of freedom for the first twenty seconds are presented in these two graphs.  54  Figure 2.5  The range and offset of the motion of the screw heads for screws inserted in a) UHMWPE, b) bone with normal insertion and c) bone with overd rilling.  56  Figure 2.6  Scatterplots of screw head translation and rotation ranges versus BMD for pedicle screws inserted with normal hole preparation.  57  Figure 2.7  Typical motion of pedicle screws inserted in the normal fashion at the screw head and screw tip during the last 5 cycles for specimens (a) with low BMD and (b) with high BMD.  58  Figure 3.1  Augmentation with laminar hook (left) and with sublaminar wires (right).  65  Figure 3.2  (a) The resultant forces on the inferior screw consisted of a compression and a flexion bending moment.  66  Figure 3.3  The motion of the screw head in the vertical direction during the last 10 cycles is presented in this figure.  67  Figure 3.4  Schematic diagram of the screw positions at three or four instances of time during one loading cycle, starting with the screw head at the superior position (position 1).  69  Figure 3.5  Mean and standard deviation of translation offsets at the screw head and screw tip for the three groups of specimens.  70  Figure 3.6  Mean and standard deviation of translation ranges at the screw head and screw tip for the three groups of specimens.  71  Figure 3.7  For the 16 screws augmented with calcium phosphate cement, the translation range at the screw tip was linearly correlated with BMD (r = 0.58, p = 0.017).  73  Figure 3.8  (a) The motion at the screw tip and screw head during the last five cycles is presented.  74  Figure 4.1  The kidney, elliptical and clover-leaf shaped indentors.  84  Figure 4.2  The test jig was designed with a pair of matching spherical concave and convex surfaces to allow for three dimensional rotations of the indentors about a centre of rotation located at the geometric centre of the indentor face (marked X).  86  Figure 4.3  Typical load-displacement curves where failure load was the maximum load before the gradient of the curve changes from positive to negative while stiffness was the slope of the linear portion of the curve before failure occurred.  87  Figure 4.4  Mean failure load for the kidney, elliptical and clover-leaf indentor groups, with 20 % and 40 % endplate coverage.  89  Figure 4.5  Mean failure strength for the kidney, elliptical and clover-leaf indentor  89 xiii  groups, with 20% and 40% endplate coverage. Figure 4.6  Mean stiffness for the kidney, elliptical and clover-leaf indentor groups, with 20% and 40% endplate coverage.  90  Figure 4.7  Images of a specimen indented with an elliptical indentor through the mid sagittal plane a) pre and b) post indentation tests using peripheral quantitative computer tomography (pQCT).  91  Figure 4.8  Density profile across the vertebral body from superior endplate (left) to inferior endplate (right) at the mid-sagittal plane.  91  Figure 5.1  Four sizes of clover-leaf and elliptical shaped indentors were used for the tests.  101  Figure 5.2  Custom-built clamps were used to secure the specimen and an adjustable vise was used to align the superior endplate in the horizontal plane.  103  Figure 5.3  Typical load-displacement curves where failure load was the maximum load before the gradient of the curve changed from positive to negative, while stiffness was the maximum slope of the curve before failure occurred.  104  Figure 5.4  Mean failure loads for elliptical and clover-leaf shaped indentors with i) pedicle screws only, ii) pedicle screws with cement, and iii) anterior vertebral body screws with cement.  105  Figure 5.5  Mean failure strengths for elliptical and clover-leaf shaped indentors with i) pedicle screws only, ii) pedicle screws with cement, and iii) anterior vertebral body screws with cement.  106  Figure 5.6  Mean stiffnesses for elliptical and clover-leaf shaped indentors with i) pedicle screws only, ii) pedicle screws with cement, and iii) anterior vertebral body screws with cement.  107  Figure 5.7  Pedicle screw insertion resulted in disruption of trabecular bone at the posterolateral region of the vertebral body and along the screw.  110  Figure 5.8  (A) Underlying trabeculae supports the applied load from the interbody device.  111  Figure 6.1  Schematic diagram of rigid instrumentation, with interbody device and pedicle screw fixation, following corpectomy.  118  Figure 6.2  Distribution of areal bone mineral density (aBMD) and volumetric bone mineral density (vBMD) among the 12 specimens in this study.  120  Figure 6.3  Schematic diagram of different surgical conditions tested in a repeated measures fashion in a T9-L3 spine segment.  121  Figure 6.4  Measured height of the expandable vertebral body replacement device excluded the spiked protuberances at both ends.  122  Figure 6.5  Extending rigid and flexible posterior rods from T12 to L1 of T9-L3  123 xiv  segments (T11-T12 in T8-L2 segments) were carried out using a "parallel rods connector" and a "transverse connector" on each side. Figure 6.6  The custom spine testing machine used to apply pure moments to the specimen.  125  Figure 6.7  Flexibility curve of the T8-L2 spine in flexion-extension (FE) of a typical specimen (3 cycles shown), following cement augmentation of all pedicle screws and with rigid posterior rod extension.  126  Figure 6.8  Mean normalized range of motion (ROM) of the destabilized level in AR.  129  Figure 6.9  Mean normalized range of motion (ROM) of the destabilized level in LB.  129  Figure 6.10  Mean normalized range of motion (ROM) of the destabilized level in FE.  130  Figure 6.11  Mean normalized range of motion (ROM) of the destabilized level in FE+FL.  131  Figure 6.12  Mean normalized range of motion (ROM) at the extension level in AR.  133  Figure 6.13  Mean normalized range of motion (ROM) at the extension level in LB.  134  Figure 6.14  Mean normalized range of motion (ROM) at the extension level in FE.  134  Figure 6.15  Mean normalized range of motion (ROM) at the extension level in FE+FL.  135  Figure 6.16  Mean normalized range of motion (ROM) and neutral zone (NZ) of each mobile level and instrumented segment in AR.  138  Figure 6.17  Mean normalized range of motion (ROM) and neutral zone (NZ) of each mobile level and instrumented segment in LB.  139  Figure 6.18  Mean normalized range of motion (ROM) and neutral zone (NZ) of each mobile level and instrumented segment in FE.  140  Figure 6.19  Mean normalized range of motion (ROM) and neutral zone (NZ) of each mobile level and instrumented segment in FE+FL.  141  Figure 6.20  Summary of changes in ROM across levels and directions, for the noncemented conditions.  144  Figure 6.21  Summary of changes in ROM across levels and directions, for the cemented conditions.  145  Figure 6.22  Schematic diagrams augmentation.  cement  148  Figure 6.26  (A) Schematic diagrams of rigid instrumentation following corpectomy, with cement augmentation of pedicle screws.  149  Figure 7.1  When pure moments were applied at the ends of a beam, the moment experienced in any section throughout the beam was the same even when a section was reinforced.  161  Figure 7.2  Flexibility curves at the third cycle for all surgical conditions were  165  of  pedicle  screws  with  and  without  xv  compared to determine the baseline (smallest) positive and negative range of motion (+ROM and - R O M ) and corresponding applied moment. 7  Figure 7.3  Two strain gauges were attached to the lateral aspects of the body and one strain gauge attached onto the left surface of the spinous process, on 2 distal vertebral bodies to the corpectomy level (T12 and L1 in T8-L2).  166  Figure 7.4  Strain data were collected for the full duration of each test.  167  Figure 7.5  Typical flexibility curves of the T8-L2 spine of a specimen under AR motion, in the intact and various surgical intervention states.  168  Figure 7.6  Mean normalized applied moment in AR at baseline ROM.  171  Figure 7.7  Mean normalized applied moment in LB at baseline ROM.  171  Figure 7.8  Mean normalized applied moment in FE at baseline ROM.  172  Figure 7.9  Mean normalized applied moment in FE with follower-load at baseline ROM.  172  Figure 7.10  Distribution of overall intersegmental range of motion (ROM) in axial rotation for specimen 1 (corresponds to data in Table 7.5).  174  Figure 7.11  Normalized intersegmental range of motion (ROM) in AR.  175  Figure 7.12  Normalized intersegmental range of motion (ROM) in LB.  175  Figure 7.13  Normalized intersegmental range of motion (ROM) in FE.  176  Figure 7.14  Normalized intersegmental range of motion (ROM) in FE+FL.  176  Figure 7.15  Mean changes in strain values between flexion and extension at respective baseline ROM.  183  Figure 7.16  Normalized changes in strain values during FE at respective baseline ROM.  184  Figure 7.17  Mean changes in strain values during LB at respective baseline ROM.  186  Figure 7.18  Normalized changes in strain values during LB at respective baseline ROM.  187  Figure 7.19  Summary of changes in ROM and strains across levels and directions, for the non-cemented conditions.  188  Figure 7.20  Summary of changes in ROM and strains across levels and directions, for the cemented conditions.  189  Figure 7.21  Normalized ROM at all levels plotted against normalized ROM at the destabilized level.  193-194  xvi  Acknowledgements I w o u l d like to thank my wife Vivian for her sacrifice, support a n d understanding, a n d my children R y a n a n d Celestine for their patience. I w o u l d also like to t h a n k Dr T o m O x l a n d for the opportunity to be his P h D student, for his d e v o t e d supervision, a n d his guidance a n d support. This research is only m a d e possible by the 57 a n o n y m o u s m e n a n d w o m e n w h o selflessly d o n a t e d their bodies to science a n d participated in this study. I k o w t o w to t h e m . I w o u l d also like to thank the spine surgeons, Drs Marcel Dvorak, Charles Fisher, Brian K w o n , Chris Bailey, S a n d e e p Singh and Robert L a b r o m , f o r s p e n d i n g their valuable time in discussion a n d carrying out surgical procedures. I also thank m y research c o m m i t t e e , Dr T o m O x l a n d , Dr Marcel Dvorak, Dr H e a t h e r M c K a y a n d Dr M o h a m e d G a d a l a , for their guidance. I a m also grateful for the special assistance given by staff a n d fellow students of the Division of O r t h o p a e d i c Engineering R e s e a r c h , in particular, S i m o n Sjovold, Dr J e s s e C h e n , Dr Q i n g - A n Z h u , C h a d Larson a n d Dr Hanspeter Frei. T h e r e are also m a n y others w h o have helped in s o m e w a y this past 5 years including Dinesh S a m a r a s e k e r a , P a m e l a Grant, A n t h o n y Choo, Dr V l a d S t a n e s c u , T h o m a s Nydegger, Darrell G o e r t z e n , Dr Peter Cripton, Leslie Bryant M a c L e a n , Christina Niosi a n d m a n y others. I w o u l d also like to a c k n o w l e d g e funding f r o m the C a n a d i a n Institutes of Health R e s e a r c h a n d f r o m the A g e n c y for Science, T e c h n o l o g y a n d R e s e a r c h , S i n g a p o r e . In addition, I thank S y n t h e s Spine, Medtronic S o f a m o r - D a n e k , D e P u y Spine, Zimmer, Biomet Merck, a n d Stryker for providing the implants, surgical tools a n d bone cement.  xvii  Co-Authorship Statement Sections of this thesis have been submitted as multi-authored papers in refereed journals. The research was performed by me except for surgical procedures which were carried out by a spine surgeon in each study. Detailed design of all studies was presented to the co-authors for approval prior to the experiments being carried out. Data analyses and manuscript preparation were carried out by me and the co-authors contributed in editing the manuscripts for submission to refereed journals. The original ideas in a Canadian Institutes of Health Research (CIHR) grant, titled "Enhancement of Osteoporotic Spine Fixation", submitted jointly by the Principal Investigators, Drs Oxland, Dvorak and Fisher in September 2000, provided the research topic arid the grant provided funding. .Juay Seng Tan Sections of this thesis have been submitted as multi-authored papers in refereed journals. Details of the authors' contribution are provided. I agree with the stated contributions of the thesis author, as indicated below. Dr. Thomas R Oxland (thesis supervisor) Published Papers 1. Tan JS, Kwon BK, Samarasekera D, Dvorak MF, Fisher CG, Oxland TR. Vertebral bone density - a critical element in the performance of spinal implants. In: Spinal Implants: Are We Evaluating Them Appropriately? American Standards for Testing and Materials (ASTM) STP 1431, Melkerson MN, Griffith SL and Kirkpatrick JS, (Eds.), ASTM International, West Conchohocken, PA, 2003:217-30. Authors' contribution: Juay Seng Tan was jointly responsible for the original ideas behind the paper, conducting the experiments, data analyses, presentation of the findings, and writing and editing of the original paper. Dr Brian Kwon conducted the surgical procedures. Mr Dinesh Samarasekera conducted a preliminary pilot study. Dr Marcel Dvorak and Dr Charles Fisher stimulated discussion. Dr Thomas Oxland was jointly responsible for the original ideas, provided supervision, and was the key editor on this paper. 2. Tan JS, Kwon BK, Dvorak MF, Fisher CG, Oxland TR. Pedicle screw motion in the osteoporotic spine after augmentation with laminar hooks, sublaminar wires, or calcium phosphate cement: a comparative analysis. Spine 2004; 29:1723-1730. Authors' contribution: Juay Seng Tan was responsible for the final proposal (based on the original ideas), conducting the experiments, data analyses, presentation of the findings, and writing and editing of the original paper. Dr Brian Kwon conducted the surgical procedures, stimulated discussion and provided editorial assistance. Dr Marcel Dvorak and Dr Charles Fisher were jointly responsible for the original ideas behind this paper, stimulated discussion and provided editorial assistance. Dr Thomas Oxland was jointly responsible for the original ideas behind this paper, provided supervision, and was the key editor on this paper.  xviii  3. Tan JS, Bailey CS, Dvorak MF, Fisher CG, Oxland TR. Interbody device shape and size are important to strengthen the vertebra-implant interface. Spine 2005; 30:638-644. Authors' contribution: Juay Seng Tan was responsible for the final proposal (based on the original ideas), designing the indentors, preparing the test setup, conducting the experiments, data analyses, presentation of the findings, and writing and editing of the original paper. Dr Christopher Bailey conducted the surgical procedures, stimulated discussion and provided editorial assistance. Dr Marcel Dvorak and Dr Charles Fisher were jointly responsible for the original ideas behind this paper, stimulated discussion and provided editorial assistance. Dr Thomas Oxland was jointly responsible for the original ideas behind this paper, provided supervision, and was the key editor on this paper. Papers Submitted 4. Tan JS, Bailey CS, Dvorak MF, Fisher CG, Cripton PA, Oxland TR. Cement augmentation of vertebral screws enhances the interface strength between interbody device and vertebral body. Submitted to Spine (7 Nov 2005). Authors' contribution: Juay Seng Tan was responsible for the final proposal (based on the original ideas), designing the indentors, preparing the test setup, conducting the experiments, data analyses, presentation of the findings, and writing and editing of the original paper. Dr Christopher Bailey conducted the surgical procedures, stimulated discussion and provided editorial assistance. Dr Marcel Dvorak and Dr Charles Fisher were jointly responsible for the original ideas behind this paper, stimulated discussion and provided editorial assistance. Dr Peter Cripton also stimulated discussion and provided editorial assistance. Dr Thomas Oxland was jointly responsible for the original ideas behind this paper, provided supervision, and was the key editor on this paper. Papers in Preparation 5. Tan JS, Singh S, Zhu QA, Dvorak MF, Fisher CG, Oxland TR. Spinal stabilization in the osteoporotic spine following posterior rod extension and cement augmentation of screws using a flexibility protocol. In preparation for submission to Spine. 6. Tan JS, Singh S, Zhu QA, Dvorak MF, Fisher CG, Oxland TR. Adjacent level effects in the osteoporotic spine following posterior rod extension and cement augmentation of screws using a hybrid protocol. In preparation for submission to Spine. Authors' contribution: Juay Seng Tan was responsible for the final proposal (based on the original ideas), overhauling the spine machine, preparing the test setup, conducting the experiments, data analyses, presentation of the findings, and writing and editing of the original paper. Dr Sandeep Singh conducted the surgical procedures, stimulated discussion and provided editorial assistance. Dr Qing-An Zhu assisted in data collection and in conducting the experiments. Dr Marcel Dvorak and Dr Charles Fisher were jointly responsible for the original ideas behind these 2 papers, stimulated discussion and provided editorial assistance. Dr Thomas Oxland was jointly responsible for the original ideas behind these 2 papers, provided supervision, and was the key editor on these papers.  xix  Chapter 1  Introduction  Chapter 1 Introduction  Overview The  populations  of t h e d e v e l o p e d  nations a r e aging.  Degenerative  conditions  such as  d e g e n e r a t i v e spondylosis, spondylolisthesis, spinal stenosis a n d d e g e n e r a t i v e spinal l u m b a r k y p h o s i s a n d scoliosis, a n d other disabling conditions s u c h a s t r a u m a o r t h e p r e s e n c e of t u m o u r s , a r e conditions in older individuals a n d they often require stabilization with spinal implants. O s t e o p o r o s i s is a condition of r e d u c e d b o n e density that could result in loss of b o n e strength a n d c a u s e pain, deformity a n d loss of function f o r t h e patient.  While  vertebral  osteoporotic fractures d o not often necessitate surgery, t h e p r e s e n c e of diminished b o n e quality c o m p l i c a t e s spinal stabilization surgery w h e n spinal implants a r e u s e d . T h e w e a k e n e d bone in the osteoporotic spine results in poor bone-implant interface properties a n d brings  about  surgical challenges s u c h a s loosening of pedicle s c r e w s a n d s u b s i d e n c e of interbody d e v i c e s . Currently available spinal instrumentation d e v i c e s that w o u l d o t h e r w i s e b e beneficial often l o o s e n w h e n u s e d in t h e aging spine.  M o r e o v e r , t h e increased stiffness within t h e f u s e d  s e g m e n t s could also lead t o early failure of adjacent levels in the elderly patients d u e to altered b i o m e c h a n i c s s u c h a s increased load a n d motion d e m a n d s , a n d m a y necessitate surgery.  further  T h e objective of this thesis w a s t o provide insights into both t h e p r o b l e m s of implant  loosening a n d adjacent level effects in the aging spine.  1  Chapter 1  Introduction  1.  Introduction  1.1  Overview The current generation of spinal implants are often not specifically designed for use in  the elderly patient. As the quality of the bone in the elderly spine is commonly compromised with degenerative conditions such as spondylosis and osteoporosis, spinal implant loosening is a common cause of failure in these patients. To compound the problem, elderly patients are prone to adjacent level disease, a spinal condition believed to be caused by altered biomechanics at the levels adjacent to the implanted levels. These two problems complicate spinal instrumentation in the osteoporotic patient. Many elderly patients who would otherwise benefit from currently available spinal instrumentation devices are denied the treatment due to their compromised bone quality. The absolute number and proportion of elderly people in the population are projected to increase in the next few decades. With the projected increase in the number of elderly patients with osteoporosis requiring spinal surgery, it is in the interest of spinal implant companies to develop devices that would cater to this rising demand in the market. There are however no initiative to develop spinal devices specifically for this patient group, according to the author's knowledge. In recent years, motion preservation devices have been conceived as the next generation of spinal implants and many implant companies are jostling for market share. Motion preservation devices could become a solution for preventing adjacent level diseases in the younger patients. Also in recent years, bone cement has become more popular to improve the implant-bone interface.  Bone cement holds much promise to address the issue of device  loosening in patients with compromised bone. The use of motion preservation devices with bone cement could thus be the future solution for preventing both device loosening and adjacent level diseases. However, further research would be necessary before they could be indicated for elderly patients with compromised bone quality. Further understanding of the biomechanics of spinal device loosening and adjacent level diseases are necessary in order to solve both issues together. A current surgical solution that could address both issues is needed for osteoporotic elderly patients. It is the goal of this thesis to provide further biomechanical input into the problems of spinal device loosening and adjacent level disease and to propose feasible solutions to these issues so as to improve quality of life for the elderly patients.  2  Chapter  1  Introduction  1.2  The Normal Spine  1.2.1 Basic Anatomy and Physiology  The human spinal column consists of 24 mobile vertebrae: 7 cervical vertebrae (C1-C7), 12 thoracic vertebrae (T1-T12), and 5 lumbar vertebral (L1-L5), and the relatively immobile sacrum (S1-S5) and coccyx.  Except for C1 and C2, the other mobile vertebrae consist  anteriorly of a vertebral body, and posteriorly of a neural arch formed by the pedicles and the laminae (Figure 1.1). Where the laminae meet at the posterior midline, a spinous process emerges. At the junction between the pedicle and the lamina, the transverse process projects laterally.  The superior and inferior articular processes connect to corresponding articular  processes of adjacent vertebral bodies, forming the facet joints. Neural Arch  / \ \ ,pedicle superior articular process lamina  spinous process  vertebral body  Vertebra <  transverse process  lamina  endplate  spinous process  Intervertebral Disc 1  ^X^S  JJJJJ^^\^f  F a c e t  J o i n t  annulus fibrosus nucleus pulposus superior articular process inferior articular process Figure 1.1  (Top) Schematic diagram of a vertebra viewed from above, showing the vertebral body and the neural arch. (Below) Schematic diagram of a functional spinal unit (FSU) viewed from the side, showing a pair of vertebrae, the intervertebral disc (cross-section), and a facet joint (the other facet joint is blocked from view). 3  Chapter 1  Introduction  The spinal column can be considered to be made up of a series of functional spinal units (FSUs). Each FSU consists of two adjoining vertebrae and its connecting soft tissues (Figure 1.1). An intervertebral disc and a pair of facet joints transmit load and allow motion of the FSU. The vertebral body is mainly made up of trabecular bone (also known as cancellous bone), with a thin shell of cortical bone on its outer surface. The posterior elements, consisting of the pedicles, transverse processes, laminae, spinous process, superior articular processes, and inferior articular processes, are mainly made up of cortical bone, with some cancellous bone within.  At the interfaces between the vertebral body and the superior and inferior  intervertebral discs are the vertebral endplates.  The endplate in the adult consists of an  osseous layer integrated with the vertebral body, and a thin cartilaginous layer in contact with the nucleus pulposus and annulus fibrosus. The endplate is not vascularized and it prevents blood flow into the intervertebral disc. The intervertebral disc consists of a gelatinous nucleus pulposus surrounded by concentric layers of annulus fibrosus (Figures 1.1 and 1.2). The gelatinous nucleus pulposus behaves like an incompressible fluid material.  The fibres in the annulus are oriented  approximately between 20° to 30° with the horizontal plane, with alternating directions between concentric layers, making it strong in resisting the hydrostatic pressure of the nucleus pulposus.  cortical shel  nucleus Figure  1.2  Schematic vertebral  diagram of a vertebral body is made up mainly  body with of trabecular  an intervertebral bone, with  disc. a thin  The cortical 4  Chapter 1  Introduction  shell. The intervertebral disc consists of a gelatinous nucleus pulposus, surrounded by concentric layers of annulus fibrous. The fibres in the annulus are oriented between 20° to 30° with the horizontal plane, with alternating directions between layers. The  p h y s i o l o g i c a l p u r p o s e o f t h e i n t e r v e r t e b r a l d i s c is t o t r a n s m i t l o a d f r o m t h e  vertebral  body  to  the  inferior v e r t e b r a l  body,  while  allowing  motion  for  the  superior  FSU.  Under  c o m p r e s s i o n , a u n i f o r m h y d r o s t a t i c p r e s s u r e is c r e a t e d in t h e n u c l e u s o f t h e h e a l t h y d i s c , w h i c h in  t u r n e x e r t s a r a d i a l s t r e s s o n t o t h e a n n u l u s f i b r o s u s ( F i g u r e 1.3).  annulus  is t h u s  inferior  to  hydrostatic bending,  the  to  enclose  intervertebral  pressure  the  and  in t h e  annulus  o n the  contain  disc,  the  nucleus concave  r e g i o n o f t h e a n n u l u s is u n d e r t e n s i o n .  the  nucleus  endplates pulposus side  is  during load  and  during placed  and  underlying motion.  O n e important role of motion.  trabecular  During  Superior  bone  sideward  under compression,  while  resist or  the  and the  forward opposite  L o a d a c t i n g f r o m t h e v e r t e b r a l b o d y in b e n d i n g is  t r a n s m i t t e d t h r o u g h t h e p r e s s u r i z e d n u c l e u s to t h e n e x t l e v e l , a n d t h e r e s u l t a n t f o r c e s a r e to p u s h the n u c l e u s p u l p o s u s a w a y f r o m the c o n c a v e s i d e [ B e n z e l 1 9 9 5 ,  the  also noted  T s a n t r i z o s et al. 2 0 0 5 ] .  D u r i n g a x i a l r o t a t i o n m o t i o n , t h e c o n c e n t r i c l a y e r s o f t h e a n n u l u s w i t h its o r i e n t e d f i b r e d i r e c t i o n s withstand the s t r e s s e s created.  L o a d s from the vertebral b o d y acting o n the intervertebral d i s c s  a r e m o r e c o m m o n l y a c o m b i n a t i o n of c o m p r e s s i o n , b e n d i n g , axial t o r s i o n , a n d s h e a r .  Together,  t h e n u c l e u s p u l p o s u s a n d t h e a n n u l u s f i b r o s u s p r o v i d e t h e s t r e n g t h f o r t h e i n t e r v e r t e b r a l d i s c to withstand compressive,  bending a n d torsional loads.  compression  uniform disc  Figure  1.3  pressure  Schematic diagram of a functional spinal unit (FSU) under compressive load and during bending. A uniform hydrostatic pressure is created in the nucleus, which in turn exerts a radial stress onto the surrounding annulus. 5  Chapter  Introduction  1  During bending, the annulus compression while the opposite  on the bending annulus is under  side is placed tension.  under  The inferior articular process of the superior vertebra and the superior articular process of the inferior body are connected by a synovial joint, and this joint is commonly called the facet joint (also the apophyseal, the zygapophyseal, or the posterior intervertebral joint) (Figure  1.1).  The facet joints connect the FSU posteriorly and together with the intervertebral disc transmit loads and allow motion. Under pure compression, the pair of facet joints in the lumbar spine withstands approximately 20% of the load while the intervertebral disc withstands approximately the remaining 80% [Lorenz et al. 1983, Yang and King 1984].  Higher amount of load is  transmitted through the intervertebral disc in forward flexion, while more load is transmitted through the facet joints in extension motion. In axial rotation, one facet joint is generally in compression while the opposing facet is generally in tension and shear. One important role of the facet joints is to prevent excessive motion of the FSU with a bone-on-bone mechanical stopper between the articular processes. The differing orientations of the facet articulartions across the cervical, thoracic and lumbar spine, and the subtle differences within each of these regions, results in different ratio of load sharing with the intervertebral disc and in different range of motion (ROM) for each level. Several ligaments connect vertically across the FSU (Figure 1.4). The role of ligaments is to enhance stability by restricting excessive motion. The anterior longitudinal ligament (ALL) attaches firmly to the anterior surface of the vertebral body and is stretched during extension motion.  The posterior longitudinal ligament (PLL) attaches to the posterior surface of the  vertebral body and lies anterior to the spinal canal. Both the interspinous ligament (ISL) and the supraspinous ligament (SSL) connect between spinous processes and are stretched during flexion motion. The ligamentum flavum (LF) connects between laminae and lies posterior to the spinal canal.  Lastly, the facet capsular ligament (CL) connects across adjacent articular  processes and prevents excessive motion of the facet joint.  Several other ligaments are  present in the thoracic spine which connects across the costovertebral joints between the rib and the vertebra.  6  Chapter 1  Figure  Introduction  1.4  Schematic diagram showing the ligaments, muscles, spinal canal and intervertebral foramen in the lumbar spine. Ligaments restrict excessive motion while muscles are the prime movers. The spinal cord is protected by bony structures of the vertebra. The spinal cord passes through the spinal cana while the spinal roots exit through the intervertebral foramen. Ligaments: anterior longitudinal ligament (ALL), posterior longitudinal ligament (PLL), ligamentum flavum (LF), capsular ligament (CL), interspinous ligament (ISL), supraspinous ligament (SSL). Muscles: psoas major (PM), quadratus lumborum (QL), iliocostalis lumborum (IL), longissimus thoracis (LT), multifidus muscle (M).  M u s c l e s are the the  extensor  group  p r i m e m o v e r s o f j o i n t s in t h e  includes  the  multifidus  i l i o c o s t a l i s l u m b o r u m (IL) ( F i g u r e 1.4). erector spinae.  Deep  intrinsic flexor m u s c l e s  abdominal muscles  Basic The  represented When does  longissimus  thoracis  that m a k e  up  (LT) and  the  include the p s o a s major ( P M ) a n d the  is t h r o u g h t h e  the  quadratus  of the a b d o m i n a l wall i n c l u d e the e x t e r n a l o b l i q u e  internal o b l i q u e (IO) a n d t h e r e c t u s a b d o m i n i s ( R A ) .  the t h o r a c o l u m b a r spine  1.2.2  (M),  T h e major muscles  T h e l a t t e r t w o ( L T a n d IL) t o g e t h e r a r e a l s o k n o w n a s  l u m b o r u m ( Q L ) , while extrinsic flexor m u s c l e s (EO),  muscle  body.  L a t e r a l b e n d i n g a n d axial rotation of  unilateral action of d e e p m u s c l e s  ( E O , IO), with c o n t r a c t i o n of m u s c l e s at t h e o p p o s i t e  (IL, L T , M , Q L ) a n d  side.  Biomechanics definition  of  stability  in  the  engineering  field  is  well  established  and  could  p i c t o r i a l l y b y a c o n e o n a flat s u r f a c e o r a b a l l o n c u r v e d o r flat p l a n e s ( F i g u r e  perturbed by a n external force, the not topple,  but c o u l d b e  displaced  unstable  a n d is t h u s  object w o u l d not s t a b l e .  'topple'.  T h e neutral  be 1.5).  object  T h e stable object would  not  7  Chapter topple  1 over  Introduction and  would  Instability of t h e s p i n e 1989].  return  however  to  its  initial  position  is h a r d t o d e f i n e  I n s t a b i l i t y is a t e r m v a g u e l y  used  a n d a single  to d e s c r i b e  involving e x c e s s i v e or a b n o r m a l m o v e m e n t s  when  some  perturbed  by  an  external  force.  definition d o e s not exist [ C r i s c o biomechanical spinal  of the s p i n e [ A d a m s et al. 2 0 0 2 ] .  dysfunction  Identification  t h e u n s t a b l e s p i n e is i m p o r t a n t c l i n i c a l l y a s it c a n b e a n i n d i c a t i o n f o r s u r g i c a l i n t e r v e n t i o n .  of  One  i n t e r p r e t a t i o n o f c l i n i c a l i n s t a b i l i t y d e f i n e d it a s "the i n a b i l i t y o f t h e s p i n e u n d e r p h y s i o l o g i c  loads  t o limit p a t t e r n s o f d i s p l a c e m e n t  roots  and, and  so  as  not to d a m a g e  o r irritate t h e s p i n a l c o r d o r n e r v e  in a d d i t i o n , t o p r e v e n t i n c a p a c i t a t i n g d e f o r m i t y o r p a i n d u e t o s t r u c t u r a l c h a n g e s "  [White  P a n j a b i 1990].  Unstable  Figure  1.5  The  Neutral  Stable  Engineering definition of stability as illustrated by a cone on a flat surface and by a ball on curved or flat surfaces. All objects are in equilibrium initially. When perturbed by an external force, the unstable object would 'topple', the neutral object and stable objects would not topple but the stable object would return to its initial position. [Adapted from Crisco 1989 and Adams et al. 2002]  three  fundamental  biomechanical  b e t w e e n l e v e l s , 2) t o t r a n s f e r l o a d a c r o s s roots [White a n d P a n j a b i 1990].  functions  of the  spine  are  l e v e l s , a n d 3) t o p r o t e c t t h e  T h e unstable  spine could b e c o m e  1)  to  allow  motion  spinal cord a n d  incompetent  spinal  in p e r f o r m i n g  o n e or m o r e of t h e s e functions. The  cervical, thoracic, lumbar a n d sacral regions  with different d e g r e e s of e m p h a s i s .  of the  spine  perform these  functions  T h e l u m b a r r e g i o n , f o r e x a m p l e , is s u b j e c t e d t o l a r g e r l o a d s  t h a n t h e t h o r a c i c a n d c e r v i c a l s p i n e , a n d p r o v i d e s m o s t o f t h e m o t i o n in f l e x i o n - e x t e n s i o n trunk [White a n d P a n j a b i 1990].  In c o n t r a s t ,  the  cervical region protects  the  central  for the nervous  8  Chapter 1  Introduction  system that connects the brain to the body and limbs, withstands a much smaller load as compared to the thoracic and lumbar spine and which in most instances is equal to the weight of the head, and provides a wide ROM for the head and neck. Motion between vertebral levels from the second cervical vertebra to the sacrum is made possible by a three-joint complex at each level, made up of the intervertebral disc and a pair of facet joints that connect adjacent vertebral bodies. Rotational motions are the predominant modes of motion across these FSUs. Translational motion in the normal spine exists as part of the coupled motions during rotational motion. For example, a mean forward translation of the lumbar FSU of between 2.3 to 3.3 mm was measured during full flexion-extension [Schneider et al. 2005], together with 2° of coupled axial rotation and 3° of coupled lateral bending [Pearcy et al. 1984, Pearcy 1985]. The rotational ROM at each FSU differs between regions of the spine and also between individual levels in each region. The lumbar spine, for example, has a rotational ROM of about 12° to 17° in flexion and extension, while the thoracic spine has about 4° to 12° [White and Panjabi 1990].  In axial rotation, the lumbar spine has a rotational ROM of about 1° to 2°  unilaterally while the thoracic spine has between 2° to 9°. The rotational ROMs in lateral bending are about the same for FSUs in the thoracic and the lumbar spine, and range between 5° to 8°. While motion across individual FSU is limited, the cumulative rotational ROM of the spinal column allows for substantial motion of the trunk and head. Load transfer across vertebral levels and with active muscle stabilization is necessary in order for the spinal column to remain erect in static and dynamic situations. Stabilization of the spinal column can be compared to the stabilization of an inverted-pendulum (Figure 1.6). The inverted pendulum, depicted as a long pole on a hinge, can be stabilized with guy wires pulling from both sides. This inverted pendulum model of the spinal column without the guy wires is unstable like the inverted cone in Figure 1.5. In the spinal column, muscle activation act like guy wires to help stabilize the column and the lost of active muscle in the asleep or dead person results in the spinal column not being capable of remaining upright. Coactivation of the agonist and antagonist muscles is necessary in order to maintain a neutral position [Cholewicki et al. 1997]. As a result, compressive loads are applied onto and transmitted across vertebral levels.  9  Chapter  Introduction  1  Figure 1.6  Analogy of the stabilization of the spinal column to an inverted pendulum. A long pole can stand vertically if stabilized with guy wires. Changes in the applied forces in each wire would result in change in inclination of the pole. In the spinal column, a combination of muscle forces creates motion in the spine. Coactivation of the agonist and antagonist muscles are necessary for motion and stability. Between vertebral levels, an intervertebral disc and a pair of facets act as spacers to provide distraction, and to allow for load transfer as well as motion. Surrounding ligaments further stabilise the spine and prevent excessive motion. The shaded arrows indicate compressive loads through the structures, as a result of tension from the guy wires, muscle forces and reaction forces. [Adapted from Daggfeldt and Thorstensson 1997 and Cholewicki et al. 1997]  Load  transfer across  spinal levels  mainly p a s s e s  facet joints a n d intervertebral disc, onto the both  agonist  a n d antagonist  prevent excessive motion. agonist muscles. relaxation muscles  of the works  [Cholewicki (extensor)  et  muscles  as  adjacent vertebrae.  muscle  muscle  allows  c o o r d i n a t i o n with  is  the  through  ligaments  to s t a b i l i z e t h e  spine  the and and  T h e e x t e r n a l f o r c e s n e c e s s a r y to c r e a t e m o t i o n a r e p r o v i d e d by the  antagonist  al. 1997].  Surrounding  provide counteracting forces  Activation of the agonist m u s c l e c r e a t e s  in  compressive forces  During agonist,  the  backward the  anterior longitudinal ligament a n d the  the  agonist  agonist  to  m o t i o n t o w a r d s it w h i l e c o n c u r r e n t effect  muscles  trunk e x t e n s i o n ,  psoas  major  (flexor)  to  movement. achieve  for e x a m p l e , muscle  the  The  antagonist  mechanical the  stability  erector  antagonist,  spinae  and  i n t e r v e r t e b r a l d i s c h e l p s t a b i l i z e t h e joint ( F i g u r e  the 1.7).  U s i n g a f r e e b o d y d i a g r a m , t h e f o r c e s a c t i n g o n t h e F S U m u s t b e in e q u i l i b r i u m [ D a g g f e l d t a n d Thorstensson  1997].  10  Chapter  Introduction  1  Extension Motion  Muscles  Muscles  Figure 1.7  The cord  Muscles  v e r t e b r a l c o l u m n p r o t e c t s t h e s p i n a l c o r d a s it d e s c e n d s f r o m t h e b r a i n . through  the  ( F i g u r e 1.4).  spinal  canal,  and  exits  as  spinal  by the t r a n s v e r s e ligamenta  p r o c e s s a n d the pedicle.  flava  roots  through  the  T h e spinal  intervertebral  B o n y p r o t e c t i o n o f t h e s p i n a l c o r d a n t e r i o r l y is p r o v i d e d b y t h e  b o d y a n d posteriorly by the s p i n o u s p r o c e s s a n d laminae.  the  Muscles  Free body diagrams of a functional spinal unit (FSU) during extension and during flexion. The extension muscles are agonist muscles during extension motion, while the flexor muscles are antagonist muscles. Other counter forces could be contributed by gravity, the ligaments (L), and the disc.  passes  foramen  Flexion Motion  lies  posterior  to  the  L a t e r a l l y , t h e s p i n a l c o r d is  vertebral protected  T h e posterior longitudinal l i g a m e n t lies anterior a n d spinal  cord,  and  they  both  help  to  provide  further  p r o t e c t i o n to the c o r d .  11  Chapter  1  1.3  The Aging Spine  1.3.1  Introduction  Epidemiology  The populations of the developed nations are aging, with a projected increase in absolute numbers and proportions of elderly people in the population. With low fertility levels (1.56 children per women in 2005-2010) and higher life expectancy (75 years in 1995-2000 and 82 years in 2045-2050), the proportion of the population in developed nations aged 60 years or above will increase from 19% in 2000 to 32% in 2050 [UN Population 2003]. In Canada, the trend of lower fertility levels and higher life expectancy reflect the world trend for developed nations. The average number of children per woman is projected to be about 1.5 from 2000 to 2025. Life expectancy at the same time is projected to increase from 78.7 years in 1995-2000 to 83.3 in 2045-2050 [UN Population 2003]. The proportion of the population above 65 years in 2000 was 12.7% and was projected to increase to 21.4% by 2026 and 25.4% by the year 2051 [Statistics Canada 2001] (Figure 1.8). The number of people aged above 65 years will increase from 3.9 million in 2000 to 7.8 million in 2026 and to 9.4 million in 2051.  At the same time, the number of people aged above 80 years will increase from 0.9  million in 2000 to 1.9 million in 2026 and 3.3 million in 2051 [Statistics Canada 2001]. Population distribution of Canada by age and sex, 1971 and 1996  1 200  1 000  800  600  4 0 0 2 0 0  0  0  2  0  0  4  0  0  8  0  0  8  0  0  1 000  1 200  For a population of 100,000 1971  Figure  1.8  •  1996  Population distribution of Canada in 1971 and 1996, from Statistics Canada (www.statcan.ca). The proportions of the Canadian population 65 years and above will reach 25.4% of the total population in 2051.  12  Chapter 1  Introduction  W i t h a n a g i n g p o p u l a t i o n , t o t a l h e a l t h e x p e n d i t u r e will b e p r o p o r t i o n a l l y h i g h e r . Canada  was  already  spending  40%  of  its t o t a l  health  spending  population a b o v e 65 y e a r s old [ A n d e r s o n a n d H u s s e y 2000]. n e e d t o b e s p e n t o n h e a l t h c a r e in t h e n e x t f e w d e c a d e s  on  the  12.8%  In  of the  1994, total  A h i g h e r p r o p o r t i o n o f G N P will  in o r d e r t o m a i n t a i n t h e c u r r e n t q u a l i t y  of c a r e .  1.3.2  Age Related There  are several  [Vernon-Roberts various  Degenerative  1992,  structures  physical degenerative  Prescher  in t h e  Pathology  1998,  F S U , such  changes  Ferguson and as  the  p u l p o s u s , a n d t h e a n n u l u s f i b r o s i s ( F i g u r e 1.9),  vertebral  t h a t c o u l d o c c u r in t h e  Steffen body,  2003]. facet  These  joints,  individually or collectively.  aging  changes  endplate,  nucleus also  section  1.3.3).  endplate concavity  Figure  affect  O s t e o p o r o s i s is  a d e g e n e r a t i v e c o n d i t i o n o f t h e a g i n g s p i n e , a n d will b e d i s c u s s e d s e p a r a t e l y in t h e n e x t (Section  spine  1.9  (5)  Degenerative changes in the aging spine could result in a intervertebral foramen space, loss of disc height, loss of vertebral and a kyphotic posture. The vertebral body is generally larger  smaller height, in size 13  Chapter 1  Introduction  anteriorly and laterally in the degenerated spine. Osteophytes (6) are commonly seen growing at the anterolateral and posterolateral regions. The vertebral body could also suffer from an osteoporotic fracture (8). Nucleus degeneration (10) would result in loss of intervertebral disc height, which could lead to bridging osteophytes (9) forming at the anterolateral vertebral rim. Vertebral fracture and loss of disc height could result in excessive loading on the facet joint, resulting in facet osteophytes (1) and facet osteoarthritis (2). Endplate degeneration and annulus lesion (7) are commonly found in the degenerated disc. If the nucleus remains fairly hydrated but the surrounding endplate or annulus is degenerated, Schmod's node (4) and disc bulge/prolapse (3) could result from nucleus migration. With a hydrated nucleus and weak adjacent bone, the endplate could become more concave (5).  Vertebral rim osteophytosis [Nathan et al. 1994a] and facet joint osteophytosis [Tischer et al. 2005] are two common degenerative conditions of the vertebra. The cause of osteophyte formation is attributed to adaptation of bone tissues to abnormal strains observed [Nathan et al. 1994b].  These osteophytes could compress surrounding nerves [Matsumoto et al. 2002],  resulting in pain, loss of function and/or loss of sensation. A classification describing four degrees of vertebral osteophytes was previously defined [Nathan et al. 1994a], ranging from only isolated points of hyperostosis to fused osteophytes across adjacent vertebrae.  The  prevalence of osteophytosis is higher in men than in women [Kinoshita et al. 1998, Pye et al. 2004]. Osteophytes with second degree or higher occurs in 90% of men and in 64% of women above 60 years old [Kinoshita et al. 1998]. The incidence of osteophytosis is also associated with increasing age [Pye et al. 2004]. Facet joint degeneration occurs after disc or vertebral body degeneration [Butler et al. 1990]. The prevalence of facet osteoarthritis occurs in 14% of men and women above 60 years old [Kinoshita et al. 1998]. In most cases, disc degeneration would occur without the presence of facet degeneration, while facet degeneration would only occur in the presence of disc degeneration. The cause of facet degeneration is possibly due to altered load patterns within the facet joint, such as loading location and increased magnitudes [Nathan et al. 1994b]. Disc degeneration also results in a change in the instantaneous axis of rotation of the FSU [Rousseau et al. 2005] and an increase in the facet forces. As a result of an increase in load magnitudes and a change in loading location in the facet joints, adaptive remodelling within the facet could lead to enlarged facets, facet osteophytes and facet osteoarthritis [Taylor and Twomey 1986, Berlemann et al. 1998]. In the vertebral endplates, fissure formation, fractures and horizontal cleft formation are some early degenerative changes observed [Vernon-Roberts 1992]. In later stages, increase in 14  Chapter 1  Introduction  vascular penetration, extension of calcification and ossification from the bony endplate, loss of cartilaginous endplate thickness and microscopic intrusion of nucleus material through the endplate are observed.  Biomechanically, decreased strength at the central location of  endplates in degenerated disc has been measured experimentally [Grant et al. 2002]. The cranial or caudal migration of nucleus material through these endplates, which would develop into a Schmorl's node, takes place in vertebrae with weakened endplates and underlying trabeculae. In a study involving 100 T1-L5 cadaveric specimens from donors aged 43 to 93 years old, Schmorl's nodes were found in 58% of the specimens, with 4 1 % of the specimens having multiple nodes [Pfirrmann and Resnick 2001]. Degeneration of the intervertebral disc is another physiological change that occurs in the aged spine. It is generally associated with a loss of disc height and is found to occur in 67% of those above 60 years old [Jones et al. 1995]. The annulus fibrosis and the nucleus pulposus could develop signs of degeneration independently of each other in the aging spine. Annulus lesions are observed as early as the third decade, before nucleus degeneration occurs, while concentric tears and delamination are also observed in later stages of the degenerated annulus [Vernon-Roberts 1992]. Disc bulge and disc prolapse (disc herniation) commonly occur at the posterolateral region, as the annulus is thinnest in this area compared to the anterior or lateral regions, while the posterior longitudinal ligament prevents prolapse through the central posterior annulus [Vernon-Roberts 1992]. Degeneration of the nucleus pulposus is considered to be due to a loss of nutrients that it receives through the vertebral endplates [Buckwalter 1995]. Subsequently, the nucleus losses water content with age [Urban and McMullin 1988], and the disc pressure is decreased. A uniform disc pressure is present in the healthy spine and is lost in the degenerated spine, resulting in higher magnitudes of pressures at the peripheral regions [McNally et al. 1996]. Nucleus degeneration and loss of disc height results in excessive load through the annulus and also the facet joints, and could trigger degenerative responses at these sites. A four group classification of the grade of degeneration of the disc has been established, based on the status of the nucleus [Nachemson 1960]. In group 1, there are no changes visible to the naked eye. In group 2, some fibrous tissue in the nucleus pulposus could be seen, while the annulus remains intact. In group 3, there are some fibrous tissues in the nucleus, with isolated fissures in the annulus fibrosis.  In group 4, fissures and cavities are present in both nucleus and  annulus, and osteophytes are commonly found in the adjoining vertebrae.  15  Chapter 1  1.3.3  Introduction  Osteoporosis  in the Aging  Spine  Osteoporosis was defined at a 1993 consensus conference as "a systemic skeletal disease characterized by low bone mass and microarchitectural deterioration of bone tissue, with a consequent increase in bone fragility and susceptibility to fracture" [Consensus 1993]. More recently, osteoporosis was defined as "a skeletal disorder characterized by compromised bone strength predisposing a person to an increased risk of fracture" [NIH 2001]. The World Health Organisation (WHO) established an operational definition of osteoporosis based on the bone mineral density (BMD) measurement compared to the mean for a normal young adult population of the same sex and race [WHO 1994, Kanis et al. 1994]. A "tscore" is assigned to the patient, where t-score corresponds to the number of standard deviations above or below the mean BMD for normal young adults. A patient is considered to be osteoporotic if the t-score is less than -2.5, and osteopenic if the t-score is between -1.0 and -2.5 (Figure 1.10).  Normal bone is defined as a t-score between +2.5 and -1.0.  These  definitions were adopted in Canada [Brown et al. 2002a], in the United States [NOF 1998, NIH 2001], and in Europe [WHO 1994, Kanis et al. 2000].  i 10  Figure  1.10  1— —i 1  20  30  1  1  1  1  r  40  50  60  70  80  Typical variation of bone mineral density (BMD) with age in a population (in bold, mean BMD). Bone accrual usually occurs up to 30 years old, is maintained from 30-50 and is rapidly lost after 50 years of age. The mean BMD of the 25 year olds is usually used as a reference point (X) and the t-score is assigned based on the number of standard deviation from the mean. Osteoporosis is defined as having a t-score less than 2.5 (below the lower dashed line). Osteopenia is defined as having a t16  Chapter 1  Introduction  score between -1 and -2.5 (between the dashed lines). Normal BMD is defined as a t-score between +2.5 and -1. [Adapted from Bartl and Frisch 2004] A strong correlation is found between osteoporosis and fracture risk [Melton et al. 1993, Ross et al. 1996, Legrand et al. 1999]. In a recent study in Canada on fracture risk of women older than 50 years, it was reported that one in four will suffer one or more osteoporotic fractures in her remaining lifetime [Lorrain et al. 2003]. The lifetime risk of sustaining a fracture from 50 years old onward is estimated at 11-18% for women and about 5-6% for men [Lips 1997, Lorrain et al. 2003]. As many osteoporotic patients are aged and frail, they have an increased rate of falls due to weakness, poor balance, poor eyesight or other co-morbidity [Lord et al. 1994, Arnold et al. 2005]. Since osteoporotic patients have a higher chance of fracture from a fall, their risk of fracture increases with age. With increased life expectancy, this lifetime fracture risk will further increase. In adults at age 50, lifetime risk of fracture of the vertebra is the highest compared to risk of fracture of the hip or of the wrist [Melton et al. 1993, NOF 1998]. Unlike fracture of the hips or wrists where the bone breaks into two and the patient usually needs to be hospitalised for treatment, the majority of vertebral fractures are unsymptomatic and are thus undiagnosed. It is estimated that only about 30% of vertebral fracture/deformities come to clinical attention [Cooper et al. 1993, Lips 1997]. Vertebral fracture is however a clinical marker for subsequent fractures [Papaioannou et al. 2002] and a sign for osteoporosis in the patient. Patients normally have complaints in the later stages of fracture when they present with kyphosis, loss of anterior body height of more than 20%, or fracture at other sites. As prevalence of vertebral fracture is associated with a five-fold increase in risk of sustaining a further vertebral fracture [Black et al. 1999], many osteoporotic patients have multiple vertebral fractures when presented at the clinic. Vertebral osteoporosis commonly presents as vertebral compression fracture on radiographs.  The prevalence of vertebral compression fracture increases with age and  osteoporosis [Cortet et al. 2002]. Vertebral compression fracture could present as 1) concaved endplate, 2) decreased vertebral height, 3) wedge shaped vertebral body, or in combination (Figure 1.11) [Genant et al. 1993, Lenchik et al. 2004]. Wedged fractures are most common from T5-L2, while an overall decrease in vertebral height are most common in L1-L4 [De Smet et al. 1988]. Vertebral compression fracture affects the overall spinal curvature, and spinal deformity such as kyphosis and scoliosis could develop following compression fracture in the osteoporotic spine [Healey and Lane 1985, Cortet et al. 2002].  17  Chapter 1  Introduction  a) c o n c a v e d  b) c o m p r e s s e d  d) c o n c a v e d a n d c o m p r e s s e d  e) w e d g e d  c)  and compressed  g) c o m p r e s s e d , c o n c a v e d a n d  Figure  1.11  wedged  f) c o n c a v e d a n d  wedged  wedged  Vertebral compression fracture commonly present with a) concaved endplates, b) decreased vertebral height, c) wedge shaped vertebral body, or in combination (d-g). [Adapted from Genant et al. 1993 and Lenchik et al. 2004]  P a i n , p h y s i c a l i m p a i r m e n t , f u n c t i o n a l i m p a i r m e n t a n d q u a l i t y o f life a r e i s s u e s with o s t e o p o r o t i c v e r t e b r a l c o m p r e s s i o n f r a c t u r e s [Hall et a l . 1 9 9 9 , 1999].  N e v i t t et a l . 1 9 9 8 ,  associated L i p s et a l .  V e r t e b r a l o s t e o p o r o t i c f r a c t u r e is t e r m e d "a s i l e n t killer" a s m i c r o f r a c t u r e s o c c u r w i t h o u t  s y m p t o m s o r s o m e t i m e s w i t h s y m p t o m a t i c p a i n t h a t will s u b s i d e  after 3-4 w e e k s .  Patients  with  v e r t e b r a l o s t e o p o r o t i c f r a c t u r e s o f t e n c o m p l a i n o f b a c k p a i n a n d r e q u i r e o n e to five d a y s o f b e d r e s t a n d a n a d d i t i o n a l 12 to 4 4 might be no obvious and  days  of limited activities  radiologic evidence  a n d patients  r e c o v e r after b e d  Conservative treatments  at this e a r l y s t a g e a r e  rest a n d l i m i t e d activity [Nevitt et a l . 1 9 9 8 ] .  i n t e r v e n t i o n is t h u s n o t n e c e s s a r y e v e n if t h e r e is e v i d e n c e  Surgical  of vertebral fracture, mainly b e c a u s e  t h e p a i n w o u l d n o r m a l l y g o a w a y a n d t h e r e is u s u a l l y n o c h r o n i c n e u r o l o g i c a l d e f i c i t . y e a r s , s o m e studies reported the u s e of p e r c u t a n e o u s vertebroplasty, w h e r e c a l c i u m c e m e n t or P M M A c e m e n t  There  o f v e r t e b r a l f r a c t u r e a t t h e initial o n s e t o f t h e f r a c t u r e  radiologic a s s e s s m e n t could be subjective.  common  p e r y e a r [Nevitt et a l . 1 9 9 8 ] .  In r e c e n t phosphate  a r e i n j e c t e d bilaterally t h r o u g h t h e p e d i c l e s into o n e o r m o r e l e v e l s , to  18  Chapter 1  Introduction  alleviate pain resulting from vertebral fracture and to prevent further vertebral fractures [Mehbod et al. 2003, Phillips 2003]. On a smaller scale, changes in the trabecular bone are observed in the aging spine [Inoue and Chao 2002, Keaveny and Yeh 2002].  Trabecular strut thickness, length and  connectivity are typically reduced in the osteoporotic spine [Barger-Lux and Recker 2002]. Thinning and loss of horizontal struts in the trabecular lattice are often also observed [Mosekilde 1988, Thomsen et al. 2002], which makes the remaining vertical struts susceptible to buckling failure [Schnitzler 2003]. To prevent osteoporosis, a lifelong approach is required. Bone accrual in the young should be maximized, bone in adults should be maintained and bone loss in elderly should be prevented.  In the young, the rate of bone mineral accrual is found to be highest before  menarche in girls [McKay et al. 1998]. This increased accumulation of bone mineral in the young is related to increased bone mass in later life and would result in reduced fracture risk [Matkovic et al. 1979, Barr et al. 1998, Khan et al. 2000]. It is recommended that children, particularly those entering and passing through puberty, should be encouraged to participate in sports [Brown et al. 2002a].  In adults, exercise is found to maintain BMD of early  postmenopausal women [Engelke et al. 2005]. Without exercise, the benefits of BMD gains in athletes are found to be lost after three to five decades of retirement [Karlsson 2004]. In the elderly, a combination of exercise and pharmacologic interventions had been shown to prevent bone loss [NIH 2001, Brown et al. 2002a, Kannus et al. 2005]. Pharmaceutical agents such as biphosphonates, calcitonin and selective estrogenreceptor modulators (SERMs), are recommended by National Agencies for inhibiting the resorption of bone in osteoporotic patients [NOF, 1998, NIH 2001, Brown et al. 2002a]. Treatment to promote bone growth in the elderly using anabolic agents such as parathyroid hormone (PTH) is a new direction to treat osteoporosis [Neer et al. 2001, Rosen and Bilezikian 2001, Rubin and Bilezikian 2002]. Osteoporosis is not considered to be an age dependent disease as it can affect people of all ages, and it can be prevented in the elderly [NIH 2001]. The proportion of the population with t-score less than -2.5, however, increases exponentially with age (Figure 1.10) [Kanis et al. 1994]. This is due, in part, to the fact that loss of bone mineral increases rapidly following menopause in women. Other factors such as genetics [Rhalston 2005], exercise and diet [Lunt et al. 2001] are important factors on BMD for the elderly patients. Studies indicate varying rates of osteoporosis in different populations [Lunt et al. 1997]. The prevalence of osteoporosis among women aged above 50 years varies from 10.9% in 19  Chapter 1  Introduction  Taiwanese women [Yang et al. 2004] to 20% in white American women [Looker et al. 1997, Mazess and Barden 1999].  Among Canadian women above 50 years old, 12.1% were  diagnosed with osteoporosis in 2000 [Tenenhouse et al. 2000]. In the United Kingdom, among all women above the age of 50, 22.5% were diagnosed as osteoporotic in the hip while the proportion increased to 58.7% among those above the age of 80 [Kanis et al. 1994]. The rate of osteoporosis in Europe is consistently higher than 50% for those above 80 years of age [Melton 1997, Kanis et al. 1994]. While men tend to have higher bone mass, they are also at risk for osteoporosis. Using a cut-off value of 0.545 g/cm in femoral neck BMD to define osteoporosis, 6% of European men 2  and 23% of women above 50 years old were estimated to be osteoporotic [Kanis et al. 1994]. Consistent with this, a separate study carried out in the United States using a cut-off of 0.56 g/cm in femoral neck BMD to define osteoporosis, estimated the rates of osteoporosis in men 2  and women above 50 years old to be 4% and 20% respectively [Looker et al. 1997]. The use of peak bone mass in the female populations to define osteoporosis in male is however controversial [Lombardi and Ross 2001]. Using the same cut-off values in male and female, the rate of osteoporosis was underestimated in men [Melton et al. 1998, Melton 2001, Lombardi and Ross 2001]. With an aging population in Canada and the rest of the world, the number of people affected by osteoporosis will definitely increase in the near future [Papadimitropoulos et al. 1997, Brown et al. 2002a].  In Canada, where 1.2 million women and 0.7 million men are  expected to be above 80 years old by 2026 [Statistics Canada 2001], 0.6 million women and a conservative estimate of 0.1 million men in this age range alone would be osteoporotic. By 2051, with 2.0 million women and 1.3 million men above 80 [Statistics Canada 2001], there will be 1.0 million osteoporotic women and at least 0.2 million osteoporotic men from this age range alone. A total of 109,502 hospitalization days due to osteoporosis and osteoporosis related vertebral fractures, involving 4,033 males and 23,657 females, were recorded in 1993 [Goeree et al. 1996]. The acute care cost for vertebral fractures was estimated at CDN$72 million (1993) dollars, while the total direct costs due to osteoporotic fractures including hospitalization, drugs, long-term and chronic care to the Canadian health care system in 1993 was estimated to be CDN$1.3 billion [Goeree et al. 1996]. The cost of health expenditure on osteoporosis related fractures will increase proportionally with the increase in absolute number of osteoporotic elderly in the population [Wiktorowicz et al. 2001].  20  Chapter 1  1.3.4  Spinal  Introduction  Degeneration  and  Osteoporosis  Osteoporosis and disc degeneration are generally mutually exclusive in the elderly spine. While osteoporosis of the spine is commonly associated with changes to the vertebral body, it could lead to other degenerative conditions involving the disc and facet joints (Figure 1.12). Alteration of stresses in the osteoporotic spine was postulated as a pathway leading to disc degeneration [Margulies et al. 1996]. For example, with a weak endplate and underlying bone, Schmorl's nodes could result from local stresses exceeding the material strength of the endplate. Increased endplate concavity, common in the osteoporotic spine, could also lead to increased stresses in the annulus fibrosus.  Spinal deformity such as kyphosis following  vertebral compression fracture, increased load in the facet and altered load distribution, and leaded to facet degeneration and facet osteoarthritis. Disc degeneration is associated with loss of disc height, facet degeneration, disc bulge, and osteophyte growth.  Disc bulge and tension in the anterior longitudinal ligament was  postulated to result in stresses in the local area, and together with adaptive bone remodelling, results in osteophyte growth [Kumaresan et al. 2001]. Disc degeneration could also lead to an altered load transmission path through the spine (Figure 1.12). In discs with a hydrated nucleus, load transmission passes through the nucleus and the central region of the vertebral. In degenerated discs, load transmission through the nucleus diminishes and load transmission is transferred through the annulus, onto the peripheral regions of the vertebral body [Horst and Brinckmann 1981, Kurowski and Kubo 1986, Adams et al. 1996].  In support of this observation, a finite element model simulating a  degenerated disc also found a shift in load path towards the peripheral regions of the vertebral body [Polikeit et al. 2004]. As a result of the different load path through the intervertebral disc in the degenerated spine, the facet joints sustain a higher compressive load during erect standing [Pollintine et al. 2004]. Stress-shielding could occur within the anterior column, and could subsequently lead to a decrease in compressive strength, stiffness and modulus of the trabecular bone at the central regions of the vertebral body [Keller et al. 1989]. Subsequently, the weakened anterior column is susceptible to a higher risk of osteoporotic compression fracture. A negative correlation between disc degeneration and low BMD was noted in the literature [Jones et al. 1995, Harada et al. 1998, Nanjo et al. 2003, Pye et al. 2006]. In other words, disc degeneration would be more common in patients with high BMD, and conversely, patients with low BMD would be less likely to have disc degeneration. Furthermore, an inverse relation was found to exist between vertebral body deformity and intervertebral disc narrowing 21  Chapter 1 [Okawa  et  reported  Introduction al.  1996].  [Miyakoshi  Also, et  an  inverse  al. 2003].  relation b e t w e e n  Therefore,  these  osteoporosis  studies  and  spondylosis  further s u p p o r t the  two  was  different  p a t h w a y s o f s p i n a l d e g e n e r a t i o n f o r t h e e l d e r l y r e p o r t e d in t h e l i t e r a t u r e ( F i g u r e 1 . 1 2 ) .  In t h e  first c o n d i t i o n , t h e r e is a h e a l t h y d i s c a n d w e a k b o n e , w h i l e in t h e s e c o n d c o n d i t i o n t h e r e is a d e g e n e r a t e d disc a n d healthy bone.  In t h e first c o n d i t i o n , s p o n d y l o s i s w a s  stages.  normal  In t h e  second  scenario  of  bone  mass  and  degenerated  n o t p r e s e n t in l a t e r disc,  spondylosis  d e v e l o p e d in t h e l a t e r s t a g e s .  Healthy bone  Figure  1.12  Osteophytes  osteophytes  Facet degeneration  Progression of spinal degeneration in the elderly spine from weak bone [Margulies et al. 1996] or from degenerated disc [Vernon-Roberts 1992]. The vertical arrows indicate load paths through the anterior and posterior regions of the functional spinal unit (FSU) in compression. (Top row): The combination of a healthy disc with weak vertebra bone could result in osteoporotic compression fracture and Schmorl's node. Further compression fracture could result in a wedged vertebra with kyphosis of the spinal column. The increased load at the facet joint could lead to facet arthritis. Disc bulge could also result with a hydrated disc. (Bottom row): A degenerated nucleus would result in a loss of disc height, increased disc bulge and a shift in load path through the annulus and facet joints. Disc bulge and tension in the anterior longitudinal ligament could be the cause for osteophytes to grow at the anterolateral regions of the vertebral rim. The increased peripheral load, as a result of loss of disc pressure, could cause buckling of the vertebral cortex. Further loss 22  Chapter 1  Introduction  of disc height and osteophyte growths could result in osteophytes. The increased load at the facet joint could again facet degeneration. A decrease in bone mass at the core of the could result from decreased loading through the nucleus adaptive bone remodelling. 1.3.5  Biomechanics  of the Aged  bridging lead to vertebral following  Spine  Biomechanics! properties of the spine, such as its kinematics and the overall load bearing capacity, are affected in the elderly patients with spinal degeneration and/or osteoporosis. These changes to the biomechanical properties could be attributed in part to changes in geometric properties (Figures 1.9 and 1.11) and also to material properties [Aspden 2003]. The three fundamental roles of allowing motion, bearing load and protecting the spinal cord could subsequently become undermined in the aged spine due to these changes in biomechanics. Changes to the kinematics of the FSU such as its stiffness, laxity (neutral zone) and ROM have been reported in the literature. Stiffness of cadaver FSUs initially decreases during early disc degeneration and increases with severe disc degeneration [Brown et al. 2002b]. In a separate animal study, the stiffness of degenerated disc is higher than in intact disc when subjected to cyclic loading [Kaigle et al. 1998]. Higher joint laxity is present in degenerated spines than in normal spines in a flexibility study of 47 cadaveric lumbar FSUs in axial rotation [Mimura et al. 1994].  In another study, the neutral zone of degenerated mouse tail discs  increases by 33% in flexion following compressive overloading to induce degeneration [Lotz et al. 1998]. ROM of degenerated discs has been found to increase with increasing severity in degeneration grades, but decrease for the worst degeneration grade in two separate cadaver studies [Fujiwara et al. 2000, Tanaka et al. 2001].  It appears that the overall change to  mechanics of the degenerated disc is an increase in stiffness and joint laxity, with changes to the ROM dependent on the grade of degeneration. Geometric changes of the vertebral body in the elderly have been reported. In general, the elderly vertebra has larger overall dimensions in the antero-posterior direction [Evans et al. 1993] and has smaller vertebral heights [Diacinti et al. 1995]. This increase in antero-posterior dimension could increase load bearing area in the elderly. Changes in geometric properties on bone strength have been studied in the long bone [Aspden 2002, Russo et al. 2006] and in the spine [Duan et al. 2001, 2005]. Age related increase in cross sectional area and decrease in volumetric BMD of the vertebral body increases the risk of vertebral fracture [Duan et al. 2001, 2005]. 23  Chapter  1  Introduction  Geometric changes of the spinal column, such as kyphosis, could also affect the load applied to the lumbar vertebrae.  The centre of gravity of the upper body is located more  ventrally in a kyphotic patient and a larger flexion moment is experienced in the lumbar region [Rohlmann et al. 2001]. This change in load pattern in the kyphotic patient affects the overall load bearing capacity of the spine. Material properties of the bone and surrounding soft tissues have been observed to degrade in elderly patients.  The trabeculae in the osteoporotic spine are oriented more  vertically while trabeculae in normal bone are more isotropic [Inoue and Chao 2002]. This results in the vertebrae of the aged being strong in axial compression only, but weak in all other loading directions. Ligament and muscle strength are also generally weaker in elderly patients. Elastic modulus of the anterior longitudinal ligament has been reported to increase by 135% with aging (from 21 to 79 years old), while it's tensile strength decreases by 55% [Neumann et al. 1994]. Similarly, the elastic modulus and tensile strength of the L4-L5 spinous process and supraspinous/interspinous ligament complex has been reported to decrease by about 50% with aging [lida et al. 2002]. Moreover, sarcopenia, the loss of skeletal muscle mass and strength in the elderly, has a 30% prevalence among the population above 60 years old [Doherty 2003].  1.3.6  Clinical  Problems  of the Aging  Spine  Back pain is a major factor affecting physical health status in many adults and remains a problem in the elderly [Hartvigsen et al. 2003]. Any sites in the spine could be the source of back pain, and it could arise from any structure that receives an innervation, including any ligament, muscle, fascia, joint or disc [Adams et al. 2002]. The three general areas in the spine where pain could originate from are: A) deep and originating in the FSU, B) superficial and emanating from skin, fascia, and muscles, and C) from the spinal roots protected by the spinal column [O'Brien 1984].  Deep pain originating from within the FSU is a problem in the  degenerated spine. The degenerated disc, facets and vertebra are potential sources of deep spinal pain [Martin et al. 2002, Benoist 2003]. Spinal cord or root compression results in severe pain, sensory loss and/or loss of function of the extremities. As a result of an impingement on the spinal cord or spinal roots, the motor and/or sensory function of the central nervous system could be compromised and surgical intervention is often necessary to resolve the problem. Impingement on the dorsal root could result in loss of sensation, an impingement on the ventral root could result in loss of motor function and impingement of more distal spinal roots could result in loss of both motor and sensory function. Surgical intervention to alleviate the pain and restore the motor and sensory 24  Chapter 1  Introduction  functions would often involve surgical disruption of some of the spine's structure in order to access the relevant impingement sites.  As a result, the normal motion and load transfer  function of the affected FSU may be compromised. Back pain causes disability and results in reduction in productivity at the work place [Kopec and Esdaile 1998]. The patient's ability to bear load could also be diminished by back pain and in some cases there is also inability to stand or walk a short distance. Neural deficit, diminished ROM and diminished load bearing roles are common in patients with back pain. Surgical intervention may alleviate the chronic pain and associated disability [Dawson and Bernbeck 1998]. Instability is another problem in the aging spine. Spondylolisthesis is an example of spinal instability that occurs commonly at the lumbosacral level (L4-L5 and L5-S1), when one vertebra shifts forward upon another. The spine is unstable, in such a case, as the ligaments, discs and facets are unable to prevent forward shearing of the superior vertebral level (L5) against the sacrum [DeWald 2002]. Degenerative conditions in the thoracolumbar spine often result in spinal pain, instability and deformity. Degenerative spondylolysis, spondylolisthesis [Bassewitz and Herkowitz 2001, Herkowitz and Kurz 1991], spinal stenosis [Sengupta and Herkowitz 2003, Postacchini 1999] and adult degenerative spinal kyphosis and scoliosis [Healey and Lane 1985] are common degenerative conditions that may affect the elderly. Degenerative spondylolysis is the condition associated with disc degeneration and growth of osteophytes in the elderly, resulting in an increase in vertebral body dimensions, posterior disc bulge and occasionally leading to spinal stenosis [Vernon-Roberts 1992]. Degenerative spondylolisthesis is the condition where there is excessive forward translation of a spinal level against the next as a result of degeneration primarily in the facet joints but also in the discs. Canal stenosis is the narrowing of the spinal canal resulting in constriction of the spinal cord or cauda equina, while intervertebral stenosis is the narrowing of the spinal canal resulting in constriction or compression of the spinal nerve roots [Cinotti et al. 2002]. Compression to the spinal cord or spinal nerve roots results in pain and loss of lower limb function and surgical intervention would be necessary to stabilize the affected segment [Sengupta and Herkowitz 2005]. Spinal deformity such as kyphosis and scoliosis could also affect the well-being of the elderly.  Adult spinal deformity often occurs with vertebral deformity involving one or more  levels. Kyphosis is generally the result of vertebral compression fracture with loss of anterior body height, while adult degenerative scoliosis results from lateral wedge vertebral compression fracture or asymmetric disc narrowing [Healey and Lane 1985, De Smet et al. 1988, Hu 1997b]. 25  Chapter  1  Introduction  Kyphosis could also result from degenerative disc disease or osteoarthritis, with no vertebral fracture or osteoporosis [Osman et al. 1994, Schneider et al. 2004]. Osteoporotic fractures, described in an earlier section, could lead to the deformity of one or more vertebral bodies in the elderly. The prevalence of at least one vertebral deformity in those aged 65 years or older was reported to be 39%, and similar in both elderly men and women [Pluijm et al. 2000].  Vertebral deformity is associated with impairment, functional  limitations, disability and poor self-perceived health [Lyles et al. 1993, Burger et al. 1997, Pluijm et al. 2000]. Vertebral deformity in older women is also associated with significant back pain [Lau et al. 1998], and also with increased risks of mortality and hospitalization [Ensrud et al. 2000]. 1.4  Surgical Management of the Aging Spine  1.4.1  Conditions  Requiring  Spinal  Stabilization  In a recent special issue of the Journal of Neurosurgery in 2005, titled "Guidelines for the performance of fusion procedures for degenerative disease of the lumbar spine", detailed guidelines were comprehensively presented. In one recommendation, for example, "Lumbar fusion is recommended as a treatment for carefully selected patients with disabling low back pain due to one- or two-level degenerative disease without stenosis or spondylolisthesis" [Resnick et.al. 2005]. The compromise of one or more of the three fundamental roles of protection, load bearing and motion of the degenerated spine are the underlying principles where surgical intervention might be necessary. Alleviation of pain and return of neurological functions for the patients are usually reasons for surgical intervention [Postacchini 1999, Bassewitz and Herkowitz 2001, Sengupta and Herkowitz 2003]. Pain is closely related to a loss in function (motion and load bearing) and could be caused by spinal root compression (loss of protection). Loss of neurological function on the other hand is a result of spinal root compression and could result in pain and also loss of function. Other conditions that are not limited to the elderly population, such as trauma [Nguyen et al. 2003] and metastatic diseases [Fisher et al. 2005], could also result in the loss of performance by the spine of its fundamental roles and would also necessitate surgical intervention. Surgical treatments are often necessary for such conditions and spinal stabilization is often indicated to provide stability. Spinal fusion in patients with degenerative spinal disorders  26  Chapter 1  Introduction  could result in improved spinal stability and improved neurological function [Vaccaro and Ball 2000]. 1.4.2  General  Surgical  Approach  Spinal arthrodesis (fusion) is a common strategy carried out in patients with severe instability for the various problems of the aging spine [Postacchini 1999, Vaccaro and Ball 2000, Bassewitz and Herkowitz 2001, Sengupta and Herkowitz 2003]. In the surgical treatment of these patients, osteophytes and any protruding disc material causing nerve root compression are identified and removed.  As patients with spondylosis often have degenerated discs,  resulting in diminished disc height and intervertebral stenosis, interbody devices are often implanted into the intervertebral disc space to restore disc height. The objective in spinal fusion is to restore intervertebral disc heights, to create bone growth across vertebral bodies of unstable levels, to return to normal neurological function, and to provide relief of pain. Cancellous bone grafts are commonly inserted into and around with the interbody devices to promote bony fusion. Bony fusion is usually completed in three to six months. Spinal stabilization with rigid instrumentation devices such as pedicle screws and rods, or anterior body screws and plates, is often necessary to provide structural support. The posterior surgical approach for example would require removal of posterior bony structures and the remaining structure would be structurally unsound following implantation with the interbody device alone.  Pedicle screw instrumentation (Figure 1.13) is thus commonly used with  interbody devices to maintain stability and alignment so that bony fusion could occur. Spinal fusion carried out with rigid fixation such as pedicle screws has been reported to promote fusion [Kim et al. 1990, Zdeblick 1993, Thomsen et al. 1997, Bjarke et al. 2002].  27  Chapter  Introduction  1  Figure  1.13  The process  Radiographs of the spine following corpectomy and insertion of a titanium vertebral body replacement device and pedicle screw and rigid rod fixation.  i m m e d i a t e p o s t - o p e r a t i v e r i g i d i t y o f t h e i n s t r u m e n t e d l e v e l is i m p o r t a n t f o r t h e f u s i o n  to t a k e  place  [Benzel  1995].  Either too  little l o a d t r a n s m i t t e d t h r o u g h o r t o o  m o t i o n o f t h e b o n e g r a f t s c o u l d r e s u l t in s u b o p t i m a l c o n d i t i o n s f o r o s t e o i n t e g r a t i o n t o t a k e [Bauer and  M u s c h l e r 2000].  Non-fusion  in t h e  long  run m a y  result  much place  in i m p l a n t l o o s e n i n g  or  f a t i g u e f a i l u r e d u e to t h e c y c l i c a l l o a d i n g s a n d r e l i a n c e o n i m p l a n t s t o b e a r l o a d s [ B a b a t e t a l . 2004].  The  challenge  of a c h i e v i n g  post-operative  r i g i d i t y is c o n f o u n d e d  in t h e  osteoporotic  p a t i e n t s d u e to t h e w e a k b o n e - i m p l a n t i n t e r f a c e .  1.4.3  Clinical Problems The  spinal 2003].  with Surgical  Management  l o w b o n e q u a l i t y in t h e o s t e o p o r o t i c e l d e r l y p a t i e n t m a k e s s u r g i c a l i n t e r v e n t i o n w i t h  implants  a  challenge  [ O x l a n d et  al.  1996,  D i w a n et  al. 2 0 0 3 b ,  Potential problems include hardware loosening a n d s u b s e q u e n t  Glassman and  Alegre  a d j a c e n t level effects.  H a r d w a r e l o o s e n i n g w a s r e p o r t e d t o r e s u l t in f a i l u r e o f t h e s u r g i c a l p r o c e d u r e [ P a r a m o r e et al. 1996, interface stresses  B a b a t et a l . 2 0 0 4 ] .  a n d is a n i s s u e in t h e at  the  bone-implant  Hardware loosening osteoporotic interface  s u r r o u n d i n g t r a b e c u l a e a n d in s u b s e q u e n t screw-bone  is c a u s e d  b y f a i l u r e at t h e  p a t i e n t [ A n d e r s s o n et al. 2 0 0 0 ] .  following  surgery  device loosening.  i n t e r f a c e c o u l d r e s u l t in p e d i c l e s c r e w  results  in  bone-implant  A n increase  localised  failure  In p a r t i c u l a r , f a i l u r e at t h e  pullout [ P a r a m o r e et al. 1996,  in of  pedicle  B a b a t et a l .  28  Chapter 1  Introduction  2 0 0 4 ] , w h i l e f a i l u r e at t h e i n t e r b o d y c a g e - b o n e i n t e r f a c e c o u l d result in c a g e s u b s i d e n c e into t h e v e r t e b r a [ G e r c e k et al. 2 0 0 3 , Polikeit et al. 2003]. Pedicle screw loosening Okuyama  et al. 2001].  is a c l i n i c a l p r o b l e m  A s cancellous  in t h e o s t e o p o r o t i c  b o n e is m o r e a f f e c t e d  pedicle screws c o m m o n l y occurs within t h e vertebral body. concentrated  at t h e distal e n d o f t h e pedicle  screw.  patient [ H u 1997b,  by osteoporosis,  loosening of  M o r e o v e r , h i g h e r s t r e s s is a l s o  T h e failure  pattern  of pedicle  screw  loosening h a s been described as a butterfly-shaped void within t h e pedicle tract [ L a w et al. 1993]. T h e pedicle acts as a fulcrum point for t h e pedicle s c r e w under toggle loads. study of 52 patients w h o underwent  In a clinical  pedicle screw fixation, t h e m e a n B M D of patients  with  radiographic s c r e w loosening w a s significantly lower than those without loosening [ O k u y a m a et al. 2 0 0 1 ] .  Moreover, t h e m e a n B M D of patients with non-union of t h e fusion procedure w a s  significantly lower than those with successful union.  In vitro s t u d i e s h a d s i m i l a r l y d e m o n s t r a t e d  the association b e t w e e n stability o f pedicle s c r e w s a n d l o w vertebral B M D [Bennett et al. 1 9 9 7 , C o e et al. 1 9 9 0 , H a l v o r s o n et al. 1 9 9 4 , O k u y a m a et al. 1 9 9 3 , Soshi et al. 1 9 9 1 , W i t t e n b e r g et al. 1993, Y a m a g a t a et al. 1992]. I n t e r b o d y d e v i c e s u b s i d e n c e is a n o t h e r p r o b l e m w i t h s p i n a l f u s i o n i n o s t e o p o r o t i c b o n e . As  in t h e c a s e f o r p e d i c l e s c r e w s , t h e p o s t o p e r a t i v e i n c r e a s e  device-bone  interface  results  in  trabecular  failure  and  in s t r e s s e s at t h e i n t e r b o d y  subsequent  device  loosening.  S u b s i d e n c e of i n t e r b o d y d e v i c e s into a d j a c e n t e n d p l a t e s h a s b e e n identified t o involve failure o f the subchondral trabecular bone interface  [ H a s a g a w a et al. 2001].  T h u s failure at t h e bone-implant  rather t h a n failure o f t h e implant results in s u b s i d e n c e  of interbody fusion  device.  Local compressive strength of t h e endplate decreases with vertebral B M D [Grant et al. 2002], a n d b o t h p a r a m e t e r s p l a y a n i m p o r t a n t role in p r e v e n t i n g i n t e r b o d y d e v i c e s u b s i d e n c e .  BMD  w a s h i g h l y c o r r e l a t e d w i t h f a i l u r e l o a d s o f i n t e r b o d y c a g e s i n t h e l u m b a r s p i n e i n a n in study [Jost et al. 1998].  vitro  O t h e r in vitro s t u d i e s a l s o f o u n d s t r o n g c o r r e l a t i o n b e t w e e n B M D a n d  failure loads of interbody d e v i c e s [Closkey et al. 1 9 9 3 , H a s e g a w a et al. 2 0 0 1 , L i m et al. 2 0 0 1 , Steffen et al. 2 0 0 0 , T s a n t r i z o s et al. 2 0 0 0 ] . Subsequent challenge  faced  failure  at t h e adjacent  by the spine  surgeon  segments  in t h e e l d e r l y  following patients  spinal  fixation  is  [ E c k et al. 1999].  another Adjacent  s e g m e n t s normally refer to o n e or t w o levels cephalad or caudal t o t h e instrumented  levels.  T h e t e r m s 'adjacent level effects' a n d 'adjacent s e g m e n t b i o m e c h a n i c s ' h a v e b e e n u s e d t o refer to t h e b i o m e c h a n i c a l c h a n g e s o b s e r v e d at t h e s e g m e n t s adjacent to a spinal fusion [ H a et al. 1993,  Kumar  et al. 2 0 0 1 , Panjabi  2002,  Panjabi  a n d Goel 2005], while  'adjacent  segment  degeneration' a n d 'adjacent s e g m e n t disease' have b e e n used to refer to t h e radiographically o r 29  Chapter 1  Introduction  clinically observed degenerative changes and related new symptoms [Eck et al. 1999, Park et al. 2004, Hilibrand and Robbins 2004]. Accelerated failure of adjacent levels has been reported following rigid spinal fixation [Lee 1988].  This was attributed to both the degenerative  progression in the elderly and to the increased demands placed on these adjacent levels following spinal fixation [Hilibrand and Robbins 2004].  Biomechanical changes such as  increased load, mobility and intradiscal pressure in adjacent segments after fusion had been shown in biomechanical and radiographic studies and outlined in recent review articles [Eck et al. 1999, Park et al. 2004]. It is the interest of the biomechanical engineer to reduce these biomechanical changes which contribute towards adjacent level effects. Pedicle screw loosening and interbody device subsidence could result in surgical failure and would require revision surgery. Accelerated failure at the adjacent segment could also necessitate revision surgery to address the new affected levels.  Further details on pedicle  screw loosening are discussed in Chapter 3, further details on interbody device subsidence are discussed in Chapters 4 and 5, further discussion on increasing overall construct stiffness of a spinal fixation are in Chapter 6, and further discussion on adjacent level effects are in Chapter 7. The rate of revision surgery following spinal fusion was reported at about 10% [Hu et al. 1997a, Diwan et al. 2003a], 13% [Greiner-Perth et al. 2004] and 18% [Malter et al. 1998]. Pseudoarthrosis at the initial fusion level was reportedly a main indication for revision surgery. Adjacent level failures, such as juxtafusion kyphosis, scoliosis or degeneration, or a combination of these problems, were also cause for revision surgery [Diwan et al. 2003a, 2003b]. Augmentation to existing fusion and extension of the fusion to adjacent segment(s) were common indications for revisions. Revision surgery in the osteoporotic spine poses a higher level of complexity than the initial surgery [Dvorak and Fisher 1999]. The presence of fibrous scar tissue, altered anatomy and poor vascularity can predispose patients to higher rates of complications.  Increased length of surgery, more extensive exposure, higher rates of dural  tears and infections complicate revision spine surgery [Diwan et al. 2003a, 2003b]. In order to reduce the number of revision surgeries in frail osteoporotic patients, an effective surgical technique needs to be identified such that the problems of implant loosening and adjacent level effect are addressed at the same time.  No technique currently available is capable of  addressing both implant loosening and adjacent level effects in the osteoporotic spine.  30  Chapter 1  1.4.4  Current  Introduction  Surgical  Strategies  There are currently two concepts to address adjacent level effects. In the first concept, internal fixation is used across multiple segments to reduce motion across all levels that could potentially develop adjacent level effects [Hu 1997b, Inaoka et al. 2002].  This concept is  currently used to prevent adjacent segment disease. The other concept is to maintain motion of the spine with spinal arthroplasty rather than spinal fusion [Robertson et al. 2005, Regan 2005]. The use of spinal arthroplasty is however a recent concept. Further discussions of this concept for the elderly spine are in the next section. Current surgical strategies adopted by spine surgeons to address device loosening can be broadly categorized into two groups. To prevent pedicle screw loosening, one technique is to increase the number of anchor points [Hu 1997b, Inaoka et al. 2002]. Another technique that has become popular is the cement augmentation of screws, which was shown to biomechanically improve fixation strength at the screw-bone interface [Cameron et al. 1975, Zindrick et al. 1986, Wittenberg et al. 1993, Lotz et al. 1997, Moore et al. 1997, Yerby et al. 1998,  Ignatius et al. 2001, Bai et al. 2001, Sarzier et al. 2002, Schultheiss et al. 2004, Cook et  al. 2004, Renner et al. 2004]. In recent years, the use of cements such as polymethylmethacrylate (PMMA) and calcium phosphate cement are increasingly used in osteoporotic patients. The use of cement in osteoporotic patients reportedly improved fixation in the osteoporotic hip [Urist 1975], knee [Larsson and Bauer 2002] and shoulder [Friedman 1998]. In the spine, cement is used in the procedures of kyphoplasty [Rhyne et al. 2004] and vertebroplasty [Barr et al. 2000]. Patients with osteoporotic compression fractures of the spine are injected with cement percutaneously with much clinical success.  A recent clinical review also reported the use of cement in  augmenting screw fixation and in preventing interbody device subsidence [Heini 2005]. The long term side effects of cement, if any, are not known. A known complication with cement use is the extrusion of cement into the foramina or blood vessels during insertion. Cement is thus contraindicated when there is a breach of the pedicle wall during screw insertion. Real-time imaging during cement injection greatly improves the safety of injecting cement into the spine. The exothermic chemical reaction during the in situ curing of the cement could also cause damage to neighbouring cells. Calcium phosphate cement is used recently in place of PMMA because of the lower curing temperature and also because calcium phosphate is the basic mineral in bone and might be more biocompatible.  31  Chapter 1  Introduction  1.4.5 New Surgical Strategies Research directions that could become part of future surgical techniques to improve fixation of the osteoporotic spine include the use of bone growth factors and motion preservation devices. Bone morphogenetic protein (BMP) had been shown to promote bone growth in animal models [Lovell et al. 1989, Alden et al. 2002]. In humans, the use of recombinant human BMP type 2 (rhBMP-2) with threaded fusion cages was also shown to result in consistent osteoinduction in single level lumbar fusion [Boden et al. 2000].  Other clinical trials with  humans also reported successful osteoinduction with fusion cages [Burkus et al. 2002, Haid et al. 2004, Boden et al. 2002]. The use of rhBMP-2 to fill the fusion cages resulted in reduced pain for the patient as it replaced the need to harvest bone grafts from the iliac crest. The use of BMP in spinal fusion has not been carried out in osteoporotic patients to date and future studies in this area would be important. The administration of rhBMP-2 to osteopenic mice was found to prevent bone loss [Turgeman et al. 2002] and thus its potential in osteoporotic patients to promote bone growth is encouraging. The use of BMP might, in the future, be a solution to prevent implant loosening in the osteoporotic patients. Spinal arthroplasty, as mentioned earlier, has been recently reported to prevent adjacent segment disease [Robertson et al. 2005]. Spinal arthroplasty can be considered to be a new surgical technique as its use has not been widespread. The Charite Artificial Disc is indicated for the treatment of degenerative disc disease and has been in use for the longest time as compared to all other spinal arthroplasty devices.  In a recent 17 years follow-up on the  performance of the Charite, the authors reported an 11 % reoperation rate in 53 patients [Putzier et al. 2005]. A prospective randomized control trial of the Charite in the US demonstrated equivalent quantitative clinical outcome as compared with the BAK fusion device and the authors recommended it as "a safe and effective alternative to fusion for the surgical treatment of symptomatic disc degeneration in properly indicated patients" [Blumenthal et al. 2005]. The use of motion preservation devices, while not currently indicated in the elderly population, might become a possible solution to prevent adjacent level disease in the near future.  14.6  Problem with Device Evaluation Current spinal devices are not designed for the osteoporotic patient group.  Implant  manufacturers design their products to meet the requirements stipulated in test standards, so as to gain approval for marketing their products.  Test standards usually include testing the  properties of the implants perse, such as static failure strength, fatigue life, and biocompatibility 32  Chapter  Introduction  1  issues.  S t a n d a r d for  O n e s u c h t e s t s t a n d a r d is t h e A S T M  F1717)  the test of p e d i c l e s c r e w s  2001].  to d e t e r m i n e failure s t r e n g t h a n d f a t i g u e l i f e s p a n [ A S T M  (ASTM  T h e rationale for s u c h a  t e s t is t h a t p e d i c l e s c r e w s s h o u l d h a v e s u f f i c i e n t l y l o n g f a t i g u e l i f e s p a n in c l i n i c a l u s e .  In o r d e r  to attain t h e r e q u i r e d f a t i g u e l i f e s p a n a n d f a i l u r e s t r e n g t h , p e d i c l e s c r e w s s y s t e m s a r e t y p i c a l l y rigid a s  c o m p a r e d to t h e  neighbouring  bone  stiffness  results  in  of n e i g h b o u r i n g b o n e .  localized  trabecular  bone  H i g h s t r e s s e s at t h e failure.  Implant  interface  loosening  with  is  thus  c o m m o n in t h e o s t e o p o r o t i c p a t i e n t g r o u p . Current implant  spinal  on the  device  interface  test  with  standards  do  not  include  For  neighbouring bone.  the  example,  investigation no  of  equivalent  f o c u s e s o n t h e effect o f rigid i m p l a n t s y s t e m s o n t h e f a i l u r e o f s u r r o u n d i n g b o n e .  rigid  test  spinal  standard  Further detail  o n t h i s i s s u e o f b o n e d e n s i t y is d i s c u s s e d in C h a p t e r 2. in 1997,  In a c o n s e n s u s m e e t i n g o n t r e a t m e n t o f o s t e o p o r o s i s osteoporotic  patients  was  considered  not a d v i s a b l e ,  w o u l d a l l o w for rigid fixation a n d b o n e [Andersson cement  et  in r e i n f o r c e m e n t  The of  healing without affecting  success  osteoporotic  of  for n e w  adjacent levels w a s  polymethylmethacrylate  vertebral fractures  instrumentation  (PMMA)  in m o r e  recent  and  that  identified  other  bone  had  been  years  2005].  p r o m i s i n g [Heini  1.5  1997].  al.  and a need  s p i n a l i n s t r u m e n t a t i o n in  Motivation Spinal  elderly.  degeneration  and  osteoporosis  can  Deformity, fracture a n d degeneration  negatively  of the  d i s a b i l i t y w h i c h r e s u l t s in a d i m i n i s h e d q u a l i t y o f life. necessary treatment  when  there  is t o a c h i e v e  is  neurological  deficit  or  spine  affect  can  the  quality of  produce substantial  surgeon  study  is  chronic  e a r l y a m b u l a t i o n a n d p a i n relief.  motivated  by  the  desire  to  in t h e i r t r e a t m e n t o f t h e e l d e r l y p a t i e n t s  pain  pain.  The  main  Unfortunately, the  provide  biomechanical  goal  of  interfaces subsidence  should or  be  and subsequent  sufficiently  loosening.  strong  Secondly,  to the  withstand surgical  the  loads  input  intervention  Firstly, the  transmitted should  not  and often  surgical  spine.  using spinal fusion techniques.  adjacent level d i s e a s e .  the  i s s u e s of i m p l a n t  to  the  and  spine  Specifically,  t h e o b j e c t i v e o f t h i s r e s e a r c h is t o e n a b l e t h e s p i n e s u r g e o n t o p r e v e n t p e d i c l e s c r e w interbody device subsidence,  of  S u r g i c a l i n t e r v e n t i o n in t h e s p i n e is  l o o s e n i n g a n d a d j a c e n t l e v e l e f f e c t s in t h e s p i n e r e m a i n a p r o b l e m in t h e a g i n g This  life  loosening,  implant-bone not  result  result in  in  further  c o m p l i c a t i o n s s u c h a s a d j a c e n t level effects with the n e c e s s i t y to follow u p with a n o t h e r s u r g e r y . The  s u r g e r y s h o u l d ideally a c h i e v e  satisfactory  w o u l d outlast the l i f e s p a n of the o s t e o p o r o t i c  r e s u l t s w i t h t h e initial s u r g i c a l i n t e r v e n t i o n t h a t  patient.  33  Chapter 1  1.6  Introduction  Objectives The  devices  main  objective  to a d d r e s s  aging spine.  of this  both the  thesis  is t o  provide insights  p r o b l e m s of i m p l a n t l o o s e n i n g  T h i s is a c c o m p l i s h e d w i t h t h e f o l l o w i n g s p e c i f i c  and  directions  and adjacent  in t h e  use  level effects  in  of the  goals:  i)  T o h i g h l i g h t t h e i m p o r t a n c e o f v e r t e b r a l b o n e d e n s i t y in p e d i c l e s c r e w f i x a t i o n ( C h a p t e r 2 ) .  ii)  To  compare  cement,  three  pedicle  laminar hooks  screws  augmentation  a n d sublaminar wires,  as  techniques, a mean  using  calcium  phosphate  of preventing implant  loosening  ( C h a p t e r 3). iii)  To  compare  subsidence iv)  two  interbody  (Chapter  sizes  and  three  shapes  as  a  mean  of  preventing  device  4).  T o d e t e r m i n e the effects of s c r e w fixation a n d c e m e n t a u g m e n t a t i o n of s c r e w s o n i n t e r b o d y d e v i c e s u b s i d e n c e ( C h a p t e r 5).  v)  T o c o m p a r e two current surgical techniques, or  using  posterior  rod extension,  as  using c e m e n t a u g m e n t a t i o n of pedicle  means  of  achieving  adequate  screws  stabilization  in  the  o s t e o p o r o t i c p a t i e n t s ( C h a p t e r 6). vi)  To  compare  three  without extension  posterior rods, as  rod extension  means  conditions,  using  of preventing a d v e r s e  rigid  rods,  accelerated  flexible  rods  degeneration  at  or the  a d j a c e n t l e v e l s ( C h a p t e r 7).  1.7  Scope T h i s t h e s i s is l i m i t e d t o t h e s t u d y o f t h e t h o r a c o l u m b a r s p i n e u s i n g  tests.  biomechanical  T h e first t w o s e c t i o n s o n p e d i c l e s c r e w s w e r e c a r r i e d o u t u s i n g s i n g l e v e r t e b r a l  with s i m u l a t e d p h y s i o l o g i c a l l o a d s a p p l i e d to t h e s c r e w h e a d s . of  in vitro  pedicle s c r e w testing  a d o p t e d by the A S T M  Chapter 2 addresses  a n d the effects of B M D .  Chapter 3  bodies  the  issue  compares  t h r e e c u r r e n t clinical t e c h n i q u e s of i m p r o v i n g p e d i c l e s c r e w fixation: c a l c i u m p h o s p h a t e  cement,  laminar h o o k s a n d s u b l a m i n a r wire augmentation. The  next two c h a p t e r s o n interbody d e v i c e s u b s i d e n c e w e r e also carried out using  vertebral bodies.  Superior endplates  axial c o m p r e s s i v e  load.  of  interbody devices  of t h o r a c o l u m b a r v e r t e b r a e w e r e  l o a d e d to failure u n d e r  C h a p t e r 4 p r o v i d e s s u g g e s t i o n s to i m p r o v e the b o n e - i m p l a n t  with different c a g e s h a p e  and sizes.  single  interface  C h a p t e r 5 provides further i d e a s  to  r e d u c e i n t e r b o d y d e v i c e s u b s i d e n c e with vertebral s c r e w s a n d c e m e n t a u g m e n t a t i o n of s c r e w s . In t h e  last two  chapters,  multisegmental  thoracolumbar spines  ( T 9 - L 3 or T 8 - L 2 )  were  used.  C h a p t e r 6 a d d r e s s e s t h e i s s u e o f m o t i o n a f t e r f i x a t i o n o f a c o r p e c t o m y in t h e  osteoporotic  spine  and  of  compares  between  cement  augmentation  of  screws  and  extension  posterior  34  Chapter  1  Introduction  instrumentation.  Chapter 7 compares adjacent level motion and strains following cement  augmentation and extension of posterior instrumentation. 1.8  References  [I]  A d a m s M A , M c N a l l y D S , D o l a n P. ' S t r e s s ' d i s t r i b u t i o n s inside intervertebral d i s c s . a g e a n d d e g e n e r a t i o n . J o u r n a l of B o n e a n d Joint S u r g e r y . 1 9 9 6 ; 7 8 B . 9 6 5 - 9 7 2 .  T h e e f f e c t s of  [2]  A d a m s M, B o g d u k N, Burton K, D o l a n P. T h e B i o m e c h a n i c s of B a c k Pain. Churchill L i v i n g s t o n e , Edinburgh, 2002.  [3]  A l d e n T D , V a r a d y P, K a l l m e s D F , J a n e J A Jr, H e l m G A . therapy. Spine 2002; 27:S87-S93.  [4]  A n d e r s o n G F , H u s s e y P S . P o p u l a t i o n a g i n g : a c o m p a r i s o n a m o n g industrialized c o u n t r i e s . Affairs 2 0 0 0 ; 1 9 : 1 9 1 - 2 0 3 .  [5]  A n d e r s s o n G B J , B o s t r o m M P G , E y r e D R , G l a s e r DL, Hu S S , L a n e J M , M e l t o n LJ III, M y e r s E R , S e e g e r LL, W e i n s t e i n J N . C o n s e n s u s s u m m a r y o n t h e d i a g n o s i s a n d t r e a t m e n t of o s t e o p o r o s i s . Spine 1997; 22:63S-65S.  [6]  A n d e r s s o n S, R o d r i g u e s M, O l e r u d C. O d o n t o i d f r a c t u r e s : high c o m p l i c a t i o n rate a s s o c i a t e d w i t h anterior s c r e w fixation in t h e elderly. E u r o p e a n S p i n e J o u r n a l 2 0 0 0 ; 9 : 5 6 - 6 0 .  [7]  A m e r i c a n S o c i e t y for T e s t i n g a n d Materials. A S T M F 1 7 1 7 . S t a n d a r d t e s t m e t h o d s for spinal implant c o n s t r u c t s in a v e r t e b r e c t o m y m o d e l . A S T M International 2 0 0 1 .  [8]  A r n o l d C M , B u s c h A J , S c h a c h t e r C L , H a r r i s o n L, O l s z y n s k i W . T h e relationship of intrinsic fall risk f a c t o r s to a r e c e n t history of falling in older w o m e n w i t h o s t e o p o r o s i s . J o u r n a l of O r t h o p a e d i c a n d Sports Physical Therapy 2005; 35:452-460.  [9]  A s p d e n R M . B i o m e c h a n i c s of o s t e o p o r o t i c b o n e . In: O r t h o p a e d i c I s s u e s in O s t e o p o r o s i s . (Ed.) C R C P r e s s , B o c a R a t o n , 2 0 0 3 ; 3 9 - 5 4 .  [10]  B a b a t L B , M c L a i n R F , B i n g a m a n W , Kalfas I, Y o u n g P, R u f o - S m i t h C. S p i n a l s u r g e r y in patients w i t h p a r k i n s o n ' s d i s e a s e : c o n s t r u c t failure a n d p r o g r e s s i v e d e f o r m i t y . S p i n e 2 0 0 4 ; 2 9 : 2 0 0 6 - 2 0 1 2 .  [II]  Bai B, K u m m e r F J , S p i v a k J. A u g m e n t a t i o n of anterior v e r t e b r a l b o d y s c r e w fixation by a n injectable, b i o d e g r a d a b l e c a l c i u m p h o s p h a t e b o n e s u b s t i t u t e . S p i n e 2 0 0 1 ; 2 6 : 2 6 7 9 - 2 6 8 3 .  [12]  Barger-Lux MJ, Recker RR. B o n e m i c r o s t r u c t u r e in o s t e o p o r o s i s : transilial h i s t o m o r p h o m e t r y . T o p i c s in M a g n e t i c R e s o n a n c e I m a g i n g 2 0 0 2 : 1 3 : 2 9 7 - 3 0 6 .  [13]  Barr S I , M c K a y HA. Nutrition, e x e r c i s e , a n d b o n e s t a t u s in y o u t h . Nutrition 1998; 8 : 1 2 4 - 1 4 2 .  [14]  Barr J D , Barr M S , L e m l e y T J , M c C a n n R M . P e r c u t a n e o u s v e r t e b r o p l a s t y for pain relief a n d spinal stabilization. S p i n e 2 0 0 0 ; 2 5 : 9 2 3 - 9 2 8 .  [15]  Bartl R, Frisch B. O s t e o p o r o s i s : D i a g n o s i s , P r e v e n t i o n , T h e o r y . 72.  [16]  Bauer TW, Muschler GF. B o n e graft materials: a n o v e r v i e w of t h e basic s c i e n c e . Orthopaedics and Related Research 2000; 371:10-27.  [17]  B a s s e w i t z H, H e r k o w i t z H. L u m b a r s t e n o s i s w i t h s p o n d y l o l i s t h e s i s : c u r r e n t c o n c e p t s of surgical t r e a t m e n t . Clinical O r t h o p a e d i c s a n d R e l a t e d R e s e a r c h 2 0 0 1 ; 3 8 4 : 5 4 - 6 0 .  [18]  B e n n e t t G J , S e r h a n HA, Sorini P M , Willis B H . Spine 1997; 22:1448-1453.  [19]  B e n o i s t M. N a t u r a l history of the a g i n g s p i n e . E u r o p e a n S p i n e J o u r n a l 2 0 0 3 ; 1 2 : S 8 6 - S 8 9 .  B o n e m o r p h o g e n e t i c protein  gene Health  An YH  biopsy  and  International J o u r n a l of S p o r t  S p r i n g e r - V e r l a g , Berlin, 2 0 0 4 ; 5 8 -  Clinical  A n e x p e r i m e n t a l s t u d y of l u m b a r d e s t a b i l i z a t i o n .  35  Chapter 1  Introduction  [20]  Benzel EC. Biomechanics of Spine Stabilization: Principles and Clinical Practice. New York, 1995.  McGraw-Hill,  [21]  Berlemann U, Jeszenszky DJ, Buhler DW, Harms J. Facet joint remodelling in degenerative spondylolisthesis: an investigation of joint orientation and tropism. European Spine Journal 1998; 7:376-380.  [22]  Bjarke CF, Stender HE, Laursen M, Thomsen K, Bunger CE. Long-term functional outcome of pedicle screw instrumentation as a support for posterolateral spinal fusion: randomized clinical study with a 5-year follow-up. Spine 2002; 27:1269-1277.  [23]  Black DM, Arden NK, Palermo L, Pearson J, Cummings SR, Study of osteoporotic fractures research group. Prevalent vertebral deformities predict hip fractures and new vertebral deformities but not wrist fractures. Journal of Bone and Mineral Research 1999; 14:821-828.  [24]  Blumenthal S, McAfee PC, Guyer RD, Hochschuler SH, Geisler FH, Holt RT, Garcia R Jr, Regan JJ, Ohnmeiss DD. A prospective, randomized, multicenter Food and Drug Administration investigational device exemptions study of lumbar total disc replacement with CHARITE artificial disc versus lumbar fusion: part I: evaluation of clinical outcomes. Spine 2005; 30:1565-1575.  [25]  Boden SD, Zdeblick TA, Sandhu HS, Heim SE. The use of rhBMP-2 in interbody fusion cages. Spine 2000; 25:376-381.  [26]  Boden SD, Kang J, Sandhu H, Heller JG. Use of recombinant human bone morphogenetic protein2 to achieve posterolateral lumbar spine fusion in humans: a prospective, randomized clinical pilot trial. Spine 2002; 27:2662-2673.  [27]  Brown JP, Josse RG, Scientific Advisory Council of the Osteoporosis Society of Canada. 2002 clinical practice guidelines for the diagnosis and management of osteoporosis in Canada. Canadian Medical Association Journal 2002a; 167(10 suppl):S1-S34.  [28]  Brown MD, Holmes DC, Heiner AD. stiffness. Spine 2002b; 27:918-922.  [29]  Buckwalter JA. Spine update: aging and degeneration of the human intervertebral disc. 1995; 20:1307-1314.  [30]  Burger H, Van Daele PLA, Grashuis K, Hofman A, Grobbee DE, Schutte HE, Birkenhager JC, Pols HAP. Vertebral deformities and functional impairment in men and women. Journal of Bone and Mineral Research 1997; 12:152-157.  [31]  Burkus JK, Transfeldt EE, Kitchel SH, Watkins RG, Balderston RA. Clinical and radiographic outcomes of anterior lumbar interbody fusion using recombinant human bone morphogenetic protein-2. Spine 2002; 27:2396-2408.  [32]  Butler D, Trafimow JH, Andersson GBJ, McNeill TW, Huckman MS. Disc degenerate before facets. Spine 1990; 15:111-113.  [33]  Cameron HU, Jacob R, Macnab I, Pilliar RM. Use of polymethylmethacrylate to enhance screw fixation in bone. Journal of Bone and Joint Surgery 1975; 57A:655-656.  [34]  Cholewicki J, Panjabi MM, Khachatryan A. Stabilizing function of trunk flexor-extensor muscles around a neutral spine posture. Spine 1997; 22:2207-2212.  [35]  Cinotti G, De Santis P, Nofroni I, Postacchini F. Stenosis of lumbar intervertebral foramen: anatomic study on predisposing factors. Spine 2002; 27:223-229.  [36]  Closkey RF, Parsons JR, Lee CK, Blacksin MF, Zimmerman MC. Mechanics of interbody spinal fusion. Analysis of critical bone graft area. Spine 1993; 18:1011-1015.  [37]  Coe JD, Warden KE, Herzig MA, McAfee PC. Influence of bone mineral density on the fixation of thoracolumbar implants: A comparative study of transpedicular screws, laminar hooks and spinous process wires. Spine 1990; 15:902-907.  Measurement of cadaver lumbar spine motion segment Spine  36  Chapter 1  Introduction  [38]  Consensus Development Conference: Diagnosis, prophylaxis, and treatment of osteoporosis. American Journal of Medicine 1993; 94:646-650.  [39]  Cook SD, Salkeld SL, Stanley T, Faciane A, Miller SD. Biomechanical study of pedicle screw fixation in severely osteoporotic bone. Spine Journal 2004; 4:402-408.  [40]  Cooper C, O'Neill T, Silman A. The epidemiology of vertebral fractures. European vertebral osteoporosis study group. Bone 1993; 14(Suppl 1):S89-S97.  [41]  Cortet B, Roches E, Logier R, Houvenagel E, Gaydier-Souquieres G, Puisieux F, Delcambre B. Evaluation of spinal curvatures after a recent osteoporotic vertebral fracture. Joint, Bone, Spine: revue du rhumatisme 2002; 69:201-208.  [42]  Crisco JJ III. The biomechanical stability of the human lumbar spine: experimental and theoretical investigations. PhD dissertation, Yale University, 1989.  [43]  Daggfeldt K, Thorstensson A. The role of intra-abdominal pressure in spinal unloading. Journal of Biomechanics 1997; 30:1149-1155.  [44]  Dawson E, Bernbeck J. The surgical treatment of low back pain. Rehabilitation Clinics of North America 1998; 9:489-495.  [45]  De Smet AA, Robinson RG, Johnson BE, Lukert BP. Spinal compression fractures in osteoporotic women: patterns and relationship to hyperkyphosis. Radiology 1988; 166:497-500.  [46]  DeWald RL. Severe spondylolisthesis: Introduction. In: Severe Spondylolisthesis. Harms J, Sturz H (Eds.), Steinkopff, Darmstadt, 2002; 1-12.  [47]  Diacinti D, Acca M, D'Erasmo E, Tomei E, Mazzuoli GF. Aging changes in vertebral morphometry. Calcified Tissue International 1995; 57:426-429.  [48]  Diwan AD, Parvartaneni H, Cammisa F. Failed degenerative lumbar spine surgery. Clinics of North America 2003a; 34:309-324.  [49]  Diwan AD, Sandhu HS, Khan SN, Parvataneni HK, Cammisa FP. Lumbar spine: Degenerative disease. In: The Failed Spine. Boden SD and Bohlman HH (Eds.), Lippincott Williams & Wilkins, Philadelphia, 2003b; 203-219.  [50]  Doherty TJ. Invited Review: aging and sarcopenia. Journal of Applied Physiology 2003; 95:17171727.  [51]  Duan Y, Seeman E, Turner CH. The biomechanical basis of vertebral body fragility in men and women. Journal of Bone and Mineral Research 2001; 16:2276-2283.  [52]  Duan Y, Wang XF, Evans A, Seeman E. Structural and biomechanical basis of racial and sex differences in vertebral fragility in Chinese and Caucasians. Bone 2005; 36:987-998.  [53]  Dvorak MF, Fisher CG. Revision spine surgery in the presence of osteoporosis. In: Revision Spine Surgery. Margulies JY, Aebi M, Farcy JPC (Eds.) Mosby, St Louis, 1999, 646-653.  [54]  Eck JC, Humphreys SC, Hodges SD. Adjacent-segment degeneration after lumbar fusion: a review of clinical, biomechanical, and radiologic studies. American Journal of Orthopedics 1999; 28:336-340.  [55]  Engelke K, Kemmler W, Lauber D, Beeskow C, Pintag R, Kalender WA. Exercise maintains bone density at spine and hip EFOPS: a 3-year longitudinal study in early postmenopausal women. Osteoporosis International 2006; 17:133-142.  [56]  Ensrud KE, Thompson DE, Cauley JA, Nevitt MC, Kado DM, Hochberg MC, Santora AC, Black DM. Prevalent vertebral deformities predict mortality and hospitalization in older women with low bone mass. Journal of the American Geriatrics Society 2000; 48:338-339.  [57]  Evans SF, Nicholson PH, Haddaway MJ, Davie MW. Vertebral morphometry in women aged 50-81 years. Bone and Mineral 1993; 21:29-40.  Physical Medicine and  Orthopedic  37  Chapter 1  Introduction  [58]  Ferguson SJ, Steffen T. 12(Suppl 2): S97-S103.  Biomechanics of the aging spine.  European Spine Journal 2003;  [59]  Fisher CG, Keynan O, Boyd MC, Dvorak MF. The surgical management of primary tumors of the spine. Initial results of an ongoing prospective cohort study. Spine 2005; 30:1899-1908.  [60]  Friedman RJ. Humeral technique in total shoulder arthroplasty. America 1998; 29:393-402.  [61]  Fujiwara A, Lim TH, An HS, Tanaka N, Jeon CH, Andersson GBJ, Haughton VM. The effect of disc degeneration and facet joint osteoarthritis on the segmental flexibility of the lumbar spine. Spine 2000; 25:3036-3044.  [62]  Genant HK, W u CY, van Kuijk C, Nevitt MC. Vertebral fracture assessment using a semiquantitative technique. Journal of Bone and Mineral Research 1993; 8:1137-1148.  [63]  Gercek E, Arlet V, Delisle J, Marchesi D. Subsidence of stand-alone cervical cages in anterior interbody fusion: warning. European Spine Journal 2003; 12:513-516.  [64]  Glassman SD, Alegre GM. Adult spinal deformity in the osteoporotic spine: options and pitfalls. Instructional Course Lectures 2003; 52:579-588.  [65]  Goeree R, O'Brien B, Pettitt D, Cuddy L, Ferraz M, Adachi J. An assessment of the burden of illness due to osteoporosis in Canada. Journal of the Society of Obstetricians and Gynaecologists of Canada 1996; 18 (7 Suppl): 15-24.  [66]  Grant JP, Oxland TR, Dvorak MF, Fisher CG. The effects of bone density and disc degeneration on the structural property distributions in the lower lumbar vertebral endplates. Journal of Orthopaedic Research 2002; 20:1115-1120.  [67]  Greiner-Perth R, Boehm H, Allam Y, Elsaghir H, Franke J. Reoperation rate after instrumented posterior lumbar interbody fusion: A report on 1680 cases. Spine 2004; 29:2516-2520.  [68]  Ha KY, Schendel MJ, Lewis JL, Ogilvie JW. Effect of immobilization and configuration on lumbar adjacent-segment biomechanics. Journal of Spinal Disorder 1993; 6:99-105.  [69]  Haid RW Jr, Brance CL Jr, Alexander JT, Burkus JK. Posterior lumbar interbody fusion using recombinant human bone morphogenetic protein type 2 with cylindrical interbody cages. Spine Journal 2004; 4:527-538.  [70]  Hall SE, Criddle RA, Comito TL, Prince RL. A case-control study of quality of life and functional impairment in women with long-standing vertebral osteoporotic fracture. Osteoporosis International 1999; 9:508-515.  [71]  Halvorson TL, Kelley LA, Thomas KA, Whitecloud TS, Cook SD. Effects of bone mineral density on pedicle screw fixation. Spine 1994; 19:2415-2420.  [72]  Harada A, Okuizumi H, Miyagi N, Genda E. Correlation between bone mineral density and intervertebral disc degeneration. Spine 1998; 23:857-862.  [73]  Hartvigsen J, Christensen K, Frederiksen H. Back pain remains a common symptom in old age. A population-based study of 4486 Danish twins aged 70-102. European Spine Journal 2003; 12:528-534.  [74]  Hasegawa K, Abe M, Washio T, Hara T. An experimental study on the interface strength between titanium mesh cage and vertebra in reference to vertebral bone mineral density. Spine 2001; 26:957-963.  [75]  Healey JH, Lane JM. Structural scoliosis in osteoporotic women. Related Research 1985; 195:216-223.  [76]  Heini PF. The current treatment - a survey of osteoporotic fracture treatment. Osteoporotic spine fractures: the spine surgeon's perspective. Osteoporosis International 2005; Suppl2:S85-S92.  Orthopedic Clinics of North  Clinical Orthopaedics and  38  Chapter 1  Introduction  [77]  Herkowitz HN, Kurz LT. Degenerative lumbar spondylolisthesis with spinal stenosis. Bone and Joint Surgery 1991; 73A:802-808.  Journal of  [78]  Hilibrand AS, Robbins M. Adjacent segment degeneration and adjacent segment disease: the consequences of spinal fusion? Spine Journal 2004; 4:190S-194S.  [79]  Horst M, Brinckmann P. 1980 Volvo award in biomechanics. Measurement of the distribution of axial stress on the end-plate of the vertebral body. Spine 1981; 6:217-232.  [80]  Hu RW, Jaglal S, Axcell T, Anderson G. A population-based study of reoperations after back surgery. Spine 1997a; 22:2265-2270.  [81]  Hu SS. Internal fixation in the osteoporotic spine. Spine 1997b; 22:43S-48S.  [82]  Ignatius AA, Augat P, Ohnmacht M, Pokinskyj P, Kock HJ, Claes LE. A new bioresorbable polymer for screw augmentation in the osteosynthesis of osteoporotic cancellous bone: a biomechanical evaluation. Journal of Biomedical Materials Research. (Part B, Applied Biomaterials) 2001; 58:254-60.  [83]  lida T, Abumi K, Kotani Y, Kaneda K. Effects of aging and spinal degeneration on mechanical properties of lumbar supraspinous and interspinous ligaments. Spine Journal 2002; 2:95-100.  [84]  Inaoka M, Tada K, Yonenobu K. Problems of posterior lumbar interbody fusion (PLIF) for the rheumatoid spondylitis of the lumbar spine. Archives of Orthopedic and Trauma Surgery 2002; 122:73-79.  [85]  Inoue N, Chao EYS. Biomechanics of osteoporotic bone and fractures. In: Internal Fixation in Osteoporotic Bone. An YH (Ed.), Thieme, New York, 2002; 9-21.  [86]  Jones G, Nguyen T, Sambrook PN, Kelly PJ, Eisman JA. A longitudinal study of the effect of spinal degenerative disease on bone density in the elderly. Journal of Rheumatology 1995; 22:932-936.  [87]  Jost B, Cripton PA, Lund T, Oxland TR, Lippuner K, Jaeger P, Nolte LP. Compressive strength of interbody cages in the lumbar spine: the effect of cage shape, posterior instrumentation and bone density. European Spine Journal 1998, 7:132-141.  [88]  Kaigle A, Ekstrom L, Holm S, Rostedt M, Hansson T. In vivo dynamic stiffness of the porcine lumbar spine exposed to cyclic loading: influence of load and degeneration. Journal of Spinal Disorders 1998; 11:65-70.  [89]  Kanis JA, Melton LJ III, Christiansen C, Johnston CC, Khaltaev N. The diagnosis of osteoporosis. Journal of Bone and Mineral Research 1994; 9:1137-1141.  [90]  Kanis JA, Torgerson D, Cooper C. Comparison of the European and USA practice guidelines for osteoporosis. Trends in Endocrinology and Metabolism: TEM 2000; 11:28-32.  [91]  Kannus P, Uusi-Rasi K, Palvanen M, Parkkari J. Non-pharmacological means to prevent fractures among older adults. Annals of Medicine 2005; 37:303-310.  [92]  Karlsson MK. Physical activity, skeletal health and fractures in a long term perspective. Journal of Musculoskeletal & Neuronal Interactions 2004; 4:12-21.  [93]  Keaveny TM, Yeh OC. Architecture and trabecular bone - Toward an improved understanding of the biomechanical effects of age, sex and osteoporosis. Journal of Musculoskeletal and Neuronal Interactions 2002; 2:205-208.  [94]  Keller TS, Hansson TH, Abram AC, Spengler DM, Panjabi MM. Regional variations in the compressive properties of lumbar vertebral trabeculae: effects of disc degeneration. Spine 1989; 14:1012-1019.  [95]  Khan K, McKay HA, Haapasalo H, Bennell KL, Forwood MR, Kannus P, Wark JD. Does childhood and adolescence provide a unique opportunity for exercise to strengthen the skeleton? Journal of Science and Medicine in Sport 2000; 3:150-164.  39  Chapter 1  Introduction  '96]  Kim SS, Denis F, Lonstein JE, Winter RB. Factors affecting fusion rate in adult spondylolisthesis. Spine 1990; 15:979-984.  97]  Kinoshita H, Tamaki T, Hashimoto T, Kasagi F. Factors influencing lumbar spine bone mineral density assessment by dual-energy x-ray absorptiometry: comparison with lumbar spine radiogram. Journal of Orthopaedic Science 1998; 3:3-9.  98]  Kopec JA, Esdaile JM. Occupational role performance in persons with back pain. Disability and Rehabilitation 1998; 20:373-379.  99]  Kumar MN, Baklanov A, Chopin D. Correlation between sagittal plane changes and adjacent segment degeneration following lumbar spine fusion. European Spine Journal 2001; 10:314-319.  400] Kumaresan S, Yoganandan N, Pintar FA, Maiman DJ, Goel VK. Contribution of disc degeneration to osteophyte formation in the cervical spine: a biomechanical investigation. Journal of Orthopaedic Research 2001; 19:977-984. 401] Kurowski P, Kubo A. The relationship of degeneration of the intervertebral disc to mechanical loading conditions on lumbar vertebrae. Spine 1986; 11:726-731. 102] Larsson S, Bauer TW. Use of injectable calcium phosphate cement for fracture fixation: a review. Clinical Orthopaedics and Related Research 2002; 395:29-32. 103] Lau EMC, Woo J, Chan H, Chan MKF, Griffith J, Chan YH, Leung PC. The health consequences of vertebral deformity in elderly Chinese men and women. Calcified Tissue International 1998; 63:1-4. 104] Law M, Tencer AF, Anderson PA. Caudo-cephalad loading of pedicle screws: mechanisms of loosening and methods of augmentation. Spine 1993; 18:2438-2443. 105] Lee CK. Accelerated degeneration of the segment adjacent to a lumbar fusion. 13:375-377.  Spine 1988;  106] Lenchik L, Rogers LF, Delmas PD, Genant HK. Diagnosis of osteoporotic vertebral fractures: importance of recognition and description by radiologists. American Journal of Roentgenology 2004; 183:949-958. 107] Legrand E, Chappard D, Pascaretti C, Duquenne M, Rondeau C, Simon Y, Rohmer V, Basle M-F, Audran M. Bone mineral density and vertebral fractures in men. Osteoporosis International 1999; 10:265-270. 108] Lim TH, Kwon H, Jeon CH, Kim JG, Sokolowski M, Natarajan R, An HS, Andersson GB. Effect of endplate conditions and bone mineral density on the compressive strength of the graft-endplate interface in anterior cervical spine fusion. Spine 2001; 26:951-956. 109] Lips P. Epidemiology and predictors of fractures associated with osteoporosis. American Journal of Medicine 1997; 103:3S-11S. 110] Lips P, Cooper C, Agnusdei D, Caulin F, Egger P, Johnell O, Kanis JA, Kellingray S, Leplege A, Liberman UA, McCloskey E, Minne H, Reeve J, Reginster JY, Scholz M, Todd C, de Vernejoul MC, Wiklund I. Quality of life in patients with vertebral fractures: validation of the quality of life questionnaire of the European foundation for osteoporosis (QUALEFFO). Osteoporosis International 1999; 10:150-160. 111] Lombardi A, Ross PD. The assessment of bone mass in men. Calcified Tissue International 2001; 69:222-224. 112] Looker AC, Orwoll ES, Johnston CC, Lindsay RL, Wahner HW, Dunn WL, Calvo MS, Harris TB, Heyse SP. Prevalence of low femoral bone density in older US adults from NHANES III. Journal of Bone and Mineral Research 1997; 12:1761-1768.  40  Chapter 1  Introduction  [113] Lord SR, Sambrook PN, Gilbert C, Kelly PJ, Nguyen T, Webster IW, Eisman JA. Postural stability, falls and fractures in the elderly: results from the Dubbo osteoporosis epidemiology study. Medical Journal of Australia 1994; 160:684-685, 688-691. [114] Lorenz M, Patwardhan A, Vanderby R. Load-bearing characteristics of lumbar facets in normal and surgically altered spinal segments. Spine 1983; 8:122-130. [115] Lorrain J, Paiement G, Chevrier N, Lalumiere G, Laflamme GH, Caron P, Fillion A. Population demographics and socioeconomic impact of osteoporotic fractures in Canada. Menopause 2003; 10:228-234. [116] Lotz JC, Hu SS, Chiu DFM, Yu M, Colliou O, Poser RD. Carbonated apatite cement augmentation of pedicle screw fixation in the lumbar spine. Spine 1997; 22:2716-2723. [117] Lotz JC, Colliou OK, Chin JR, Duncan NA, Liebenberg E. Compression-induced degeneration of the intervertebral disc: an in vivo mouse model and finite-element study. Spine 1998; 23:24932506. [118] Lovell TP, Dawson EG, Nilsson OS, Urist MR. Augmentation of spinal fusion with bone morphogenetic protein in dogs. Clinical Orthopaedics and Related Research 1989; 243:266-274. [119] Lunt M, Felsenberg D, Adams J, Benevolenskaya L, Cannata J, Dequeker J, Dodenhof C, Falch JA, Johnell O, Khaw KT, Masaryk P, Pols H, Poor G, Reid D, Scheidt-Nave C, Weber K, Silman AJ, Reeve J. Population-based geographic variations in DXA bone density in Europe: the EVOS study. Osteoporosis International 1997; 7:175-189. [120] Lunt M, Masaryk P, Scheidt-Nave C, Nijs J, Poor G, Pols H, Falch JA, Hammermeister G, Reid DM, Benevolenskaya L, Weber K, Cannata J, O'Neill TW, Felsenberg D, Silman AJ, Reeve J. The effects of lifestyle, dietary dairy intake and diabetes on bone density and vertebral deformity prevalence: the EVOS study. Osteoporosis International 2001; 12:688-698. [121] Lyles KW, Gold DT, Shipp KM, Pieper CF, Martinez S, Mulhausen PL. Association of osteoporotic vertebral compression fractures with impaired functional status. American Journal of Medicine 1993; 94:595-601. [122] Malter AD, McNeney B, Loeser JD, Deyo RA. 5-year reoperation rates after different types of lumbar spine surgery. Spine 1998; 23:814-820. [123] Margulies JY, Payzer A, Nyska M, Neuwirth MG, Floman Y, Robin GC. The relationship between degenerative changes and osteoporosis in the lumbar spine. Clinical Orthopaedics and Related Research 1996; 324:145-152. [124] Martin MD, Boxell CM, Malone DG. Pathophysiology of lumbar disc degeneration: a review of the literature. Neurosurgical Focus 2002; 13:1-6. [125] Matkovic V, Kostial K, Sirmonovic I, Buzina R, Brodarec A, Nordin BEC. Bone status and fracture rates in two regions in Yugoslavia. American Journal of Clinical Nutrition 1979; 32:540-549. [126] Matsumoto M, Chiba K, Nojiri K, Ishikawa M, Toyama Y, Nishikawa Y. Extraforaminal entrapment of the fifth lumbar spinal nerve by osteophytes of the lumbosacral spine: anatomic study and a report of four cases. Spine 2002; 27:E169-E173. [127] Mazess RB, Barden H. Bone density of the spine and femur in adult white females. Tissue International 1999; 65:91-99.  Calcified  [128] McKay HA, Bailey DA, Mirwald RL, Davison KS, Faulkner RA. Peak bone mineral accrual and age at menarche in adolescent girls: a 6-year longitudinal study. Journal of Pediatrics 1998; 133:682687. [129] McNally DS, Shackleford IM, Goodship AE, Mulholland RC. In vivo stress management can predict pain on discography. Spine 1996; 21:2580-2587.  41  Chapter 1  Introduction  130] Mehbod A, Aunoble S, Le Huec JC. Vertebroplasty for osteoporotic spine fracture: prevention and treatment. European Spine Journal 2003; 12(Suppl.2):S155-S162. 131] Melton LJ III, Atkinson EJ, O'Fallon M, Warmer HW, Riggs L. Long-term fracture prediction by bone mineral assessed at different skeletal sites. Journal of Bone and Mineral Research 1993; 8:1227-1233. 132] Melton LJ III. Epidemiology of spinal osteoporosis. Spine 1997; 22:2S-11S. 133] Melton LJ III, Atkinson EJ, O'Connor MK, O'Fallon W M , Riggs BL. Bone density and fracture risk in men. Journal of Bone and Mineral Research 1998; 13:1915-1923. 134] Melton LJ III. The prevalence of osteoporosis: gender and racial comparison. International 2001; 69:179-181.  Calcified Tissue  135] Mimura M, Panjabi MM, Oxland TR, Crisco JJ, Yamamoto I, Vasavada A. Disc degeneration affects the multidirectional flexibility of the lumbar spine. Spine 1994; 19:1371-1380. 436] Miyakoshi N, Itoi E, Murai H, Wakabayashi I, Ito H, Minato T. Inverse relation between osteoporosis and spondylosis in postmenopausal women as evaluated by bone mineral density and semiquantitative scoring of spinal degeneration. Spine 2003; 28:492-495. 137] Moore DC, Maitra RS, Farjo LA, Graziano GP, Goldstein SA. Restoration of pedicle screw fixation with an in situ setting calcium phosphate cement. Spine 1997; 22:1696-1705. 138] Mosekilde L. Age-related changes in vertebral trabecular bone architecture - Assessed by a new method. Bone 1988; 9:247-250. 139] Nachemson A. Lumbar intradiscal pressure. Experimental studies on post-mortem material. Acta Orthopaedica Scandinavica 1960; Suppl43:1-104. 140] Nanjo Y, Morio Y, Nagashima H, Hagino H, Teshima R. Correlation between bone mineral density and intervertebral disk degeneration in pre- and postmenopausal women. Journal of Bone and Mineral Metabolism 2003; 21:22-27. 141] Nathan M, Pope MH, Grobler LJ. Osteophyte formation in the vertebral column: a review of the etiologic factors, Part 1. Contemporary Orthopaedics 1994a; 29:31-37. 142] Nathan M, Pope MH, Grobler LJ. Osteophyte formation in the vertebral column: a review of the etiologic factors, Part 2. Contemporary Orthopaedics 1994b; 29:113-119. 143] National Osteoporosis Foundation (NOF). Osteoporosis: review of the evidence for prevention, diagnosis, and treatment and cost-effectiveness analysis. Osteoporosis International 1998; Suppl4:S1-S88. 144] Neer RM, Arnaud CD, Zanchetta JR, Prince R, Gaich GA, Reginster JY, Hodsman AB, Eriksen EF, Ish-Shalom S, Genant HK, Wang OH, Mitlak BH. Effect of parathyroid hormone (1-34) on fractures and bone mineral density in postmenopausal women with osteoporosis. New England Journal of Medicine 2001; 344:1434-1441. 145] Neumann P, Ekstrom A, Keller TS, Perry L, Hansson TH. Aging, vertebral density, and disc degeneration alter the tensile stress-strain characteristics of the human anterior longitudinal ligament. Journal of Orthopaedic Research 1994; 12:103-112. 146] Nevitt MC, Ettinger B, Black DM, Stone K, Jamal SA, Ensrud K, Segal M, Genant HK, Cummings SR. The association of radiographically detected vertebral fractures with back pain and function: a prospective study. Annals of Internal Medicine 1998; 128:793-800. 147] Nguyen HV, Ludwig S, Gelb D. Osteoporotic vertebral burst fractures with neurologic compromise. Journal of Spinal Disorders and Techniques 2003; 16:10-19. 148] NIH Consensus Development Panel on Osteoporosis Prevention, Diagnosis, and Therapy. Osteoporosis prevention, diagnosis, and therapy. American Medical Association 2001; 285:785795. 42  Chapter 1  Introduction  149] O'Brien JP. Mechanisms of spinal pain. In: Textbook of Pain. Wall PD, Melzack R (Eds.), Churchill Livingstone, Edinburgh, 1984; 240-251. 150] Okawa A, Shinomiya K, Furuya K, Ukegawa Y, Miyazawa A. Inverse relation between vertebral body deformity and intervertebral disk narrowing in lumbar spine of elderly women. Journal of Spinal Disorder 1996; 9:23-31. 151] Okuyama K, Sato K, Abe E, Inaba H, Shimada Y, Murai H. Stability of transpedicle screwing for the osteoporotic spine. An in vitro study of the mechanical stability. Spine 1993; 18:2240-2245. 152] Okuyama K, Abe E, Suzuki T, Tamura Y, Chiba M, Sato K. Influence of bone mineral density of pedicle screw fixation: a study of pedicle screw fixation augmenting posterior lumbar interbody fusion in elderly patients. Spine Journal 2001; 1:402-407. 153] Osman AAH, bassiouni H, Koutri R, Nijs J, Geusens P, Dequeker J. Aging of the thoracic spine: distinction between wedging in osteoarthritis and fracture in osteoporosis - A cross-sectional and longitudinal study. Bone 1994; 15:437-442. 154] Oxland TR, Lund T, Jost B, Cripton P, Lippuner K, Jaeger P, Nolte LP. The relative importance of vertebral bone density and disc degeneration in spinal flexibility and interbody implant performance: an in vitro study. Spine 1996; 21:2558-2569. 155] Panjabi MM. Biomechanical testing to quantify adjacent-level effects. In Proceedings: 4th World Congress of Biomechanics 2002; 4-9 Aug, Calgary, Alberta, Canada. 156] Panjabi MM, Goel VK. Adjacent-level effects: design of a new test protocol and finite element model simulations of disc replacement. Roundtables in Spine Surgery 2005; 1:45-55. 157] Papadimitropoulos EA, Coyte PC, Josse RG, Greenwood CE. Current and projected rates of hip fractures in Canada. Canadian Medical Association Journal 1997; 157:1357-1363. 158] Papaioannou A, Watts NB, Kendler DL, Yuen CK, Adachi JD, Ferko N. Diagnosis and management of vertebral fractures in elderly adults. American Journal of Medicine 2002; 113:220-228. 159] Paramore CG, Dickman CA, Sonntag VK. Radiographic and clinical follow-up review of Caspar plates in 49 patients. Journal of Neurosurgery 1996; 84:957-961. 160] Park P, Gallon HJ, Gala VC, Hoff JT, McGillicuddy JE. Adjacent segment disease after lumbar or lumbosacral fusion: review of the literature. Spine 2004; 29:1938-1944. 161] Pearcy M, Portek I, Shepherd J. Three-dimensional x-ray analysis of normal movement in the lumbar spine. Spine 1984; 9:294-297. 162] Pearcy MJ. Stereo radiography of lumbar spine motion. Supplementum 1985; 2212:1-45.  Acta Orthopaedica Scandinavica  163] Phillips FM. Minimally invasive treatments of osteoporotic vertebral compression fractures. Spine 2003; 28:S45-S53. 164] Pfirrmann CWA, Resnick D. Schmorl nodes of the thoracic and lumbar spine: Radiologicpathologic study of prevalence, characterization, and correlation with degenerative changes of 1,650 spinal levels in 100 cadavers. Radiology 2001; 219:368-374. 165] Pluijm SMF, Tromp AM, Smit JH, Deeg DJH, Lips P. Consequences of vertebral deformities in older men and women. Journal of Bone and Mineral Research 2000; 15:1564-1572. 166] Polikeit A, Ferguson SJ, Nolte LP, Orr TE. The importance of the endplate for interbody cages in the lumbar spine. European Spine Journal 2003; 12:556-561. 167] Polikeit A, Nolte LP, Ferguson SJ. Simulated influence of osteoporosis and disc degeneration on the load transfer in a lumbar functional spinal unit. Journal of Biomechanics 2004; 37:1061-1069.  43  Chapter 1  Introduction  [168] Pollintine P, Dolan P, Tobias JH, Adams MA. Intervertebral disc degeneration can lead to "stressshielding" of the anterior vertebral body: a cause of osteoporotic vertebral fracture? Spine 2004 29:774-782. [169] Postacchini F. Spine update: surgical management of lumbar spinal stenosis. 24:1043-1047. [170] Prescher A. Anatomy and pathology of the aging spine. 27:181-195.  Spine 1999  European Journal of Radiology 1998  [171] Putzier M, Funk JF, Schneider SV, Gross C, Tohtz SW, Khodadadyan-Klostermann C, Perka C, Kandziora F. Charite total disc replacement - clinical and radiographical results after an average follow-up of 17 years. European Spine Journal 2006; 15:183-195. [172] Pye SR, Reid DM, Smith R, Adams JE, Nelson K, Silman AJ, O'Neill TW. Radiographic features of lumbar disc degeneration and self-reported back pain. Journal of Rheumatology 2004; 31:753758. [173] Pye SR, Reid DM, Adams JE, Silman AJ, O'Neill TW. Radiographic features of lumbar disc degeneration and bone mineral density in men and women. Annals of the Rheumatic Diseases 2006; 65:234-238. [174] Regan JJ. Clinical results of charite lumbar total disc replacement. America 2005; 36:323-340.  Orthopedic Clinics of North  [175] Renner SM, Lim TH, Kim WJ, Katolik L, An HS, Andersson GBJ. Augmentation of pedicle screw fixation strength using an injectable calcium phosphate cement as a function of injection time and method. Spine 2004; 29:E212-E216. [176] Resnick DK, Choudhri TF, Dailey AT, Groff MW, Khoo L, Matz PG, Mummaneni P, Watters W C III, Wang J, Walters BC, Hadley MN. Guidelines for the performance of fusion procedures for degenerative disease of the lumbar spine. Parts 1-17. Journal of Neurosurgery (Spine 2) 2005; 637-740. [177] Rhalston SH. 17:475-479.  Genetic determinants of osteoporosis.  Current Opinion in Rheumatology 2005;  [178] Rhyne A III, Banit D, Laxer E, Odum S, Nussman D. Kyphoplasty: report of eighty-two thoracolumbar osteoporotic vertebral fractures. Journal of Orthopaedic Trauma 2004; 18:294299. [179] Robertson JT, Papadopoulos SM, Traynelis VC. Assessment of adjacent-segment disease in patients treated with cervical fusion or arthroplasty: a prospective 2-year study. Journal of Neurosurgery (Spine) 2005; 3:417-423. [180] Rohlmann A, Klockner C, Bergmann G. [The biomechanics of kyphosis]. Orthopade 2001; 30:915918 [Article in German]. [181] Rosen CJ, Bilezikian JP. Clinical review 123: Anabolic therapy for osteoporosis. Journal of Clinical Endocrinology & Metabolism 2001; 86:957-964. [182] Ross PD, Kim S, Wasnich RD. Bone density predicts vertebral fracture risk in both men and women: a prospective study. Journal of Bone and Mineral Research 1996; 11(Suppl):S127. [183] Rousseau MA, Bradford DS, Hadi TM, Pedersen KL, Lotz JC. The instant axis of rotation influences facet forces at L5/S1 during flexion/extension and lateral bending. European Spine Journal 2006; 15:299-307. [184] Rubin MR, Bilezikian JP. New anabolic therapies Rheumatology 2002; 14:433-440.  in osteoporosis.  Current Opinion in  [185] Russo CR, Lauretani F, Seeman E, Bartali B, Bandinello S, Di lorio A, Guralnik J, Ferrucci L. Structural adaptations to bone loss in aging men and women. Bone 2006; 38:112-118. 44  Chapter 1  Introduction  [186] Sarzier JS, Evans AJ, Cahill DW. Increased pedicle screw strength with vertebroplasty augmentation in osteoporotic spines. Journal of Neurosurgery 2002; 96S:309-312. [187] Schneider DL, von Muhlen D, Barrett-Connor E, Sartoris DJ. Kyphosis does not equal vertebral fractures: the Rancho Bernardo study. Journal of Rheumatology 2004; 31:747-752. [188] Schneider G, Pearcy MJ, Bogduk N. Abnormal motion in spondylolytic Spondylolisthesis. 2005; 30:1159-1164.  Spine  [189] Schnitzler CM. Histomorphology of osteoporosis. In: Orthopaedic Issues in Osteoporosis. An YH (Ed.), CRC Press, Boca Raton, 2003; 19-37. [190] Schultheiss M, Hartwig E, Claes L, Kinzl L, Wilke HJ. Influence of screw-cement enhancement on the stability of anterior thoracolumbar fracture stabilization with circumferential instability. European Spine Journal 2004; 13:598-604. [191] Sengupta DK, Herkowitz HN. Lumbar spinal stenosis: Treatment strategies and indications for surgery. Orthopedic Clinics North America 2003; 34:281-295. [192] Sengupta DK, Herkowitz HN. Degenerative spondylolisthesis: review of current trends and controversies. Spine 2005; 30:S71-S81. [193] Soshi S, Shiba R, Kondo H, Murota K. An experimental study on transpedicular screw fixation in relation to osteoporosis of the lumbar spine. Spine 1991; 16:1335-1341. [194] Statistics Canada Demography Division. Population Projections for Canada, Provinces and Territories, 2000-2026. Catalogue No. 91-520. 2001. [195] Steffen T, Tsantrizos A, Aebi M. Effect of implant design and endplate preparation on the compressive strength of interbody fusion constructs. Spine 2000; 25:1077-1084. [196] Tanaka N, An HS, Lim TH, Fujiwara A, Jeon CH, Haughton VM. The relationshop between disc degeneration and flexibility of the lumbar spine. Spine Journal 2001; 1:47-56. [197] Taylor JR, Twomey LT. Age changes in lumbar zygapophyseal joints: observations on structure and function. Spine 1986; 11:739-745. [198] Tenenhouse A, Joseph L, Kreiger N, Poliquin S, Murray TM, Blondeau L, Berger C, Hanley DA, Prior JC, CaMos Research Group. Estimation of the prevalence of low bone density in Canadian women and men using a population-specific DXA reference standard: the Canadian multicentre osteoporosis study (CaMos). Osteoporosis International 2000; 11:897-904. [199] Thomsen K, Christensen FB, Eiskjaer SP, Hansen ES, Fruensgaard S, Bunger CE. The effect of pedicle screw instrumentation on functional outcome and fusion rates in posterolateral lumbar spinal fusion: a prospective, randomized clinical study. Spine 1997; 22:2813-2822. [200] Thomsen JS, Ebbesen EN, Moskilde L. Age-related differences between thinning of horizontal and vertical trabeculae in human lumbar bone as assessed by a new computerized method. Bone 2002; 31:136-142. [201] Tischer T, Aktas T, Milz S, Putz RV. Detailed pathological changes of human lumbar facet joints L1-L5 in elderly individuals. European Spine Journal 2006; 15:308-315. [202] Tsantrizos A, Baramki HG, Zeidman S, Steffen T. Segmental stability and compressive strength of posterior lumbar interbody fusion implants. Spine 2000; 25:1899-1907. [203] Tsantrizos A, Ito K, Aebi M, Steffen T. Internal strains in healthy and degenerated lumbar intervertebral discs. Spine 2005; 30:2129-2137. [204] Turgeman G, Zilberman Y, Zhou S, Kelly P, Moutsatsos IK, Kharode YP, Borella LE, Bex FJ, Komm BS, Bodine PV, Gazit D. Systemically administered rhBMP-2 promotes MSC activity and reverses bone and cartilage loss in osteopenic mice. Journal of Cellular Biochemistry 2002; 86:461-474.  45  Chapter 1  Introduction  [205] United Nations Population Division of the Department of Economic and Social Affairs. World Population Prospects: The 2002 Revision (http://www.un.org/esa/population/unpop.htm). Feb 2003. [206] Urban JPG, McMullin JF. Swelling pressure of the lumbar intervertebral discs: influence of age, spinal level, composition, and degeneration. Spine 1988; 13:179-187. [207] Urist MR. Acrylic cement stabilized joint replacements. Current Problems in Surgery 1975; 1-54. [208] Vaccaro AR, Ball ST. Indications for instrumentation in degenerative lumbar spinal disorders. Orthopedics 2000; 23:260-271. [209] Vernon-Roberts B. Age related and degenerative pathology of intervertebral discs and apophyseal joints. In: The Lumbar Spine and Back Pain. 4th ed. Jayson MIV and Dixon ASJ (Eds.), Churchill Livingston, Edinburgh, 1992; 17-41. [210] White AA, Panjabi MM. Clinical Biomechanics of the Spine. 2nd ed. Lippincott, Philadelphia, 1990. [211] Wiktorowicz ME, Goeree R, Papaioannou A, Adachi JD, Papadimitropoulos E. Economic implications of hip fracture: health service use, institutional care and cost in Canada. Osteoporosis International 2001; 12:271-278. [212] Wittenberg RH, Lee KS, Shea M, White AA, Hayes WC. Effect of screw diameter, insertion technique, and bone cement augmentation of pedicular screw fixation strength. Clinical Orthopaedics and Related Research 1993; 296:278-287. [213] World Health Organization (WHO). Assessment of Fracture Risk and its Application to Screening for Postmenopausal Osteoporosis. WHO Technical Report Series 1994; 843:1-101. [214] Yamagata M, Kitahara H, Minami S, Takahashi K, Isobe K, Moriya H, Tamaki T. Mechanical stability of the pedicle screw fixation systems for the lumbar spine. Spine 1992; 17:S51-S54. [215] Yang KH, King Al. Mechanism of facet load transmission as a hypothesis for low-back pain. Spine 1984; 9:557-565. [216] Yang TS, Chen YR, Chen YJ, Chang CY, Ng HT. Osteoporosis: prevalence in Taiwanese women. Osteoporosis International 2004; 15:345-347. [217] Yerby SA, Toh E, McLain RF. Revision of failed pedicle screws using hydroxyapatite cement: a biomechanical analysis. Spine 1998; 23:1657-1661. [218] Zdeblick TA. A propective, randomized study of lumbar fusion. Preliminary results. Spine 1993; 18:983-991. [219] Zindrick MR, Wiltse LL, Widell EH, Thomas JC, Holland WR, Field BT, Spencer CW. A biomechanical study of intrapeduncular screw fixation in the lumbosacral spine. Clinical Orthopaedics and Related Research 1986; 203:99-112.  46  Chapter 2  Bone Density and Pedicle Screw Testing  Chapter 2 Importance of Vertebral Bone Density in Pedicle Screw Testing  Overview This chapter highlights the importance of vertebral bone density in pedicle screw fixation and testing.  Current ASTM assessment methods, particularly the ASTM F1717, use synthetic  elements as vertebral surrogates and do not address important in vivo performance and failure characteristics. The effectiveness of spinal implants in fixation is dependent upon the boneimplant interface, and thus on vertebral bone density. The results in this study was presented to the American Standards of Testing and Materials (ASTM) Committee F04 on Medical and Surgical Materials and Devices at the "Symposium on Spinal Implants: Are We Evaluating Them Appropriately?" held in Dallas, Texas, Nov 6-7 2001. It was subsequently published as a peer reviewed article: Tan JS, Kwon BK, Samarasekera D, Dvorak MF, Fisher CG, Oxland TR. Vertebral bone density - a critical element in the performance of spinal implants. In: Spinal Implants: Are We Evaluating Them Appropriately? American Standards for Testing and Materials (ASTM) STP 1431, Melkerson MN, Griffith SL and Kirkpatrick JS, (Eds.), ASTM International, West Conchohocken, PA, 2003:217-30.  47  Chapter  2  Bone Density  and Pedicle  Screw  Testing  2.  Importance of Vertebral Bone Density in Pedicle Screws Testing  2.1  Introduction Spinal instrumentation is commonly used to stabilize the spine in conditions of trauma,  tumor, deformity, and degeneration. Various implant types are in existence, including pedicle screws, pedicle and laminar hooks, wires, cables, rods and plates, as well as anterior devices such as interbody cages. The pre-clinical assessment of these devices may involve many different types of testing, as described previously [Panjabi 1988, Ashman et al. 1989]. Within the last decade, ASTM standard test methods for spinal implants have been developed. One such Standard is the ASTM Test Methods for Static and Fatigue for Spinal Implant Constructs in a Corpectomy Model (F1717) [ASTM 2001]. The focus of this article is to address the relevance of F1717 to the clinical situation. In general, spinal instrumentation is effective in correcting deformity and stabilizing the spine, however, failures do occur. Clinical mechanisms of instrumentation failure are varied, but include  intra-operative,  or early  post-operative  fixation  failure,  late  development of  pseudoarthrosis with concomitant implant loosening, and disc or vertebra failure adjacent to the instrumentation [Paramore et al. 1996, McAfee et al. 1999, Eck et al. 2001, Uzi et al. 2001]. Ideally, testing methods would address these failure mechanisms, but this can be a significant challenge. For example, adjacent segment effects represent both a biological and mechanical problem that may be difficult to simulate in a pre-clinical model. However, early fixation failure represents a series of failure modes that can be evaluated in a pre-clinical model, including both failure of the specific device or the interface between the device and the spine.  Clinical  investigation suggests that it is the interface between the device and the spine that is the most common site of early failure [Paramore et al. 1996, Eck et al. 2001]. Essentially, this represents failure of the bone-implant interface. Biomechanical studies have demonstrated that bone mineral density, and more so the presence of osteoporosis, is a critical factor in the success of achieving rigid fixation both for pedicle screws [Zindrick et al. 1986, Soshi et al. 1991, Wittenberg et al. 1991, Yamagata et al. 1992, Okuyama et al. 1993, Halvorson et al. 1994, Pfeiffer et al. 1996] and interbody cages [Oxland et al. 1996, Lund et al. 1998]. In general, pedicle screw pull-out strength and toggle amplitude as well as interbody cage subsidence resistance and overall stabilisation have been shown to be linearly correlated to BMD. The F1717 uses synthetic elements as vertebral surrogates and therefore does not address the bone-implant interface. The F1717 does evaluate the mechanical properties of the screw-rod constructs, particularly in fatigue failure.  However, such failure is not a common 48  Chapter 2  Bone Density and Pedicle Screw Testing  mode of clinical failure [Paramore et al. 1996, McAfee et al. 1999, Eck et al. 2001, Uzi et al. 2001]. Since cancellous bone has an elastic modulus that ranges between 0.01 to 1.0 GPa, whereas UHMWPE has a modulus of 1 GPa, the construct would be expected to behave differently if the bone was included in the assessment. The primary purpose of this study was to contrast the mechanical behaviour of pedicle screws under short-term cyclic physiologic loads when inserted in cadaveric vertebrae versus synthetic surrogates.  Secondarily, the effect of bone mineral density on the pedicle screw  behaviour was investigated, as was a novel model of simulating a situation where the screwbone interface is recognised as being poor. In essence, we are questioning whether the fixation at the bone-implant interface can be measured during short-term cyclic testing.  2.2  Materials and Methods  2.2.1  Specimen Preparation Eight lower lumbar cadaveric vertebrae (three L3s, three L4s, two L5s) from a previous  endplate indentation study [Grant et al. 2001] were used in this study. All specimens had intact pedicles. Lateral DEXA bone mineral density (BMD) ranged from 0.30 to 1.04 g/cm with a 2  mean of 0.70 g/cm . 2  Pedicle screws of uniform size (Universal Spinal System, USS diameter 6 mm and length 45 mm, Stainless Steel, Synthes Spine, Paoli, PA) were inserted into each pedicle. A single spine surgeon (Dr Brian Kwon) carried out all screw insertions for the specimens. Two types of pedicle screw hole preparation were carried out in a randomised manner for each specimen. One side was prepared in the normal recommended fashion using a blunt pedicle probe to a depth of approximately 50 mm. The other side was prepared with an overdrilled hole using a 15/64 inch (5.95 mm) drill bit to a depth of 50 mm. The latter method was an attempt to create a "loosened" screw model. Accuracy of insertion was checked by direct palpation of the pedicle walls.  Radiographs were taken in the cephalo-caudad (axial) direction to verify screw  placement through the pedicle. Specimens were defrosted at room temperature for 4 to 5 hours before screw insertions were carried out. After the screws had been inserted, the specimens were soaked in saline solution and submerged in a water bath maintained at 37 °C for 16 to 24 hours before being tested.  2.2.2  Pedicle Screw Loading A custom built jig was used to test the pedicle screws with loads simulating physiological  conditions.  The basic test configuration was similar to the ASTM standard (F1717), but a 49  Chapter 2  Bone Density and Pedicle Screw Testing  cadaveric vertebra replaced the inferior block and the dimensions were modified to reflect recent in vivo load data [Rohlmann et al. 1997, Rohlmann et al. 2000]. To allow comparison between screw fixation in a vertebra and UHMWPE, an inferior block of UMHWPE was used in other tests. A vertical connecting rod of length 140 mm was attached to the two pedicles screws, which were inserted in a UHMWPE block and in the pedicle of a cadaveric specimen respectively (Figure 2.1). The inferior blocks and specimens were clamped rigidly to the base of the testing machine, which is a different setup from the ASTM standard. This fixation method was selected to prevent the specimen breakdown that would probably occur if a single loading point had been used for the vertebra. The design of a rigid inferior mount results in somewhat different moments being applied to the screw-vertebra interface, depending upon the vertebral material. A two-dimensional finite element analysis demonstrated this effect (Figure 2.2).  (a) Figure  2.1  (b)  (a) Lateral and (b) frontal view of the testing jig. A single vertical rod connects the two pedicle screws, which in turn are inserted into a UHMWPE block superiorly and into the pedicle of a cadaveric specimen inferiorly. Other tests included a UHMWPE block inferiorly as well. The pedicle screws were placed parallel to each other and the connecting rod was aligned vertically before every test. The vertebral body was securely clamped to the base of the testing machine, whereas the UHMWPE was allowed to pivot about an axis 35 mm away from the vertical rod. 50  Chapter 2  Bone Density and Pedicle Screw Testing  |  Figure  2.2  Elastic M o d u l u s (MPa)  A two-dimensional finite element analysis determined the variation ground reaction moments as a factor of elastic modulus of the block. The lower block was rigidly fixed in translation and rotation the upper block was allowed to rotate freely.  |  in lower while  The initial position of the specimen was such that the pedicle screw was in a horizontal position (Figure 2.1). The upper segment of the testing apparatus was attached to the actuator arm of the testing machine and was hinged to the UHMWPE block, allowing the block to freely rotate about a single axis. The horizontal distance between the hinge and the vertical rod was 35 mm. Liquid levels were used to ensure that the vertical connecting rod was in an absolute vertical position prior to toggle tests. A single UHMWPE block was used and the same pedicle screw was inserted in the block throughout all the tests. The vertical load acting on the setup was cycled between an axial compression of 300 N and tension of 30 N at a rate of 0.5 Hz for 100 cycles using a servohydraulic test system (Instron 8874, Instron Corporation, Canton, MA). The moment arm of 35 mm resulted in a bending moment (300 N x 35 mm = 10.5 Nm) in the connecting rod in addition to the cephalocaudad force.  This ratio of axial load to bending moment was based upon in  vivo  measurements [Rohlmann et al. 1997, Rohlmann et al. 2000]. 2.2.3  Kinematics  Motion of the pedicle screws with respect to the inferior UHMWPE and the cadaveric specimens was detected using a high precision optoelectronic system (Optotrak 3020, Northern Digital Inc., Waterloo Canada) at a rate of 10 Hz, which gave twenty data points for each cycle. Two sets of marker carriers, each with four infrared LEDs, were mounted onto the rigid bodies: 51  Chapter 2  Bone Density and Pedicle Screw Testing  pedicle screw and bone or pedicle screw and UHMWPE block (Figure 2.3). The set of marker carriers on the bone was mounted onto the posterior elements, such as the lamina or the superior articular process, so as to be close to the screw and pedicle. The bone was assumed to be moving as a rigid body. The three dimensional motion of these two rigid bodies with respect to one another was post processed.  (b) Figure  2.3  Two marker carriers, each with four infrared LEDs, were mounted onto the two rigid bodies, (a) pedicle screw and vertebral specimen, and (b) pedicle screw and UHMWPE block. A high precision optoelectronic system was used to capture the motion of each rigid body with time. Subsequent post processing determines the relative motion of the pedicle screw in bone and of the pedicle screw in UHMWPE. 52  Chapter 2  Bone Density and Pedicle Screw Testing  A Cartesian local coordinate system was specified to describe the motion of the pedicle screws. The origin of the local coordinate system laid along the axis of the screw at the point where the connecting rod was attached to the pedicle screw. The z-axis always pointed away from the screw tip. The y-axis was parallel to the connecting rod and pointed upwards toward the direction of the superior pedicle screw. The direction of the x-axis was the vector cross product of the y- and z-axes. The relative motions between the pedicle screw and the UHMWPE and between the pedicle screw and the vertebral bone were post processed from the three-dimensional motion data using custom software. The range and offset of the motion of the pedicle screw head were obtained from the average of the last five cycles (96th to 100th). These parameters gave an indication of dynamic motion under repetitive loading (range) and of settling or subsidence (offset). As the applied loads were in the sagittal plane (parallel to the pedicle screws), the resultant motions were primarily y translation and x rotation (Figure 2.4).  Furthermore, the  translation of the screw tip was also calculated, which when combined with the translation of the screw head, allowed for an estimation of the instantaneous axis of rotation to be made. 2.2.4  Statistical Analysis  The range and offset of the motion of the pedicle screw head were compared between the three types of insertion: UHMWPE, normal insertion and overdrilling. An unpaired t-test was used to determine if there was any difference between the translation and rotation ranges and offsets between screws in UHMWPE and screws in bone inserted in the normal fashion. A paired t-test was used to compare between pedicle screws in normal and in overdrilled holes. Effects of BMD on translation and rotation offsets and ranges were determined for both normal insertion and overdrilling by calculating Pearson product-moment correlation coefficients between these factors and BMD. Statistical significance was assumed to be at the 95 % level. i  53  Chapter  Bone Density  2  and Pedicle  Screw  Testing  -X translation  Y translation  •Z translation  0.5  20 s TV-  E E,  u  C  E-0.5 u> o ro  a  !!  !!  U i! i!  I!  V V VU V VV U  J  u> -1 H  -1.5  M  I  a>  Offset  Range  Time (seconds)  J  (a) X rotation  • Y rotation  •Z rotation  2.5  ? 1.5 2  5.  1  i'0.5 < OH -0.5  ,1 A A Ai Ai\ A A U 11  II  i  1  Range  11  -I  I i I I I i I I i M I i Ii 1 I l I I i I i 1i I 1  ,  '  1  '  1  1  1  I  1  I  )  1  I  /  1  1  '  OTTI  Time (seconds)  i  Offset  20  (b) Figure 2.4  The (a) relative translations and (b) rotations between pedicle screw and bone in all six degrees of freedom for the first twenty seconds are presented in these two graphs. Translations in y of the screw head and rotations in x were the main motions measured, since the applied loads were in that plane. The other translations and rotations were not deemed significant as their magnitudes were much smaller. The range and offset of the y translation and x rotation are as shown on the graphs. Range and offset of the motion of the pedicle screw head were obtained from the average of the last five cycles (96th to 100th).  54  Chapter 2  Bone Density and Pedicle Screw Testing  2.3  Results  2. 3.1  Motion  of Pedicle  Screw  The motions reported here consist of the sagittal plane rotation of the screw with respect to the vertebra and the axial translation of the screw head. The other rotations and translations were measured, but were very small and therefore not deemed significant. The range of motion of the pedicle screw in the UHMWPE block was small, averaging 0.2 mm and 0.9°. The range of motion in the bone was significantly higher, averaging 1.4 mm and 2.4° for normally inserted screws and 2.4 mm and 2.7° for screws in overdrilled holes (Table 2.1 and Figure 2.5). Table  2.1  Mean s.d. Max Min Mean s.d. Max Min  Range  and offset  of translation  and rotation  Y Translation-Range (mm) Screw in Normal Overdrilled UHMWPE Insertion (n=8) (n=8) (n=8) 0.2 1.4 2.4 0.02 1.0 0.9 0.3 3.1 3.9 0.2 0.5 1.5 Y Translation-Offset (mm) -0.1 -0.5 -1.2 0.02 0.5 1.3 -0.1 0.0 0.0 -0.1 -1.6 -4.1  in the sagittal  plane.  XRota tion-Range (degrees) Screw in Normal Overdrilled UHMWPE Insertion (n=8) (n=8) (n=8) 2.4 0.9 2.7 0.02 0.7 1.0 0.9 3.9 4.1 0.9 1.3 1.8 X Rotation-Offset (degrees) 0.4 0.4 0.3 0.05 0.8 1.6 0.4 1.7 2.7 0.3 -2.2 -1.3  The offset of translation of pedicle screws in the UHMWPE block was -0.1 mm whereas the offset of the pedicle screws in bone was -0.5 mm and in the overdrilled holes was -1.2 mm. The offset of the rotation of pedicle screws in UHMWPE was 0.4° whereas the offset of the pedicle screws in bone was 0.4° and in overdrilled holes was 0.3°. Offset of the pedicle screws in bone resulted in enlarged insertion holes. There was a significant difference in translation range (p < 0.005), translation offset (p < 0.05) and rotation range (p < 0.001) between motion of screw in UHMWPE and bone, while rotation offsets were not different (p > 0.95). There was no significant difference in translation range (p = 0.15), translation offset (p = 0.47), rotation range (p = 0.33) and rotation offset (p = 0.89) between pedicle screws in normal and overdrilled holes.  55  Bone Density and Pedicle Screw Testing  Chapter 2  • translation-range • translation-offset • rotation-range • rotation-offset  o> '—I 5-  3 A  ro  K] UHMWPE  Vertebra (normal)  Vertebra (overdrilled)  Figure  2.5  The range and offset of the motion of the screw heads for screws inserted in a) UHMWPE, b) bone with normal insertion and c) bone with overdrilling. Error bars denote the standard deviation. N = 8 for each group.  2.3.2  Motion  of Pedicle  Screw  in Bone;  Effects  of BMD  The correlation between bone mineral density and the range and offset of motion of the pedicle screw inserted in the normal fashion (Table 2.2 and Figure 2.6) was not statistically significant. There was, however, a trend that the specimens with higher bone mineral density generally had lower translation and rotation ranges, whereas these ranges tended to be higher for specimens with lower bone mineral densities. No direct correlation between magnitude of translation or rotation with BMD was found for the pedicle screws inserted into the overdrilled holes. Table  2.2  Correlation BMD.  between  range  and offset  of translation  and rotation  against  Correlation coefficient, r 0.37 No 0.15 Translation range No 0.24 0.26 Translation offset No 0.35 0.12 Rotation range 0.03 No 0.69 Rotation offset A 95 % confidence level (p < 0.05) was used to decide if the data were significantly correlated. p value  Correlation  3  2  a  56  Chapter 2  Bone Density and Pedicle Screw Testing  • Translation-Range  • ,  4.5 4  • Rotation-Range  •  3.5 3  o> a> 2.5 S c o 2 ~  • •  5  • • •  1.5 (2  •  1  •  0.5 0  1  0.2  0.4  0.6 BMD (g/cm )  0.8  1.2  2  Figure  2.6  Scatterplots of screw head translation and rotation ranges versus BMD for pedicle screws inserted with normal hole preparation. No significant linear correlations in translation range (p=0.15, r =0.369) or rotation range (p=0.12, ^=0.350) with BMD were found but there was a trend of lower translations and rotations for specimens with higher BMD. No correlations were found for pedicle screws inserted in overdrilled holes. 2  2.3.3  Center  of  Rotation  The kinematics of the pedicle screw in bone could be broadly categorised into two groups. For specimens with higher bone mineral density, the screws were pivoting about a point located somewhere between the screw tip and the screw head.  Translations and rotations were  generally smaller (Figures 2.6 and 2.7). For specimens with lower bone mineral density, the motion of the screws underwent two stages. In the first stage, screw motion was characterized by vertical rigid body translation, with the screw head and screw tip both translating along the yaxis together. In the second stage, the screw demonstrated a rotational motion, with the screw head and tip moving in opposite directions, indicative of a center of rotation somewhere between the screw tip and the screw head. Overdrilling did not alter the kinematic patterns of the pedicle screws in the specimens. With overdrilling, pedicle screws in specimens with higher bone mineral density were observed to be pivoting about a point, without the translations as observed in the overdrilled specimens with lower bone mineral density. This was in spite of the observation that the magnitudes of range and offset in the overdrilled holes were not correlated with BMD.  57  Chapter  2  Bone Density  and Pedicle  Screw  Testing  screwhead  Figure  2.7  Typical motion of pedicle screws inserted in the normal fashion at screw head and screw tip during the last 5 cycles for specimens (a) low BMD and (b) with high BMD. The motion pattern for screws in BMD bone was in two stages, a rigid body translation and a rigid rotation. While in high BMD bone, the screws were mainly in rigid rotation, oscillating about a point some distance b e t w e e n the screw and screw tip.  2.4  Discussion  the with low body body head  The testing methodology in this study applied similar loads and boundary conditions onto the two pedicle screws, one inserted in bone and the other inserted in UHMWPE. The results of this study showed clearly that the motion of the pedicle screw with respect to bone was significantly higher than its motion in a homogeneous block of UHMWPE. The higher stiffness of the UHMWPE results in less rotation of the screws and rod construct.  Inspection of the  specimens indicated that the pedicle screws caused permanent enlargement of the insertion 58  Chapter 2  Bone Density and Pedicle Screw Testing  holes of the pedicles. On the other hand, the pedicle screws inserted in the UHMWPE did not result in any observable damage to the latter. The motion of the pedicle screw with respect to the UHMWPE at the screw head could be attributed to a) motion of the screw within the UHMWPE and b) bending of the screw outside the UHMWPE. Indeed, a combination of both scenarios, motion within and bending outside the UHMWPE, could have occurred. While no gross permanent damage to the UHMWPE was observed, such motion could also be partially attributed to load application within the elastic limit of the UHMWPE. As the portion of the pedicle screw embedded within the UHMWPE was not substantially loaded, it could be postulated that fatigue fracture of pedicle screws tested in UHMWPE would occur at the neck of the screw. Motion of the pedicle screw with respect to the vertebra was largely attributed to the rigid body motion of the screws within the bone and not to the bending of the screw itself. This rigid body motion of the screw against the bone resulted in enlargement of the insertion hole and permanent damage to the internal trabecular structure of the vertebral body. The kinematics of the pedicle screws in the bone was consistent with the presence of a fulcrum between the head and tip about which the pedicle screws were oscillating. Description of this clinically relevant mode of failure may aid in the development of improved techniques of spinal fixation. Bone mineral density appeared to have an effect on the kinematics of the pedicle screws. Although translation and rotation ranges were not significantly correlated with bone mineral density in this study, there was a trend to suggest this association. The small sample size in this study could have contributed to the failure to demonstrate a statistically significant correlation. Moreover, the BMD values used in the statistical analysis were of the vertebral bodies and did not include the quality of bone in the pedicle region. The same trend against BMD was observed in pull-out tests by many researchers [Soshi et al. 1991, Wittenberg et al. 1991, Yamagata et al. 1992, Okuyama et al. 1993, Halvorson et al. 1994, Pfeiffer et al. 1996]. In those studies, specimens with higher BMD had higher pull-out force, while in the current study, there was a trend of lower translation and rotation ranges for specimens with higher BMD. By studying the motion of the screw tip and screw head, it was revealed that for screws in vertebrae with higher bone mineral density, the screws were pivoting about a point between the screw head and the screw tip. For screws in vertebrae with lower bone mineral density, the screws underwent a rigid body translation followed by a pivoting motion. The former group of pedicle screws inserted in vertebrae with higher BMD could be considered to have achieved satisfactory early bone-implant fixation and they, in addition to the screws inserted in the 59  Chapter 2  Bone Density and Pedicle Screw Testing  UHMWPE, would have passed the ASTM test protocol. The latter group of pedicle screws inserted in vertebrae with lower BMD mimicked the clinical scenario of intra-operative or early post-operative bone-implant interface failure. Thus, bone mineral density indeed influenced short-term fixation and early failure of pedicle screws. This would not have been detected using the F1717 test protocol [ASTM 2001]. The damaging effects of metallic implants on bone have not been fully characterized or addressed in the F1717. It is possible that some designs of pedicle screws could result in more damage to the trabecular structure than others. Therefore the fatigue life of an implant does not appear to be correlated to the ability of an implant to successfully develop a strong bone implant interface. Overdrilling did not result in significantly higher translation or rotation ranges and offsets as compared to the screws inserted in the normal fashion. The kinematics of the pedicle screws was also not affected by overdrilling.  Overdrilling furthermore caused the translation and  rotation ranges and offsets to be independent of BMD, eliminating the trend of BMD effect observed for the normal screw insertions. Overdrilling the insertion holes resulted in a loosened screw model which otherwise could be achieved by a high number of cyclic motions on a screw inserted in a normal fashion. 2.5  Conclusion  This study contrasted, under short-term cyclic physiologic loads, the mechanical behavior of pedicle screws inserted in cadaveric vertebrae versus synthetic surrogates. The kinematics of the screws inserted in bone and in UHMWPE were found to be different in terms of the range of motion, pivoting and bending points of the screws and in terms of the effects of bone mineral density. The kinematics of the screws in bone is more relevant to clinical modes of failure. The fixation at the bone-implant interface can be quantified during short-term cyclic testing when an appropriate model is used, as demonstrated in this study. 2.6  References  [1]  Ashman RB, Bechtold JE, Edwards WT, Johnston II CE, McAfee PC, Tencer AF. In vitro spinal arthrodesis implant mechanical testing protocols. Journal of Spinal Disorders 1989; 2:274-281.  [2]  American Society for Testing and Materials. ASTM F1717. Standard test methods for spinal implant constructs in a vertebrectomy model. ASTM International 2001.  [3]  Eck KR, Bridwell KH, Ungacta FF, Riew D, Lapp MA, Lenke LG, Baldus C, Blanke K. Complications and results of long adult deformity fusions down to L4, L5, and the sacrum. Spine 2001; 26:E182-E192.  [4]  Grant JP, Oxland TR, Dvorak MF. Mapping the structural properties of the lumbosacral vertebral endplates. Spine 2001; 26:889-896. 60  Chapter 2  Bone Density and Pedicle Screw Testing  [5]  Halvorson TL, Kelley LA, Thomas KA, Whitecloud III TS, Cook SD. Effects of bone mineral density on pedicle screw fixation. Spine 1994; 19:2415-2420.  [6]  Lund T, Oxland TR, Jost B, Cripton PA, Grassmann S, Etter C, Nolte L-P. Interbody cage stabilisation in the lumbar spine. Journal of Bone and Joint Surgery 1998; 80B:351-359.  [7]  McAfee PC, Cunningham BW, Lee GA, Orbegoso CM, Haggerty CJ, Fedder IL, Griffith SL. Revision strategies for salvaging or improving failed cylindrical cages. Spine 1999; 24:21472153.  [8]  Okuyama K, Sato K, Abe E, Inaba H, Shimada Y, Murai H. Stability of transpedicle screwing for the osteoporotic spine. Spine 1993; 18:2240-2245.  [9]  Oxland TR, Lund T, Jost B, Cripton PA, Lippuner K, Jaeger Ph, Nolte L-P. The relative importance of vertebral bone density and disc degeneration in spinal flexibility and interbody implant performance. Spine 1996; 21:2558-2569.  [10]  Panjabi MM. Biomechanical evaluation of spinal fixation devices. Spine 1988; 13:1129-1134.  [11]  Paramore CG, Dickman CA, Sonntag VK. Radiographic and clinical follow-up review of Casper plates in 49 patients. Journal of Neurosurgery 1996; 84:957-961.  [12]  Pfeiffer M, Gilbertson LG, Goel VK, Griss P, Keller JC, Ryken TC, Hoffman HE. Effect of specimen fixation method on pullout tests of pedicle screws. Spine 1996; 21:1037-1044.  [13]  Rohlmann A, Bergmann G, Graichen F. Loads on an internal spinal fixation device during walking. Journal of Biomechanics 1997; 30:41-47.  [14]  Rohlmann A, Graichen F, Weber U, Bergmann G. Monitoring in vivo implant loads with a telemeterized internal spinal fixation device. Spine 2000; 25:2981-2986.  [15]  Soshi S, Shiba R, Kondo H, Murota K. An experimental study on transpedicular screw fixation in relation to osteoporosis of the lumbar spine. Spine 1991; 16:1335-1341.  [16]  Uzi EA, Dabby D, Tolessa E, Finkelstein JA. Early retropulsion of titanium-threaded cages after posterior lumbar interbody fusion. Spine 2001; 26:1073-1075.  [17]  Wittenberg RH, Shea M, Swartz DE, Lee KS, White AA, Hayes WC. density in instrumented spine fusions. Spine 1991; 16:647-652.  [18]  Yamagata M, Kitahara H, Minami S, Takahashi K, Isobe K, Moriya H, Tamaki T. Mechanical stability of the pedicle screw fixation systems for the lumbar spine. Spine 1992; 17:S51-S54.  [19]  Zindrick MR, Wiltse LL, Widell EH, Thomas JC, Holland WR, Field BT, Spencer CW. A biomechanical study of intrapeduncular screw fixation in the lumbosacral spine. Clinical Orthopaedics and Related Research 1986; 203:99-112.  Importance of bone mineral  61  Chapter  3  Pedicle  Screw  Augmentation  Chapter 3 Augmentation of Pedicle Screws in Osteoporotic Spine  Overview This chapter addresses the issue of pedicle screw loosening in the osteoporotic spine by comparing three existing techniques of pedicle screw augmentation in the osteoporotic spine. The motion patterns and motion magnitudes of pedicle screws following augmentation with laminar hooks, sublaminar wires or calcium phosphate cement were contrasted. The results in this study were presented at national and international conferences: 1) 33rd Annual Orthopaedic Research Day, Nov 29 2002, Vancouver, BC, Canada. 2) 30th Annual Meeting of the International Society for the Study of the Lumbar Spine, May 1317, 2003, Vancouver, BC, Canada. 3) 37th Annual Meeting of the Canadian Orthopaedic Research Society, Oct 2-5, 2003, Winnipeg, Manitoba, Canada. It was also recently published in a refereed journal: Tan JS, Kwon BK, Dvorak MF, Fisher CG, Oxland TR. Pedicle screw motion in the osteoporotic spine after augmentation with laminar hooks, sublaminar wires, or calcium phosphate cement: a comparative analysis. Spine 2004; 29:1723-1730.  62  Chapter 3  Pedicle Screw  Augmentation  3.  Augmentation of Pedicle Screws in Osteoporotic Spine  3.1  Introduction Pedicle screws are commonly used for a variety of indications to achieve segmental  fixation within the lumbosacral spine [Lin et al. 1983, Cloward 1985, Lin 1999, Lowe and Tahernia 2002]. Despite being the most rigid form of posterior instrumentation, pedicle screws sometimes achieve poor initial fixation primarily in patients with low bone mineral density (BMD) [Okuyama et al. 2001]. Furthermore pedicle screw constructs in the lumbosacral spine can loosen with time, particularly at the ends of a long fusion and in weaker bone. Augmentation of pedicle screws in patients with poor bone quality may be necessary to enhance the rigidity of the construct [Weinstein et al. 1992, Bostrom and Lane 1997, Okuyama et al. 2001]. The common clinical methods to improve the fixation of pedicle screws within the osteoporotic spine include the addition of laminar hooks or sublaminar wires to the fixation construct, or the supplementation of the pedicle channel with polymethylmethacrylate (PMMA) or calcium phosphate cement [An et al. 2002]. Other novel augmentation methods suggested in the literature include the use of a bushing, a plate [Law et al. 1993], an expansive pedicle screw design [Cook et al. 2000, Cook et al. 2001], an expandable anchor [McKoy and An 2001a] and an interlocking screw [McKoy et al. 2001b]. The use of laminar hooks concurrently with pedicle screws in vitro has been shown to enhance the rigidity of pedicle screw fixation [Halvorson et al. 1994, Hilibrand et al. 1996, Hasegawa et al. 1997, Liljenqvist et al. 2001].  The use of  sublaminar wires in a variety of configurations has also been shown to improve fixation in the osteoporotic spine [Coe et al. 1990, Butler et al. 1994, Hu 1997, Heller et al. 1998]. In recent years, various bone cements have been shown to increase pedicle screw fixation strength [Soshi et al. 1991, Wittenberg et al. 1993, Lotz et al. 1997, Moore et al. 1997, Yerby et al. 1998, Ignatius et al. 2001, Sarzier et al. 2002].  Few comparative studies between these various  methods have been conducted. The fixation strengths of sublaminar wires, hooks, and screws, when used alone, have been contrasted [Coe et al. 1990, Butler et al. 1994]. However the use of sublaminar wires, laminar hooks, and cement to augment pedicle screw fixation have not been compared. The purpose of this study was to compare three methods of pedicle screw augmentation (laminar hooks, sublaminar wires, and calcium phosphate cement) in an in vitro biomechanical model of a loose pedicle screw in an osteoporotic spine. Specifically, we sought to determine the effects of these augmentation techniques on both the pattern and the magnitude of screw motion under physiologic loading conditions.  63  Chapter 3  Pedicle Screw  3.2  Methods  3.2.7  Specimens  Augmentation  Twenty-four lumbar vertebrae (8-L3, 7-L4 and 9-L5) from eleven human cadavers were used in this study. The ages ranged between 48 to 90 years, with a mean of 74.5 years. Three were female and eight were male. All specimens were previously utilized in an experiment to determine the structural properties of the endplate [Grant et al. 2001]. Although the testing protocol of this previous study involved only the endplates, each specimen was carefully visually inspected for structural damage. All specimens had intact posterior elements, and in particular, the pedicles were intact. Bone mineral density from lateral DXA scans in the twenty-four specimens ranged between 0.28 and 1.30 g/cm . The specimens were arranged into three groups of eight to 2  ensure a comparable distribution of bone mineral densities amongst the groups (i.e. each group averaged approximately 0.7 g/cm ). 2  3.2.2  Augmentation Methods  Within each group of eight specimens, two augmentation methods were compared in a paired experimental design. The first group compared wire versus hook, the second compared hook versus calcium phosphate cement, and the third compared calcium phosphate cement versus wire. The loosening of pedicle screws within an osteoporotic spine was simulated by drilling an oversized pedicle channel, thus creating "loose screws" without subjecting the specimens to large numbers of cyclic loads. This technique of mimicking loose pedicle screws was validated in a previous study (Chapter 2) [Tan et al. 2003]. Some of the specimens from this initial validation study were used in the current study (wire-hook group). Using the standard posterior landmarks for pedicle screw placement, the pedicle was drilled to a depth of 50 mm with an oversized drill bit measuring 6.35 mm (1/4 inch) in diameter. For the specimens in which hooks and wires were tested, a stainless steel pedicle screw, 6.0 mm in diameter, 45.0 mm in length (Universal Spine System, Synthes, Paoli, PA) was then inserted into the pedicle and connected to a 6.0 mm by 140 mm stainless steel hard rod with the standard caps (Synthes Spine, Paoli PA). For sublaminar wire augmentation, two smooth 18 gauge Luque wires (Zimmer, Warsaw, IN), were wrapped around the pedicle screw and rod construct and tightened independently to secure the screw and rod construct to the laminar (Figure 3.1). For laminar hook augmentation, a medium or large laminar hook was attached to the rod distal to the screw with a 15 or 25 mm transverse connector (Synthes Spine, 64  Chapter 3  Pedicle Screw Augmentation  Paoli, PA) and then compressed against the pedicle screw (Figure 3.1).  For the calcium  phosphate cement augmentation, the cement (Norian CRS, Norian Corporation, Cupertino, CA) was mixed in the manufacturer's pneumatic mixer, and approximately 2 ml were injected in a retrograde fashion into the oversized pedicle drill hole. A 6.0 mm by 45.0 mm stainless steel pedicle screw was inserted into the cement-filled pedicle channel, and the specimen was then immediately immersed into a 37 °C isotonic saline bath for 16 to 24 hours to allow the cement to harden completely and achieve its maximal compressive strength [Lotz et al. 1997]. Specimens with sublaminar wire or laminar hook augmentation were also placed in the saline bath so that consistency in the condition of the vertebral bodies was maintained. All pedicle screw insertions and augmentation procedures were carried out by a single spine surgeon (Dr Brian Kwon) in order to minimize variation in technique.  Figure  3.1  Augmentation with laminar hook (left) and with sublaminar wires (right). The wires were twisted separately (inset). Compression b e t w e e n the pedicle screw and the transverse connector of the laminar hook pulled the head of the pedicle screw down inferiorly (rotating the tip of the screw in a superior direction). Tightening of the sublaminar wires pulled the screw head medially, rotating the screw tip laterally. The rotation of the screw in the vertical plane by this external force creates a "three-point contact" along the length of the screw. 65  Chapter 3  Pedicle Screw Augmentation  Radiographs of each specimen were taken in the axial and lateral directions to verify position of the implants. Drilling to a depth of 50 mm did not perforate the anterior cortical wall and none of the 45 mm screws penetrated the anterior cortex.  (a) Figure  3.2  3.2.3  Mechanical  (b)  (a) The resultant forces on the inferior screw consisted of a compression and a flexion bending moment, (b) Two sets of marker carriers with four infrared LEDs each were rigidly attached to the pedicle screw and the superior articular process so as to capture the rigid body screw motion during cyclic test. The marker carrier was attached to the pedicle screw using a 4 mm bolt. Rigid connection b e t w e e n the superior articular process and marker carrier was provided by a 4-prong stiff braided wire, and attached using a bone screw. Testing  Testing methodology used in this study was the same as in the previous study (Chapter 2).  The pedicle screws were subjected to a flexion bending moment by applying a cephalo-  caudad force with an offset to the rigidly connected vertical rods (Figure 3.2a). The magnitudes of these forces were analogous to loads experienced during walking [Rohlmann et al. 1997, Rohlmann et al. 2000].  The vertebral bodies with the augmented pedicle screw and rod  constructs were rigidly clamped to the base of the testing machine during tests. The rod was 66  Chapter  Pedicle  3  Screw  Augmentation  connected 120 mm superiorly to a pedicle screw inserted into a polyethylene (PE) block. The PE block was hinged 35 mm anterior to the rod. A servo-hydraulic testing machine (Instron 8874, Instron Corporation, Canton, MA) was used to cyclically apply forces between 300 N of compression and 30 N of tension using a sinusoidal waveform. This load transformed into a compressive force (300 N) and a bending moment (10.5 Nm) at the inferior pedicle screw. Each screw was tested for 100 cycles at a rate of 0.5 Hz. At the completion of these tests, the fixation was removed from the specimen, and the process of pedicle screw insertion and augmentation was repeated on the contralateral pedicle with new screws and the comparison augmentation technique. An optoelectronic camera system (Optotrak 3020, Northern Digital Inc., Waterloo, Canada) was used to monitor the rigid body motion of the pedicle screws and the vertebral bodies. Two "marker carriers", each mounted with four infrared LEDs, were rigidly attached to the head of the pedicle screw and to the superior articular process of the vertebra being tested (Figure 3.2b). The capture rate of the camera system was at 10 Hz, which gave 20 data points for each loading cycle.  0.8 n  mean of last 5 cycles I  I  Range  I I  Offset  Q  mean from first 5 cycles 0.1  -0.1 Figure  3.3  H J  <  20 seconds (10 cycles, 0.5 Hz)  The motion of the screw head in the vertical direction during the last 10 cycles is presented in this figure. The translation range was defined as the difference between the maximum and minimum displacement between the 96th to the 100th cycles. Translation offset was defined as the change in mean position between the first five to the last five cycles.  67  Chapter  3  Pedicle  Screw  Augmentation  3.2.4 Data Analyses The translation range and translation offset were analyzed for the screw tip and screw head with respect to the vertebral body. All analyses, particularly the determination of screw tip motion, assumed that the pedicle screws underwent rigid body motion. Motion of the screws during the first and last five cycles of the total 100 cycles was used to determine the translation range and translation offset (Figure 3.3). The translation range was defined as the difference between the maximum and minimum displacement in the axial (Y) direction averaged between the 96th to 100th cycles (Chapter 2) [Tan et al. 2003]. Translation offset was defined as the change in mean position between the first five to the last five cycles. Both the translation range and offset were analysed for motion at the screw tip and screw head for all forty-eight screws. Since the direction of applied loads was restricted to the vertical plane alone, the primary motions observed were translation and rotation within that plane. The translation and rotation motion subsequently discussed referred to motion in this vertical plane only. As well as comparing the magnitudes of translation at the screw head and screw tip, the patterns of motion for each screw were also categorised as a) pure rotation, b) translation, c) rotation-translation or d) double rotation (Figure 3.4) by inferring from the motion of the screw head and screw tip over time.  Frequencies of occurrence of these motion patterns were  compared between augmentation types. 3.2.5  Statistical Analysis Non-parametric analyses using Wilcoxon signed-rank tests were carried out within each  group of specimens to determine whether the two different augmentation methods resulted in a difference in translation offsets and ranges at both the screw head and screw tip. Pearson product-moment correlations between translation offsets and ranges with BMD were determined for all three augmentation methods. Chi-squared tests were used to compare the observed frequencies of occurrence of each motion pattern between augmentation methods. Analysis of variance was carried out to compare the current study with augmentation against a previous study (Chapter 2) [Tan et al. 2003] with normally inserted pedicle screws (without overdrilling). Statistical significance was assumed at the 95% level.  68  Chapter  3  Pedicle  Screw  Augmentation  (Vertical motion exaggerated) Figure 3.4  Schematic diagram of the screw positions at three or four instances of time during one loading cycle, starting with the screw head at the superior position (position 1). The screw head was displaced downwards through position 2 to the inferior position (positions 3 or 4) and back through position 2 to position 1. Four patterns of motion in the vertical plane were observed for the screws after augmentation with one of three techniques and these could be broadly characterised as a) pure rotation, b) translation, c) rotation-translation or d) double rotation. In (a), (b) and (c) the motions were mostly planar while in (d) there was out of plane motion.  69  Chapter 3  3.3  Pedicle Screw  Augmentation  Results One specimen in the wire-hook group was rejected as a result of loosening of the  transverse bar connecting the laminar hook to the vertical rod during cyclic testing. Thus, the sample size was seven for the wire-hook group and eight for the hook-cement and wire-cement groups. The screw motions after augmentation resulted in a maximum translation offset and range of -3.2 mm and 3.0 mm respectively (Figures 3.5 and 3.6).  Translation offset at screw head 0.8  -i  0.6  -  E E r  0.4  -  i  o.o  0  o  I wire  • hook  cement  0.2  isQ_  -0.2  Q  -0.4  tfl  -0.6 -0.8  J  Wire vs Hook  Hook vs Cement  Wire vs Cement  Translation offset at screw tip  0.4 -,  .<2 -1.6 -  Q  -2.0  -  -2.4 Wire vs Hook  Figure  3.5  Hook vs Cement  Wire vs Cement  Mean and standard deviation of translation offsets at the screw head and screw tip for the three groups of specimens. There were no significant differences between any of the pairs within each group, using the Wilcoxon signed-rank test.  70  Chapter  3  Pedicle  Screw  Augmentation  The mean translation offset and range for all forty-seven screws following augmentation were 0.0 mm (s.d. = 0.4) and 0.7 mm (s.d. = 0.7 mm) at the screw head and were -0.4 mm (s.d. = 0.7) and 0.7 mm (s.d. = 0.5) at the screw tip respectively. In our previous study (Chapter 2) [Tan et al. 2003], the corresponding translation offset and range at the screw head without any augmentation were -1.2 mm (s.d. = 1.3) and 2.4 mm (s.d. = 0.9) with an overdrilled hole and were -0.5 mm (s.d. = 0.5) and 1.4 mm (s.d. = 1.0) with a normal screw hole preparation using a standard blunt pedicle probing technique.  Translation range at screw head  2.4  I wire  • hook  cement  2.0 H |  1.6  I 1-2 8 as  0.8  I  CL  .<« 0.4 D  0.0 -0.4  Wire vs Hook  Hook vs Cement  Wire vs Cement  Translation range at screw tip  2.4  n  2.0 1.6 -  £ 1.2 A o -5. 0.8 H w  b  0.4 0.0  Figure 3.6  Wire vs Hook  Hook vs Cement  •  Wire vs Cement  Mean and standard deviation of translation ranges at the screw head and screw tip for the three groups of specimens. There were no significant differences between any of the pairs within each group using the Wilcoxon signed-rank test.  71  Chapter 3  Pedicle Screw  Table  P-values between  3.1  Translation Translation Translation Translation  Augmentation  analysed using two augmentation  offset, screw head offset, screw tip range, screw head range, screw tip  the non-parametric methods within  Wires - Hooks (n = 7) 0.47 0.11 0.34 0.15  Wilcoxon the same  signed-rank specimen.  Hooks - Cement (n = 8) 0.19 0.37 0.32 0.32  tests  Wire - Cement (n = 8) 0.42 0.53 0.10 0.47  Comparisons were made between translation offsets and ranges at the screw heads and screw tips. No statistically significant differences were determined between augmentation methods, at the 95% significance level. The small sample sizes in each group could have contributed to a Type II error. 3.3.1  Paired  Comparison  between  Augmentation  Methods  Comparing between the two augmentation methods for each of the three groups of specimens (wire-hook, hook-cement, wire-cement), no significant differences were observed between translation ranges and translation offsets at the screw head and screw tip (Figures 3.5 and 3.6). P-values for the Wilcoxon analyses are presented in Table 3.1. 3.3.2  Correlation  with  BMD  After augmentation with sublaminar wires, there was poor correlation between BMD and the translation ranges and offsets at the screw tip and screw head (Table 3.2).  Following  augmentation with calcium phosphate cement, there was a linear correlation (p = 0.017) between BMD and translation ranges at the screw tip. The positive correlation coefficient (r) of 0.58 (Figure 3.7) indicated that higher translation ranges were observed with higher bone mineral densities. Subsequent to augmentation with laminar hooks, there was a negative linear correlation (p = 0.049) between BMD and translation ranges at the screw head. Table  3.2  Translation Translation Translation Translation  Pearson BMD.  correlation  coefficient,  offset  and range  Laminar Hooks (n = 15)  Sublaminar Wires (n = 16)  0.02 0.39 -0.51* -0.35  0.17 -0.03 -0.13 -0.09  offset, screw head offset, screw tip range, screw head range, screw tip  * p < 0.05, indicating linear sublaminar wires or calcium  r, of translation  correlation. phosphate  Screws cement  were  pooled into augmentation.  against  Calcium Phosphate Cement (n = 16) 0.23 0.17 -0.24 0.58* laminar  hooks,  72  Chapter 3  Pedicle Screw Augmentation  0.8  •  i  — 0.7 E E 0.6 -  o '"S 0.2  •  A  •  0.0  H  0.0  0.2  0.4  0.6 BMD  Figure  3.7  3.3.3  Motion  0.8  1.0  1.2  (g/cm 2)  For the 16 screws augmented with translation range at the screw tip was 0.58, p = 0.017). Specimens with lower and vice versa. This could be the result phosphate cement into the more porous with lower BMD. All translation ranges small (less than 0.8 mm).  A  calcium phosphate cement, the linearly correlated with BMD (r = BMD had lower translation range of better infiltration of the calcium cancellous bone for specimens at the screw tip were relatively  Patterns  The motion pattern of thirteen of the sixteen (81 %) screws augmented with calcium phosphate cement was that of pure rotation (Table 3.3 & Figure 3.8). Chi-squared analyses determined this observation to be significant (p < 0.01) (Table 3.3). In contrast, pure rotation was seen in only seven of sixteen screws (44 %) augmented with sublaminar wires and three of fifteen screws (20 %) augmented with laminar hooks. None of the sixteen screws augmented with calcium phosphate cement demonstrated a translational motion pattern, while this was seen in seven of fifteen screws (47 %) augmented with laminar hooks and four of sixteen screws (25 %) augmented with sublaminar wires.  73  Chapter 3  Pedicle Screw  Augmentation  Table 3.3  Frequency and percentage of the number of screws that underwent pure rotation, translation, rotation-translation and double rotation. Laminar Sublaminar Calcium Hooks Wires Phosphate (n = 15) (n = 16) Cement (n = 16) Pure Rotation 7 (44 %) 3 (20 %)** 13 (81 % ) * Translation 7 (47 %)* 4 (25 %) 0 (0 %)** Rotation-Translation 5 (33 %) 2(13%) 3 (19%) Double Rotation 0 (0 %) 3 (19%)* 0 (0 %) ** Significantly lower number of occurrence than expected (p < 0.05), * Significantly higher number of occurrence than expected (p < 0.05), Chisquared test. Screws were pooled into laminar hooks, sublaminar wires or calcium phosphate cement augmentation.  Time of 10 seconds (5 cycles, 0.5 Hz) (a) Figure 3.8  Time of 10 seconds (5 cycles, 0.5 Hz) (b)  (a) The motion at the screw tip and screw head during the last five cycles is presented. Different modes of screw motion are demonstrated in these two specimens. The figure above exemplifies a pedicle screw undergoing rigid body rotation, with the screw head and screw tip being 180 degrees out of phase. The figure below demonstrates a screw undergoing mainly rigid body translation, with the screw head and screw tip moving in phase. Notice that the absolute motions are small (less than 1 mm). Screws augmented with calcium phosphate cement generally demonstrated a rotational pattern of rigid body motion (81%) with higher translation magnitude at the screw tip (in 8 of 13 pure rotations). While those with 74  Chapter  3  Pedicle  Screw  Augmentation  hooks often demonstrated rigid body translation (47%) with similar translation magnitudes at the screw head and screw tip (3 cases with higher magnitude at screw head, 2 cases with higher magnitude at screw tip and 2 cases with somewhat similar magnitudes at screw head and screw tip), (b) Two other modes of motion were identified. They were categorised as the rotation-translation and the double rotation modes. The screws in the rotation-translation mode moved in rigid motion translation for some period until an impingement (possibly within the pedicle) caused the screw tip to move in the opposite direction as the screw head. All three augmentation methods presented with some cases of this mode. The double rotation mode was mainly observed with wire augmentation. Loosened sublaminar wires could have resulted in loss of ability to pull the screw head medially and allowed for this out of plane motion. At the extreme positions of the screw, the wires were again taut and the screws were pulled medially.  3.4  Discussion Osteoporosis and poor bone mineral density present a significant challenge for internal  fixation within the lumbosacral spine. A number of augmentation strategies exist to enhance the rigidity of the pedicle screw-bone interface, but comparative evaluations of their biomechanical efficacy are rare.  In the current study, we evaluated one innovative and two common  augmentation techniques and compared both the magnitudes and patterns of motion at the screw head and tip under loading conditions that simulate those experienced in vivo.  Using a  similar biomechanical testing paradigm, we have compared the motion of pedicle screws inserted in the normal fashion, to the motion of pedicle screws in an overdrilled hole but without any form of augmentation (Chapter 2) [Tan et al. 2003]. The motion of the screws observed in the current study of three augmentation methods was significantly less than that observed previously in normally inserted pedicle screws (without overdrilling) and with overdrilling. The reduction in motion after augmentation is supported by the fact that the eight specimens in the previous study were used for the wire-hook group in the current study.  Furthermore the  reduction in motion that we observed with these augmentation techniques is consistent with previously published studies on individual augmentation methods [Coe et al. 1990, Soshi et al. 1991, Wittenberg et al. 1993, Butler et al. 1994, Halvorson et al. 1994, Hilibrand et al. 1996, Hu 1997, Lotz et al. 1997, Moore et al. 1997, Hasegawa et al. 1997, Yerby et al. 1998, Heller et al. 1998, Ignatius et al. 2001, Liljenqvist et al. 2001, Sarzieret al. 2002]. In the current study, paired comparisons of the displacement magnitudes at the screw head and screw tip did not identify any one of the three augmentation methods to be superior to the others.  It is recognised that our observation of no significant differences in translation  75  Chapter 3  Pedicle Screw Augmentation  offsets and ranges amongst the three augmentation techniques could represent a Type II error related to the sample size of each group. When motion patterns were analysed, screws augmented with calcium phosphate cement revealed more frequent rotational motion with minimal translation. On the other hand, translation was the primary motion observed for screws augmented with sublaminar wires (25 %) or laminar hooks (47 %). Determining the clinical significance of these differences in translation and rotation is somewhat speculative, however we note from our previous study (Chapter 2) [Tan et al. 2003] that screw motion in specimens with higher bone mineral densities tends to involve primarily rotation while screw motion in lower bone mineral density is both rotational and translational (similar to rotation-translation in Figure 3.8b). Characterizing the patterns of pedicle screw motion within osteoporotic vertebrae under physiologic loading may be important to our understanding of how such fixation loosens in the osteoporotic spine and how to most effectively prevent this occurrence.  By testing pedicle  screws under cephalo-caudad loads, Law et al (1993) demonstrated a rotational pattern of pedicle screw motion, with a fulcrum point at the base of the pedicle. This created localized failure of the cancellous bone within the vertebral body and pedicle, creating a "butterflyshaped" bony defect. More recently, we have described the motion of pedicle screws under cyclic axial compression and flexion bending moment to consist of both rigid body translation and rotation, with a center of rotation located somewhere between the screw tip and screw head (Chapter 2) [Tan et al. 2003]. These detailed motion patterns of the screw within the vertebral body may provide substantial insight into the effectiveness of different augmentation methods, but to date, the clinical significance of these patterns of movement at the bone-implant interface are unknown. The in vitro simulation of osteoporosis in human cadaveric vertebral bodies (beyond the pre-existing attenuation in bone mineral density of elderly donors) is difficult to do in a clinically meaningful manner. In this regard, the overdrilling of the pedicle canal (6.35 mm drill for a 6.0 mm diameter screw) creates a model of a loosened pedicle screw, representing a somewhat extreme form of bone compromise at the screw-bone interface. While a loosened screw model could be generated by cyclically loading a screw over an extended period of time, overdrilling creates a more consistent bony deficit which facilitates the comparison of different augmentation techniques. It is reasonable to assume that a pedicle screw inserted into a merely osteoporotic vertebral body would have greater contact with cancellous bone and thus less motion initially than if inserted into an overdrilled hole. Over time, however, with failure of the surrounding trabecular bone, this distinction between true osteoporosis and an overdrilled hole would be 76  Chapter 3  Pedicle Screw Augmentation  expected to diminish. The motion of screws in the current study may therefore be a reasonable model of the long-term motion patterns of screws in osteoporotic bone in vivo. Previous work on pedicle screw motion suggested an association between poor bone mineral density and the magnitude of motion, reflecting the intuitive notion that the biomechanical loading of a pedicle screw in osteoporotic bone will result in greater motion than in robust bone [Coe et al. 1990, Soshi et al. 1991, Halvorson et al. 1994, Bennett et al. 1997, Okuyama et al. 2001, Tan et al. 2003].  After augmentation with sublaminar wires, the  magnitude of pedicle screw motion did not correlate significantly with the bone mineral density of the specimen. This lack of correlation between magnitudes of motion and BMD implies that this technique negated the deleterious effect of osteoporosis on pedicle screw fixation rigidity. Interestingly, we noted a positive correlation between BMD and the translation range at the screw tip after calcium phosphate cement augmentation, suggesting that the motion was actually less in the more osteoporotic bone. This somewhat counter-intuitive observation may be attributable to the more osteoporotic and porous cancellous bone allowing for better, and possibly more distal, cement infiltration upon injection, which could provide a more stable boneimplant interface when hardened. While this result implies that calcium phosphate cement augmentation might provide better fixation in osteoporotic bone than in denser bone, the differences between true osteoporosis and the overdrilled pedicle should be considered before arriving at strong conclusions about this observation. Numerous studies that have assessed the fixation strength of pedicle screws in vitro have utilized the pullout test [Coe et al. 1990, Soshi et al. 1991, Butler et al. 1994, Halvorson et al. 1994, Hilibrand et al. 1996, Moore et al. 1997, Yerby et al. 1998, Cook et al. 2000, Liljenqvist et al. 2001, McKoy and An 2001a, Sarzier et al. 2002]. It is well recognized, however, that the loads applied to pedicle screws in pullout tests are not entirely representative of the loads that lumbar pedicle screws are subjected to in vivo. Rohlmann et al. (1997, 2000) measured the in vivo loads transmitted through spinal fixators during walking and determined that the main load components transmitted to the pedicle screw head consisted of an axial compressive force together with a flexion bending moment. Such in vivo loads could be simulated in vitro, as utilized in this study, by applying a cephalo-caudad compressive force with an offset so that the resultant forces acting on the pedicle screw head consist both of axial compression and a flexion bending moment [Cunningham et al. 1993, Tan et al. 2003]. The American Society for Testing and Materials (ASTM) has since proposed a similar testing methodology in their assessment of pedicle screw constructs [ASTM 2001]. These applied loads however did not  77  Chapter 3  Pedicle Screw Augmentation  include loads on the pedicle screw system during axial rotational motion, a possible mode of motion for the bedridden patient in recovery after surgery. Within the context of comparing the motion magnitudes and patterns between the three augmentation techniques, it is worth noting the differences in how each method augments the fixation of the loosened screw.  The sublaminar wires pull the pedicle screw head medially,  while compression between the screw and laminar hook pulls the screw head inferiorly towards the transverse bar (Figure 3.1).  This application of an external force (albeit from different  directions) onto the pedicle screw head creates a "three-point contact" along the length of the screw. As for augmentation using calcium phosphate cement, the cement is in contact with the screw presumably along its entire length, enhancing the interface between the screw threads and the cancellous bone. screw head.  The cement does not, however, provide any direct fixation to the  In view of these differences, one could speculate that cement augmentation  together with hook or wires might be the best way to deal with all forces to enhance fixation.  3.5  Conclusion This study compared the effectiveness of three augmentation methods in stabilizing a  pedicle screw in a model of osteoporotic screw loosening.  Compared to the results of the  previous study (Chapter 2) [Tan et al. 2003] of screw motion without augmentation, all three of the augmentation methods were successful in stabilizing the fixation, as reflected by a reduction in the magnitude of pedicle screw motion.  There were, however, no statistically significant  differences in the magnitudes of pedicle screw motion when the three augmentation methods were compared. The deleterious effect of low bone mineral density was offset by the addition of sublaminar wires. Differences in the motion patterns suggested that calcium phosphate cement augmentation might provide enhanced fixation over the other two techniques. The applicability of these findings to the clinical management of osteoporotic patients undergoing pedicle screw fixation in the lumbosacral spine, however, requires further in vivo study.  3.6  References  [1]  An YH, Burgoyne CR, Crum MS, Glaser JA. Current methods and trends in fixation of osteoporotic bone. In: Internal fixation in osteoporotic bone. An YH (Ed.). New York: Thieme, 2002:73-107.  [2]  American Society for Testing and Materials. ASTM F1717. Standard test methods for spinal implant constructs in a vertebrectomy model. ASTM International 2001.  [3]  Bennett GJ, Serhan HA, Sorini PM, Willis BH. An experimental study of lumbar destabilization. Spine 1997;22:1448-53.  [4]  Bostrom MPG, Lane JM. Future directions: augmentation of osteoporotic vertebral bodies. Spine 1997;24S:38S-42S. 78  Chapter 3  Pedicle Screw Augmentation  [5]  Butler TE, Asher MA, Jayaraman G, Nunley PD, Robinson RG. The strength and stiffness of thoracic implant anchors in osteoporotic spines. Spine 1994;19:1956-62.  [6]  Cloward RB. Posterior lumbar interbody fusion updated. Clin Orthop 1985;193:16-9.  [7]  Coe JD, Warden KE, Herzig MA, McAfee PC. Influence of bone mineral density on the fixation of thoracolumbar implants: a comparative study of transpedicular screws, laminar hooks and spinous process wires. Spine 1990;15:902-7.  [8]  Cook SD, Salkeld SL, Whitecloud TS, Barbera J. Biomechanical evaluation and preliminary clinical experience with an expansive pedicle screw design. J Spinal Disorder 2000;13:230-6  [9]  Cook SD, Barbera J, Rubi M, Salkeld SL, Whitecloud TS. Lumbosacral fixation using expandable pedicle screws: an alternative in reoperation and osteoporosis. Spine Journal 2001;1:109-14.  [10]  Cunningham BW, Setter JC, Shono Y, McAfee PC. Static and cyclical biomechanical analysis of pedicle screw spinal constructs. Spine 1993;18:1677-88.  [11]  Grant JP, Oxland TR, Dvorak MF. Mapping the structural properties of the lumbosacral vertebral endplates. Spine 2001;26:889-96.  [12]  Halvorson TL, Kelley LA, Thomas KA, Whitecloud TS, Cook SD. Effects of bone mineral density on pedicle screw fixation. Spine 1994;19:2415-20.  [13]  Hasegawa K, Takahashi HE, Uchiyama S, et al. An experimental study of a combination method using a pedicle screw and laminar hook for the osteoporotic spine. Spine 1997;22:958-63.  [14]  Heller KD, Prescher A, Schneider T, Block FR, Forst R. Stability of different wiring techniques in segmental spinal instrumentation: an experimental study. Arch Orthop Trauma Surg 1998;117:96-9.  [15]  Hilibrand AS, Moore DC, Graziano GP. pedicle bone. Spine 1996;21:445-51.  [16]  Hu SS. Internal fixation in the osteoporotic spine. Spine 1997;22:43S-8S.  [17]  Ignatius AA, Augat P, Ohnmacht M, Pokinskyj P, Kock HJ, Claes LE. A new bioresorbable polymer for screw augmentation in the osteosynthesis of osteoporotic cancellous bone: a biomechanical evaluation. J Biomed Mater Res 2001;58:254-60.  [18]  Law M, Tencer AF, Anderson PA. Caudo-cephalad loading of pedicle screws: mechanisms of loosening and methods of augmentation. Spine 1993;18:2438-43.  [19]  Liljenqvist U, Hackenberg L, Link T, Halm H. Pullout strength of pedicle screws versus pedicle and laminar hooks in the thoracic spine. Acta Orthop Belg 2001;67:157-63.  [20]  Lin PM, Cautilli RA, Joyce MF. Posterior lumbar interbody fusion. Clin Orthop 1983;180:154-68.  [21]  Lin PM. Posterior lumbar interbody fusion (PLIF): past, present and future. 1999;47:470-82.  [22]  Lotz JC, Hu SS, Chiu DFM, Yu Maria, Colliou O, Poser RD. Carbonated apatite cement augmentation of pedicle screw fixation in the lumbar spine. Spine 1997;22:2716-23.  [23]  Lowe TG, Tahernia AD. 2002;394:64-72.  [24]  McKoy BE, An YH. 2001a;19:545-7.  [25]  McKoy BE, Connor GS, An YH. An interlocking screw for fixation in osteoporotic bone. annual meeting Orthopaedic Research Society 2001b: 109.  [26]  Moore DC, Maitra RS, Farjo LA, Graziano GP, Goldstein SA. Restoration of pedicle screw fixation with an in situ setting calcium phosphate cement. Spine 1997;22:1696-705.  The role of pediculolaminar fixation in compromised  Clin Neurosurg  Unilateral transforaminal posterior lumbar interbody fusion. Clin Orthop  An expandable anchor for fixation in osteoporotic bone.  J Orthop Res 47th  79  Chapter 3  Pedicle Screw Augmentation  [27]  Okuyama K, Abe E, Suzuki T, Tamura Y, Chiba M, Sato K. Influence of bone mineral density of pedicle screw fixation: a study of pedicle screw fixation augmenting posterior lumbar interbody fusion in elderly patients. Spine Journal 2001;1:402-7.  [28]  Rohlmann A, Bergmann G, Graichen F. Loads on an internal spinal fixation device during walking. J Biomech 1997;30:41-7.  [29]  Rohlmann A, Graichen F, Weber U, Bergmann G. Monitoring in vivo implant loads with a telemetreized internal spinal fixation device. Spine 2000;25:2981-6.  [30]  Sarzier JS, Evans AJ, Cahill DW. Increased pedicle screw strength with augmentation in osteoporotic spines. J Neurosurg 2002;96:309-12.  [31]  Soshi S, Shiba R, Kondo H, Murota K. An experimental study on transpedicular screw fixation in relation to osteoporosis of the lumbar spine. Spine 1991;16:1335-41.  [32]  Tan JS, Kwon BK, Samarasekera D, Dvorak MF, Fisher CG, Oxland TR. Vertebral bone density a critical element in the performance of spinal implants. In: Spinal implants: are we evaluating them appropriately? ASTM STP 1431, Melkerson MN, Griffith SL and Kirkpatrick JS, (Eds.), ASTM International, West Conchohocken, PA, 2003:217-30.  [33]  Weinstein JN, Rydevik BL, Rauschning W. Anatomic and technical considerations of pedicle screw fixation. Clin Orthop 1992;284:34-46.  [34]  Wittenberg RH, Lee KS, Shea M, White AA, Hayes WC. Effect of screw diameter, insertion technique, and bone cement augmentation of pedicular screw fixation strength. Clin Orthop 1993;296:278-87.  [35]  Yerby SA, Toh E, McLain RF. Revision of failed pedicle screws using hydroxyapatite cement: a biomechanical analysis. Spine 1998;23:1657-61.  vertebroplasty  80  Chapter  4  Interbody  Device  Shape  and Size on Interface  Strength  Chapter 4 Interbody Device Shape and Size on Bone-Implant Interface Strength  Overview This chapter addresses the interbody device subsidence in the osteoporotic spine by comparing interface strength of interbody device with different shapes and sizes. Interbody devices that are larger in size and extend to the stronger peripheral regions of the endplates were found to result in stronger interface properties. Failure of underlying trabeculae was found to be isolated to the immediate region under the interbody devices.  The results in this study were presented at national and international conferences: 1) First Annual Meeting of the Alberta Provincial CIHR Training Program in Bone and Joint Health, Oct 24, 2003, Banff, Alberta, Canada. 2) 50th Annual Meeting of the Orthopaedic Research Society, Mar 7-10, 2004, San Francisco, California, USA. 3) First Asian Pacific Conference on Biomechanics, Mar 25-28, 2004, Osaka, Japan. 4) 31 st Annual Meeting of the International Society for the Study of the Lumbar Spine, May 30Jun 5, 2004, Portugal, Spain. 5) 38th Annual Meeting of the Canadian Orthopaedic Research Society, Jun 18-20, 2004, Calgary, Alberta, Canada. It was also recently published in a refereed journal: Tan JS, Bailey CS, Dvorak MF, Fisher CG, Oxland TR. Interbody device shape and size are important to strengthen the vertebra-implant interface. Spine 2005; 30:638-644.  81  Chapter  4  Interbody  Device  Shape  and Size on Interface  Strength  4.  Interbody Device Shape and Size on Bone-Implant Interface Strength  4.1  Introduction The use of interbody devices for spinal fusion procedures is well established [Weiner  and Fraser 1998, Zdeblick and Phillips 2003]. Implant subsidence is a concern with the majority of these devices, particularly in the osteoporotic patient, where lower bone mineral density (BMD) is associated with lower bone strength [Jost et al. 1998, Steffen et al. 2000, Lim et al. 2001, Grant et al. 2002]. To avoid implant subsidence, the strength of the vertebra-device interface must exceed the applied loads. Recent studies on regional endplate strength of lumbar and thoracic specimens have determined some consistent variability, with higher strength posterolateral^ and closer to the periphery [Grant et al. 2001, 2002, Bailey 2003]. In addition, a recent finite element model concluded that the placement of implants in the central area of the vertebral body may lead to early failure [Polikeit et al. 2003]. The same study proposed that cages should be designed to rely on the strong peripheral part of the endplate to reduce the risk of subsidence. Placement of interbody fusion devices on these stronger regions of the endplate would result in a stronger bone-implant interface. A study comparing three different cage positions on the surface of the end-plate, suggested that cages placed in the posterolateral aspect of the endplate were 20% stronger than cages placed centrally [Labrom 2002].  Rather than varying the position of  interbody fusion devices, it may also be possible to increase the vertebra-device interface strength by designing devices to lie on the strongest regions of the vertebrae.  Previous  biomechanical comparisons between interbody device shapes [Rapoff et al. 1997, Jost et al. 1998, Lund et al. 1998, Tsantrizos et al. 2000, Steffen et al. 2000, Krammer et al. 2001, Murakami et al. 2001] have found no differences between designs; however, these studies did not specifically focus on the effect of different cross-sectional shapes. A better understanding of some of the factors that lead to interbody device subsidence is necessary in order to address the problem effectively.  Interbody fusion device subsidence  results from failure at the bone-implant interface, rather than failure of the implant per se. We postulated that the region of bony failure under the various device shapes would provide clues regarding future strategies to prevent this complication. We hypothesize that interbody fusion devices with novel cross sectional shapes that extend over the stronger regions of the endplates will result in a stronger bone-implant interface. The primary purpose of this study, therefore, was to compare the in vitro failure load, strength, and stiffness of three cage shapes with comparable cross-sectional areas when compressed axially into superior thoracolumbar vertebral endplates. Two novel cage shapes, kidney and 82  Chapter 4  Interbody Device Shape and Size on Interface Strength  clover-leaf, were compared against a standard elliptical shape. The secondary purpose of this study was to document the pattern of failure of the endplate and its adjacent trabecular bone. We anticipated that there would be zones of trabeculae densification in the vertebral body, signifying the location of bone failure as it was compressed under the interbody devices.  4.2  Methods  4.2.7  Specimens  Forty-eight thoracolumbar vertebrae (T9-L2) were used in this study. These specimens were dissected from 19 cadavers and stored fresh frozen at -20 °C. There were 11 male and 7 female (one unknown gender) donors with a mean age of 77 years (max = 93, min = 58). Specimens were chosen to represent osteoporotic vertebrae in an attempt to create a 'worst case scenario' for our indentor testing. Radiographs and BMD scans (DXA, QDR 4500W, Hologic Inc., Bedford, MA) from both anterior-posterior and lateral projections were carried out on dissected T9-L2 spinal segments. The average lateral BMD of the vertebral bodies was 0.46 g/cm (max = 0.92, min = 0.25). 2  Exclusion criteria for specimens were 1) acute fracture, 2) vertebral body lysis or deformity secondary to remote fracture or other pathology (tumor), and 3) end-plate irregularity such as a significant Schmorl's node. 4.2.2  Test Groups  The three indentor shapes compared in this study were termed: a) kidney, b) elliptical, and c) clover-leaf.  Five sizes of each indentor shape were custom-manufactured.  These  indentors had solid indentation surfaces and the shapes were designed with equivalent crosssectional areas (Figure 4.1, Table 4.1).  The cross-sectional area of each indentor was  measured using imaging software (Image Pro 4.5, Media Cybernetics, Silver Spring, MD) and areas between indentor shapes were within 2% of each other, except for sizes 1 and 2 for the clover-leaf indentors. The higher area for these two indentors was due to the limitation of machining the desired inner radii for the smaller clover-leaf indentors. The forty-eight vertebrae were allocated into six groups of eight specimens. Each group had a similar mean lateral DXA BMD value of approximately 0.46 g/cm . 2  In the first three  groups, only vertebral bodies from T10, T12 and L2 were included (Table 4.2). In the last three groups, vertebral bodies from levels T9 to L2 were included.  83  Chapter 4  Interbody Device Shape and Size on Interface Strength  10mm Figure  4.1  The kidney, elliptical and clover-leaf shaped indentors. All indentors have a solid flat indentation surface. The indentors shown are all "Size 3", with an average cross-sectional area of 323 mm . 2  Table 4.1  Kidney Elliptical Clover-leaf  Average  cross-sectional  areas (mm ) 2  Size 1 125 122 141  of each solic f indentor.  Size 3 326 320 324  Size 2 219 221 232  For each size, the cross-sectional areas between indentor 2% of each other except for sizes 1 & 2 clover-leaf. Table 4.2  The 48 vertebral  Group 1; 20%, kidney Group 2; 20%, elliptical Group 3; 20%, clover-leaf Group 4; 40%, kidney Group 5; 40%, elliptical Group 6; 40% clover-leaf  bodies distributed  T9 1 1 2  Size 4 417 418 418 shapes  Size 5 512 513 524 were  within  into 6 groups.  Number of specimens T11 T12 L1 T10 2 3 3 3 2 3 1 2 2 1 2 2 1 1 2 1 1 1  L2 3 2 3 1 1 1  Average BMD (g/cm ) 0.463 0.462 0.468 0.460 0.444 0.460 2  Groups 1-3 consisted of only T10, T12 and L2 and had comparable numbers of each level distributed. Groups 4-6 consisted of 7 9 , T10, T11, T12, L1 and L2. The lateral DXA BMD in each group averaged between 0.44 and 0.47 g/cm 2  In the comparison between cage shapes, two percentages of coverage area of the indentors over the endplates were included: 20% and 40%. The endplate areas were calculated from the maximum AP and lateral dimensions by approximating the endplate as an elliptical shape [Panjabi et al. 1991, 1992, Higgins 2003].  Endplate areas of the 48 thoracolumbar  vertebrae ranged between 477 mm to 1602 mm with an average of 895 mm . The indentor 2  2  2  size which best approximated the desired endplate coverage percentage was chosen and the distribution in each test group is shown in Table 4.3.  84  Chapter 4  Interbody Device Shape and Size on Interface Strength  Table 4.3  Distribution of indentor size used in each test group.  Group 1: 20%, Group 2: 20%, Group 3: 20%, Group 4: 40%, Group 5: 40%, Group 6: 40%,  Kidney Elliptical Clover-leaf Kidney Elliptical Clover-leaf  Size 1 XXX XXX XXXX  Size 2 XXXX XXXX XXX XX XX XX  ndentor Size Size 3  Size 4  Size 5  X X X  X X  X XXXX XXXX XXXXX  Each 'X' represents one test. There were 8 specimens in each Group. specimens were rejected and not included in this Table. 4.2.3  Two  Test Protocol  Specimens were thawed in room temperature for between 4 to 8 hours before tests were carried out.  Both superior and inferior discs were sharply excised without damaging the  underlying endplates.  The cartilaginous endplates were removed.  Each vertebra was  individually secured in a custom built clamping device with the superior endplate aligned to the horizontal plane (Figure 4.2). The specimen was then centered within the testing machine using a weighted bob such that the geometric centre of the indentor face was aligned to the centre of the vertebral body. This centre was determined from the mid-point of the maximum anteriorposterior and lateral dimensions. The test jig was designed to allow for free rotation (frontal, lateral and axial rotations) of the indentors, about a centre of rotation located at the same geometric centre of the indentor face. This free rotation of the indentors will facilitate initial settling of the indentors onto the uneven geometry of the endplate, which will be reflected as an initial toe region in the load-displacement curve. The free rotation was achieved by a pair of opposing smooth articulating spherical concave and convex surfaces within the test jig (Figure 4.2). The concave surface was made out of UHMWPE and the convex surface was stainless steel. The indentor was compressed axially over the endplate at a rate of 0.2 mm/s to a depth of 20% of the vertebral body height, using a servohydraulic testing machine (Instron 8874, Instron Corporation, Canton, MA). Axial compressive force and displacement were measured. All endplates failed within this indentation depth.  85  Chapter 4  Interbody Device Shape and Size on Interface Strength  Figure 4.2  The test jig was designed with a pair of matching spherical concave and convex surfaces to allow for three dimensional rotations of the indentors about a centre of rotation located at the geometric centre of the indentor face (marked X). The specimen was rigidly clamped with the superior endplate lying in the horizontal plane. A rigid post (A) inserted through the vertebral foramen restricted posterior motion while a V-shape block (B) prevented anterior and lateral motion.  Failure load (in Newtons, N) and construct stiffness (in N/mm) were extracted from the load-displacement curves (Figure 4.3). Failure load was the maximum load before the gradient of the curve changes from positive to negative while stiffness was the slope of the linear portion of the load-displacement curve before failure occurred.  Failure strength (in, N/mm ) was 2  determined by dividing the failure load by the indentor area. Motion of the indentor and the vertebral body were measured during tests using an optoelectronic camera system (Optotrak 3020, Northern Digital Inc., Waterloo, Canada). The relative rotations of the indentor with respect to the vertebra were resolved into the anatomical planes of the specimen and defined as flexion-extension, axial rotation and lateral bending angles. These relative motions were summarised as rotation ranges in each of these planes, where rotation ranges were defined as the difference between the maximum and minimum angles for the test duration.  86  Chapter 4  Interbody Device Shape and Size on Interface Strength 1000 -i  1400 -.  800 H  _600 H T3  ro o  400 H  200 H  0  0 0  1 2 3 4 Displacement (mm)  5  0  (a) Figure 4.3  4.2.4  1 2 3 4 Displacement (mm)  5  (b)  Typical load-displacement curves where failure load was the maximum load before the gradient of the curve changes from positive to negative while stiffness was the slope of the linear portion of the curve before failure occurred. In 73% of the vertebrae tested, failure load was also the maximum load (a), while higher load beyond failure was measured in the other 27% of the tests (b).  Peripheral Quantitative Computer Tomography  Peripheral quantitative computer tomography (pQCT) scans (XCT2000, Stratec Medisintechnik, Pforzheim, Germany) were carried out on all specimens pre and post indentation test. A single 2.5 mm wide cut was scanned through the mid-sagittal plane at a resolution of 0.2 mm by 0.2 mm. The density profile in the vertebral body from the superior to the inferior endplates within this mid-sagittal slice was compared graphically between pre and post test pQCT scans to identify zones of trabeculae densification. Trabecular densification indicated permanent failure of the trabeculae.  The percentage increases in bone mineral densities in these zones  corresponding to a 20% applied strain were quantified and compared between indentor sizes. A pilot study carried out previously determined that the scan protocol and analysis was repeatable within a 4% coefficient of variation. In order to distinguish the effect of indentor size on permanent loss of vertebral height, we compared the changes in vertebral height following indentation tests. All specimens were 87  Chapter 4  Interbody Device Shape and Size on Interface Strength  indented to a strain of 20% and when unloaded some amount of permanent deformation resulted. Percentages of permanent deformation were calculated and compared between the two indentor sizes (20% and 40%) for all three indentor shapes.  Vertebral heights were  measured from pQCT images. 4.2.5  Statistical Analysis  Two-way factorial ANOVA was used to determine significant differences between cage shape and size. Post hoc Student's Newman-Keuls tests were carried out to determine the significant factors, with a significance level of 95%. Unpaired Student's t-tests were used to compare the increase in 1) bone mineral densities and 2) permanent deformation between tests with 20% and 40% indentor coverage area over endplate area, with a significance level of 5%. 4.3  Results  Failure load was significantly affected by cage shape (p = 0.0005) and percentage of endplate covered by the indentor (p = 0.000003) (Figure 4.4). The failure loads for the larger clover-leaf indentor (40% endplate coverage) were observed to be 54% and 45% higher when compared to the kidney and elliptical shapes respectively. The failure loads for the smaller clover-leaf indentor (20% endplate coverage) were observed to be 62% and 60% higher when compared to the kidney and elliptical shapes respectively. No significant difference in failure load was found between the kidney and the elliptical shapes (p = 0.72) for both indentor sizes. The failure loads observed with the larger indentor were 75% greater than those for the smaller indentors. This was most significant for the elliptical shape (83%) and less so for the kidney (75%) and the clover-leaf (66%). Failure strength, defined as the failure load divided by the indentor area, was not significantly different between the large and small indentors for any of the three indentor shapes (p = 0.65) (Figure 4.5).  Cage shape remained a significant factor (p = 0.0004), with a  significantly higher failure strength for the clover-leaf shaped indentor. The failure strength for the larger clover-leaf indentor were observed to be 59% and 49% higher compared to the kidney and the elliptical shapes respectively. The failure strength for the smaller clover-leaf indentor were observed to be 79% and 85% higher compared to the kidney and the elliptical shapes respectively. No significant difference in failure strength was found between the kidney and the elliptical shapes (p = 0.87).  88  Chapter 4  Interbody Device Shape and Size on Interface Strength  p=0.01 p=0.04 p=0.03  p=0.003 0  20%  • 40%  kidney Figure  4.4  elliptical  clover-leaf  Mean failure load for the kidney, elliptical and clover-leaf indentor groups, with 20% and 40% endplate coverage. Failure load for the clover-leaf shaped indentors were significantly higher than both kidney and elliptical shapes for both indentor sizes. Failure load was significantly higher for larger indentors. Error bars indicate standard deviation.  p=0.05  E  Z, t  j|5  co  9.0 8.0 7.0 6.0 5-0 4.0 3.0  p=0.02  -I -1  H 4.4 2.7  2.7  2.9  0  20%  • 40%  CD  2.0 CD 1.0 LL 0.0 kidney Figure 4.5  elliptical  clover-leaf  Mean failure strength for the kidney, elliptical and clover-leaf indentor groups, with 20% and 40% endplate coverage. The clover-leaf shaped indentors have significantly higher failure strength than both kidney and elliptical shapes. Error bars indicate standard deviation.  Stiffness of the construct was significantly affected by cage shapes (p = 0.0004) but not by the surface area of end-plate coverage by the indentor (p = 0.13) (Figure 4.6). Consistent with failure load and failure stress, the clover-leaf shaped indentor resulted in higher construct 89  Chapter 4 stiffness  Interbody Device Shape and Size on Interface Strength  over the  kidney a n d the  indentor were observed s h a p e s respectively. 79%  and  35%  to b e 4 7 %  and 50%  T h e stiffnesses  for the  h i g h e r c o m p a r e d to the  larger  clover-leaf  kidney a n d the  elliptical  T h e s t i f f n e s s e s for the s m a l l e r c l o v e r - l e a f i n d e n t o r w e r e o b s e r v e d to  higher  significant difference  elliptical s h a p e s .  compared  to  in s t i f f n e s s w a s  the  kidney  and  the  f o u n d b e t w e e n the  elliptical  shapes  be  respectively.  No  k i d n e y a n d t h e e l l i p t i c a l s h a p e s (p  =  0.32).  p=0.04 1200  I «  1000  800  600  |  400  200  H  805 708  549  H H H  525  536  m  20%  •  40%  0 kidney  Figure  4.6  elliptical  Mean stiffness for the kidney, elliptical with 20% and 40% endplate coverage. with clover-leaf shaped indentors. Error  Peripheral-QCT  results  indicated  that  clover-leaf  and clover-leaf indentor groups, Higher stiffnesses were found bars indicate standard deviation.  densification  of  trabeculae  occurred  directly  b e n e a t h the indentors, leaving the o s s e o u s endplate a n d immediate underlying cancellous  bone  intact ( F i g u r e 4.7).  cage  shapes  had  quantified  by  a  T h e z o n e with d e n s i f i e d  semi-elliptical comparing  indentation ( F i g u r e 4.8). for  shape.  densities Increase  trabeculae  Increase  within  the  in  in t h e s a g i t t a l  bone  boxed  mineral  region  in b o n e m i n e r a l d e n s i t y w a s  the s m a l l e r indentors with 2 0 % e n d p l a t e c o v e r a g e  i n d e n t o r s with 4 0 % e n d p l a t e c o v e r a g e (by 24%, s . d . =  ( b y 39%,  in  p l a n e f o r all t h r e e  density Figure  in t h i s  4.7  before  region and  was after  s i g n i f i c a n t l y h i g h e r (p = 0 . 0 0 0 1 ) s . d . = 14%)  t h a n for the  larger  10%).  90  Chapter  Interbody  4  Device  a) Pre-lndentation Test  S: Superior Figure 4.7  Shape  and Size on Interface  b) Post-Indentation Test  c) 2 m m slice of same specimen  ( i n d e n t e d ) e n d p l a t e , I: I n f e r i o r e n d p l a t e .  Images of a specimen indented with an elliptical indentor through the mid sagittal plane a) pre and b) post indentation tests using peripheral quantitative computer tomography (pQCT). Zones of trabecular densifications occurred beneath the indented endplate for all three indentor shapes in a semi-elliptical pattern. The osseous endplate and immediate underlying cancellous bone was subsided but remained intact. Density in the boxed region was compared quantitatively, c) A 2 mm slice of the same specimen with a semi-elliptical trabecular densification pattern (arrows) under the indented endplate.  Indented e n d p l a t e  Figure 4.8  Strength  Inferior e n d p l a t e  Density profile across the vertebral body from superior endplate (left) to inferior endplate (right) at the mid-sagittal plane. Each point on the graph represents the average density of a column of pQCT data points within the boxed region. Densification of trabeculae (circled) was observed 91  Chapter 4  Interbody Device Shape and Size on Intedace Strength under the indented endplate only and did not involve other regions in the vertebral body. This observation was consistent for all three indentor shapes.  The permanent deformation of the vertebrae under the smaller indentors (20% endplate coverage) was significantly higher (p < 0.02) than with larger indentors (40% coverage) after 20% applied strain for all three indentor shapes (Table 4.4). This higher permanent deformation for the smaller indentors occurred with associated lower failure loads (Figure 4.4). Permanent deformation for kidney and clover-leaf shaped indentors were not significantly different from the elliptical indentors for both percentages of endplate coverage. Table 4.4  Permanent deformation after 20% applied strain, determined by comparing the vertebral height before and after indentation tests at the mid sagittal plane from pre and post test pQCT scan data.  20% coverage 40% coverage  Kidney 16.6%* (1.8) 9.9% (5.7)  Elliptical 14.2%* (3.0) 9.1% (4.1)  Clover-leaf 12.9%* (3.8) 6.8% (4.0)  p-value 0.01 0.01 0.02 * significantly higher than corresponding value in 40% coverage, at 95% level. Permanent deformation is presented as the percentage change in vertebral height from the original height. Values in parenthesis denote standard deviation. Permanent deformations were significantly higher with the smaller indentors (p < 0.02). Note that the applied load was different in each test.  The rotational motions of the indentors were not significantly different between cage shapes or between sizes. The rotation in the three planes of motion were small, with indentors rotating most profoundly in the sagittal plane; an average of 1.5° (flexion-extension), with no preferential direction between flexion or extension. Axial rotation of the indentor with respect to the vertebral body averaged 0.4° and was much less than flexion/extension or lateral bending. Lateral bending of the indentors averaged 0.9°, with a trend to less motion when the kidney shaped cages were tested (0.6°). The results from two vertebrae (same donor, female, age = 58) were rejected as the failure loads (2516 N and >3000 N) were 500% and 587% higher than the means of the respective groups (504 N ± 157 N and 511 N ± 203 N). These two were clear outliers as their failure loads were at least twelve standard deviations above the respective means. These two specimens had the highest BMD amongst the 48 vertebrae and lowest age amongst the donors and clearly did not fit our inclusion criterion which was to include vertebral bodies that were somewhat osteoporotic. 92  Chapter  4  Interbody  4.4  Discussion  Device  Shape  and Size on Interface  Strength  The purpose of this study was to compare two novel cage shapes with a standard elliptical shape using an axial compressive test model. The current study indicates that the novel clover-leaf shape results in improved bone-implant interface properties in vitro. These properties of the clover-leaf design may modify the clinical behaviour of an intervertebral implant with this foot-print, and may potentially reduce implant subsidence when compared to the standard elliptical shape. The average failure loads of 504-1358 N for the various indentors in this study were comparable to the values reported by Hollowell et al. (1996) (262-1473 N) and Hasegawa et al. (2001) (510-1335 N). The average construct stiffness of 396-805 N/mm in this current study also compared reasonably with the values reported by Hasegawa et al. (2001) (385-541 N/mm). Jost et al. (1998) and Lund et al. (1998) both reported greater amounts of flexion-extension and less axial rotation of the interbody devices they tested. This is consistent with our results. The lower values of indentor motion in our study may be due to the single vertebrae to which we applied the indentor and the alignment carried out before each test. Whereas in the studies of Jost et al. (1998) and Lund et al. (1998) an interbody device was placed within a functional spinal unit and thus one would expect a larger amount of motion. A secondary purpose of this study was to document the pattern of failure of the endplate and/or sub-endplate trabecular bone when compressed by an indentor. It is by understanding the pattern of trabeculae failure that we can begin to address the issue of implant subsidence. The current study documented the pattern of failure of the trabecular bone beneath the indentor device and identified a semi-elliptical zone of trabecular densification and permanent damage occurring beneath an intact osseous end-plate and beneath some intact sub-cortical trabecular bone. Hasegawa et al. (2001) observed trabeculae failure directly beneath the cage. Their result was consistent with our study, as the zone of trabecular densification did not extend into the central region of the vertebral body. A similar semi-ellipsoidal shaped zone of trabecular densification under the endplate was previously described in an in vitro model with titanium mesh cages and pedicle screws [Labrom 2002]. That model used functional spinal units and thus mimicked the in vivo situation more closely. The shape of the failure zone in Labrom's study is similar to our findings. The mode of trabecular failure is similar to that seen when there is shear failure of horizontal struts under the edge of the indentors together with buckling of vertical struts under the indentors. This zone of trabecular failure could potentially be reinforced in selected patients to increase failure load and reduce implant subsidence. 93  Chapter 4  Interbody Device Shape and Size on Interface Strength  We recognize several limitations in the current study. In vitro biomechanical models such as ours do not take into account the remodelling that occurs following surgery. In vivo bone growth or resorption occurs depending on the load experience by the bone. Thus the current study is applicable to the immediate post-operative period before any biological changes have occurred. Furthermore, our model, which utilizes individual vertebral bodies does not take into account the dynamic load sharing which occurs within a functional spinal unit in vivo or from the addition of posterior instrumentation. While we used a compression model with unrestricted rotation, the in vivo load could be more compression-flexion or compression-extension etc., depending on the level, use of posterior instrumentation or patient's activity.  Despite these  limitations, we believe that the compressive model in this study served its intended purpose of a biomechanical comparison between indentor shapes. In order to focus on our primary research question and limit the number of variables, we decided on the use of solid indentors. A recent study assessed the axial compressive strength of interbody devices with full as opposed to peripheral contact and found similar mechanical strength [Steffen et al. 2000]. Using thin walled indentors with synthetic bone graft surrogates would most closely represent the in vivo situation, but such indentors would add several variables to the comparison, such as the indentor surface area, graft area, and thickness of the wall of the hollow indentor. The comparison between indentor shapes was carried out with indentors of the same cross-sectional areas in order to control the potentially confounding influence of indentor area. To address the effect of varying the area of the indentor, we included indentors with surface areas of 20% and 40% of the endplate area. Moreover, we compared failure strength, where failure strength is failure load normalised to indentor areas. By using percentage of endplate coverage, we also took into consideration the fact that endplate sizes varied between individuals and between spinal levels.  Our results indeed indicated that larger indentor surface area  resulted in higher failure loads. The cross sectional area of the largest clover-leaf cages which can be placed on the endplates is smaller that the cross sectional area of the largest elliptical cage. Thus one could argue that the failure load of the largest elliptical cage (assuming a size equivalent to 60% endplate coverage) could approach or even exceed the failure load of the largest clover-leaf cage (Figure 4.4). Failure strength data, however, indicated that the clover-leaf shape devices exhibited higher strength, regardless of coverage area (Figure 4.5). Attempts to minimize the contact area of the intervertebral device serve to increase the end-plate area outside the device available for bone grafting and eventual fusion. When intervertebral device designs attempt to 94  Chapter  4  Interbody  Device  Shape  and Size on Interface  Strength  balance the strength profile of the device by maximizing contact area and the opportunity for bone growth on the outside of the intervertebral device, the true benefits of this novel device shape may provide some assistance. However it is not sure if the fusion process would be compromised with the use of a clover-leaf shaped device with peripheral excursion. Moreover, the increased construct stiffness with this device shape (Figure 4.6) might be detrimental to the fusion process. 4.5  Conclusion  The results of the current study support an earlier proposal [Polikeit et al. 2003] that cages should be designed with more peripheral excursion to sit above the stronger regions of the endplates. We conclude that a clover-leaf shaped indentor was biomechanically superior to both elliptical and kidney shaped indentors of similar surface areas, and could potentially influence implant subsidence.  Moreover, larger devices should be used to increase contact  area and subsequently increase failure load and therefore reduce device subsidence. Failure of the underlying trabeculae was isolated to the immediate region under the indentor and thus it seems logical that future intervertebral devices should extend over the stronger regions of the endplate to reduce implant subsidence. 4.6  References  [1]  Bailey CS. Local strength and regional bone mineral density profiles of the thoracolumbar endplate. M.Sc.(Surg) Thesis 2003, University of British Columbia.  [2]  Grant JP, Oxland TR, Dvorak MF. Mapping the structural properties of the lumbosacral vertebral endplates. Spine 2001, 26:889-896.  [3]  Grant JP, Oxland TR, Dvorak MF, Fisher CG. The effects of bone density and disc degeneration on the structural property distributions in the lower lumbar vertebral endplates. J Orthop Res 2002, 20:1115-1120.  [4]  Hasegawa K, Abe M, Washio T, Hara T. An experimental study on the interface strength between titanium mesh cage and vertebra in reference to vertebral bone mineral density. Spine 2001, 26:957-963.  [5]  Higgins KB. The biomechanics of osteoporotic spine: evaluation of compression fracture intervention. State University of New Jersey, PhD dissertation 2003.  [6]  Hollowell JP, Vollmer DG, Wilson CR, Pintar FA, Yoganandan N. Biomechanical analysis of thoracolumbar interbody constructs. How important is the endplate? Spine 1996, 21:1032-1036.  [7]  Jost B, Cripton PA, Lund T, Oxland TR, Lippuner K, Jaeger P, Nolte LP. Compressive strength of interbody cages in the lumbar spine: the effect of cage shape, posterior instrumentation and bone density. Eur Spine J 1998, 7:132-141.  [8]  Krammer M, Dietl R, Lumenta CB, Kettler A, Wilke HJ, Buttner A, Claes L. Resistance of the lumbar spine against axial compression forces after implantation of three different posterior lumbar interbody cages. Acta Neurochir (Wien) 2001, 143:1217-1222. 95  Chapter 4  Interbody Device Shape and Size on Interface Strength  [9]  Labrom RD. The effect of cage positioning on lumbosacral vertebral endplate failure in compression. M.Sc.(Surg) Thesis 2002, University of British Columbia.  [10]  Lim TH, Kwon H, Jeon CH, Kim JG, Sokolowski M, Natarajan R, An HS, Andersson GB. Effect of endplate conditions and bone mineral density on the compressive strength of the graft-endplate interface in anterior cervical spine fusion. Spine 2001, 26:951-956.  [11]  Lund T, Oxland TR, Jost B, Cripton PA, Grassmann S, Etter C, Nolte LP. Interbody cage stabilisation in the lumbar spine: biomechanical evaluation of cage design, posterior instrumentation and bone density. J Bone Joint Surg Br 1998, 80:351-359.  [12]  Murakami H, Boden SD, Hutton WC. Anterior lumbar interbody fusion using a barbell-shaped cage: a biomechanical comparison. J Spinal Disord 2001, 14:385-392.  [13]  Panjabi MM, Takata K, Goel V, Federico D, Oxland TR, Duranceau J, Krag M. Thoracic human vertebrae: quantitative three-dimensional anatomy. Spine 1991, 16:888-901.  [14]  Panjabi MM, Goel V, Oxland TR, Takata K, Duranceau J, Krag M, Price M. vertebrae: quantitative three-dimensional anatomy. Spine 1992, 17:299-306.  [15]  Polikeit A, Ferguson SJ, Nolte LP, Orr TE. Factors influencing stresses in the lumbar spine after the insertion of intervertebral cages: finite element analysis. Eur Spine J 2003, 12:413-420.  [16]  Rapoff AJ, Ghanayem AJ, Zdeblick TA. Biomechanical comparison of posterior lumbar interbody fusion cages. Spine 1997, 22:2375-2379.  [17]  Steffen T, Tsantrizos A, Aebi M. Effect of implant design and endplate preparation on the compressive strength of interbody fusion constructs. Spine 2000, 25:1077-1084.  [18]  Tsantrizos A, Baramki HG, Zeidman S, Steffen T. Segmental stability and compressive strength of posterior lumbar interbody fusion implants. Spine 2000, 25:1899-1907.  [19]  Weiner BK, Fraser RD. Spine Update: Lumbar interbody cages. Spine 1998, 23:634-640.  [20]  Zdeblick TA, Phillips FM. Interbody cage devices. Spine 2003, 28S:S2-S7.  Human lumbar  96  Chapter  5  Screws  and Cement  on Interbody  Device  Interface  Strength  Chapter 5 Cement Augmentation of Vertebral Screws Enhances the Resistance to Subsidence of Interbody Device  Overview  This chapter addresses the effects of cement augmentation of screws on interbody devicevertebral body interface properties. Pedicle screw insertion was found to disrupt the underlying trabecular bone and reduce the interbody device-vertebral body interface properties. Cement augmentation of pedicle screws restored and further improved cage-vertebra interface properties.  No differences in interface properties were found between cement augmented  anterior vertebral body screws and pedicle screws.  Cement augmentation of screws could  reduce interbody device subsidence in the osteoporotic spine. The results of this study were presented at: 1) 5th Combined Meeting of the Orthopaedic Research Societies of the USA, Canada, Japan and Europe, Oct 10-13, 2004, Banff, Alberta, Canada. 2) 33rd Annual Meeting of the International Society for the Study of the Lumbar Spine, Jun 1217, 2006, Bergen, Norway. It was also submitted for publication in a refereed journal: JS Tan, CS Bailey, MF Dvorak, CG Fisher, PA Cripton, TR Oxland. Cement augmentation of vertebral screws enhances the interface strength between interbody device and vertebral body. Submitted to Spine (7 Nov 2005).  97  Chapter  5.  5  Screws  and Cement  on Interbody  Device  Interface  Strength  Cement Augmentation of Vertebral Screws Enhances the Resistance to Subsidence of Interbody Devices  5.1  Introduction The prevalence of osteoporosis, vertebral fracture [EPOS 2002] and spinal deformity  [Jackson et al. 200 0] is known to be associated with age.  With the percentage of total  population aged 65 and older projected to increase by 24% to 54% between 2000 and 2020 in the world's industrialized nations [Anderson and Hussey 2000], the number of patients presenting with osteoporotic fracture and deformity will become more frequent.  Surgical  stabilization of the osteoporotic spine is indicated in the presence of significant neurological deficit, instability, deformity, or in revision surgery [Dvorak and Fisher 1999]. Both pedicle screw and anterior vertebral body screw fixation [Wuisman et al. 2000, Kostuik and Shapiro 2003, Heini 2005], have been used in these situations to supplement anterior interbody devices. It is well known that the performance of most spinal implants is compromised in the presence of low bone mineral density (BMD). Pedicle screw loosening [Coe et al. 1990, Soshi et al. 1991, Yamagata et al. 1992, Okuyama et al. 1993, 2001, Halvorson et al. 1994, Pfeiffer et al. 1996, Hasegawa et al. 1997] and interbody device subsidence [Oxland et al. 1996, Lund et al. 1998, Jost et al. 1998, Tsantrizos et al. 2000] are common modes of failure of spinal fixation in the presence of osteoporosis. Numerous biomechanical studies have shown that cement augmentation of pedicle screws [Zindrick et al. 1986, Soshi et al. 1991, Wittenberg et al. 1993, Lotz et al. 1997, Moore et al. 1997, Yerby et al. 1998, Sarzier et al. 2002, Cook et al. 2004, Renner et al. 2004, Tan et al. 2004] and anterior vertebral body screws [Ignatius et al. 2001, Bai et al. 2001, Schultheiss et al. 2004] results in improved screw-bone interface strength. It is not surprising that the clinical use of bone cement to augment pedicle screws and anterior vertebral body screws in osteoporotic patients has gained popularity in recent years [Wuisman et al. 2000, Kostuik and Shapiro 2003, Heini 2005]. Surgical strategies to enhance the resistance of interbody devices (cages) to subsidence in the osteoporotic spine have received less attention than has augmentation of screw fixation. We recently reported that cage-vertebra interface failure loads could be increased by greater than 45% by using uniquely shaped interbody devices that engaged the peripheral regions of the endplate [Tan et al. 2005 (Chapter 4)]. In that study, trabecular bone failure occurred in a semi-elliptical zone underlying the interbody devices. That failure region led us to hypothesize that local cement augmentation could increase the strength at the cage-vertebra interface. The effects of pedicle screws, anterior vertebral body screws or cement placed within the vicinity of that zone of damage on cage-vertebra interface strength has not been previously studied. 98  Chapter  5  Screws  and Cement  on Interbody  Device  Interface  Strength  Clinically, cement has been injected into vertebral bodies with the intention of reducing interbody device subsidence [Heini 2005]. However, we are not aware of any biomechanical study that addresses cage-vertebra interface strength when underlying pedicle or vertebral body screw are inserted with or without cement augmentation. The main purpose of this study was to determine if cement augmentation of pedicle screws could improve interface properties of interbody devices, and therefore their resistance to subsidence. The secondary purpose was to determine any differences between pedicle screws with cement augmentation and anterior vertebral body screws with cement augmentation. We investigated these questions with both elliptical and clover-leaf shaped interbody devices, for three screws and cement configurations: 1) pedicle screws only, 2) pedicle screws with cement and 3) anterior vertebral body screws with cement. 5.2  Methods  5.2.7  Experimental Protocol  Individual vertebra was prepared with 1 of 3 different screw and cement configuration and axially compressed with 1 of 2 shaped indentor onto the superior endplate until failure. Failure load and failure strength were compared. 5.2.2  Specimens  Forty-eight thoracolumbar vertebrae (T9-L2) from 21 human cadavers were used in this study. The donors' ages ranged between 62 and 93 years with a mean age of 77 years. There were 13 male and 7 female donors (one unknown gender). All specimens were kept frozen at 20 C after harvesting from donors in the University of British Columbia Tissue Donation 0  Program. This study was approved by our Institutional Clinical Research Ethics Board. Areal bone mineral densities (aBMD) of the vertebral bodies were measured from the lateral projection of dissected T9-L2 spinal segments using a dual-energy x-ray absorptiometry (DXA) machine (QDR 4500W, Hologic Inc., Bedford, MA) with methodology previously described by our group in detail [Sran et al. 2005]. The specimens were sectioned and divided into six groups with 8 vertebral bodies per group such that each group had similar mean BMD, standard deviation and comparable distribution of spinal levels (Table 5.1). Mean BMD for the 48 vertebrae was 0.46 g/cm and ranged between 0.29 and 0.72 g/cm . 2  2  Exclusion criteria for specimens were 1) acute fracture, 2) vertebral body lysis or deformity secondary to remote fracture or other pathology (tumor), and 3) end-plate irregularity such as a significant Schmorl's node or fracture. 99  Chapter 5  Screws and Cement on Interbody Device Interface Strength  Table 5.1  Distribution of spinal each group.  Screw Configuration Pedicle screw alone Pedicle screw alone Pedicle screw + cement Pedicle screw + cement Anterior body screw + cement Anterior body screw + cement  Test  and mean  bone  mineral  Number of Specimens T10 T11 T12 L1 1 3 3  density  Indentor Shape elliptical  T9 1  cloverleaf  1  1  2  1  2  1  elliptical  2  -  3  -  3  -  cloverleaf  -  2  2  1  3  -  elliptical  1  3  1  2  -  1  cloverleaf  2  -  3  -  3  -  s.d. = standard 5.2.3  levels  L2 -  (BMD) in  Average BMD, g/cm (s.d.) 0.467 (0.10) 0.462 (0.14) 0.462 (0.12) 0.463 (0.12) 0.469 (0.10) 0.463 (0.08) 2  deviation  Groups  The experiment was designed to test the main effects and two way interaction between three different screw-cement combinations: i) pedicle screws only, ii) pedicle screws with cement, and iii) anterior vertebral body screws with cement, and two different interbody device shapes: i) elliptical, and ii) clover-leaf. The 6 groups of vertebrae (Table 5.1) were assigned to a screw-cement configuration and cage shape. 5.2.4  Compression  Failure  Tests  The specimens were tested under compression with a shaped indentor over the superior endplate of each vertebra. The test setup and protocol used was similar to our previous study on interbody device shape and size [Tan et al. 2005 (Chapter 4)]. Four different sizes of each custom-shaped stainless steel indentor, with cross sectional areas of approximately 519, 418, 322 and 227 mm , were used (Figure 5.1). Indentor size for each test was chosen such that 2  they covered an area as close to 40% of the endplate area as possible. Endplate areas were calculated from the maximum anterior-posterior (AP) and lateral dimensions by approximating the endplate as an elliptical shape [Panjabi et al. 1991, 1992, Higgins 2003]. Cross-sectional areas of the superior endplate of the 48 thoracolumbar vertebrae ranged between 517 mm to 2  1493 mm with an average of 951 mm . The distribution of indentor size used for the tests was 2  2  similar between the siz groups (Table 5.2).  Overall mean percentage endplate coverage  achieved was 40% with means in the 6 groups ranging from 38% to 42%.  100  Chapter 5  Screws and Cement on Interbody Device Interface Strength  10 llill|ll!l  Figure 5.1  5.2.5  III!  mm  i l l ! JM  I! 11J1111  Ull|  Four sizes of clover-leaf and elliptical shaped indentors were used for the tests. Cross-sectional areas were similar for each size.  Test Procedure Specimens were thawed at room temperature (23°C ± 2°C) for between 4 to 8 hours  before screw insertion and cement injection.  Both superior and inferior discs were sharply  excised without damaging the underlying endplates, and the cartilaginous endplates were dissected away. Anterior vertebral body screws and pedicle screws (CD Horizon, Medtronic Sofamor-Danek and Moss Miami, DePuy Spine) were implanted by a fellowship trained spine surgeon (Dr Chris Bailey) in clinically relevant orientations. Screw diameters were 6 mm and screw lengths were 40-55 mm. Screws from both manufacturers were randomly used in all groups. Anterior vertebral body screws were inserted with bicortical purchase on lateral cortices  101  Chapter  5  Screws  and Cement  on Interbody  Device  Intedace  Strength  while pedicle screws were inserted to 80% of the vertebral body depth. Radiographs in the axial and lateral projections confirmed proper positioning of screws. Table 5.2  Distribution of indentor size used in each test group and mean percentage area of endplate coverage achieved. Screw Indentor Indentor Size Configuration Shape (Indentor Area) Size 2 Size 3 Size 4 Size 5 (227 mm ) (322 mm ) (418 mm ) (519 mm )  Mean Percentage Area of Endplate Coverage (s.d.) Pedicle screw elliptical XX XXX 41% XXX alone (4%) Pedicle screw X XXX XXX X cloverleaf 40% alone (5%) Pedicle screw + elliptical X XXX XX XX 38% cement (3%) Pedicle screw + cloverleaf X XX XXXX 42% cement (3%) Anterior body elliptical XXX XX X XX 40% screw + cement (5%) Anterior body XXX XXXX X 42% cloverleaf screw + cement (1%) Each 'X' represents one test. There were 8 specimens in each group. One test was rejected as a result of metal-on-metal contact between the indentor and the custom built specimen clamps and was not included in this Table. 2  2  2  2  In the cemented groups, approximately 2 ml of polymethylmethacrylate (PMMA) bone cement (Osteobond, Zimmer, Warsaw, IN) was injected into the screw tracts after removal of a normally inserted screw. The same screw was immediately reinserted after cement injection. This method of cement injection into a screw tract is used clinically [Wuisman et al. 2000] and has been applied experimentally [Soshi et al. 1991, Lotz et al. 1997]. The cement was allowed to cure for 1 to 2 hours before indentation tests were carried out. The vertebral bodies were consistently considered as the inferior body of a fusion construct where the interbody device was in contact with the superior endplate of the vertebral body being tested. For each test, the vertebra was secured in a custom built clamp and the superior endplate was aligned in the horizontal plane with the aid of an adjustable vise (Figure 5.2). Following that, the specimen was centred within the testing machine using a weighted bob such that the geometric centre of the indentor face was aligned with the centre of the vertebral body. The centre of the vertebral body was determined from the mid-point of the maximum AP dimension in the mid-sagittal plane. The test jig was designed to simulate a spherical joint and 102  Chapter  5  Screws  and Cement  on Interbody  Device  Interface  Strength  to allow for unconstrained rotation in the frontal, lateral and transverse planes of the indentor, about a centre point located at the geometric centre of the indentor face. A pair of opposing smooth articulating spherical concave and convex surfaces within the test jig made up the spherical joint. The concave surface was made out of UHMWPE and the convex surface was stainless steel. The compression stiffness of this indentor setup was extablished experimentally to be 5140 N/mm. Free rotation of the indentor during application of the compression load facilitated initial settling of the indentors onto the uneven geometry of the endplate, which was reflected as an initial toe region in the load-displacement curve.  Figure 5.2  Custom-built clamps were used to secure the specimen and an adjustable vise was used to align the superior endplate in the horizontal plane. A pair of concave-convex surfaces was used to create a nonconstraining spherical joint through which the compression was applied.  All tests were carried out using a servohydraulic testing machine (Instron 8874, Instron Corporation, Canton, MA) and axial compressive force and displacement were collected by the machine at 20 Hz. The indentor was compressed axially at a constant rate of 0.2 mm/s to a depth of 30% of the vertebral body height. Vertebral body height was measured directly using a pair of Vernier calipers. 103  Chapter  5  Screws  and Cement  on Interbody  Device  Interface  Strength  Failure load (in Newtons, N) and construct stiffness (N/mm) were extracted from the load-displacement curves (Figure 5.3). Failure load was interpreted as the maximum load before the slope on the load-displacement curve first changed from positive to negative. All specimens failed within the indentation depth of 30% of vertebral body height. Failure strength (MPa) was determined by dividing the failure load by the indentor area.  Stiffness was  determined to be the maximum slope of the load-displacement curve before failure occurred. To account for the fact that failure load was assumed to be a factor of BMD, specimen size and indentor size, we distributed the specimens such that their BMD and spinal level characteristics were evenly distributed among the 6 groups, and we matched indentor size to specimen size to obtain 40% of endplate coverage. To further account for differences associated with specimen and indentor size, we compared normalized failure load with indentor area (failure strength). 5.2.6  Statistical Analysis Two-way factorial ANOVA was used to determine significant differences due to cage  shape and method of fixation. Post hoc Student's Newman-Keuls tests were carried out to determine the differences between groups, with a significance level of 95%. 2000 i  Displacement (mm) Figure 5.3  Typical load-displacement curves where failure load was the maximum load before the gradient of the curve changed from positive to negative, while stiffness was the maximum slope of the curve before failure occurred. In 55% of the vertebrae tested, failure load was also the maximum load while a higher load beyond failure was measured in the other 45% of the tests. 104  Screws and Cement on Interbody Device Interface Strength  Chapter 5  5.3  Results Mean failure loads ranged from 741 N for specimens inserted with pedicle screws alone  to 1919 N for specimens inserted with anterior vertebral body screws augmented with cement, both tested with clover-leaf shaped indentors (Figure 5.4).  Failure load was significantly  affected by the method of fixation (p = 0.001) and not by indentor shape (p = 0.20), and there was a significant 2-way interaction (p = 0.009). In the post-hoc analyses between groups with pedicle screws, the use of cement significantly increased mean failure load (p = 0.002), with mean increases of 54% for elliptical shaped indentors and 121% for clover-leaf shaped indentors.  There was no significant difference in failure load between pedicle screws with  cement and anterior vertebral body screws with cement (p = 0.41).  The significant 2-way  interaction was attributed to the 89% higher mean failure load for clover-leaf shaped indentors compared to elliptical shaped indentors with anterior vertebral body screws (p = 0.02), but not with pedicle screws.  3000 2500  • Elliptical Shape • Clover-leaf Shape 1919  §2000  1616  "O  1641  o 1500 _ | 1000 'ro ^  500 0  Figure 5.4  741  Pedicle Screws Only  Pedicle Screws + Cement  Anterior Screws + Cement  Mean failure loads for elliptical and clover-leaf shaped indentors with i) pedicle screws only, ii) pedicle screws with cement, and iii) anterior vertebral body screws with cement. Error bars denote standard deviation.  Mean failure strengths ranged from 1.9 MPa for specimens inserted with pedicle screws alone to 4.6 MPa for specimens inserted with anterior vertebral body screws augmented with cement, both tested with clover-leaf shaped indentors (Figure 5.5).  Similar to failure load, 105  Chapter  5  Screws  and Cement  on Interbody  Device  Interface  Strength  failure strength was also significantly affected by the method of fixation (p = 0.0001) and not by indentor shape (p = 0.33), with a borderline significant 2-way interaction (p = 0.05). In the posthoc analyses between groups with pedicle screws with and without cement, use of cement significantly increased mean failure strength (p = 0.0002), with mean increases of 69% for elliptical shaped indentors and 137% for clover-leaf shaped indentors. There was no significant difference in failure strength between pedicle screws with cement and anterior vertebral body screws with cement (p = 0.16). Similar to failure load, the significant 2-way interaction was attributed to a higher (53%, p = 0.08) mean failure strength for clover-leaf shaped indentors compared to elliptical shaped indentors with anterior vertebral body screws, but not with pedicle screws.  :i  • Elliptical Shape • Clover-leaf Shape 4.4  4.6  4.5 3.0 T  1.9  i  Pedicle Screws Only Figure 5.5  Pedicle Screws + Cement  Anterior Screws + Cement  Mean failure strengths for elliptical and clover-leaf shaped indentors with i) pedicle screws only, ii) pedicle screws with cement, and iii) anterior vertebral body screws with cement. Error bars denote standard deviation.  Mean stiffnesses ranged from 476 N/mm to 703 N/mm for all groups (Figure 5.6). Stiffness was not significantly affected by method of fixation (p = 0.72) or by indentor shape (p = 0.47).  There was a significant 2-way interaction (p = 0.04) caused by the trend of higher  stiffness with clover-leaf shaped devices when used with anterior vertebral body screws, but a trend of lower stiffness with pedicle screws.  106  Chapter  Screws  5  and Cement  on Interbody  Device  Interface  Strength  Results from one test were rejected as metal-on-metal contact of the test jig with the specimen holder was noted during test.  1000  n  800 A E Z  600 A  co  CO CD  c  • Elliptical Shape  692  • Clover-leaf Shape  703 564  476  524  400 A  00 200  J  0 Pedicle Screws Only Figure 5.6  5.4  Pedicle Screws + Cement  Anterior Screws + Cement  Mean stiffnesses for elliptical and clover-leaf shaped indentors with i) pedicle screws only, ii) pedicle screws with cement, and iii) anterior vertebral body screws with cement. Error bars denote standard deviation.  Discussion The purpose of this study was to determine the effects of cement augmentation of  screws on cage-vertebra interface properties.  Results from this study clearly indicated an  increase in cage-vertebra strength when cement was used with pedicle or anterior vertebral body screws versus with using the screws alone. These results suggest that interbody device subsidence may be reduced in osteoporotic patients if cement is used to augment pedicle screws or anterior vertebral body screws. The improved cage-vertebra interface strength in this study was determined for the superior vertebral endplates, and similar results could not be implied for the inferior endplates as pedicle screws are placed further from the inferior endplate than superior endplate. Another limitation of this study was the use of a single vertebra experimental model, where the pedicle screws were not connected to posterior rods and thus the screw heads were not under load. Based on current literature, it is unclear whether this difference from the clinical scenario would have an effect on our results as there is contradictory information of the effect of posterior 107  Chapter 5  Screws and Cement on Interbody Device Interface Strength  instrumentation on the compressive failure load of functional spinal units with interbody fusion devices [Jost et al. 1998, Tsantrizos et al. 2000, Cripton et al. 2000]. Finally, only compression loads were applied in this study while loads in vivo include bending moments and shear forces. However, compression load is the predominant load expected to be transmitted through interbody devices, and thus we find that this is clinically germane. Clinical usage of bone cement in the spine has been primarily focused on the techniques of vertebroplasty and kyphoplasty as treatments for osteoporotic compression fracture [Barr et al. 2000, Kostuik and Shapiro 2003, Phillips 2003, Heini 2005]. Biomechanically, vertebroplasty resulted in increased vertebral compression failure load ranging from 100% [Bai et al. 1999] to 300% [Belkoff et al. 2000]. Clinically, significant pain relief has been seen in numerous studies [Barr et al. 2000, Wuisman et al. 2000, Fourney et al. 2003, Kostuik and Shapiro 2003, Heini 2005]. The use of cement as an augmentation technique for spinal implants in osteoporotic patients has been described for pedicle screws [Wuisman et al. 2000, Kostuik and Shapiro 2003, Heini 2005] in an attempt to address concerns of poor screws purchase, screws loosening, and screws pullout. The ability to improve local screw purchase, and thus achieve adequate construct stability in patients with low bone mass, has made cement use a viable option in osteoporotic spine patients. The effect of pedicle screws and/or cement on cage-vertebra interface strength has not previously been reported to our knowledge. Compared to our earlier study where no screws or cement were used [Tan et al. 2005 (Chapter 4)], failure load and failure strength after inclusion of the pedicle screws were not significantly affected for elliptical shaped devices, but were significantly decreased with clover-leaf shaped devices (Table 5.3). Cement augmentation of vertebral screws restored or even increased failure load and failure strength.  A direct  comparison could be made to results from this previous study as the specimen characteristics, such as age, vertebral levels and BMD range were similar, and the test protocol used was identical. We speculate that the decrease in failure strength with pedicle screws only for the clover-leaf devices was due to the disruption of the trabecular structure of underlying bone beneath the interbody device.  Due to different seating of the interbody devices over the  endplate, the disruption of bone by the screws had a variable effect on cage-vertebra interface strength (Figure 5.7). Trabecular disruption associated with pedicle screw insertion included the posterolateral region of the vertebral body, which was the same region where the clover-leaf shaped indentor harnessed the higher interface strength. Pedicle screw insertion therefore had a direct effect on reducing interface strength for clover-leaf shaped indentors. Cage-vertebra 108  Chapter 5  Screws and Cement on Interbody Device Interface Strength  interface strength for the elliptical shaped device was less affected with pedicle screw insertion, due to the smaller overlap of the elliptical indentor with the pedicle screw trajectory. Table 5.3  Effects of pedicle screws insertion on failure load and failure strength.  Failure Load, N (s.d.) Elliptical 936 (238) 1051 (199) p* = 0.31  No Screw/ Cement Pedicle Screws Inserted 3  Clover-leaf 1358 (327) 741 (328) p* = 0.002  Failure Streng th, MPa (s.d.) Elliptical Clover-leaf 4.4 (0.8) 2.9 (1.0) 1.9(0.4) 2.6 (0.8) p* = 0.45 p* = 0.00002  No Screw/ Cement Pedicle Screws Inserted 3  From previous study [Tan et al. 2005] *p-value using un-paired t-tests. s.d. = standard deviation  a  Cement injection into the pedicle tracts resulted in structural reinforcement to surrounding trabeculae. We speculate that the stiffer cement-bone composite block together with its closer vicinity to the superior endplate, redistributed load onto larger volumes of supporting trabecular bone beneath the screws and cement (Figure 5.8), and increased failure strength. As an ex situ DXA measurements of 0.29 g/cm could differ by 43% from in situ scans 2  with soft tissue present [Svendsen et al. 1995], the clinical definitions of osteoporosis and osteopenia, and the use of t-scores, are not applicable on ex situ specimens. Thus the bone quality of our dissected specimens could not be classified as normal, osteopenic or osteoporotic. We could, however, infer from the donors' age range of 62 to 93 years that the specimens would most likely be osteopenic or osteoporotic. In comparison to previous studies, the mean AP DXA BMD in a study with 16 T9-L4 specimens aged above 65 was 0.486 g/cm  2  (range: 0.416 to 0.586 g/cm ) [Bai et al. 2001], the mean AP DXA BMD in another study 2  involving 16 T12-L1 specimens aged 60 to 96 was 0.56 g/cm [Belkoff et al. 2001b], while the 2  mean lateral DXA BMD in a third study with 9 L3-S1 specimens aged 72 to 93 was 0.486 g/cm  2  (range: 0.185 to 1.025 g/cm ) [Labrom et al. 2005]. Accounting for differences due to specimen 2  levels, age range, scan direction and scan machines, the DXA BMD range in these previous studies were comparable to the current study.  109  Chapter  5  Figure 5.7  Screws  and Cement  on Interbody  Device  Interface  Strength  Pedicle screw insertion resulted in disruption of trabecular bone at the posterolateral region of the vertebral body and along the screw. Without pedicle screws, trabecular bone failure was found to occur in a semielliptical zone underlying the interbody devices (hatched area) [Tan et al. 2005 (Chapter 4)]. Due to the different seating of the elliptical and clover-leaf shaped devices over the endplate and over the screw trajectories, the disruption of bone by the screws had different outcome on device-vertebral body interface strength.  In a recent biomechanical study the mean compressive failure load of vertebral bodies following kyphoplasty or vertebroplasty with PMMA and calcium phosphate cement ranged between 1083 N to 1533 N [Tomita et al. 2003]. In a separate study [Bai et al. 1999], the mean compressive failure load after vertebroplasty with calcium phosphate cement was 1063 N while with PMMA it was 1036 N. Both of these studies compressed the vertebral bodies via a movable hinge to create wedge fractures.  Comparatively, compressive failure load for  vertebrae with cement augmentation in the current study ranged with means of 1013 N to 1919 N.  In spite of the differences in compressive test setup, cement type, cement volume and  specimens related characteristics, the range of failure loads obtained in the current study was 110  Chapter 5  Screws and Cement on Interbody Device Interface Strength  comparable to these earlier studies.  Studies with full compression over the endplate area  reported much higher failure loads (2500N-8000N) [Belkoff et al. 1999, 2000, 2001a, 2001b, Heini et al. 2001, Lim et al. 2002, Tomita et al. 2004]. Our failure loads were lower as we used indentors that covered over only 40% of the overall endplate area and thus did not load onto the cortical regions of the vertebral body.  •O ^ ^  ^  ^  ^  ^  Interbody device  3  Pedicle screws  Cement  Figure 5.8  (A) Underlying trabeculae supports the applied load from the interbody device. Failure strength is lower in the weaker osteoporotic bone. The hatched area represents the critical zone of trabeculae found to collapse at failure. (B) Pedicle screws disrupt the trabecular architecture in this critical zone beneath the interbody device, resulting in lower failure strength. (C) Cement injection into the pedicle tracts results in structural reinforcement to the trabeculae. The injected cement could penetrate towards the superior endplates and interbody device and the cured cement-bone composite blocks are stiffer than the surrounding trabecular bone. A larger volume of bone around this composite block prevents its subsidence, which in turn prevents subsidence of the interbody device. A 111  Chapter  5  Screws  and Cement  on Interbody  Device  Interface  Strength  combination of these factors results in higher failure strength following cement augmentation of pedicle screws.  5.5  Conclusion In conclusion, cement augmentation of pedicle screws was found to increase cage-  vertebra interface strength. There were no significant differences between pedicle and anterior vertebral body screws when cement augmentation was used.  With the use of cement  augmented screws, the increased cage-vertebra interface strength could reduce interbody device subsidence in the osteoporotic spine. 5.6  References  [I]  Anderson GF, Hussey PS. Population aging: A comparison among industrialized countries. Health Affairs (Millwood) 2000; 19:191-203.  [2]  Bai B, Jazrawi LM, Kummer FJ, Spivak JM. The use of an injectable, biodegradable calcium phosphate bone substitute for the prophylactic augmentation of osteoporotic vertebrae and the management of vertebral compression fractures. Spine 1999; 24:1521-1526.  [3]  Bai B, Kummer FJ, Spivak J. Augmentation of anterior vertebral body screw fixation by an injectable, biodegradable calcium phosphate bone substitute. Spine 2001; 26:2679-2683.  [4]  Barr JD, Barr MS, Lemley TJ, McCann RM. Percutaneous vertebroplasty for pain relief and spinal stabilization. Spine 2000; 25:923-928.  [5]  Belkoff SM, Maroney M, Fenton DC, Mathis JM. An in vitro biomechanical evaluation of bone cements used in percutaneous vertebroplasty. Bone 1999; 25:23S-26S.  [6]  Belkoff SM, Mathis JM, Erbe EM, Fenton DC. Biomechanical evaluation of a new bone cement for use in vertebroplasty. Spine 2000; 25:1061-1064.  [7]  Belkoff SM, Mathis J M , Fenton DC, Scribner RM, Reiley ME, Talmadge K. An ex vivo biomechanical evaluation of an inflatable bone tamp used in the treatment of compression fracture. Spine 2001a; 26:151-156.  [8]  Belkoff SM, Mathis JM, Jasper LE, Deramond H. An ex vivo biomechanical evaluation of a hydroxyapatite cement for use with vertebroplasty. Spine 2001b; 26:1542-1546.  [9]  Coe JD, Warden KE, Herzig MA, McAfee PC. Influence of bone mineral density on the fixation of thoracolumbar implants: A comparative study of transpedicular screws, laminar hooks and spinous process wires. Spine 1990; 15:902-907.  [10]  Cook SD, Salkeld SL, Stanley T, Faciane A, Miller SD. Biomechanical study of pedicle screw fixation in severely osteoporotic bone. Spine Journal 2004; 4:402-408.  [II]  Cripton PA, Jain GM, Wittenberg RH, Nolte LP. Load-sharing characteristics of stabilized lumbar spine segments. Spine 2000; 25:170-179.  [12]  Dvorak MF, Fisher CG. Revision spine surgery in the presence of osteoporosis. In: Revision Spine Surgery. Margulies JY, Aebi M, Farcy JPC (Eds.), Mosby, St Louis, 1999; 646-653.  [13]  European Prospective Osteoporosis Study (EPOS) Group. Incidence of vertebral fracture in Europe: results from the European prospective osteoporosis study (EPOS). Journal of Bone and Mineral Research 2002; 17:716-724.  112  Chapter 5  Screws and Cement on Interbody Device Interface Strength  [14]  Fourney DR, Schomer DF, Nader R, Chlan-Fourney J, Suki D, Ahrar K, Rhines LD, Gokaslan ZL. Percutaneous vertebroplasty and kyphoplasty for painful vertebral body fractures in cancer patients. Journal of Neurosurgery 2003; 98S:21-30.  [15]  Halvorson TL, Kelley LA, Thomas KA, Whitecloud TS, Cook SD. Effects of bone mineral density on pedicle screw fixation. Spine 1994; 19:2415-2420.  [16]  Hasegawa K, Takahashi HE, Uchiyama S, Hirano T, Hara T, Washio T, Sugiura T, Youkaichiya M, Ikeda M. An experimental study of a combination method using a pedicle screw and laminar hook for the osteoporotic spine. Spine 1997; 22:958-963.  [17]  Heini PF, Berlemann U, Kaufmann M, Lippuner K, Fankhauser C, van Landuyt P. Augmentation of mechanical properties in osteoporotic vertebral bones - A biomechanical investigation of vertebroplasty efficacy with different bone cements. European Spine Journal 2001; 10:164-171.  [18]  Heini PF. The current treatment - a survey of osteoporotic fracture treatment. Osteoporotic spine fractures: the spine surgeon's perspective. Osteoporosis International 2005; 16:S85-S92.  [19]  Higgins KB. The biomechanics of osteoporotic spine: evaluation of compression intervention. State University of New Jersey, PhD dissertation 2003.  [20]  Ignatius AA, Augat P, Ohnmacht M, Pokinskyj P, Kock HJ, Claes LE. A new bioresorbable polymer for screw augmentation in the osteosynthesis of osteoporotic cancellous bone: a biomechanical evaluation. Journal of Biomedical Materials Research. (Part B, Applied Biomaterials) 2001; 58:254-260.  [21]  Jackson SA, Tenenhouse A, Robertson L, CaMos Study Group. Vertebral fracture definition from population-based data: Preliminary results from the Canadian multicenter osteoporosis study (CaMos). Osteoporosis International 2000; 11:680-687.  [22]  Jost B, Cripton PA, Lund T, Oxland TR, Lippuner K, Jaeger P, Nolte LP. Compressive strength of interbody cages in the lumbar spine: the effect of cage shape, posterior instrumentation and bone density. European Spine Journal 1998; 7:132-141.  [23]  Kostuik JP, Shapiro MB. Open surgical treatment of osteoporotic fractures and deformity of the spine. Instructional Course Lectures 2003; 52:569-578.  [24]  Labrom RD, Tan JS, Reilly CW, Tredwell SJ, Fisher CG, Oxland TR. The effect of interbody cage positioning on lumbosacral vertebral endplate failure in compression. Spine 2005; 30:E556E561.  [25]  Lim TH, Brebach GT, Renner SM, Kim W J , Kim JG, Lee RE, Andersson GBJ, An HS. Biomechanical evaluation of an injectable calcium phosphate cement for vertebroplasty. Spine 2002; 27:1297-1302.  [26]  Lotz JC, Hu SS, Chiu DFM, Yu M, Colliou O, Poser RD. Carbonated apatite cement augmentation of pedicle screw fixation in the lumbar spine. Spine 1997; 22:2716-2723.  [27]  Lund T, Oxland TR, Jost B, Cripton PA, Grassmann S, Etter C, Nolte LP. Interbody cage stabilisation in the lumbar spine. Journal of Bone and Joint Surgery 1998; 80B:351-359.  [28]  Moore DC, Maitra RS, Farjo LA, Graziano GP, Goldstein SA. Restoration of pedicle screw fixation with an in situ setting calcium phosphate cement. Spine 1997; 22:1696-1705.  [29]  Okuyama K, Sato K, Abe E, Inaba H, Shimada Y, Murai H. Stability of transpedicle screwing for the osteoporotic spine: An in vitro study of the mechanical stability. Spine 1993; 18:2240-2245.  [30]  Okuyama K, Abe E, Suzuki T, Tamura Y, Chiba M, Sato K. Influence of bone mineral density on pedicle screw fixation: A study of pedicle screw fixation augmenting posterior lumbar interbody fusion in elderly patients. Spine Journal 2001; 1:402-407.  fracture  113  Chapter 5  Screws and Cement on Interbody Device Interface Strength  [31]  Oxland TR, Lund T, Jost B, Cripton PA, Lippuner K, Jaeger P, Nolte LP. The relative importance of vertebral bone density and disc degeneration in spinal flexibility and interbody implant performance: An in vitro study. Spine 1996; 21:2558-2569.  [32]  Panjabi MM, Takata K, Goel V, Federico D, Oxland T, Duranceau J, Krag M. vertebrae: Quantitative three-dimensional anatomy. Spine 1991; 16:888-901.  [33]  Panjabi MM, Goel, V, Oxland TR, Takata K, Duranceau J, Krag M, Price M. vertebrae: Quantitative three-dimensional anatomy. Spine 1992; 17:299-306.  [34]  Pfeiffer M, Gilbertson LG, Goel VK, Griss P, Keller JC, Ryken TC, Hoffman HE. Effect of specimen fixation method on pullout tests of pedicle screws. Spine 1996; 21:1037-1044.  [35]  Phillips FM. Minimally invasive treatments of osteoporotic vertebral compression fractures. Spine 2003; 15S:S45-S53.  [36]  Renner SM, Lim TH, Kim WJ, Katolik L, An HS, Andersson GBJ. Augmentation of pedicle screw fixation strength using an injectable calcium phosphate cement as a function of injection timing and method. Spine 2004; 29:E212-E216.  [37]  Sarzier JS, Evans AJ, Cahill DW. Increased pedicle screw strength with vertebroplasty augmentation in osteoporotic spines. Journal of Neurosurgery 2002; 96S.309-312.  [38]  Schultheiss M, Hartwig E, Claes L, Kinzl L, Wilke HJ. Influence of screw-cement enhancement on the stability of anterior thoracolumbar fracture stabilization with circumferential instability. European Spine Journal 2004; 13:598-604.  [39]  Soshi S, Shiba R, Kondo H, Murota K. An experimental study on transpedicular screw fixation in relation to osteoporosis of the lumbar spine. Spine 1991; 16:1335-1341.  [40]  Sran MM, Khan KM, Keiver K, Chew JB, McKay HA, Oxland TR. Accuracy of DXA scanning of the thoracic spine: cadaveric studies comparing BMC, areal BMD and geometric estimates of volumetric BMD against ash weight and CT measures of bone volume. European Spine Journal 2005; 14:971-976.  [41]  Svendsen OL, Hassager C, Skodt V, Christiansen C. Impact of soft tissue on in vivo accuracy of bone mineral measurements in the spine, hip, and forearm: a human cadaver study. Journal of Bone and Mineral Research 1995; 10:868-873.  [42]  Tan JS, Kwon BK, Dvorak MF, Fisher CG, Oxland TR. Pedicle screw motion in the osteoporotic spine after augmentation with laminar hooks, sublaminar wires, or calcium phosphate cement: A comparative analysis. Spine 2004; 29:1723-1730.  [43]  Tan JS, Bailey CS, Dvorak MF, Fisher CG, Oxland TR. Interbody device shape and size are important to strengthen the vertebra-implant interface. Spine 2005; 30:638-644.  [44]  Tomita S, Kin A, Yazu M, Abe M. Biomechanical evaluation of kyphoplasty and vertebroplasty with calcium phosphate cement in a simulated osteoporotic compression fracture. Journal of Orthopaedic Science 2003; 8:192-197.  [45]  Tomita S, Molloy S, Jasper LE, Abe M, Belkoff SM. Biomechanical comparison of kyphoplasty with different bone cements. Spine 2004; 29:1203-1207.  [46]  Tsantrios A, Baramki HG, Zeidman S, Steffen T. Segmental stability and compressive strength of posterior lumbar interbody fusion implants. Spine 2000; 25:1899-1907.  [47]  Wittenberg RH, Lee KS, Shea M, White AA, Hayes WC. Effect of screw diameter, insertion technique, and bone cement augmentation of pedicular screw fixation strength. Clinical Orthopaedics and Related Research 1993; 296:278-287.  [48]  Wuisman PIJM, Van Dijk M, Staal H, Van Royen BJ. Augmentation of (pedicle) screws with calcium apatite cement in patients with severe progressive osteoporotic spinal deformities: an innovative technique. European Spine Journal 2000; 9:528-533.  Thoracic human Human lumbar  114  Chapter 5  Screws and Cement on Interbody Device Interface Strength  [49]  Yamagata M, Kitahara H, Minami S, Takahashi K, Isobe K, Moriya H, Tamaki T. Mechanical stability of the pedicle screw fixation systems for the lumbar spine. Spine 1992; 17:S51-S54.  [50]  Yerby SA, Toh E, McLain RF. Revision of failed pedicle screws using hydroxyapatite cement: A biomechanical analysis. Spine 1998; 23:1657-1661.  [51]  Zindrick MR, Wiltse LL, Widell EH, Thomas JC, Holland WR, Field BT, Spencer CW. A biomechanical study of intrapeduncular screw fixation in the lumbosacral spine. Clinical Orthopaedics and Related Research 1986; 203:99-112.  115  Chapter 6  Rod Extension and Cement on Destabilized  Level  Chapter 6 E n h a n c i n g Fixation of the O s t e o p o r o t i c S p i n e in a C o r p e c t o m y M o d e l : The Effect of Cement A u g m e n t a t i o n and Posterior Instrumentation Extension  Overview  This chapter compares the enhancement of osteoporotic spine fixation by 1) cement augmentation of pedicle screws and 2) extension of posterior instrumentation, using the flexibility test protocol. Both techniques are available to surgeons to obtain improved fixation in the osteoporotic spine. Cement augmentation was found to consistently reduce intersegmental motion at the destabilized level while posterior rod extension reduced intersegmental motion in the absence of cement. The next study (Chapter 7) was carried out in parallel with the current study.  The results of this study were presented at these national and international conferences: 1) 52nd Annual Meeting of the Orthopaedic Research Society, Mar 19-22, 2006, Chicago, Illinois, USA. 2) 40th Annual Meeting of the Canadian Orthopaedics Research Society, Jun 2-4, 2006, Toronto, Ontario, Canada. 3) 33rd Annual Meeting of the International Society for the Study of the Lumbar Spine, Jun 1317, 2006, Bergen, Norway. A manuscript on the results of this study is being prepared for publication in the refereed journal Spine.  116  Chapter 6  6  Rod Extension and Cement on Destabilized Level  Enhancing Fixation of the Osteoporotic Spine in a Corpectomy Model: The Effect of Cement Augmentation and Posterior Instrumentation Extension  6.1  Introduction Reconstruction of the anterior spinal column is necessary after vertebral body resection  in the thoracic and lumbar spine in cases with substantial instability such as following severe trauma or resection of a tumor [Vahldiek and Panjabi 1998, Vieweg et al. 2003, Dvorak et al. 2003, Thongtrangan et al. 2003, Pflugmacher et al. 2004, Fisher et al. 2005, Holman et al. 2005]. Structural allografts [Bridwell et al. 1995, Molinari et al. 1999], autografts [Buttermann et al. 1997], bone cement [Harrington 1988], titanium cages [Akamura et al. 2002] or expandable cages [Vieweg et al. 2003, Thongtrangan et al. 2003] are used clinically to reconstruct the anterior column. Supplementary posterior or anterior instrumentation is necessary to enhance the structural stability of the spinal column [Vahldiek and Panjabi 1998]. Immediately following surgery, the implants are expected to provide sufficient stability for spinal fusion to occur. The implants are thus subjected to full load bearing during the immediate post-operative period, with minimal load transmitted through the remaining facet joints. The bone-implant interfaces are consequently placed under high stresses during this immediate post-operative period. The challenge of maintaining bone-implant interfaces and initial rigidity is even greater when such surgery is to be carried out in patients with low bone density [Vieweg et al. 2003, Babat et al. 2004]. Failure at the pedicle screw-bone interface, and/or the interbody device-vertebra interface could result in failure of bony fusion and the loss of correction in spinal alignment (Figure 6.1). To achieve initial rigidity, the surgeon aims to prevent both pedicle screw loosening and cage-vertebra subsidence. One technique to prevent pedicle screw loosening is to increase the number of anchor points [Hu 1997, Hasegawa and Hirano 2002], such as extending the pedicle screws two or more levels above and below the immediate fusion level. Another technique that has become popular is the cement augmentation of screws, which was shown to biomechanically improve fixation strength at the screw-bone interface [Cameron et al. 1975, Zindrick et al. 1986, Wittenberg et al. 1993, Lotz et al. 1997, Moore et al. 1997, Yerby et al. 1998, Ignatius et al. 2001, Bai et al. 2001, Sarzier et al. 2002, Schultheiss et al. 2004, Cook et al. 2004, Renner et al. 2004, Tan et al. 2004]. No biomechanical study has directly compared these two approaches to our knowledge.  117  Chapter 6  Figure 6.1  Rod Extension and Cement on Destabilized Level  Schematic diagram of rigid instrumentation, with interbody device and pedicle screw fixation, following corpectomy. Failure at the screw-bone interface resulted in pedicle screw loosening (A & B), while failure at the interbody device-endplate interface resulted in device subsidence (C & D). Increased relative motion at the instrumented levels could result from failure at any of these 4 implant-bone interfaces. Loss of fixation rigidity in the osteoporotic spine is a result of poor interface strength. Excessive motion may result in non-union of the bone graft.  Increased motion at adjacent non-instrumented levels following rigid spinal fixation has been suggested to be the cause of adjacent segment diseases [Lee 1988, Shono et al. 1998]. Adjacent segment failure in 4 out of 7 patients included progressive kyphosis with pseudoarthrosis following instrumented fusion in a retrospective study on patients with Parkinson's disease [Babat et al. 2004]. One solution utilised by surgeons to prevent adjacent segment disease is to extend the rigid posterior instrumentation to adjacent levels [Hu 1997]. Extending the rigid posterior instrumentation, however, could simply bring the adjacent segment disease to the next level where the extension rod ends. To prevent such adjacent segment effect, we conceptualized that this extension of instrumentation to an adjacent level should result in a more gradual gradient in resultant motion from the fusion construct, through the 118  Chapter  6  Rod Extension  and Cement  on Destabilized  Level  extension level, and to the non-instrumented levels. With no sudden increase in motion at the junction to a fusion construct or extension level, mechanically induced adjacent segment disease could possibly be reduced. One technique to achieve this gradual gradient in resultant motion would be to use flexible extension rods of appropriate flexural modulus in place of rigid rods. In a study investigating change in range of motion (ROM) at the adjacent segment after a solid fusion and after implantation of a flexible stabilization device (Dynesys, Centerpulse Orthopedics) at the adjacent segment, ROM at L4-L5 increased following rigid instrumentation at L5-S1, but reduced with implantation of a flexible transition device at L4-L5 [Goertzen et al. 2003]. The effect on ROM at the immediate surgical level (L5-S1) in that study was however not reported. This study is the first of a two- part study on addressing both initial construct fixation and adjacent level effects in the osteoporotic spine. In Part I (Chapter 6), enhancement of initial construct fixation with cement augmentation of pedicle screws is contrasted against extension of posterior instrumentation to the next level. In Part II (Chapter 7), adjacent level kinematics is contrasted between cement augmentation and posterior instrumentation extension.  The  purposes of Part I and Part II are to determine if cement augmentation and/or posterior rod extension could 1) enhance stability at the immediate surgical level, and 2) modulate motion at the adjacent vertebral level, respectively. For the current study, it is hypothesized that both cement augmentation of pedicle screws and extension of posterior fixation to the next level could enhance stabilization to the immediate surgical level. The kinematics of the immediate surgical level (T10-T12) in a T9-L3 cadaveric model with T11 corpectomy following: 1) cement augmentation of pedicle screws, and 2) extension with flexible or rigid posterior rods to L1, under the same applied moment of 5Nm, were contrasted.  6.2  Methods  6.2.1 Specimens Twelve 7-vertebrae thoracolumbar cadaveric specimens (T8-L2 or T9-L3) were utilised in this study. The donors' ages ranged between 64 and 84 years with a mean of 75 years (ages of 4 donors unknown). There were 3 male and 7 female donors (gender of 2 unknown). All specimens were harvested from unembalmed donors from the University of British Columbia Tissue Donation Program and stored frozen at -20 °C.  This study was approved by our  Institution's Clinical Research Ethics Board. Surrounding soft tissues such as muscles were carefully dissected, without damaging the intervertebral ligaments and discs.  All ribs were removed by dissecting through the 119  Chapter 6  Rod Extension and Cement on Destabilized Level  costovertebral and costotransverse joints. Thereafter, areal bone mineral densities (aBMD) of the 5 middle vertebral bodies (T9-L1 of T8-L2 segments and T10-L2 of T9-L3 segments) were measured from the lateral projection of dissected (T8-L2 or T9-L3) spinal segments using a dual-energy x-ray absorptiometry (DXA) machine (QDR 4500W, Hologic Inc., Bedford, MA) and using methodology previously described in detail [Sran et al. 2005]. The lateral dimension of each vertebral body was physically measured using a pair of vernier calipers and the volumetric BMD (vBMD) was subsequently calculated by approximating the vertebral body as an elliptical cylinder [Sran et al. 2005]. Average aBMD and vBMD for each spinal segment was calculated from the 5 middle vertebral levels. Mean aBMD and vBMD for the 12 spines was 0.420 g/cm  2  (s.d. = 0.090) and 0.151 g/cm (s.d. = 0.034) respectively (Figure 6.2). Specimens with 1) 3  bridging osteophytes resulting in bony fusion, and 2) vertebral body fracture, were excluded from this study. Two specimens included in the study had third degree (bird's beak shaped) osteophytes [Nathan et al. 1994] at T10 and T11, while one specimen had third degree osteophytes at L1 and L2.  1.0 n  -r 0.25  • aBMD • vBMD  •- 0.2 CO  £  -- 0.15 o 2 -- 0.1 Q CQ  - - 0.05 >  1  2  3  4  5  6  7  8  9 10 11 12  Specimen Figure 6.2  Distribution of areal bone mineral density (aBMD) and volumetric bone mineral density (vBMD) among the 12 specimens in this study. The aBMD data were obtained directly from the DXA machine output, derived from dividing bone mineral content (BMC) by the projection area (2dimensional area). The vBMD data were calculated from dividing the BMC by the estimated volume of the vertebral body. aBMD is the traditional technique of presenting BMD data but is less accurate as it uses the projected area in its calculation while vBMD is more accurate as it considers the total volume.  120  Chapter 6  Rod Extension and Cement on Destabilized Level  6.2.2 Anterior and Posterior Instrumentation Configurations  The experiment was conducted in a repeated measures fashion to test two main effects with interactions, between  1) cement augmentation of screw fixation and 2) posterior rod  extension (rigid and flexible). The three surgical conditions in a T9-L3 (T8-L2) spinal segment (Figure 6.3) involved: (A) T11 (T10) corpectomy, interbody implant bridging across T10 to T12 (T9 to T11), pedicle screws and rigid rod fixation bridging across T10 to T12 (T9 to T11) (Figure 6.3A), (B) surgical condition A) and with rigid rod bridging across T10 to L1 (T9 to T12) (Figure 6.3B), and (C) surgical condition A), with rigid rod bridging across T10 and T12 (T9 to T11) and with flexible rod bridging across T12 and L1 (T11 to T12) (Figure 6.3C).  (I) Figure 6.3  (A)  (B)  (C)  Schematic diagram of different surgical conditions tested in a repeated measures fashion in a T9-L3 spine segment. (I): Intact spine. (A): T11 corpectomy, a vertebral body replacement device inserted and rigid posterior instrumentation across T10 and T12. (B): Previous condition, with rigid rod across T10 to L1. (C): Previous condition, but with flexible rod between T12 and L1. The T10 screw fixations were augmented with bone cement in all surgical test conditions.  Stainless steel pedicle screws (USS, Synthes Spine, Mississauga, ON) were implanted by a single spine surgeon (Dr Sandeep Singh), by the following clinical protocol. Screws of uniform diameter (5mm), with lengths of 35 mm, 40 mm, 45 mm or 50 mm, were used. The 121  Chapter 6  Rod Extension and Cement on Destabilized Level  pedicle screws were inserted to approximately 80% of the vertebral body depth and the length of the pedicle screws used varied between specimen and spinal level. An expandable vertebral body replacement device (Synex, Synthes Spine, Mississauga, ON) was used to bridge across the T10-T12 (T9-T11) space following corpectomy to restore and maintain the intervertebral height.  The device was inserted in the un-expanded state,  aligned and expanded in situ with the accompanying tool (Synthes Spine), and engaged the endplates at both ends. A smaller vertebral body replacement device expandable from 23-31 mm was used in 9 spines, and a larger device expandable from 26-36 mm was used in the remaining 3 spines.  Measured heights at the anterior edges (Figure 6.4) of the expanded  devices ranged from 27 mm to 36 mm, with a median height of 29.5 mm. Both sizes of the device had endplate contact area of 25x28 mm and zero degrees of endplate angulation. No bone grafts were inserted into or around the vertebral body replacement device.  Figure 6.4  Measured height of the expandable vertebral body replacement device excluded the spiked protuberances at both ends. The end dimensions of the device contact areas are 25x28 mm. The devices included in this study had zero degrees of endplate angulation.  122  Chapter 6  Rod Extension and Cement on Destabilized Level  Connection of the extending rigid and flexible rods to the next distal level was carried out using a "parallel rod connector" and a "transverse connector" (Synthes Spine) (Figure 6.5). The average length of the extended rod segments was 76% (s.d. = 7%) of the length of the T10-T12 (T9-T11) rods.  Similar lengths of rigid and flexible extension rods were used in surgical  conditions (B) and (C). Using three-point bending tests, the ratio of the stiffness of the flexible rods (diameter 6.35 mm) to the stiffness of the rigid stainless steel rods (diameter 6 mm) was determined to be 1:37. The flexible rods are made of polyacetal (Trade name Delrin®, Industrial Plastics and Paints, Vancouver, BC) a biocompatible material used in heart valves and hip prostheses [McKellop et al. 1996]. The use of these rods for this application in the spine is not clinically approved to our knowledge. They are used in this study as a matched comparison to the rigid rods to test the concept of flexible posterior instrumentation.  Figure 6.5  Extending rigid and flexible posterior rods from T12 to L1 of T9-L3 segments (T11-T12 in T8-L2 segments) were carried out using a "parallel rods connector" and a "transverse connector" on each side. Both rigid and flexible extending rods' lengths were adjusted to three-quarter of the length of the 110-112 (19-111) rod. Ihe length of the 110-112 rod was measured between screw heads while the length of the extension rod was measured between the parallel and transverse connectors. Ihe extension rods (black in colour) shown in this figure are the 'flexible rods' made of polyacetal. 123  Chapter 6  6.2.3  Rod Extension and Cement on Destabilized Level  Test Protocol Intact flexibility tests in axial rotation (AR), lateral bending (LB) and flexion-extension  (FE) were carried out in a randomized order with applied pure moments ranging between +5 Nm and -5 Nm. FE flexibility tests were carried out with and without 400 N of compressive follower-load (FL) based on the method previously described [Patwardhan et al. 1999]. The FL apparatus consists of components attached individually to each vertebral (Figure 6.6) and the compressive load remains tangential to the curvature of the spine during motion. A FL was not used in AR and LB as the current concept of using bilateral cables to simulate muscle loads is not generally accepted in these directions [Goel 2005]. All tests were conducted using a custom spine testing machine (Figure 6.6) capable of applying pure moments [Goertzen et al. 2004]. This spine tester was retrofitted in March 2005 with a new actuator arm that has a lower torsional compliance (0.6 degree/Nm), and with new universal joints to eliminate the issues of slippage previously experienced [Niosi et al. 2005]. The actuator was rotated at a rate of 2 degrees/second between a maximum of 5 Nm and a minimum of -5 Nm for 3 cycles. A servohydraulic actuator (Instron A591-4, Instron Corporation, Canton, MA) was used to apply and maintain the constant 400 N of FL. Following testing in the intact condition, the three surgical conditions were carried out in a randomized order by a trained surgeon (Dr Sandeep Singh). Cement was used to augment only the T10 (T9 in T8-L2 segments) screws at this stage. In order to simulate the worst case scenario of loosened screws in the osteoporotic spine, the T12 and L1 (T11 and T12) pedicle screw tracts were over-drilled with a 5 mm diameter drill prior to the insertion of the pedicle screws [Tan et al. 2004 (Chapter 3)]. Flexibility tests similar to those in the intact state were repeated following each of the three surgical conditions. The flexibility tests in AR, LB, FE and FE+FL, were again carried out in a randomized order. After all tests for the non-cemented state were completed, the T12 and L1 (T11 and T12) pedicle screws were augmented with cement, and the tests for the cemented state carried out. Each pedicle screw was removed, then about 2 cm of cement (Palacos R, Biomet Merck, 3  Kerzers, Switzerland or Simplex, Stryker, Hamilton, ON) was injected in the pedicle screw tract and lastly the same screw was reinserted. This method of cement injection into a screw tract is used clinically [Wuisman et al. 2000] and was evaluated experimentally [Soshi et al. 1991]. Both cements were randomly used among the 12 specimens. The cement was allowed to cure and the same three surgical conditions (A), (B) and (C) were then randomly carried out, and with flexibility tests conducted in a randomized order for each surgical condition. 124  Chapter 6  Figure 6.6  Rod Extension and Cement on Destabilized Level  The custom spine testing machine used to apply pure moments to the specimen. The follower load (FL) apparatus consists of components attached to each vertebra and with bilateral eyelets where the compressive load is guided through.  Three-dimensional kinematics of all vertebral bodies (except the corpectomy level) was collected during each flexibility test using an optoelectronic camera system (Optotrak 3020, Northern Digital Inc., Waterloo, Canada).  The data f o r motion and applied moment was  synchronised and recorded at 20 Hz for the duration of each test. Intersegmental and overall 125  Chapter 6  Rod Extension and Cement on Destabilized Level  range of motion (ROM) and neutral zones (NZ) (Figure 6.7) were determined from the flexibility curves at the third (last) cycle.  A lower ROM at the destabilized level between surgical  techniques would indicate better fixation stabilization. An increase in ROM at the adjacent level indicates potential accelerated adjacent level effects. A constant ROM at the non-instrumented level would indicate application of pure moments.  Increase in NZ across spinal segment(s)  indicates injury [Oxland and Panjabi 1992] and increased instability at the level(s). 6.2.4  Statistical Analysis  One-way repeated measures ANOVA, with Newman-Keuls post-hoc tests, were carried out to determine significant differences between intact and surgical conditions.  Two-way  repeated measures ANOVA, with Newman-Keuls post-hoc analyses, were carried out to determine significant effects of cement augmentation (factor 1) and posterior rod extension (factor 2). ROM and NZ were normalized against its intact condition for the two-way ANOVA to reduce effects of inter-specimen differences. An alpha of 0.01 was chosen as multiple two-way ANOVAs were carried out independently across 5 spinal levels and 3 test directions.  10  -10  n  J  Moment (Nm)  Figure 6.7  Flexibility curve of the T8-L2 spine in flexion-extension (FE) of a specimen (3 cycles shown), following cement augmentation of all screws and with rigid posterior rod extension. Overall range of (ROM) and neutral zone (NZ) were determined from the third (last)  typical pedicle motion cycle.  126  Chapter 6  6.3  Rod Extension and Cement on Destabilized Level  Results  6.3.1 Flexibility across Destabilized Level  The mean ROMs across the destabilized level in the intact state (T10-T12 in T9-L3 segments) in AR, LB and FE were 9.3°, 7.8° and 6.2° respectively (Table 6.1). The corresponding mean NZs in AR, LB and FE were 0.5°, 0.7° and 0.4° respectively. In the 1-way ANOVA analyses, AR, LB and FE ROMs and NZs were significantly affected by surgical intervention (all p < 0.001). The mean ROMs at the destabilized level were significantly reduced with the basic surgical procedures carried out (no cement, no extension), in LB to 5.1° (by 35%, p < 0.001) and in FE to 4.5° (27%, p < 0.001), but with a marginal reduction in AR to 7.9° (16%, p = 0.06). Conversely, surgical intervention without cement or rod extension increased mean NZ at the destabilized level from its intact state in AR by 248% to 1.6° (p < 0.001), in LB by 40% to 1.0° (p = 0.02), and in FE by 92% to 0.8° (p < 0.001). Table 6.1  Range of motion and neutral zone across destabilized levels for different surgical conditions in the respective test directions.  Intact Not No extension cemented Flexible extension Rigid extension No extension Cemented Flexible extension Rigid extension p-values from 1-way ANOVA  AR 9.3° (3.0°) 7.9° (4.2°) *6.8° (3.3°) *5.4° (2.5°) *4.4° (1.6°) *4.2° (1.6°) *3.5° (1.5°) < 0.001  Intact No extension Flexible extension Rigid extension No extension Cemented Flexible extension Rigid extension p-values 1from 1-way ANOVA  AR 0.5° (0.3°) *1.6° (1.5°) 1.1° (0.7°) 1.0° (0.7°) 0.5° (0.4°) 0.5° (0.3°) 0.4° (0.3°) < 0.001  Not cemented  Range of Motion, mean (s.d.) LB FE 7.8° (2.8°) 6.2° (1.9°) *5.1° (2.2°) *4.5° (2.5°) *3.0° (2.3°) *3.3° (1.9°) *2.2° (1.8°) *2.2° (1.4°) *2.3° (1.1°) *1.9° (1.1°) *1.8° (0.9°) *1.3° (1.3°) *1.5° (0.7°) *1.2° (1.1°) < 0.001 < 0.001 Neutral Zone, mean (s.d.) LB FE 0.7° (0.4°) 0.4° (0.3°) 1.0° (0.6°) *0.8° (0.7°) 0.6° (0.6°) 0.4° (0.3°) 0.5° (0.5°) 0.3° (0.2°) *0.3° (0.2°) 0.2° (0.1°) *0.3° (0.3°) 0.1° (0.1°) 0.1° (0.08°) *0.2° (0.3°) < 0.001 < 0.001  FE+FL 6.2° (1.8°) *2.0° (1.7°) *0.9° (0.7°) *0.6° (0.5°) *0.9° (0.4°) *0.6° (0.3°) *0.6° (0.3°) < 0.001 FE+FL 0.4° (0.3°) 0.3° (0.4°) *0.1° (0.1°) *0.08° (0.1°) *0.07° (0.07°) *0.05° (0.04°) *0.04° (0.04°) < 0.001  * - significantly different from corresponding intact value, alpha = 0.01.  In the 2-way ANOVA analyses, mean ROMs and NZs at the destabilized level in all directions were reduced significantly by cement augmentation and rod extension (Table 6.2, Figures 6.8 to 6.11). In general, cement augmentation resulted in greater percentage decrease 127  Chapter 6  Rod Extension and Cement on Destabilized Level  in ROM than did flexible posterior rod extension in all loading directions, and greater than or similar percentage decrease in ROM than did rigid posterior rod extension. In all cases, rigid posterior rod extension resulted in greater reduction in ROM at the destabilized level than flexible rod extension. There was a significant 2-way interaction between cement and posterior rod extension on ROM in all directions and on NZ in LB and FE. Without cement augmentation of screws, rod extension significantly reduced ROM and NZ at the destabilized level; however, the effects of rod extension were not or were less significant in the presence of cement augmentation. Table 6.2  2-way ANOVA results on normalized ROM and NZ between cement (factor 1) and rod extension (factor 2) at destabilized levels.  Factor 1: Cement Factor 2: Extension Interactions  AR 0.002 < 0.001 0.01  Factor 1: Cement Factor 2: Extension Interactions  AR 0.01 0.004 0.05  Range of Motion (ROM) LB FE 0.003 0.003 < 0.001 < 0.001 < 0.001 < 0.001 Neutral Zone (NZ) LB FE 0.002 0.003 < 0.001 < 0.001 < 0.001 < 0.001  FE+FL 0.06 < 0.001 0.002 FE+FL 0.06 0.01 0.01  In AR, cement augmentation with no rod extension resulted in a 38% decrease in normalized ROM (from 87% of intact ROM to 49%). Rigid rod extension decreased ROM by 10% (p = 0.03) with cement and by 26% (p < 0.001) without, while flexible rod extension decreased ROM by only 2% with cement and by 13% without. Overall, cement augmentation resulted in higher percentage decreases in ROM (22% to 38%) than posterior rod extension (2% to 26%).  A borderline interaction between cement augmentation and posterior rod  extension (p = 0.01) was the result of significant effects of posterior rod extension when cement was not used and a non significant effect of posterior rod extension with cement. Of note was that cement augmentation of the baseline condition (no cement, no extension) resulted in smaller ROM (49% of intact) than with rigid (61%) rod extension (p = 0.003).  128  Chapter 6  Rod Extension and Cement on Destabilized Level Axial Rotation  1.4  • Not Cemented • Cemented  p< 0.001  1.2 u  2  p = 0.003  £ 1.0 o +J  p = 0.001  "S 0.8 .N "re  E 0.6  001  0.4  O OC  p = 0.7  0.2 -|  p = 0.02 p = 0.03  0.0 No extension  Figure 6.8  Rexible extension  Rigid extension  Mean normalized range of motion (ROM) of the destabilized level in AR. Error bars indicate standard deviation.  Lateral Bending 1.2  +->  •Not Cemented  u  I  • Cemented  p< 0.001  1.0 H p< 0.001  p = 0.002  0.8  N "re  E 0.4 o  z  S  o  0C  p< 0.001  0.2 p = 0.5 1—  0.0 -0.2  Figure 6.9  J  p = 0.01 No extension  Rexible extension  1  -<^>^p< 0.001  Rigid extension  Mean normalized range of motion (ROM) of the destabilized level in LB. Error bars indicate standard deviation.  In LB, cement augmentation decreased ROM between 15% (with rigid rod) and 44% (without rod extension), while rigid rod extension resulted in 38% decrease without cement but only 9% with cement. A significant interaction between cement augmentation and posterior rod extension was observed, wherein posterior rod extension significantly decreased ROM only 129  Chapter 6  Rod Extension and Cement on Destabilized Level  when cement was not used. Of note was that cement augmentation of the baseline condition resulted in smaller ROM (27% of intact) than with rigid rod extension (33%), although the difference was not significant (p = 0.07). In FE, cement augmentation decreased ROM between 12% (with rigid rod) and 35% (without rod extension), while rigid rod extension resulted in 37% decrease without cement and in 14% with cement. A significant interaction between cement augmentation and posterior rod extension was observed, and this was again due to the differences between effects of posterior rod extension with and without cement.  Cement augmentation of the baseline condition  resulted in a ROM (38% of intact) similar to ROM with rigid rod extension (36%) (p = 0.61).  Flexion-Extension  1.2  -Not  p < 0.001  o  I o  0.8  • C e m e n t e d  p < 0.001  ~ i.o H  C e m e n t e d  p < 0.001  H  TJ  o> =E 0-6 ^ i_ 0.4 -| 2o  O  = 0.002  0.2 p = 0.001  0.0 N o  Figure 6.10  extension  Flexible  extension  Rigid  extension  Mean normalized range of motion (ROM) of the destabilized level in FE. Error bars indicate standard deviation.  130  Chapter 6  Rod Extension and Cement on Destabilized Level  Rexion-Extension with Follower-Load  No extension  Figure 6.11  Rexible extension  Rigid extension  Mean normalized range of motion (ROM) of the destabilized level in FE+FL. Error bars indicate standard deviation.  6.3.2 Flexibility across Rod Extension Level  The mean ROM across the rod extension level (T12-L1 in T9-L3 segments) in the intact state in AR, LB and FE were 2.7°, 4.6° and 4.0° respectively (Table 6.3). The corresponding mean NZs at the rod extension level were 0.1°, 0.4° and 0.3° respectively. In the 1-way ANOVA analyses, AR, LB and FE ROMs were significantly affected by surgical intervention (p < 0.001). With no posterior rod extension, with or without cement, the mean ROMs at the rod extension level were not significantly different from intact, while rod extension significantly reduced ROMs in all test directions (Table 6.3). NZs at the rod extension level were generally unaffected as no injury was created at that level, such as removal of disc matter, when posterior rod extension was included. In the 2-way ANOVA analyses, mean ROMs at the rod extension level in all flexibility test directions were significantly reduced after inclusion of extension rods, regardless on the presence of cement (Tables 6.3 and 6.4, Figures 6.12 to 6.15).  In general, posterior rod  extension resulted in greater percentage decrease in ROM across the rod extension level than did cement augmentation in all loading directions.  A noteworthy point was that cement  augmentation without posterior rod extension did not result in increased motion at the rod extension level. In all cases, rigid posterior rod extension resulted in greater reduction in ROM at the rod extension level than flexible rod extension. There was no significant 2-way interaction between cement augmentation and posterior rod extension on ROM or NZ in all directions. 131  Chapter 6  Rod Extension and Cement on Destabilized Level  Table 6.3  Range of motion and neutral zone across extension level for different surgical conditions in the respective test directions.  Intact Not No extension cemented Flexible extension Rigid extension No extension Cemented Flexible extension Rigid extension p-values from 1-way ANOVA  AR 2.7° (2.1°) 2.7° (2.0°) 2.0° (1.4°) *1.5° (1.0°) 2.4° (1.6°) *1.6° (1.1°) *1.1° (0.8°) < 0.001  Intact No extension Flexible extension Rigid extension No extension Cemented Flexible extension Rigid extension p-values from 1-way ANOVA  AR 0.1° (0.2°) 0.2° (0.3°) 0.2° (0.2°) 0.1° (0.1°) 0.2° (0.2°) 0.2° (0.1°) 0.1° (0.1°) 0.6  Not cemented  Range of Motion, mean (s.d.) LB FE 4.6° (1.9°) 4.0° (1.8°) 5.0° (1.6°) 4.4° (1.8°) *2.8° (1.1°) *2.3° (1.0°) *2.3° (1.1°) *1.8° (1.0°) 4.6° (1.3°) 3.7° (1.3°) *1.7° (0.5°) *1.3° (0.6°) *1.2° (0.7°) *0.9° (0.5°) < 0.001 < 0.001 Neutral Zone, mean (s.d.) LB FE 0.4° (0.4°) 0.3° (0.2°) 0.4° (0.3°) 0.3° (0.2°) 0.3° (0.2°) 0.2° (0.1°) 0.4° (0.2°) 0.2° (0.1°) 0.5° (0.3°) 0.3° (0.2°) 0.3° (0.1°) 0.1° (0.05°) 0.3° (0.1°) *0.1° (0.07°) 0.04 < 0.001  FE+FL 3.9° (2.0°) 3.7° (1.8°) *2.1° (1.1°) *1.6° (0.9°) 3.2° (1.3°) *1.2° (0.7°) *0.9° (0.5°) < 0.001 FE+FL 0.3° (0.3°) 0.3° (0.3°) 0.2° (0.2°) 0.1° (0.08°) 0.3° (0.3°) 0.1° (0.05°) 0.1° (0.08°) 0.008  * - significantly different from corresponding intact value, alpha = 0.01. Table 6.4  2-way ANOVA results on normalized ROM and NZ between cement (factor 1) and rod extension (factor 2) at the rod extension level (below destabilized level).  Factor 1: Cement Factor 2: Extension Interactions  AR 0.01 < 0.001 0.5  Factor 1: Cement Factor 2: Extension Interactions  AR 0.8 0.2 0.6  Range of Motion (ROM) LB FE 0.003 < 0.001 < 0.001 < 0.001 0.02 0.3 Neutral Zone (NZ) LB FE 0.7 0.3 0.4 < 0.001 0.5 0.01  FE+FL < 0.001 < 0.001 0.08 FE+FL 0.9 0.001 0.4  In AR, cement augmentation with no extension resulted in an 11% decrease in normalized ROM (from 105% of intact ROM to 94%). Rigid rod extension decreased ROM by 49% with cement and by 45% without, while flexible rod extension decreased ROM by 31% with cement and 26% without.  Overall, posterior rod extension resulted in higher percentage  decreases in ROM (26% to 49%) than cement augmentation (11% to 16%). There was no interaction between cement augmentation and posterior rod extension.  132  Chapter 6  Rod Extension and Cement on Destabilized Level Axial Rotation  No extension  Figure 6.12  Rexible extension  Rigid extension  Mean normalized range of motion (ROM) at the extension level in AR. Error bars indicate standard deviation.  In LB, the trend of higher percentage decrease in ROM with posterior rod extension (50% to 76%) than with cement augmentation (8% to 22%) was again observed.  Cement  augmentation decreased ROM by 8% without rod extension, by 2 1 % with flexible rod and by 22% with rigid rod. Rigid rod extension decreased ROM by 62% without cement and by 76% with cement, while flexible rod extension decreased ROM by 50% without cement and by 63% with cement.  There was no interaction between cement augmentation and posterior rod  extension. In FE, posterior rod extension again resulted in higher percentage decrease in ROM (53% to 72%) than cement augmentation (15% to 23%). Cement augmentation decreased ROM by 15% without rod extension, by 23% with flexible rod and by 21% with rigid rod. Rigid rod extension decreased ROM by 66% without cement and by 72% with cement, while flexible rod extension decreased ROM by 53% without cement and by 61% with cement. There was no interaction between cement augmentation and posterior rod extension.  133  Chapter 6  Rod Extension and Cement on Destabilized Level Lateral Bending p < 0.001  1.4 -i  p = 0.003  1.2 • p = 0.07 ^  o  f  • Not Cemented  i.o -  o "g 0.8 •  • Cemented  .N «  E 0.6 • o  z  5 O  }  p < 0.001  0.4 -  p < 0.001  cc  p <  0.2 -  No e x t e n s i o n  0.001  Rexible e x t e n s i o n  Rigid e x t e n s i o n  0.0 -  Figure 6.13  Mean normalized range of motion (ROM) at the extension level in LB. Error bars indicate standard deviation.  Rexion-Extension p < 0.001  1.4 p < 0.001  1.2 o  p = 0.003  p = 0.001  8  c 1.0  •Not Cemented  u o  -Cemented  0.8  N  E 0.6 S 0.4 O  p < 0.001  0.2  P<Q,QQ1 p <  0.0 No e x t e n s i o n  Figure 6.14  0.001  Rexible e x t e n s i o n  Rigid e x t e n s i o n  Mean normalized range of motion (ROM) at the extension level in FE. Error bars indicate standard deviation.  134  Chapter 6  Rod Extension and Cement on Destabilized Level Hexion-Extension with Follower-Load p < 0.001  1.4  p = 0.004  p < 0.001  1.2 u  I  1.0  i  • Not Cemented  = 0.05 {_  • Cemented  o  "g 0.8 N  E 0.6 o O  p = 0.002  0.4 0.2  P<0001 p<  0.0 No extension  Figure 6.15  6.3.3  0.001  Flexible extension  Rigid extension  Mean normalized range of motion (ROM) at the extension level in FE+FL. Error bars indicate standard deviation.  Flexibility across Non-Instrumented Levels  At the non-instrumented levels of T9-T10 (1 level above destabilized level), AR, LB and FE ROMs were significantly increased by surgical intervention (p < 0.001) (Table 6.5). Corresponding NZs at T9-T10 were also increased in LB (p < 0.001), and FE (p < 0.001), but not in AR (p = 0.2). Compared to intact values, the basic surgical procedures carried out (no cement, no extension) significantly increased mean ROMs at T9-T10, in AR from 5.1° to 6.1° (18 %, p = 0.002), in LB from 4.5° to 5.6° (25 %, p < 0.001), and in FE from 3.0° to 4.0° (35 %, p < 0.001) (Table 6.5, Figures 6.16 - 6.19).  These increases in ROMs and NZs were  unexpected, as pure moments were applied. On the other hand, ROMs and NZs at the distal non-instrumented levels of L1-L2 (2 levels below) and L2-L3 (3 levels below) were generally unaffected by surgical intervention (Tables 6.6 and 6.7, Figures 6.16 - 6.19).  135  Chapter 6  Rod Extension and Cement on Destabilized Level  Table 6.5  Range of motion and neutral zone across level above destabilized level  Intact Not No extension cemented Flexible extension Rigid extension No extension Cemented Flexible extension Rigid extension p-values from 1-way ANOVA  AR 5.1° (1.3°) *6.1° (1.5°) *6.0° (1.5°) *6.0° (1.5°) *6.2° (1.5°) *6.2° (1.5°) *6.2° (1.7°) < 0.001  Intact No extension Flexible extension Rigid extension No extension Cemented Flexible extension Rigid extension p-values from 1-way ANOVA  AR 0.5° (0.3°) 0.6° (0.5°) 0.6° (0.5°) 0.5° (0.4°) 0.6° (0.4°) 0.6° (0.3°) 0.6° (0.4°) 0.2  Not cemented  Range of Motion, mean (s.d.) LB FE FE+FL 4.5° (1.7°) 3.0° (1.1°) 2.7° (1.1°) *5.6° (1.7°) *4.0° (1.2°) *3.2° (1.1°) *5.7° (1.8°) *4.0° (1.2°) *3.2° (1.1°) *5.5° (1.6°) *3.9° (1.1°) *3.1° (0.9°) *5.9° (1.8°) *4.3° (1.5°) *3.3° (1.1°) *5.7° (1.9°) *4.0° (1.2°) *3.1° (1.0°) *5.8° (1.7°) *4.0° (1.2°) *3.3° (1.1°) < 0.001 < 0.001 < 0.001 Neutral Zone, mean (s.d.) LB FE FE+FL 0.4° (0.2°) 0.2° (0.2°) 0.2° (0.1°) 0.6° (0.3°) 0.3° (0.2°) 0.2° (0.1°) 0.5° (0.3°) *0.3° (0.2°) 0.2° (0.1°) 0.5° (0.3°) 0.3° (0.1°) 0.2° (0.1°) *0.6° (0.4°) *0.3° (0.2°) 0.2° (0.1°) *0.6° (0.4°) *0.4° (0.2°) 0.2° (0.1°) *0.3° (0.2°) 0.6° (0.3°) 0.2° (0.1°) < 0.001 < 0.001 0.2  * - significantly different from corresponding intact value, alpha = 0.01. Table 6.6  Range of motion and neutral zone at 2 levels below destabilized level (L1L2) for different surgical conditions in the respective test directions.  Intact No extension Not cemented Flexible extension Rigid extension No extension Cemented Flexible extension Rigid extension p-values from 1-way ANOVA  AR 2.3° (1.7°) 1.9° (1.8°) 1.9° (1.7°) 1.9° (1.7°) 2.1° (1.7°) 1.9° (1.6°) 1.8° (1.6°) 0.2  Intact Not No extension cemented Flexible extension Rigid extension No extension Cemented Flexible extension Rigid extension p-values from 1-way ANOVA  AR 0.1° (0.2°) 0.1° (0.2°) 0.1° (0.2°) 0.1° (0.2°) 0.2° (0.4°) 0.1° (0.3°) 0.2° (0.3°) 0.5  Range of Motion, mean (s.d.) LB FE 5.7° (1.9°) 4.9° (1.7°) 5.9° (2.0°) 5.1° (2.0°) 6.0° (2.0°) 5.1° (1.9°) 6.1° (2.0°) 5.2° (1.9°) 6.1° (2.1°) 5.0° (1.9°) 6.0° (2.0°) 5.0° (1.9°) *6.2° (2.1°) 5.1° (1.9°) 0.05 0.3 Neutral Zone, mean (s.d.) LB FE 0.4° (0.4°) 0.6° (0.5°) 0.6° (0.4°) 0.4° (0.5°) 0.4° (0.5°) 0.5° (0.3°) 0.5° (0.4°) 0.4° (0.3°) 0.6° (0.3°) 0.5° (0.5°) 0.5° (0.5°) 0.6° (0.3°) 0.5° (0.3°) 0.4° (0.4°) 0.6 0.4  FE+FL 4.7° (1.7°) 4.6° (2.2°) 4.7° (2.2°) 4.7° (1.6°) 4.6° (1.9°) 4.7° (2.0°) 4.7° (2.0°) 1.0 FE+FL 0.3° (0.3°) 0.3° (0.3°) 0.3° (0.4°) 0.2° (0.1°) 0.3° (0.4°) 0.4° (0.4°) 0.3° (0.3°) 0.005  * - significantly different from corresponding intact value, alpha = 0.01.  136  Chapter 6  Rod Extension and Cement on Destabilized Level  Table 6.7  Range of motion and neutral zone at 3 levels below destabilized level (L2L3) for different surgical conditions in the respective test directions.  Intact Not No extension cemented Flexible extension Rigid extension No extension Cemented Flexible extension Rigid extension p-values from 1-way ANOVA  AR 3.0° (1.5°) 2.7° (1.5°) 2.6° (1.6°) 2.7° (1.5°) 3.1° (1.9°) 2.9° (1.5°) 2.8° (1.6°) 0.1  Intact No extension Flexible extension Rigid extension No extension Cemented Flexible extension Rigid extension p-values from 1-way ANOVA  AR 0.2° (0.1°) 0.1° (0.1°) 0.2° (0.1°) 0.1° (0.1°) 0.2° (0.3°) 0.2° (0.2°) 0.2° (0.2°) 0.07  Not cemented  Range of Motion LB FE 5.0° (2.7°) 7.4° (2.5°) 7.6° (2.6°) 5.3° (2.5°) 7.7° (2.7°) 5.3° (2.6°) 7.6° (2.6°) 5.2° (2.3°) 7.7° (2.8°) *5.6° (2.7°) 7.8° (2.7°) 5.5° (2.7°) 7.7° (2.8°) 5.4° (2.5°) 0.5 0.002 Neutral Zone LB FE 0.5° (0.4°) 0.9° (0.8°) 0.8° (0.6°) 0.6° (0.6°) 0.8° (0.6°) 0.6° (0.6°) 0.6° (0.4°) 0.8° (0.5°) 0.9° (0.6°) 0.7° (0.5°) 0.9° (0.5°) 0.8° (0.6°) 0.9° (0.5°) 0.6° (0.3°) 0.9 0.1  FE+FL 4.6° (2.8°) 4.7° (2.6°) 4.7° (2.6°) 4.4° (2.2°) 4.9° (2.7°) 4.8° (2.6°) 4.8° (2.6°) 0.1 FE+FL 0.3° (0.2°) 0.3° (0.2°) 0.3° (0.2°) 0.2° (0.2°) 0.3° (0.2°) 0.3° (0.2°) 0.3° (0.2°) 0.3  Values in parenthesis denote standard deviation. * - significantly different from corresponding intact value, alpha = 0.01.  In the 2-way ANOVA analyses, mean AR, LB and FE ROMs and NZs at all 3 noninstrumented levels were not significantly different between surgical procedures carried out at the instrumented and extension levels (Figures 6.16 - 6.19).  137  Chapter 6  Rod Extension and Cement on Destabilized Level  Axial Rotation ROM  1.8 1.6 & 1.4 _c o 1.2 +-> x> /ft  |  1  0.8  z 0.6 S O 0.4 0.2 0  1  1  ilil  1 level a b o v e destabilized level  destabilized level  1 level below destabilized level  2 levels b e l o w destabilized level  3 levels b e l o w destabilized level  Axial Rotation NZ  •  1 level a b o v e destabilized level 1 level below 2 levels below 3 levels below destabilized level destabilized level destabilized level destabilized level • No extension  0 Flexible extension  • Rigid extension  H Cemented, no ext  B C e m e n t e d , flex ext  D) C e m e n t e d , rig ext  * - significant effect of cement and significant effect of rod extension. Using 2-way repeated measures ANOVA, alpha = 0.01. Error bars indicate standard deviation. Figure 6.16  Mean normalized range of motion (ROM) and neutral zone (NZ) of each mobile level and instrumented segment in AR. Posterior rod extension involved the first level below the destabilized level. No significant differences were found at the non-instrumented levels (1 level above, 2 below or 3 below from destabilized level) between surgical conditions.  138  Chapter 6  Rod Extension and Cement on Destabilized Level  1 level above destabilized level  1 level above destabilized level  destabilized level  1 level below destabilized level  2 levels below destabilized level  3 levels below destabilized level  destabilized level  1 level below destabilized level  2 levels below destabilized level  3 levels below destabilized level  • No extension  • Flexible extension  • Rigid extension  H Cemented, noext  B Cemented, flexext  • Cemented, rig ext  * - significant effect of cement and significant effect of rod extension. Using 2-way repeated measures ANOVA, alpha = 0.01. Error bars indicate standard deviation. Figure 6.17  Mean normalized range of motion (ROM) and neutral zone (NZ) of each mobile level and instrumented segment in LB. There were no significant differences in ROM between surgical conditions, at the non-instrumented levels.  139  Chapter 6  Rod Extension and Cement on Destabilized Level  Flexion-Extension NZ  1 level above destabilized level  destabilized level  1 level below destabilized level  2 levels below destabilized level  3 levels below destabilized level  • No extension  • Flexible extension  • Rigid extension  E Cemented, no ext  B Cemented, flex ext  • Cemented, rig ext  * - significant effect of cement and significant effect of rod extension, ** - significant effect of extension only, within the level. Using 2-way repeated measures ANOVA, alpha = 0.01. Error bars indicate standard deviation. Figure 6.18  Mean normalized range of motion (ROM) and neutral zone (NZ) of each mobile level and instrumented segment in FE. There were no significant differences in ROM between surgical conditions, at the non-instrumented levels.  140  Chapter 6  Rod Extension and Cement on Destabilized Level Flexion-Extension (w/FL) ROM  2 -, 1.8 1.6 1.4 TJ  1.2  (0 O  z £ O  or  1 0.8 -I 0.6 -| 0.4 0.2 0 1 level above destabilized level  destabilized level  4  1 level below destabilized level  2 levels below destabilized level  3 levels below destabilized level  Flexion-Extension (w/FL) NZ  3.5  a  3  c  2 2.5  TJ *  =1 15  I  2  ro  z N Z  III  1 0.5 0 1 level above destabilized level  destabilized level  • No extension S Cemented, noext  1 level below destabilized level  • Flexible extension B Cemented, flexext  2 levels below destabilized level  3 levels below destabilized level  • Rigid extension O Cemented, rig ext  * - significant effect of cement and significant effect of rod extension, ** - significant effect of extension only, within the level. Using 2-way repeated measures ANOVA, alpha = 0.05. Error bars indicate standard deviation. Figure 6.19  Mean normalized range of motion (ROM) and neutral zone (NZ) of each mobile level and instrumented segment in FE+FL. There were no significant differences in ROM between surgical conditions, at the noninstrumented levels.  141  Chapter 6  6.4  Rod Extension and Cement on Destabilized Level  Discussion The main purpose of this study was to determine the relative effect of cement  augmentation of screws and posterior rod extension on spinal motion in the osteoporotic spine. This study focused on the destabilized level, but other levels were also addressed. An in vitro multi-segmental spine model was adopted, so as to compare the immediate post-operative kinematics following different surgical conditions.  Less motion at the destabilized level is  desirable so as to achieve better fusion, while motion at the adjacent levels should remain similar to the intact state so as to avoid adjacent level effects. 6.4.1  Limitations A common limitation associated with repeated measures in vitro studies where extensive  surgical procedures and tests were carried out on the same specimen was the deterioration of the tissues over the duration of the test, which could result in device loosening or soft tissues damage. This was addressed by randomizing the sequence of rod extension configurations and the order of flexibility tests.  To mitigate the effects of differences between specimens,  comparisons were made within specimens and normalized against its intact values.  As a  prudent measure, all tests on a specimen were carried out on the same day to minimise the effect of refreezing and the specimens were sprayed with saline and water at intervals to keep the tissues moist during tests. To further alleviate the issue of tissue deterioration, cement augmentation of screws were conducted in the later half of each test day, which resulted in significantly lower ROM at the instrumented and extension levels, and mitigated any adverse effects of device loosening over time. The fact that ROMs at the non-instrumented levels were not significantly affected over time demonstrated minimal damage to its biomechanical integrity. Although the testing machine was designed to apply pure moments of the same magnitudes in each test, there were some limitations inherent with the spine machine (Figure 6.6). The balance weight of 4.3 N located 0.14 m from the midline was balanced for the neutral position. During operation, the length of the ball spline extends and retracts by as much as 0.02 m in each direction. As a result, the change in moment by weight of the actuator arm could vary by about 0.08 Nm. This moment acts in an orthogonal plane to the plane of applied moment. For example in flexion-extension tests, this 0.08 Nm is applied in the lateral bending direction. A second limitation with the testing machine is with the counter weight system, which hung on a passive swivel system that has some inherent friction. From the angle of the hanging wires and the mass of the counterweight, it can be estimated that a 2 N force was necessary to overcome the friction in the swivel. This resulted in a counteracting moment, with smaller magnitudes 142  Chapter 6  Rod Extension and Cement on Destabilized Level  (about 0.2 Nm) applied to the upper most spinal segment and higher magnitudes (about 0.5 Nm) to the lowest spinal segment. These errors were however consistent in every specimen and surgical condition, and should not affect the overall result and conclusion. 6.4.2  Destabilized Level Posterior instrumentation together with vertebral body replacement cage has been found  to provide greater rigidity at the destabilized levels than the intact state in vertebral body replacement surgeries [Vahldiek and Panjabi 1998, Pflugmacher et al. 2004, Schultheiss et al. 2004]. The ROM following stabilization with vertebral bone replacement cages and posterior instrumentation in an in vitro study was about 10% of intact values in FE and LB and about 40% in AR [Pflugmacher et al. 2004]. In another in vitro study, the ROM was 12.5% of intact values in FE, 40% in LB and was not different from intact in AR [Vahldiek and Panjabi 1998]. In the current study, the minimal surgical configuration without cement or posterior rod extension did not substantially reduce ROM in FE (73% of intact), LB (65%) and AR (84%). This was an expected findings as the condition in an osteoporotic spine was simulated with loosened screws at T12. Following cement augmentation and/or posterior rod extension, ROM at the destabilized level was reduced. With cement augmentation, ROMs were reduced in FE (37% of intact), LB (24%) and AR (47%).  With both cement augmentation and rigid rod  extension, ROMs were further reduced in FE (24% of intact), LB (15%) and AR (38%). These values were comparable to the literature [Vahldiek and Panjabi 1998, Pflugmacher et al. 2004], and thus the data supports the use of cement and/or posterior rod extension to improve fixation in the osteoporotic spine. Figures 6.20 and 6.21 summaries the results in this study. 6.4.3  Rod Extension Level Increased biomechanical demand at levels adjacent to a rigid construct were previously  found to cause significant degenerative changes at these adjacent levels in clinical studies [Axelsson et al. 1997, Oda et al. 1999, Kumar et al. 2001a, 2001b]. At the rod extension level in this study, ROMs were found to be significantly reduced with the use of rigid or flexible rod extension, but less so with cement. The use of posterior rod extension could thus be a feasible technique to suppress the potential increase in demands following rigid fixation. The use of rigid posterior rod extension resulted in 45% to 55% reduction in motion at the rod extension level in AR, LB and FE. Flexible rod extension on the other hand resulted in 26% reduction in motion in AR, 39% in LB and 42% in FE. No related data is available in the published literature for comparison, according to our knowledge. 143  Chapter 6  Rod Extension and Cement on Destabilized Level  Axial Rotation  Lateral Bending  Flexion-Extension  (b) no c e m e n t , flexible rods extension  Legend Increased ROM Decreased ROM No change in ROM  (c) no cement, rigid rods e x t e n s i o n  Figure 6.20  Summary of changes in ROM across levels and directions, for the noncemented conditions. 144  Chapter 6  Rod Extension and Cement on Destabilized Level  Axial Rotation  Lateral Bending  Flexion-Extension  (d) cemented, no extension  (e) cemented, flexible rods extension  (f) cemented, rigid rods extension Figure 6.21  Summary of changes in ROM across levels and directions, for the cemented conditions.  145  Chapter 6  6.4.4  Rod Extension and Cement on Destabilized Level  Other Levels Compared to the respective intact states, the initial surgical intervention resulted in 23%  increase in AR ROM, 29% in LB and 42% in FE at the upper most level (T9-T10) (Table 6.5). These increases were unexpected as pure moments were applied. Injury to the T9-T10 motion segment during the initial surgery could have caused these increases in ROMs.  As the  dimensions of T9-T10 were the smallest within the test specimen, the 5 Nm applied moment could also have caused soft tissue damage at this level but not at the more distal levels. A further check of the data revealed that 2 of 12 specimens had higher increases in ROM with the initial surgical condition than the average of the other 10 specimens, which inflated the overall average increase. One specimen had 151% increase in AR, a second specimen had 94% increase in LB, and in FE these two specimens had increases of 215% and 115% respectively. Omitting the data from these outliers, the average increase in ROMs would have been more moderate at 11% in AR, 23% in LB and 29% in FE. Other factors that could have affected kinematics at T9-T10 included changes within the motion segment. The injection of cement into T10 followed by insertion of pedicle screws could have affected normal deflection of the superior T10 endplate, and changes to the T9-T10 disc pressure.  As cement was used in all T10  screws, it could have consistently affected kinematics at the T9-T10 level in all specimens. In a finite element study, cement injection of vertebral bodies was found to increase disc pressure in adjacent discs, endplate deflection, stresses and strains [Polikeit et al. 2003]. Cement augmentation or posterior rod extension did not result in changes in ROMs at the all non-instrumented levels, and were consistent with expectation when the flexibility protocol was used. 6.4.5  Spinal Fusion and Adjacent Level Effect The role of spinal instrumentation and interbody fusion devices are to provide initial  rigidity for bony fusion to occur [Evans 1985].  It is known that too much motion could be  detrimental to the fusion process, and granulation tissue and fibrosis would develop from the bone grafts [Bauer and Muschler 2000], leading to pseudoarthrosis, implant loosening, and possibly in subsequent fatigue failure of instrumentation [Babat et al.2004]. Wolff's law on bone remodelling or the interfragmentary strain theory used in fracture fixation of long bones [Chao and Aro 1997] could probably be used to describe the bone remodelling or growth that takes place within the bone grafts in spinal fusion. In short, the bone grafts need to be loaded in order for bone remodelling to occur, while too much motion may inhibit the fusion process. Less motion at the destabilized level is therefore desirable in order to achieve better fusion. 146  Chapter 6  Rod Extension and Cement on Destabilized Level  One solution to address adjacent level effects was to increase the extent of the fixation to include multiple segments [Hu 1997, Babat et al. 2004]. There was concern, however, that rod extension might result in postponement of adverse adjacent level effects to the level beyond the extension rods [Hu 1997, Diwan et al. 2003, Kostuik and Shapiro 2003, Babat et al. 2004]. The use of flexible rod extension to maintain motion at the extension level theoretically could result in no further increase in demand at remaining non-instrumented levels and might resolve this issue of postponement of adverse adjacent level effects. Ideally, the flexible extension rods would provide a transition zone between the rigid spinal fixation to the non-instrumented levels, from minimal motion at the fixation level and with gradual increase in motion at the flexible rod extension level(s) to the first non-instrumented level. Motion at the adjacent levels should thus remain similar to the intact state in order to avoid adjacent level effects. This study demonstrated that flexible rod extension resulted in moderate reduction in motion at the adjacent level when compared with rigid rod extension. The optimal range of stiffness and other physical properties of the flexible extension rod to most successfully address adjacent level effects, however, would require further investigation, and this was not within the scope of the thesis. The use of both cement augmentation of pedicle screws and extension of posterior rods to adjacent level could potentially result in an optimal fixation construct in the degenerated and/or osteoporotic spine, addressing both issues of device loosening and adverse adjacent level effects. 6.4.6  Testing Methodology This study used the flexibility protocol of testing. This method uses a pure moment  applied to a spine specimen with the resultant vertebral motions measured.  The applied  moment across all segments remained a pure moment of the same magnitude and thus the flexibility protocol could not be used to ascertain biomechanical changes at the adjacent levels following rigid spinal fixation. The hybrid flexibility-stiffness protocol should be used [Panjabi and Goel 2005], and it will be presented and discussed in more detail in Chapter 7. In this study however, it was sufficiently demonstrated using the flexibility protocol that the concept of posterior rod extension could reduce motion at the adjacent level. 6.4.7  Cement Augmentation versus Posterior Rod Extension Cement augmentation of pedicle screws has been found previously to enhance fixation  of screws in bone [Cameron et al. 1975, Zindrick et al. 1986, Wittenberg et al. 1993, Lotz et al. 147  Chapter 6  Rod Extension and Cement on Destabilized Level  1997, Moore et al. 1997, Yerby et al. 1998, Ignatius et al. 2001, Bai et al. 2001, Sarzier et al. 2002, Schultheiss et al. 2004, Cook et al. 2004, Renner et al. 2004, Tan et al. 2004]. The injected cement penetrated through the porous trabeculae, and formed a stronger composite of bone and cement when the cement hardened.  The rigid screw and surrounding cement-  impregnated bone composite together can be considered as a single "implant", with enlarged size and larger surface area for interface area with surrounding trabeculae.  As a wider  infiltration of cement was found to be possible when injected into more porous low density bone, the size of the "implant" would be larger. Moreover, the screw-composite bone interface could be assumed to be rigid as both materials have high stiffness as compared to bone. As the previously weak screw-bone interface in low density bone was replaced by a larger "implant", and with a stronger "implant"-bone interface following cement augmentation, better fixation strength was thus achieved (Figure 6.22). Pedicle screw  Cement-bone composite  Bone-"implant" interface Figure 6.22  Schematic diagrams of pedicle screws with and without cement augmentation. The rigid composite of cement and bone created a new bone-"implant" interface with surrounding trabecular bone. Wider infiltration of cement was achieved when impregnated into more porous low density bone, and a larger "implant" created.  In the combined anterior-posterior fixation construct, the overall structural stability of the reconstructed spinal column depended on both the anterior interbody device's interface and the posterior pedicle screw instrumentation's interface with the neighbouring bone (Figure 6.1). Posterior rod extension results in an increase in the number of fixation points for the rigid posterior rods in order to improve fixation (Figure 6.23B). As the posterior instrumentation does not directly improve the interbody device-vertebral interface, bending of the long rods could result in subsidence of the cage into the endplates. Not only does posterior instrumentation not improve fixation strength of the anterior construct, pedicle screw insertion was found in our 148  Chapter 6  Rod Extension and Cement on Destabilized Level  earlier study to disrupt trabecular bone and decrease interbody device-bone interface strength [Chapter 5]. The use of posterior rod extension thus directly enhanced 2 out of 4 of the interfaces (A and B in Figure 6.1) in the fusion construct. Cement augmentation of pedicle screws, on the other hand, directly improved overall structural stability to the posterior [Tan et al. 2004 (Chapter 3)] device-bone interfaces. Cement augmentation of pedicle screws was also found in our previous study to increase interface strength at the interbody device-inferior vertebral bone interface [Chapter 5] (interface D in Figure 6.1). The use of cement augmentation of pedicle screws thus enhanced at least 3 out of 4 of the interfaces (A, B and D in Figure 6.1) in the combined anterior-posterior fusion construct (Figures 6.1 and 6.23A). Cement augmentation should thus, theoretically, result in more stable constructs over posterior rod extension, and probably better condition for bony fusion.  (A) Figure 6.23  (B)  (A) Schematic diagrams of rigid instrumentation following corpectomy, with cement augmentation of pedicle screws. Cement could directly reduce pedicle screw loosening, and reduce interbody device subsidence. (B) Schematic diagrams of rigid instrumentation following corpectomy, with extension of posterior instrumentation to the inferior adjacent vertebral level. Increasing the number of fixation points for the posterior instrumentation directly reduced pedicle screw loosening. Posterior instrumentation does not directly improve cage-vertebra interface and bending of the long rods under load could result in cage subsidence.  149  Chapter 6  6.4.8  Rod Extension and Cement on Destabilized Level  Advantages and Disadvantages Cement use in the spine is commonly associated with the techniques of vertebroplasty  and kyphoplasty in patients with vertebral compression fractures [Barr et al. 2000, Kostuik and Shapiro 2003, Phillips 2003, Berlemann et al. 2004, Rhyne et al. 2004, Heini 2005]. Alleviation of pain and correction of deformity were achieved. Cement was also used in revision surgeries [Yerby et al. 1998, Dvorak and Fisher 1999]. At other sites, cement was also used to improve screw fixation in bone with low bone mass [Cameron et al. 1975, Larsson 2002, Kwon et al. 2002]. The disadvantages and complications associated with the use of cement were previously highlighted.  The heat created during the exothermic polymerization of PMMA  cement and presence of toxic non-polymerized monomer could cause necrosis of neighbouring cells [Kim et al. 2004] and affect normal bone healing [Santin et al. 2004]. This problem was reduced with the use of alternate bone cement, made of materials such as calcium phosphate [Bai et al. 1999, Lim et al. 2002]. Cement leakage was reportedly a potential issue during injection, which could result in neurological deficit [Ratliff et al. 2001, Shapiro et al. 2003, Nakano et al. 2005]. The use of radiologic guidance during injection of cement improved safety of the technique [Barr et al. 2000, Laredo and Hamze 2005].  Cement also has the  disadvantage of not being easily removable after impregnation into porous trabecular bone, which could be problematic in cases with severe infection and the need for implant removal. Drilling and tapping a screw hole in cement is nearly impossible and patients with prior vertebroplasty have fewer surgical options in follow-up. The effect of the cement impregnated bone on bone growth and spinal fusion is also unclear. As bone needs vascularization for growth, it is not known currently if cement within the vertebrae would affect the fusion process at the adjacent fusion mass. Long posterior rods were most commonly used in fixation of deformity such as scoliosis and kyhosis. Extension rods were also used in patients with low bone mass, to improve spinal fixation in the osteoporotic patients [Hu 1997, Kostuik and Shapiro 2003, Babat et al. 2004]. They were used to bridge across unsymptomatic degenerated adjacent levels to prevent juxtafusion degeneration or kyphosis in spinal fixation [Hu 1997, Babat et al. 2004] and in revision surgeries [Dvorak and Fisher 1999, Diwan et al. 2003].  The disadvantages and  complications associated with the use of long rigid posterior rod extension included the postponement of adjacent level effects to the levels where the rod extension ends [Babat et al. 2004], and the disadvantage of immobilization of otherwise functional motion segments.  150  Chapter 6  6.5  Rod Extension and Cement on Destabilized Level  Conclusion Cement augmentation resulted in better fixation of the osteoporotic spine at the  immediate surgical level than with posterior rod extension. On the other hand, adjacent level motion was significantly reduced with posterior rod extension and less so with cement augmentation. Flexible rod extension resulted in significantly lower ROM at the extension level than without extension and higher ROM than with rigid extension. In conclusion, a combination of cement augmentation and flexible rod extension could provide the optimum solution to both issues of device loosening in the osteoporotic spine and adjacent level effects. 6.6  References  [I]  Akamaru T, Kawahara N, Tsuchiya H, Kobayashi T, Murakami H, Tomita K. Healing of autologous bone in a titanium mesh cage used in anterior column reconstruction after total spondylectomy. Spine 2002; 27:E329-E333.  [2]  Axelsson P, Johnsson R, Stromqvist B. The spondylolytic vertebra and its adjacent segment. Mobility measured before and after posterolateral fusion. Spine 1997; 22:414-417.  [3]  Babat LB, McLain RF, Bingaman W, Kalfas I, Young P, Rufo-Smith C. Spinal surgery in patients with parkinson's disease: construct failure and progressive deformity. Spine 2004; 29:2006-2012.  [4]  Bai B, Jazrawi LM, Kummer FJ, Spivak JM. The use of an injectable, biodegradable calcium phosphate bone substitute for the prophylactic augmentation of osteoporotic vertebrae and the management of vertebral compression fractures. Spine 1999; 24:1521-1526.  [5]  Bai B, Kummer FJ, Spivak J. Augmentation of anterior vertebral body screw fixation by an injectable, biodegradable calcium phosphate bone substitute. Spine 2001; 26:2679-2683.  [6]  Barr JD, Barr MS, Lemley TJ, McCann RM. Percutaneous vertebroplasty for pain relief and spinal stabilization. Spine 2000; 25:923-928.  [7]  Bauer TW, Muschler GF. Bone graft materials: An overview of the basic science. Orthopaedics and Related Research 2000; 371:10-27.  [8]  Berlemann U, Franz T, Oder R, Heini PF. Kyphoplasty for treatment of osteoporotic vertebral fractures: a prospective non-randomized study. European Spine Journal 2004; 13:496-501.  [9]  Bridwell KH, Lenke LG, McEnery KW, Baldus C, Blanke K. Anterior fresh frozen structural allografts in the thoracic and lumbar spine. Do they work if combined with posterior fusion and instrumentation in adult patients with kyphosis or anterior column defects? Spine 1995; 20:14101418.  [10]  Butterman GR, Glazer PA, Hu SS, Bradford DS. Revision of failed lumbar fusions. A comparison of anterior autograft and allograft. Spine 1997; 22:2748-2755.  [II]  Cameron HU, Jacob R, Macnab I, Pilliar RM. Use of polymethylmethacrylate to enhance screw fixation in bone. Journal of Bone and Joint Surgery 1975; 57A:655-656.  [12]  Chao EYS, Aro HT. Biomechanics of fracture fixation. In: Basic Orthopaedic Biomechanics, 2nd ed., edited by Mow VC and Hayes WC, Lippincott-Raven Publishers, Philadelphia 1997, 317-351.  [13]  Cook SD, Salkeld SL, Stanley T, Faciane A, Miller SD. Biomechanical study of pedicle screw fixation in severely osteoporotic bone. Spine J 2004; 4:402-408.  [14]  Diwan AD, Parvartaneni H, Cammisa F. Failed degenerative lumbar spine surgery. Clinics of North America 2003; 34:309-324.  Clinical  Orthopedic  151  Chapter 6  Rod Extension and Cement on Destabilized Level  [15]  Dvorak MF, Fisher CG. Revision spine surgery in the presence of osteoporosis. In: Revision Spine Surgery. Margulies JY, Aebi M, Farcy JPC (Eds.) Mosby, St Louis, 1999, 646-653.  [16]  Dvorak MF, Kwon BK, Fisher CG, Eiserloh III HL, Boyd M, Wing PC. Effectiveness of titanium mesh cylindrical cages in anterior column reconstruction after thoracic and lumbar vertebral body resection. Spine 2003, 28:902-908.  [17]  Evans JH. Biomechanics of lumbar fusion. 193:38-46.  [18]  Fisher CG, Keynan O, Boyd MC, Dvorak MF. The surgical management of primary tumors of the spine: Initial results of an ongoing prospective cohort study. Spine 2005; 30:1899-1908.  [19]  Goel VK (Moderator). Panelists: Cripton PA, Dooris AP, Goel VK, Ledet EH, Moumene M, Natarajan RN, Panjabi MM, Patwardhan G, Dengupta DK, Serhan HA, Wilke HJ. Roundtable discussion with the experts. Roundtables in Spine Surgery 2005; 1:59-95.  [20]  Goertzen D, Neller S, Putzier M, Tohtz S, Froehlich M. The effect of a flexible stabilization device on motion at the segment adjacent to a lumbar fusion. In Proceedings: 30th Annual Meeting of the International Society for the Study of the Lumbar Spine 2003, May 13-17, Vancouver, Canada.  [21]  Goertzen DJ, Lane C, Oxland TR. Neutral zone and range of motion in the spine are greater with stepwise loading than with a continuous loading protocol. An in vitro porcine investigation. J Biomech 2004; 37:257-261.  [22]  Harrington KD. Anterior decompression and stabilization of the spine as a treatment for vertebral collapse and spinal cord compression from metastatic malignancy. Clinical Orthopaedics and Related Research 1988; 233:177-197.  [23]  Hasegawa K, Hirano T. Internal fixation of osteoporotic spine. In: Internal Fixation in Osteoporotic Bone. An YH (Ed.) Thieme, New York, 2002, 128-136.  [24]  Heini PF. The current treatment - a survey of osteoporotic fracture treatment. Osteoporotic spine fractures: the spine surgeon's perspective. Osteoporosis International 2005; 16: S85-S92.  [25]  Holman PJ, Suki D, McCutcheon I, Wolinsky JP, Rhines LD, Gokaslan ZL. Surgical management of metastatic disease of the lumbar spine: experience with 139 patients. Journal of Neurosurgery, Spine 2005; 2:550-563.  [26]  H u S S . Internal fixation in the osteoporotic spine. Spine 1997; 22:43S-8S.  [27]  Ignatius AA, Augat P, Ohnmacht M, Pokinskyj P, Kock HJ, Claes LE. A new bioresorbable polymer for screw augmentation in the osteosynthesis of osteoporotic cancellous bone: a biomechanical evaluation. Journal of Biomedical Materials Research. (Part B, Applied Biomaterials) 2001; 58:254-60.  [28]  Kim CW, Minocha J, Wahl CE, Garfin SR. Response of fractured osteoporotic bone to polymethylmethacrylate after vertebroplasty: case report. Spine Journal 2004; 4:709-712.  [29]  Kostuik JP, Shapiro MB. Open surgical treatment of osteoporotic fractures and deformity of the spine. Instructional Course Lectures 2003, 52:569-578.  [30]  Kumar MN, Baklanov A, Chopin D. Correlation between sagittal plane changes and adjacent segment degeneration following lumbar spine fusion. European Spine Journal 2001; 10:314-319.  [31]  Kumar MN, Jacquot F, Hall H. Long-term follow-up of functional outcomes and radiographic changes at adjacent levels following lumbar spine fusion for degenerative disc disease. European Spine Journal 2001; 10:309-313.  [32]  Kwon BK, Goertzen DJ, O'Brien PJ, Broekhuyse HM, Oxland TR. Biomechanical evaluation of proximal humeral fracture fixation supplemented with calcium phosphate cement. Journal of Bone and Joint Surgery 2002; 84A:951-961.  Clinical Orthopaedics and Related Research 1985;  152  Chapter 6  Rod Extension and Cement on Destabilized Level  [33]  Laredo JD, Hamze B. Complications of percutaneous vertebroplasty and their prevention. Seminars in Ultrasound, CT and MRI 2005; 26:65-80.  [34]  Larsson S. Treatment of osteoporotic fractures. Scandinavian Journal of Surgery 2002; 91:140146.  [35]  Lee CK. Accelerated degeneration of the segment adjacent to a lumbar fusion. 13:375-377.  [36]  Lim TH, Brebach GT, Renner SM, Kim WJ, Kim JG, Lee RE, Andersson GBJ, An HS. Biomechanical evaluation of an injectable calcium phosphate cement for vertebroplasty. Spine 2002; 27:1297-1302.  [37]  Lotz JC, Hu SS, Chiu DFM, Yu M, Colliou O, Poser RD. Carbonated apatite cement augmentation of pedicle screw fixation in the lumbar spine. Spine 1997; 22:2716-2723.  [38]  McKellop HA, Milligan HL, Rostlund T. Long term biostability of polyacetal (Delrin®) implants. Journal of Heart Valve Disease 1996; 5Suppl2:S238-S242.  [39]  Molinari RW, Bridwell KH, Klepps SJ, Baldus C. Minimum 5-year follow-up of anterior column structural allografts in the thoracic and lumbar spine. Spine 1999; 24:967-972.  [40]  Moore DC, Maitra RS, Farjo LA, Graziano GP, Goldstein SA. Restoration of pedicle screw fixation with an in situ setting calcium phosphate cement. Spine 1997; 22:1696-1705.  [41]  Nakano M, Hirano N, Ishihara H, Kawaguchi Y, Matsuura K. Calcium phosphate cement leakage after percutaneous vertebroplasty for osteoporotic vertebral fractures: risk factor analysis for cement leakage. Journal of Neurosurgery: Spine 2005; 2:27-33.  [42]  Nathan M, Pope MH, Grobler LJ. Osteophyte formation in the vertebral column: A review of the etiologic factors, Part 1. Comtemporary Orthopaedics 1994; 29:31-37.  [43]  Niosi CA, Zhu QA, Wilson DC, Keynan O, Wilson DR, Oxland TR. Biomechanical characterization of the three-dimensional kinematic behaviour of the Dynesys dynamic stabilization system: an in vitro study. European Spine Journal 2005; Oct 11 [Electronic publication, ahead of print].  [44]  Oda I, Cunningham BW, Buckley RA, Goebel MJ, Haggerty CJ, Orbegoso CM, McAfee PC. Does spinal kyphotic deformity influence the biomechanical characteristics of the adjacent motion segments? An in vivo animal model. Spine 1999; 24:2139-2146.  [45]  Oxland TR, Panjabi MM. The onset and progression of spinal injury: a demonstration of neutral zone sensitivity. Journal of Biomechanics 1992; 25:1165-1172.  [46]  Panjabi MM, Goel VK. Adjacent-level effects: design of a new test protocol and finite element model simulations of disc replacement. Roundtables in Spine Surgery 2005; 1:45-55.  [47]  Patwardhan AG, Havey RM, Meade KP, Lee B, Dunlap B. A follower load increases the loadcarrying capacity of the lumbar spine in compression. Spine 1999; 24:1003-9.  [48]  Pflugmacher R, Schleicher P, Schaefer J, Scholz M, Ludwig K, Khodadadyan-Klostermann C, Haas NP, Kandziora F. Biomechanical comparison of expandable cages for vertebral body replacement in the thoracolumbar spine. Spine 2004; 29:1413-1419.  [49]  Phillips FM. Minimally invasive treatments of osteoporotic vertebral compression fractures. Spine 2003; 15S:S45-S53.  [50]  Polikeit A, Nolte LP, Ferguson SJ. The effect of cement augmentation on the load transfer in an osteoporotic functional spinal unit. Finite element analysis. Spine 2003; 28:991-996.  [51]  Ratliff J, Nguyen T, Heiss J. Root and spinal cord compression from vertebroplasty: Spine 2001; 26:E300-E302.  Spine 1988;  methylmethacrylate  153  Chapter 6  Rod Extension and Cement on Destabilized Level  [52]  Renner SM, Lim TH, Kim WJ, Katolik L, An HS, Andersson GBJ. Augmentation of pedicle screw fixation strength using an injectable calcium phosphate cement as a function of injection timing and method. Spine 2004; 29:E212-E216.  [53]  Rhyne A 3rd, Banit D, Laxer E, Odum S, Nussman D. Kyphoplasty: Report of eight-two thoracolumbar osteoporotic vertebral fractures. Journal of Orthopaedic Trauma 2004; 18:294299.  [54]  Santin M, Motta A, Borzachiello A, Nicolais L, Ambrosio L. Effect of PMMA cement radical polymerization on the inflammatory response. Journal of Material Science: Materials in Medicine 2004; 15:1175-1180.  [55]  Sarzier JS, Evans AJ, Cahill DW. Increased pedicle screw strength with vertebroplasty augmentation in osteoporotic spines. Journal of Neurosurgery 2002; 96S:309-312.  [56]  Schultheiss M, Hartwig E, Claes L, Kinzl L, Wilke HJ. Influence of screw-cement enhancement on the stability of anterior thoracolumbar fracture stabilization with circumferential instability. European Spine Journal 2004; 13:598-604.  [57]  Shapiro S, Abel T, Purvines S. Surgical removal of epidural and intradural polymethylmethacrylate extravasation complicating percutaneous vertebroplasty for an osteoporotic lumbar compression fracture. Case report. Journal of Neurosurgery 2003; 98(1 Suppl):90-92.  [58]  Shono Y, Kanada K, Abumi K, McAfee PC, Cunningham BW. Stability of posterior spinal instrumentation and its effects on adjacent motion segments in the lumbosacral spine. Spine 1998; 23:1550-1558.  [59]  Soshi S, Shiba R, Kondo H, Murota K. An experimental study on transpedicular screw fixation in relation to osteoporosis of the lumbar spine. Spine 1991; 16:1335-1341.  [60]  Sran MM, Khan KM, Keiver K, Chew JB, McKay HA, Oxland TR. Accuracy of DXA scanning of the thoracic spine: cadaveric studies comparing BMC, areal BMD and geometric estimates of volumetric BMD against ash weight and CT measures of bone volume. European Spine Journal 2005; 14:971-976.  [61]  Tan JS, Kwon BK, Dvorak MF, Fisher CG, Oxland TR. Pedicle screw motion in the osteoporotic spine after augmentation with laminar hooks, sublaminar wires, or calcium phosphate cement: a comparative analysis. Spine 2004; 29:1723-1730.  [62]  Thongtrangan I, Balabhadra RSV, Le H, Park J, Kim DH. Vertebral body replacement with an expandable cage for reconstruction after spinal tumor resection. Neurosurgical Focus 2003; 15(5):E8.  [63]  Vahldiek MJ, Panjabi MM. Stability potential of spinal instrumentations in tumor vertebral body replacement surgery. Spine 1998; 23:543-550.  [64]  Vieweg U, Solch O, Kalff R. Vertebral body replacement system Synex in unstable burst fractures of the thoracic and lumbar spine - a retrospective study with 30 patients. [Article in German] Zentralbl Neurochir2003; 64:58-64.  [65]  Wittenberg RH, Lee KS, Shea M, White AA, Hayes WC. Effect of screw diameter, insertion technique, and bone cement augmentation of pedicular screw fixation strength. Clinical Orthopaedics and Related Research 1993; 296:278-287.  [66]  Wuisman PIJM, Van Dijk M, Staal H, Van Royen BJ. Augmentation of (pedicle) screws with calcium apatite cement in patients with severe progressive osteoporotic spinal deformities: an innovative technique. European Spine Journal 2000; 9:528-533.  [67]  Yerby SA, Toh E, McLain RF. Revision of failed pedicle screws using hydroxy apatite cement: A biomechanical analysis. Spine 1998; 23:1657-1661.  154  Chapter 6  Rod Extension and Cement on Destabilized Level  [68] Zindrick MR, Wiltse LL, Widell EH, Thomas JC, Holland WR, Field BT, Spencer CW. A biomechanical study of intrapeduncular screw fixation in the lumbosacral spine. Clinical Orthopaedics and Related Research 1986; 203:99-112.  155  Chapter  7  Adjacent  Level  Effects  Chapter 7 Adjacent Level Effects: Consequence of Extension of Posterior Instrumentation and Cement Augmentation of Pedicle Screws  Overview  This chapter addresses the issue of adjacent level effects following extension of posterior instrumentation and cement augmentation of pedicle screws, using the hybrid flexibility-stiffness test protocol. Both surgical techniques are available to surgeons to obtain improved fixation in the osteoporotic spine, however, adverse adjacent level effects could result following rigid fixation. Effects of both cement augmentation and posterior rod extension were evaluated. The effects of flexible extension rods on the adjacent level were compared against rigid extension rods and against no extension rods.  Cement augmentation of pedicle screws resulted in  increased ROM and vertebral strain at adjacent levels.  Posterior rod extension resulted in  reduced ROM at the extension level but increased ROM and strain at the next non-instrumented level. Flexible extension rod resulted in lesser degree of adjacent level effects at the remaining non-instrumented levels than with rigid rods. Extension with flexible rods might be a solution to prevent adjacent level effect at the extension level while minimizing the effect at remaining noninstrumented levels. This study was carried out in parallel with the previous study (Chapter 6), using the same 12 specimens. The results of this study, together with results in Chapter 6, were presented as a single study at these national and international conferences: 1) 52nd Annual Meeting of the Orthopaedic Research Society, Mar 19-22, 2006, Chicago, Illinois, USA. 2) 40th Annual Meeting of the Canadian Orthopaedics Research Society, Jun 2-4, 2006, Toronto, Ontario, Canada. 3) 33rd Annual Meeting of the International Society for the Study of the Lumbar Spine, Jun 1317, 2006, Bergen, Norway. A manuscript on the results of this study is being prepared as the second of a two-part paper for publication in the refereed journal Spine. 156  Adjacent  Level  Effects  Chapter  7  7  Adjacent Level Effects: Consequence of Extension of Posterior Instrumentation and Cement Augmentation of Pedicle Screws  7.1  Introduction Degenerative changes at adjacent segments following spinal fusion surgery have been  observed in patients for many years [Eck et al. 1999, Park et al. 2004, Hilibrand and Robbins 2004].  The most common clinical complication after spinal fusion at the adjacent segment  appears to be disc degeneration and vertebral fracture, while other clinical observations such as lysthesis, hypertrophic facet joint arthritis, herniated nucleus pulposus and stenosis are also attributed to fusion [Park et al. 2004]. The etiology of these clinical observations has been attributed to two main factors: 1) biomechanical changes and 2) natural progression of degeneration [Eck et al. 1999, Park et al. 2004, Hilibrand and Robbins 2004]. The focus of this study was on determining the biomechanical changes at the adjacent levels. The terms adjacent level effects and adjacent segment biomechanics have been used to refer to the altered load bearing and motion characteristics observed at the segments adjacent to a spinal fusion [Ha et al. 1993, Kumar et al. 2001, Panjabi 2002], while adjacent segment degeneration and adjacent segment disease have been used to refer to the radiographically or clinically observed degenerative changes and related new symptoms [Eck et al. 1999, Park et al. 2004, Hilibrand and Robbins 2004]. Many experimental studies have provided supporting evidence to show increased motion, increased intradiscal pressure, and altered stress states to the adjacent spinal segment(s) after fusion, which might explain the early degenerative changes observed radiographically or clinically. Past biomechanical studies using human cadaveric models have demonstrated that fusion was associated with increased disc pressure [Weinhoffer et al. 1995, Chow et al. 1996, Cunningham et al. 1997, Eck et al. 2002], segmental motion [Quinnell and Stockdale 1981, Chow et al. 1996, Esses et al. 1996, Fuller et al. 1998, Bastian et al. 2001, Eck et al. 2002, Akamaru et al. 2003, DiAngelo et al. 2003], facet loads [Lee and Langrana 1984] and vertebral strain at the adjacent segments. However, a critical review of these studies has not been carried out. Cunningham et al (1997) found increased intradiscal pressures at adjacent segments (L2-L3 and L4-L5) after instrumentation at L3-L4 when eleven L1-S1 specimens were tested under 12.5 degrees of flexion and extension. Similarly, Weinhoffer et al (1995) found increased intradiscal  pressures  at  adjacent  non-instrumented  discs  (L3-L4  and  L4-L5)  after  instrumentation from L5-S1 and from L4-S1 when six specimens (L2-S1) were tested under flexion to 20 degrees. Chow et al (1996) also carried out single level (L4-L5) and double levels 157  Chapter  7  Adjacent  Level  Effects  (L4-S1) instrumentation and found increased intradiscal pressures at adjacent non-instrumented discs (L1-L2, L2-L3, L3-L4 and L5-S1) and increased segmental motions when six specimens (L1-S1) were tested under 15 degrees of flexion and 10 degrees of extension. Eck et al (2002) compared intradiscal pressure and segmental motion at C4-C5 and C6-C7 under 20 degrees of flexion and 15 degrees of extension between intact specimens and after anterior cervical plating at C5-C6 using six cervical spines (C3-T1).  These studies demonstrated biomechanical  changes at adjacent spinal segments following rigid fixation that might cause adjacent segment diseases. It is believed that these altered biomechanics at the adjacent level affect the natural progression of the disease at the adjacent levels.  Cyclic pressure differentials in the  intervertebral discs are believed to be associated with metabolic production and fluid exchanges, and increased intradiscal pressure associated with early disc degeneration [Cunningham et al. 1997].  Increased segmental motions were associated with increased  vertebral strain and facet loads and could lead to early degenerative changes such as trabecular fracture and hypertrophic facet joint arthritis. Changes to segmental rotation centre [Lee and Langrana 1984] and lordotic angles [Umehara et al. 2000, Akamaru et al. 2003, Sudo et al. 2003] were some other biomechanical factors attributed as related causes for the radiographically or clinically observed adjacent level effects.  Adjacent segments that were  beginning to develop or were in the early stages of natural degenerative changes might be more adversely affected by the biomechanical changes than other healthy segments. The abnormal loading conditions on the adjacent degenerated segments can produce tissue trauma and/or adaptive changes that may result in more severe degenerative stages [Stokes and latridis 2004]. The effect of altered disc metabolism and vertebral strain on the adjacent segment with early degenerative changes could also be factors that lead to an accelerated degenerative process. These same abnormal loading conditions might have a slower rate of effect or no adverse effect on the tissues in healthy neighbouring segments. The natural progression of degeneration and biomechanical changes at adjacent segments are two inseparable factors that could be equally important to adjacent level disease. In contrast to earlier studies, some recent biomechanical studies using cadaveric specimens claimed no adverse biomechanical effects at the adjacent spinal segments [Rohlmann et al. 2001, Lindsey et al. 2003, Swanson et al. 2003, Schmoelz et al. 2003,]. These recent studies typically used load controlled testing methodologies in which pure moments of the same magnitude were applied to the intact and instrumented models and parameters at the adjacent non-instrumented levels were compared. Of note are the findings in Chapter 6 where 158  Chapter  7  Adjacent  Level  Effects  kinematics at the inferior non-instrumented levels under the same applied loads were not significantly different between the intact and various instrumented conditions. Similarly, in finite element studies with the same applied muscle loads to the intact and instrumented model, no significant adverse biomechanical effects at the adjacent segments were found [Rohlmann et al. 1999, Zander et al. 2003]. Lindsey et al (2003) and Swanson et al (2003) tested seven L2-L5 specimens under flexion-extension (FE), axial rotation (AR) and lateral bending (LB) with 7.5 Nm bending moment and 700 N compression in both the intact state and after implantation of an interspinous device at L3-L4. They found no significant difference in range of motion (ROM) [Lindsey et al. 2003] and intradiscal pressure [Swanson et al. 2003] in all three rotations between intact and implant states for both adjacent non-instrumented levels (L2-L3 and L4-L5) and concluded that the implant did not significantly affect the ROM and intradiscal pressure at adjacent levels. Schmoelz et al (2003) tested six L2-L5 specimens with pure moments of 10 Nm in FE, AR and LB in 1) the intact state, 2) after implantation of a dynamic stabilizing device and 3) after implantation of an internal fixator, across L3-L4. The ROM and neutral zone of the adjacent segments (L2-L3 and L4-L5) were not found to be affected under the applied loading conditions. Rohlmann et al (2001) tested seven lumbar specimens (4 L1-L5 and 3 T12-L4) under pure moments of 3.75 Nm in FE, AR and LB in the intact state and after fixation of the middle three vertebrae leaving one intact segment above and below the implant. The changes in intradiscal pressure and intersegmental rotation at the adjacent levels were small under the applied load-controlled (pure moments) conditions. Using a non-linear finite element model of the lumbosacral spine (L2-S1), Zander et al (2003) evaluated the stresses in the intervertebral discs, facet joint forces and intradiscal pressure at the adjacent level (L3-L4) after simulated facetectomy and laminectomy at L4 and L5.  Muscle forces were applied to simulate  physiological loads experienced while standing and forward bending. With the same loading conditions in the intact and surgical states, they found negligible effect on these parameters at L3-L4. In an earlier similar finite element study by Rohlmann et al (1999) on the lumbar spine (L1-L5), stresses in adjacent intervertebral discs (L1-L2 and L4-L5) after fixation of the L2-L4 levels were compared against the intact state. Muscle forces were simulated for standing and flexion postures.  The result was also similar, in that, the internal fixators had only minor  influence on stresses at the adjacent discs. The application of pure moments was the common denominator in these studies and was why no significant differences were found at the adjacent non-instrumented levels. Various other different testing protocols were used in the past to show these biomechanical adjacent level effects. The strengths, limitations and underlying assumptions of 159  Chapter 7  Adjacent Level Effects  t h e s e testing m e t h o d s h o w e v e r w e r e not clear to the clinicians a n d b i o m e c h a n i c a l alike.  engineers  T h e technique of applying a pure m o m e n t a n d m e a s u r i n g the resultant three translations  a n d t h r e e r o t a t i o n s w a s r e f e r r e d t o a s t h e 'flexibility' p r o t o c o l [ P a n j a b i a n d G o e l 2 0 0 5 ] .  The  f l e x i b i l i t y p r o t o c o l h a d b e e n u s e d t o d e t e r m i n e if a p a r t i c u l a r i n s t r u m e n t a t i o n w a s o v e r l y f l e x i b l e immediately  post  surgery  as  compared  to  its  intact  state  or  t e c h n i q u e s [ P a n j a b i et al. 1 9 7 6 , P a n j a b i 1 9 8 8 , A b u m i et al. 1 9 9 0 ] . in t h r e e o r t h o g o n a l d i r e c t i o n s in f l e x i b i l i t y t e s t s : c o m m o n l y  against  established  surgical  Pure m o m e n t s were applied  in F E , A R a n d L B .  The  applied  m o m e n t s w e r e usually of t h e s a m e m a g n i t u d e s o that t h e resultant d i s p l a c e m e n t s c o u l d directly c o m p a r e d b e t w e e n intact a n d i n s t r u m e n t e d state.  be  S p e c i m e n s with a single functional  s p i n a l unit ( F S U ) w e r e c o m m o n l y u s e d f o r s u c h t e s t s [ O x l a n d et al. 1 9 9 6 , L u n d et al. 1 9 9 8 , 2 0 0 0 , N y d e g g e r et al. 2 0 0 1 , L e H u e c et al. 2 0 0 2 ] .  T h e flexibility test w a s well suited to c o m p a r e  flexibility at t h e i m p l a n t e d level b e t w e e n intact a n d i n s t r u m e n t e d s t a t e s a n d a l s o f o r c o m p a r i s o n between  novel  techniques  against  existing  instrumentation  techniques.  Following  the  r e c o m m e n d a t i o n s t h a t l e n g t h of s p e c i m e n s s h o u l d h a v e at l e a s t o n e f r e e s e g m e n t o n e i t h e r e n d of t h e i n s t r u m e n t e d levels [ W i l k e et al. 1 9 9 8 ] , r e c e n t t e s t s w e r e c a r r i e d o u t w i t h t h r e e o r m o r e FSUs.  W i t h t h e l o n g e r t e s t s p e c i m e n s in m o r e r e c e n t s t u d i e s , t h e f l e x i b i l i t y o f t h e  adjacent  s e g m e n t s w a s i n c l u d e d in t h e r e s u l t s o f t h e s e s t u d i e s [ R o h l m a n n e t a l . 2 0 0 1 , L i n d s e y e t a l . 2 0 0 3 , S c h m o e l z et al. 2 0 0 3 , C h a p t e r 6].  R e a d e r s of these studies could interpret the results for  t h e i n s i g n i f i c a n t e f f e c t a t t h e a d j a c e n t s e g m e n t s in t w o w a y s .  Firstly, t h e r e a d e r c o u l d w r o n g l y  i n t e r p r e t t h a t t h e i n s t r u m e n t a t i o n e l i m i n a t e d t h e a d v e r s e b i o m e c h a n i c a l e f f e c t s at t h e segments.  adjacent  A l t e r n a t i v e l y , t h e r e a d e r in u n d e r s t a n d i n g t h a t p u r e m o m e n t w a s a p p l i e d in t h e i n t a c t  a n d i n s t r u m e n t e d states, w o u l d e x p e c t t h e b i o m e c h a n i c a l p a r a m e t e r s at t h e a d j a c e n t s e g m e n t s to be the same.  A n d a s t h e r e s u l t s s h o w e d i n s i g n i f i c a n t e f f e c t a t t h e a d j a c e n t s e g m e n t s , it  inversely implies that pure m o m e n t w a s successfully applied. A n u n d e r s t a n d i n g of the u n d e r l y i n g m e c h a n i c s w o u l d allow t h e r e a d e r ( a n d a u t h o r s ) to a p p r e c i a t e t h e latter interpretation as correct (Figure 7.1).  In F i g u r e 7 . 1 a , w h e n p u r e  moments  w e r e a p p l i e d at the e n d s of a long b e a m , a n y s e c t i o n t h r o u g h o u t t h e b e a m e x p e r i e n c e d same moment.  Following reinforcement of a section with the s a m e m o m e n t applied  7 . 1 b ) , a n y s e c t i o n t h r o u g h o u t t h e b e a m w o u l d still e x p e r i e n c e t h e s a m e m o m e n t .  above mentioned  in vitro  pure In t h e  studies w h e r e the s a m e pure m o m e n t w a s applied to the intact a n d  instrumented states, the m o m e n t s experienced magnitude.  (Figure  With  m o m e n t s , t h e o n l y d i f f e r e n c e w o u l d b e a r e d u c t i o n in m o t i o n a t t h e r e i n f o r c e d s e c t i o n .  the  at t h e a d j a c e n t s e g m e n t s w e r e of t h e  same  O n e w o u l d thus e x p e c t no significant c h a n g e s to adjacent level kinematics  under  these s a m e loading conditions.  T h e s e s t u d i e s i n d e e d w e r e in a g r e e m e n t w i t h t h e o r y a n d t h u s  160  Chapter 7  Adjacent Level Effects  we knew that pure moments were successfully applied.  The flexibility protocol is thus  unsuitable for assessing gross kinematic changes in the adjacent segments. Some studies on the other hand supposedly applied 'pure moments' of same magnitudes to the specimens in the intact and instrumented state but found adjacent level biomechanical changes at the noninstrumented adjacent levels [Bastian et al. 2001, Untch et al. 2004]. It can only be assumed that the moments applied might not be 'pure' and the test setup might not have provided 'unrestricted' motion as claimed.  Figure 7.1  When pure moments were applied at the ends of a beam, the moment experienced in any section throughout the beam was the same even when a section was reinforced, (a) Bending moments applied at the ends of a uniform beam. The moments in any section of the beam were the same, (c) A section of the same beam was now reinforced and the same pure moment applied at the ends of the beam. The moment experienced at any section of the reinforced beam remained the same.  Another test protocol called the 'stiffness' protocol was used to assess if a particular instrumentation was overly less stiff immediately post surgery as compared to its intact state or against established surgical techniques.  In the stiffness protocol, individual translations and  rotations in three orthogonal directions are applied and the resulting six load components are measured. A problem with using the stiffness protocol surrounds the difficulty of applying a 161  Chapter 7  Adjacent Level Effects  single component motion while disallowing the other five motion components [Panjabi 2002]. As coupled motions are commonly seen in the FSU, the stiffness test in applying a pure rotation is too restrictive and can result in excessive resultant loads. Another issue is in pre-determining the centre of rotation of the spine for the mode of motion under test as the centre of rotation shifts with rotation angle [Yoshioka et al. 1990]. Furthermore, the problem of locating a centre of rotation would be more pronounced if the test segment was much longer such as in the study by Stanley et al. (2004) and in the current study.  For studying adjacent level effects, the  stiffness protocol was difficult to implement and less repeatable between different laboratories. Many studies mentioned earlier that showed increased intradiscal pressure or motion at the adjacent segments made use of a simplified stiffness protocol [Lee and Langrana 1984, Weinhoffer et al. 1995, Chow et al. 1996, Eck et al. 2002]. These studies applied the same magnitude of flexion/extension rotation by displacing an anterior/posterior offset point by a fixed vertical displacement. In other studies, moments of same magnitudes were applied using the same technique of applying a point load with a fixed anterior/posterior offset [Esses et al. 1996, Shimamoto et al. 2001, Akamaru et al. 2003]. The problem of locating a centre of rotation was bypassed as the centre of the vertebral body was used as the reference point for the anterior/posterior offset. The motion of the specimen however was found to not reflect in vivo motion, especially when more segments were tested.  Buckling and instability of the test  specimen was also a problem and was more pronounced when longer specimens were tested. More importantly, the moment applied at each segment became ambigious as the moment arm changed with motion of the specimen [Panjabi 1988, Panjabi and Goel 2005]. With the inherent problems associated with both the flexibility and the stiffness protocols, a new protocol called the hybrid flexibility-stiffness protocol was recently proposed [Panjabi 2002]. The hybrid protocol has an underlying assumption that patients would assume the same overall ROM in their daily routine after a solid fusion. This hybrid protocol was further described in detail recently [Panjabi and Goel 2005]. In the hybrid protocol, the application of a pure moment in the flexibility protocol was married to the application of a fixed displacement in the stiffness protocol. Flexibility tests were first carried out on the intact specimen to determine the ROM of the whole segment under a physiologic moment.  After instrumentation, the same  flexibility tests were carried out until the same main motion corresponding to the test direction was achieved. The main difference of the hybrid protocol from the stiffness protocol was that the test was carried out with applied pure moments instead of displacement control, while the difference from the flexibility protocol was that the hybrid test was carried out until the intact ROM was achieved instead of to the same moment. The intradiscal pressures, vertebral strains 162  Chapter 7  Adjacent Level Effects  and/or kinematics of the adjacent segments at the same overall ROM were compared between intact and instrumented states. However, the assumption of the same overall ROM after fusion with the hybrid protocol would remain a controversial topic for discussion [Schmoelz et al. 2003]. It was countered that the application of pure moments as suggested in the hybrid protocol in a study would not provide an explanation to the clinical observations of adjacent level degeneration occurring at the immediate one or two adjacent segments to the instrumentation. Based on the underlying assumption of same overall ROM, the reduced motion at the instrumented level would have to be contributed by the other mobile levels. In order to achieve this higher motion at the other mobile levels with rigid fixation, the necessary applied moment would be substantially higher. As all adjacent non-instrumented segments in the test model would be subjected to the same increase in applied pure moment, a proportional increase in measured biomechanical parameters would be expected. Another problem with the hybrid protocol as proposed [Panjabi and Goel 2005] is the possibility of excessive loads necessary to reach the same baseline ROM especially with the inclusion of rigid fixations. Our solution is to apply the same maximum pure moments (3.75 Nm for cervical, 5 Nm for thoracic and 7.5 Nm for lumbar) to the various instrumentation configurations in the experimental stage, but to make comparisons at the same baseline ROM during data analyses. The baseline ROM would correspond to the stiffest construct within the study. A similar analysis protocol and hybrid test protocol has been recently reported in a conference abstract [Goertzen et al. 2003]. Despite the above mentioned problems associated with the hybrid protocol, it is the current state of the art in in vitro spine testing for studying adjacent level effects, and is a more coherent methodology than either the flexibility or the stiffness protocols. Moreover, the hybrid protocol remains a technique that has not been published in refereed journal.  The initial  objective of this study was to demonstrate that the proposed hybrid test protocol could be used to assess biomechanical changes at adjacent levels. The main purpose of this study was to determine, using the hybrid protocol, if flexible posterior extension rods in combination with cement augmentation of pedicle screws could reduce adjacent biomechanical effects following spinal fixation. Motion and strains at adjacent segments were compared between flexible, rigid and no posterior extension rods, and with and without cement augmentation of pedicle screws.  163  Chapter  7.2  7  Adjacent  Level  Effects  Methods The experiment in Chapters 6 and 7 were carried out concurrently, on the same 12  specimens. Specimen conditions were described in Section 6.2.1. The same repeated measures experimental protocol to test the two way interaction between cement augmentation and posterior rod extension was also carried out concurrently for both studies. Details on experimental protocol were described in Section 6.2.2. The test protocol, described in Section 6.2.3, was also carried out concurrently for both studies.  Post-test analyses using the hybrid protocol of the kinematic data was used to  determine adjacent level effects. In addition to the measurement of kinematic data, vertebral body strains were also measured. 7.2.1  Hybrid Flexibility-Stiffness Analyses Motions in three-dimensional space of all vertebral bodies (T9, T10, T12, L1, L2 and L3  in a T9-L3 segment, except the corpectomy level) were collected using an optoelectronic camera system (Optotrak 3020, Northern Digital Inc., Waterloo, Canada). Data for motion and applied moment was synchronised and recorded at 20 Hz for the duration of each test. In the post-test analyses, overall (T9-L3) AR, LB, and FE ROM in both positive and negative directions were first determined from the flexibility curves at the third (last) cycle for all surgical conditions. The smallest ROM under +/-5 Nm applied moment (in both positive and negative directions), which corresponded to the most rigid fixation, was identified for each flexibility test direction (Figure 7.2). In the example in Figure 7.2, cement augmentation and rigid rod extension resulted in the most rigid fixation and the baselines overall +ROM and -ROM were 16.4° and -11.5°. Corresponding parameters, such as applied moment, vertebral strains and intradiscal pressure, between surgical conditions were compared at these baseline ROMs in the hybrid protocol. In order to compare changes at the adjacent levels using the hybrid protocol, the applied moments (in both positive and negative directions) necessary to achieve these baseline ROMs for each surgical condition needed to be first determined. For the example in Figure 7.2, the moments applied to the intact spinal segment necessary to achieve the baseline ROMs were 0.6 Nm and -0.6 Nm respectively. Corresponding vertebral strains and segmental ROMs were determined at these moments for the intact state and compared against other surgical states. These moments for each surgical condition were different and needed to be individually determined for all flexibility test directions and surgical conditions.  164  Chapter 7  Adjacent Level Effects  Axial Rotation - H1022 ' Intact No Extension Flexible Extension  25  4  20 H baseline +ROM  Rigid Extension Cemented, No Ext  Ui  Cemented, Flex Ext  c  < c o  3  •Cemented, Rig Ext  6 ~ tL -6 r  4  e  6  A baseline - R O M  -20  J  Moment (Nm)  Figure  7.2  Flexibility curves at the third cycle for all surgical conditions were compared to determine the baseline (smallest) positive and negative range of motion (+ROM and -ROM) and corresponding applied moment. This baseline ROM was used for further comparison, and was based on the underlying assumption in the hybrid flexibility-stiffness protocol that the patient would assume the same overall ROM in their daily routine after fusion surgery. In this specimen, the baseline +ROM and -ROM were 16.4" and -11.5° respectively, with cemented screws and rigid posterior extension. The positive moments (in increasing order) necessary to create this baseline +ROM for each surgical condition were 0.6, 1.3, 1.9, 2.2, 3.1 and 4.1 Nm respectively. The negative moments (in decreasing order) necessary to create this baseline -ROM for each surgical condition were -0.6, -1.4, -1.7, -2.4, -2.9 and -4.4 Nm respectively.  Using these moments, the corresponding intersegmental ROMs (both positive and negative) could be determined from the intersegmental flexibility curves. The positive and negative intersegmental ROMs at each level were summed to obtain the overall (in both positive and negative directions) intersegmental ROM for the level.  Intersegmental ROM was  normalized against its intact condition so as to reduce effects of inter-specimen differences. The normalized intersegmental ROMs at the adjacent levels were compared between surgical conditions. These changes at the adjacent levels, under the same overall ROM, were the  165  Chapter 7  Adjacent Level Effects  underlying assumptions in the hybrid protocol. Changes in vertebral body strains using this analysis protocol were also carried out in this study. 7.2.2  Vertebral Body Strains  Single axis strain gauges (TML FLG-02-23, Tokyo Sokki Kenkyujo, Tokyo, Japan) were attached to 2 distal vertebral bodies in all tests. The L1 and L2 in T9-L3 spinal segments and the T12 and L1 vertebral bodies in T8-L2 segments were each attached with 3 strain gauges: two on the lateral aspects of each vertebral body and one on the left surface of each spinous process (Figure 7.3). Strain data was recorded in synchronisation with applied moment and motion data at 20 Hz for the duration of each test (Figure 7.4).  Figure 7.3  Two strain gauges were attached onto the lateral aspects of the body and one strain gauge attached onto the left surface of the spinous process, on 2 distal vertebral bodies to the corpectomy level (T12 and L1 in T8-L2). Gauges 1-4 were attached to the vertebral bodies and gauges 5 and 6 to the spinous processes. Positive strain gauge values represented tensile strain.  During post-test analyses, the hybrid protocol was used and strain values corresponding to the positive and negative baseline ROMs were determined. The overall change in strain from the positive baseline ROM to the negative baseline ROM was calculated and normalized against the corresponding intact ranges for each test. Changes in strain readings at the same overall ROM indicate possible adjacent level effects following cement augmentation or posterior extension. 166  Chapter 7  Adjacent Level Effects  1500 n  Strain Gauge 1 Strain Gauge 2 Strain Gauge 3 Strain Gauge 4 Strain Gauge 5 Strain Gauge 6  -500  J  Time (sec)  Figure 7.4  Strain data were collected for the full duration of each test. The data was passed through a low pass filter, which removed noise in the data.  For FE tests, all strain gauges were compared between the different surgical conditions. In LB, only gauges attached to the vertebral bodies were compared. Gauges attached to the spinous process were ignored in LB as they were not sensitive to changes during LB tests. Similarly, in AR motion, the changes in all gauges were not sensitive to this test direction and thus were not compared. As gauges 1 and 3 were mounted on the same vertebra (T12) and gauges 2 and 4 on the adjacent vertebra (L1), some symmetry in results could be expected in FE.  7.2.3  Statistical Analysis One-way repeated measures ANOVA with Newman-Keuls post-hoc tests were carried  out to determine significant difference within specimens between intact and surgical conditions. Two-way repeated measures ANOVA analyses, with Newman-Keuls post-hoc analyses, were carried out to determine the effect of cement augmentation (factor 1) and posterior rod extension (factor 2) on (i) normalized applied moment, (ii) intersegmental ROMs and (iii) normalized vertebral strains, for each flexibility test direction. An alpha of 0.01 was chosen.  167  Chapter 7 7.3  Adjacent Level Effects  Results  7.3.1 Overall Flexibility The overall ROMs (across T9-L3 or T8-L2) in both positive and negative directions in the intact and surgical states in AR, LB and FE were determined from the flexibility curves (Figure 7.5). The surgical configuration with the smallest ROM in each specimen and test direction was used for baseline comparison. Of note, the mean overall ROMs were the smallest with cement augmentation and rigid extension in all directions (Table 7.1), but within individual specimen overall ROM was not always the smallest with any particular surgical configuration.  Axial Rotation  -H1022  Intact  Moment (Nm)  Figure 7.5  Typical flexibility curves of the T8-L2 spine of a specimen under AR motion, in the intact and various surgical intervention states.  7.3.2 Baseline Range of Motion and Minimum Moment Baseline ROM corresponded to the stiffest surgical construct within each specimen at 5 Nm of applied moment. The baseline ROM+ and ROM- occurred most frequently with rigid extension rods and cement (70%), followed by rigid extension rods without cement (17%) and by flexible extension rods with cement (9%) (Table 7.2).  168  Chapter  7  Table 7.1  Adjacent  Level  Effects  Mean overall range of motion (across T9-L3 or T8-L2) for different surgical conditions in the respective test directions at 5 Nm applied moment. R a n g e of M o t i o n , m e a n (s.d.) AR Intact  Not  No  cemented  Cemented  Table 7.2  extension  Flexible  extension  Rigid  extension  No  extension  Flexible  extension  Rigid  extension  1 2 3 4 5 6 7 8 9 10 11 12  19.4° 17.3° 17.5° 16.4° 15.1°  25.1° 23.7° 26.1° 22.6° 22.0°  (7.2°) (5.6°) (5.6°) (4.9°) (4.8°)  23.1° 23.3° 19.9° 18.2° 20.8° 17.7° 16.8°  (6.5°) (5.6°) (6.4°) (5.5°) (5.4°)  FE+FL  (6.7°) (8.1°) (6.9°) (6.1°) (6.1°) (5.5°) (4.9°)  22.0°(7.1°) 18.2° (7.1°) 15.5° 14.4° 16.8° 14.3° 14.0°  (5.9°) (4.4°) (5.8°) (4.9°) (4.7°)  Baseline ROM+ and ROM- for the 12 specimens. LB  ROM+ 3  16.4° 6.6° ay go 3  6.6° 8.6° 4.3° ag o 3  b  A  2  9.3° ag o a  2  a  3  a  10.3° 7.7° 6.1°  8.3° (3.0°) Corresponding Corresponding Corresponding  Mean (s.d.) a  b  0  The  FE  30.1° (7.8°) 2 9 . 2 ° (7.1°)  A R Specimen  LB  2 2 . 0 ° (6.6°) 21.1°(8.4°)  ROM-  F  ROM+  MI.5° -6.2° -6.1° -6.0° -7.6° a .3 40  ROM-  18.0° 10.6° 10.7° 8.8° b o c  3  3  b  3  3  -13.6° -9.3° °-8.3° a _ 30 c  3  12.7° 7.5° bygo  3  4.0° -10.3° 8.8° -8.7° 10.8° 8.5° c 40 -10.0° 11.1° -16.9° a o a 30 -7.4° -13.6° a g o a.8.1° 11.0° -10.3° -6.5° 6.2° 11.9° -9.3° -6.7° 11.3° -9.4° 6.0° ag o 7.2° -5.8° -7.0° -7.1° 11.7° -10.3° 8.8° (2.9°) (3.0°) (3.0°) (2.1°) to rigid extension rods with cement. to rigid extension rods without cement. to flexible extension rods with cement. 3  3  b  1  3  7  b  2  3  3  b  3  1 6 i 1  C  b  3  3  3  a  1  4  1  2  c  3  3  3  3  a  A  2  3  b  3  3  3  FE+FL  ROM+  2  ROM-  M I . 6° -8.4° -8.2° -5.8° -6.4° a .4 30 3  3  b  -9.7° -12.1° -8.7° -8.2° 3  b  ROM+  a  ROM-  o  1 6 2  6.6° 4.9° 5.8° 7.6° 7.6° c o a go 3  b  3  1  2  1  4  2  -7.0° -5.7°  9.5° 6.2° 6.2° ag o  -8.0° (2.3°)  8.9° (3.7°)  -4.9° -3.3° -4.0° a_ B  C  C  3  3  3  3  3  b  3  3  B  3  2  0  2  3  3  3  ?  -7.3° a _ 50  3  -7.1° -5.9° -4.0° -8.2° -4.8° -4.8° -5.0° (1.8°)  a p p l i e d m o m e n t n e c e s s a r y to a c h i e v e t h e b a s e l i n e + R O M a n d - R O M for t h e stiffest  construct would be  +5  N m a n d -5  a c h i e v e the s a m e baseline R O M s d e p e n d o n its o v e r a l l s t i f f n e s s .  N m respectively,  while the  applied moment  necessary  to  o n other surgical conditions within the s a m e s p e c i m e n w o u l d  In e a c h s p e c i m e n , t h e s e a p p l i e d m o m e n t s d i f f e r e d f o r t h e i n t a c t  spine a n d e a c h surgical condition (Table  7.3).  169  Chapter 7  Adjacent Level Effects  Table 7.3  Applied moments that created the baseline ROMs in specimen 1. AR+ ARLBFlexion LB+ Baseline ROMs from Table 7.2 16.4° 12.7° -11.5° 18.0° -13.6°  Not cemented Cemented  No Flexible Rigid No Flexible Rigid  Intact extension extension extension extension extension extension  1.9 0.6 1.3 2.2 3.1 4.1 5.0  -1.7 -0.6 -1.4 -2.9 -2.4 -4.4 -5.0  Applied Moment (Nm) 1.6 -1.0 1.9 -0.7 -1.7 3.0 -2.1 3.9 3.0 -1.5 5.0 -4.0 4.5 -5.0  1.1 0.6 1.8 1.7 0.9 3.0 5.0  Extension -11.6° -1.1 -0.4 -1.3 -2.1 -1.8 -3.3 -5.0  In the 2-way ANOVA analyses, the minimum applied moments at the same baseline ROMs were generally higher for surgical conditions with cement augmented screws and with posterior extension rods, in both the positive and negative directions (Table 7.3, Figures 7.6 to 7.9). Applied moments were significantly higher with cement augmentation in AR+ (by between 74% to 94% of intact), AR- (71-77%) and LB+ (50-62%), but not in LB- (30-57%), FE+ (28-53%) and FE- (21-61%) (Table 7.4, Figures 7.6 to 7.9). Applied moments were also significantly affected by configuration of posterior rod extension. Applied moments were significantly higher with rigid rods in AR+ (72-92%), AR- (84-90%), LB+ (92-104%), LB- (102-126%), FE+ (128142%) and FE- (77-117%). Applied moments were also significantly higher with flexible rods in AR+ (36-40%), AR- (32% with cement), LB+ (66-76%), LB- (61-88%), FE+ (67-92%) and FE(51-70%). As rigid extension rods resulted in stiffer constructs than with flexible extension rods, significantly higher moments were necessary in order to achieve the same baseline ROM. Table 7.4  2-way ANOVA results on normalized moment between cement (factor 1) and posterior rod extension (factor 2) under the same baseline ROM. AR+ LB+ Flexion Flexion+FL Factor 1: Cement < 0.001 < 0.001 0.04 0.03 Factor 2: Extension < 0.001 < 0.001 < 0.001 < 0.001 Interactions 0.1 0.7 0.7 0.3 ARLBExtension Extension+FL Factor 1: Cement 0.003 0.02 0.06 0.13 Factor 2: Extension < 0.001 < 0.001 < 0.001 < 0.001 Interactions 0.2 0.2 0.9 0.3 AR+: positive axial rotation, AR-: negative axial rotation, LB+: positive lateral bending, LB-: negative lateral bending, FL: Follower Load.  170  Chapter  Adjacent  7  Level  Effects  Axial Rotation (  Axial Rotation (+) u 3 c TJ  4.5 -.  4.5  -Cemented  4.0 •  4.0  -Not Cemented  3.5 •  3.5  O  ,N  3.0 re E 2.5 u. O 2.0 c u E 1.5 o 2 1.0  •  3.0  •  2.5  •  2.0  •  1.5  -  1.0  0.5 -  0.5  p = 0.02  TJ 9  Q. Q. < 0.0 •  0.0 No e x t e n s i o n  * p < 0.001  Figure 7.6  Rexible extension  No e x t e n s i o n  Rigid extension  Rexible extension  Mean normalized applied moment in AR at baseline ROM. Error bars indicate standard deviation. Lateral Bending (-)  Lateral Bending (+) t3 4.5 n 3 i 4.0 A TJ 3.5  4.5  -Cemented  4.0  -Not Cemented  3.0  3.0  d>  "re  2.5 2.0  1-5  1.5  1.0  1.0  o.  0.0  0.0 No e x t e n s i o n  * p < 0.001  Figure 7.7  p = 0.007  0.5  *0.5H <  p = 0.002  3.5  E 2.5 L. o Zc 2.0 E o 2  Rigid extension  Rexible extension  Rigid extension  No e x t e n s i o n  Rexible extension  Rigid extension  Mean normalized applied moment in LB at baseline ROM. Error bars indicate standard deviation.  171  Chapter  Adjacent  7  Level  Effects  Extension  Flexion 4.5 p = 0.04  • Cemented  4.0  • Not Cemented  3.5 3.0 2.5 2.0 1.5 1.0 0.5 0.0 No extension  Flexible extension  *p< 0.001  Figure 7.8  Rigid extension  No extension  1  Extension with Follower-Load  73 0) N  4.5-, 4.0  -I  3.5  -I  15 3.0 E 2.5 u  Rigid extension  Mean normalized applied moment in FE at baseline ROM. Error bars indicate standard deviation.  Rexion with Follower-Load *  Rexible extension  p = 0.004  4.5  • Cemented  4.0  • Not Cemented  3.5  -I  3.0 2.5  p = 0.02  O  c  CD  E  •o  2.0  2.0  1.5  1-5 1.0  1.0  p = 0.1  Q. < 0.0  i No extension  *p< 0.001  Figure 7.9  p = 0.002  0.5  cu  i  •  Rexible  Rigid  extension  extension  0.0  i  1 No extension  i  Rexible  Rigid  extension  extension  Mean normalized applied moment in FE with follower-load at baseline ROM. Error bars indicate standard deviation.  172  Chapter  7.3.3  7  Adjacent  Level  Effects  Intersegmental Range of Motion  At the same baseline ROM, intersegmental ROMs were not evenly distributed across levels in the thoracolumbar segments, and ROM across each level was affected by different surgical configurations (Table 7.5, Figure 7.10).  Magnitudes of intersegmental ROM in the  positive and negative directions were not symmetrical but trends of changes following cement or rod extension were similar. The overall intersegmental ROM was summed from the magnitudes of the intersegmental ROM+ and ROM-, and normalized against the intact overall intersegmental ROM for further comparison.  Table 7.5  Not cemented Cemented  Not cemented Cemented  Not cemented Cemented  Intersegmental ROM+, intersegmental ROM- and overall intersegmental ROM in axial rotation for specimen 1. Intersegmental ROM+ and ROMcorresponded to baseline ROM+ and ROM-. Overall intersegmental ROM was summed from the intersegmental ROM+ and ROM-  Applied Moment (Nm)  1 level above D (T9-T10)  Intact No ext Flex ext Rigid ext No ext Flex ext Rigid ext  1.9 0.6 1.3 2.2 3.1 4.1 5.0  3.7° 3.0° 3.8° 4.1° 4.8° 5.1° 5.5°  Intact No ext Flex ext Rigid ext No ext Flex ext Rigid ext  -1.7 -0.6 -1.4 -2.9 -2.4 -4.4 -5.0  -2.5° -2.2° -2.3° -3.1° -3.2° -3.4° -3.5°  Intact No ext Flex ext Rigid ext No ext Flex ext Rigid ext  6.3° 5.2° 6.1° 7.2° 8.0° 8.5° 9.0°  Intersegmental ROM+ D Rod 2 levels (T10-12) extension below D level (L1-L2) (T12-L1) 5.6° 1.0° 2.8° 7.9° 0.1° 2.0° 6.5° 0.2° 2.5° 0.2° 2.9° 5.0° 3.2° 2.5° 1.3° 2.5° 0.7° 3.4° 2.3° 0.5° 3.5° Intersegmental ROM-1.2° -3.8° -2.3° -5.9° -0.6° -2.0° -4.4° -0.6° -2.3° -3.2° -0.6° -2.6° -1.4° -2.6° -2.0° -1.1° -2.9° -2.0° -1.8° -0.8° -3.0° Overall intersegmentai ROM 9.4° 2.2° 5.0° 0.7° 13.8° 4.0° 10.9° 0.9° 4.8° 8.2° 0.9° 5.5° 2.7° 4.5° 5.9° 1.9° 4.6° 6.3° 4.1° 1.3° 6.5°  3 levels below D (L2-L3) 3.1° 2.7° 3.2° 3.5° 3.9° 4.0° 4.1° -2.3° -1.5° -2.1° -2.5° -2.8° -2.8° -2.9° 5.3° 4.3° 5.3° 6.1° 6.7° 6.8° 7.0°  D: destabilized level, No ext: No extension rods, Flex ext: Flexible extension rods, Rigid ext: rigid extension rods. 173  Chapter 7  O  CC  Adjacent Level Effects  Axial Rotation:  30  I  25  -  20  -  H1022  • 0  1 5  D+1 D  H D - 1 10  •  D-2  •  D-3  5H  T  Intact D: destabilized  Figure  7.10  level  •  •  r  N o  Flexible  Rigid  C e m e n t e d ,  C e m e n t e d ,  Extension  Extension  Extension  no ext  flex ext  C e m e n t e d , rig  ext  Distribution of overall intersegmental range of motion (ROM) in axial rotation for specimen 1 (corresponds to data in Table 7.5). Increase in ROMs at non-instrumented levels and decrease in ROMs at instrumented levels were generally observed.  Mean normalized intersegmental ROMs across all levels for the 12 specimens were affected by cement augmentation and/or posterior rod extension (Figures 7.11 to 7.14). Reflecting the results for specimen 1 (Figure 7.10), mean normalized intersegmental ROM at the destabilized and extension level were reduced by posterior rod extension and cement augmentation of screws, while intersegmental motions at the non-instrumented levels were increased in order to maintain the overall ROM. Detailed 2-way ANOVA analyses for each level and flexibility test direction are presented in the Appendix.  174  Chapter 7  Adjacent Level Effects Axial Rotation ROM using Hybrid Flexibility-Stiffness Protocol  3n  tS 8c  H  2.5  2H •= 1.5 H  A  0.5  1 level above destabilized level  *significant Error Figure  bars  7.11  destabilized level  1 level below destabilized level  2 levels below destabilized level  3 levels below destabilized level  • No extension  • Flexible extension  • Rigid extension  Q Cemented, no ext  B Cemented, flex ext  • Cemented, rig ext  effects indicate  of cement standard  Normalized ANOVA  and  *significant  effect  of extension  only.  deviation.  intersegmental analyses  extension.  range  for each  of motion  level  presented  (ROM) in AR. in  Detailed  2-way  Appendix.  Lateral Bending ROM using Hybrid Flexibility-Stiffness Protocol  2.5  1  u S c •o 1.5 o  .2  |  0.5  1 level above destabilized level  destabilized level  • No extension Q Cemented, no ext *significant Error Figure  7.12  bars  effects indicate  Normalized ANOVA  of cement  • Rigid extension  • Cemented, flex ext  • Cemented, rig ext  **significant  effect  of extension  only.  deviation.  intersegmental analyses  3 levels below destabilized level  0 Flexible extension  and extension.  standard  2 levels below destabilized level  1 level below destabilized level  for each  range level  of motion presented  (ROM) in LB. in  Detailed  2-way  Appendix.  175  Chapter  7  Adjacent  Level  Effects  Flexion-Extension ROM usingHybrid Flexibility-Stiffness Protocol  4  3.5 3 2.5 2 1.5 • 10.5 • 0  i1l l I•  1 level above destabilized level  destabilized level  • No extension H Cemented, no ext *significant Error Figure  bars  7.13  effects indicate  Normalized ANOVA  of cement  2 levels below destabilized level  • Flexible extension B Cemented, flex ext and  standard  extension.  3 levels below destabilized level  • Rigid extension • Cemented, rig ext  *significant  effect  of extension  only.  deviation.  intersegmental analyses  1 level below destabilized level  range  for each  level  of motion presented  (ROM) in  in FE.  Detailed  2-way  Appendix.  Flexion-Extension +Follower Load, ROM usingHybrid Flexibility-Stiffness Protocol  4  3.5 3 2.5 2  11  1.5 1 -r 0.5 0  1 level above destabilized level  destabilized level  • No extension • Cemented, no ext *significant Error Figure  7.14  bars  effects indicate  Normalized way ANOVA  of cement  2 levels below destabilized level  E9 Flexible extension B Cemented, flex ext and  standard  extension.  3 levels below destabilized level  • Rigid extension • Cemented, rig ext  *significant  effect  of extension  only.  deviation.  intersegmental analyses  1 level below destabilized level  range  for each  of motion  level  presented  (ROM) in  in FE+FL.  Detailed  2-  Appendix.  176  Chapter  7  Adjacent  Level  Effects  Rod Extension Level In the 1-way ANOVA analyses, ROMs at the rod extension level (1 level below destabilized level, T12-L1 in T9-L3 segments) were significantly affected by surgical intervention in all directions (Table 7.6). The initial surgical configuration (no cement, no rod extension) was not significantly different from intact in all directions, while posterior rod extension significantly reduced ROMs from corresponding intact values in LB and FE (Tables 7.6 and 7.7). In the 2-way ANOVA analyses, cement augmentation significantly affected ROM at the rod extension level in FE only, but not in AR and LB, while posterior rod extension significantly reduced ROM in all directions (Table 7.8, Figures A.9 to A. 12 in Appendix A). There were significant 2-way interactions in all directions, which were probably due to higher ROMs in the cemented and with no rod extension state. Table 7.6  Mean ROM across rod extension level for different surgical conditions using the hybrid protocol. Range of Motion, mean (s.d.) AR LB FE FE+FL Intact 1.6° (1.4°) 3.2° (1.3°) 2.7° (1.3°) 2.4° (1.5°) Not No extension 1.5° (1.3°) 3.4° (1.2°) 2.9° (1.3°) 2.7° (1.3°) cemented Flexible extension 1.3° (1.0°) *2.4° (0.9°) *1.7° (0.7°) *1.7° (0.8°) Rigid extension 1.2° (0.8°) *1.5° (0.9°) *2.2°(1.1°) *1.5° (0.9°) No extension 2.0° (1.3°) 3.7° (1.0°) 2.9° (1.0°) 2.6° (1.2°) Cemented Flexible extension 1.5° (1.0°) *1.7° (0.5°) *1.2° (0.6°) *1.1°(0.7°) Rigid extension 1.2° (0.9°) *1.2° (0.7°) *0.9° (0.5°) *0.8° (0.5°) p-values from 1-way ANOVA < 0.001 < 0.001 < 0.001 < 0.001 * - significantly different from corresponding intact value, alpha = 0.01.  Table 7.7  Not cemented  Cemented  Percentage difference in ROM from intact and statistical significance, for rod extension level. Percentage Difference from Intact, p-values AR LB FE FE+FL No extension -11%, 0.5 6%, 0.5 7%, 0.4 14%, 0.4 Flexible extension -19%, 0.3 -25%, 0.005 -36%,<0.001 -28%, 0.007 Rigid extension -26%, 0.09 -32%, 0.001 -42%,<0.001 -39%,<0.001 No extension 22%, 0.04 16%, 0.2 8%, 0.6 7%, 0.5 Flexible extension -9%, 0.4 -47%,<0.001 -55%,<0.001 -53%,<0.001 Rigid extension -27%, 0.1 -62%,<0.001 -67%,<0.001 -67%,<0.001  2-way ANOVA results on normalized ROM between cement (factor 1) and posterior rod extension (factor 2) using the hybrid protocol for rod extension level AR LB FE FE+FL Factor 1: Cement 0.3 0.1 0.002 0.008 Factor 2: Rod extension < 0.001 < 0.001 < 0.001 < 0.001 Interactions < 0.001 < 0.001 0.006 0.006  7ao/e 7.8  177  Chapter 7  Adjacent Level Effects  2 Levels below Destabilized Level  In the 1-way ANOVA analyses, ROMs at 2 levels below destabilized level (L1-L2 in T9L3 segments) were significantly affected by surgical intervention in LB and FE (Table 7.9). The initial surgical configuration (no cement, no extension) was not significantly different from intact in all directions, while posterior rod extension and cement augmentation significantly increased ROMs from corresponding intact values in LB and FE (Tables 7.9 and 7.10). In the 2-way ANOVA analyses, cement augmentation marginally increased ROM at 2 levels below destabilized level in LB, but not in AR and FE (Table 7.11, Figures A. 13 to A.16 in Appendix A). Posterior rod extension significantly increased ROM in all directions. There were no significant 2-way interactions. Table 7.9  Mean ROM at 2 levels below destabilized level (L1-L2) for different surgical conditions using the hybrid protocol.  Intact Not No extension cemented Flexible extension Rigid extension No extension Cemented Flexible extension Rigid extension p-values from 1-way ANOVA  AR 1.5° (1.4°) 1.2° (1.2°) 1.4° (1.4°) 1.6° (1.5°) 1.7° (1.5°) 1.7° (1.7°) 1.8° (1.7°) 0.03  Range of Motion, mean (s.d.) LB FE FE+FL 3.6° (1.2°) 2.9° (1.2°) 4.0° (1.5°) 3.5° (1.2°) 4.4° (1.6°) *3.4°(1.5°) *4.3°(1.4°) *4.1°(1.6°) *5.2°(1.9°) *5.6° (2.0°) *4.8°(1.6°) *4.5°(1.7°) *5.2° (1.9°) *4.2° (4.6°) *3.7°(1.5°) *4.8°(1.8°) *5.8° (2.0°) *4.4°(1.8°) *6.2° (2.1°) *5.1°(2.0°) *4.5°(1.8°) < 0.001 < 0.001 < 0.001  * - significantly different from corresponding intact value, alpha = 0.01. Table 7.10  Not cemented Cemented  Table 7.11  Percentage difference in ROM from intact and statistical significance, for 2 levels below destabilized level.  Percentage Difference AR LB No extension -17%, 0.4 9%, 0.2 Flexible extension -6%, 0.6 29%,<0.001 Rigid extension 10%, 0.4 38%,<0.001 No extension 17%, 0.5 29%,<0.001 Flexible extension 12%, 0.6 45%,<0.001 Rigid extension 23%, 0.3 55%,<0.001  from Intact, p-values FE FE+FL -4%, 0.5 18%, 0.002 19%, 0.005 40%,<0.001 33%,<0.001 53%,<0.001 16%, 0.007 28%,<0.001 31%,<0.001 49%,<0.001 40%,<0.001 55%,<0.001  2-way ANOVA results on normalized ROM between cement (factor 1) and posterior rod extension (factor 2) using the hybrid protocol, for 2 levels below destabilized level.  Factor 1: Cement Factor 2: Rod extension Interactions  AR 0.06 < 0.001 0.2  LB 0.01 < 0.001 0.5  FE 0.02 < 0.001 0.02  FE+FL 0.09 < 0.001 0.06 178  Chapter 7  Adjacent Level Effects  3 Levels below Destabilized Level In the 1-way ANOVA analyses, ROMs at 3 levels below destabilized level (L2-L3 in T9L3 segments) were significantly affected by surgical intervention in all directions (Table 7.12). The initial surgical configuration (no cement, no rod extension) was not significantly different from intact in all directions, while posterior rod extension and cement augmentation generally increased ROMs from intact values in LB and FE (Tables 7.12 and 7.13). In the 2-way ANOVA analyses, ROMs at 3 levels below destabilized level were generally increased with cement augmentation and posterior rod extension, except in FE (Table 7.14, Figures A. 17 to A.20 in Appendix A). There were no significant 2-way interactions. Table 7.12  Mean ROM at 3 levels below destabilized level (L2-L3) for different surgical conditions using the hybrid protocol. Range of Motion, mean (s.d.) AR LB FE FE+FL Intact 2.1° (1.6°) 5.9° (2.2°) 4.1° (2.2°) 3.2° (2.1°) Not No extension 1.8° (1.0°) 5.9° (2.2°) 4.2° (1.9°) *3.8° (2.2°) cemented Flexible extension 2.0° (1.2°) *6.8° (2.5°) *4.8° (2.3°) *4.4° (2.4°) Rigid extension 2.3° (1.4°) *7.1°(2.6°) *5.0° (2.2°) *4.3° (2.2°) No extension 2.7° (1.8°) 6.6° (2.6°) *4.9° (2.4°) *4.3° (2.3°) Cemented Flexible extension 2.6° (1.5°) *7.6° (2.7°) *5.4° (2.6°) *4.8° (2.6°) Rigid extension 2.9° (1.6°) *7.7° (2.7°) *5.5° (2.5°) *4.9° (2.6°) p-values from 1-way ANOVA < 0.001 < 0.001 < 0.001 < 0.001 * - significantly different from corresponding intact value, alpha = 0.01.  Table 7.13  Not cemented Cemented  Percentage difference in ROM from intact and statistical significance, for 3 levels below destabilized level. Percentage Difference from Intact, p-values AR LB FE FE+FL No extension -17%, 0.3 - 1 % , 0.9 4%, 0.4 20%, 0.008 Flexible extension -6%, 0.6 15%,0.007 18%,<0.001 39%,<0.001 Rigid extension 7%, 0.5 20%,<